VAMC Oklahoma City - Annual Medical Gas Inspection SEPT 2025 JDa 02.docx

DOCX document 25 MB Posted

Attached to
H365--Medical Gas Annual Inspection Services for the OKC VA Medical Center and Lawton/Ft. Sill
Solicitation number
36C25926Q0722
Issued by

About this file

This document is an Annual Medical Gas and Medical Surgical Vacuum Inspection Report for the VA Oklahoma City Health Care System dated September 22-24, 2025. The inspection was conducted by inspectors Al and Chip Moon (MGPHO and CMGV certifications) in accordance with NFPA 99 2012 Edition standards and follows a comprehensive assessment methodology covering master alarm systems, gas supply sources, medical air compressor systems, medical-surgical vacuum systems, alarm functionality, valve operations, and gas concentration testing across all facility zones.

The inspection identified multiple deficiencies requiring corrective action across the 921 NE 13th Street facility, including: outlet/inlet stations with leaks, broken latches, and gauges requiring repair; improperly labeled zone valve boxes and area alarms; failed alarm signals for medical vacuum pumps and reserve gas systems at multiple panel locations (boiler room and master alarm panels); and a medical-surgical vacuum pump system requiring replacement due to age (over 40 years old) and non-compliance with current NFPA 99 code. Additional issues documented include bad gauges in operating rooms (OR #8, OR #7, OR #6), leaking WAGD (waste anesthetic gas disposal) connections in ORs #4 and #2, leaking vacuum units in OR #1, oxygen leaks in various patient care areas, low medical air pressure readings at multiple alarm panels, and inoperable oxygen flowmeters. The report provides detailed zone valve box data sheets, station outlet/inlet data, and alarm panel specifications for all facility areas from basement through seventh floor, documenting gas pressures, flow rates, outlet styles, and equipment make/models across operating rooms, intensive care units, emergency departments, surgical areas, imaging departments, and specialty clinics. Testing shows oxygen outlets consistently delivering 99+ percent purity, medical air at 19.5-23.5 percent oxygen, and vacuum systems at specified negative pressures, with pass indicators (P) or not-tested (NT) notations for individual stations.

View the file

Other files

File Type Posted
36C25926Q0722 0001.docx DOCX document
36C25926Q0722.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SEPT. 22-24, 2025

Annual Medical Gas and Medical Surgical Vacuum Inspection

VA Oklahoma City Health Care System 921 NE 13th Street Oklahoma City, OK 73104

On SEPT. 22-24, 2025 the Annual Medical Gas and Vacuum Inspection was performed at your facility. Below you will find our findings and recommendations to any issues found during this inspection. Please feel free to contact me with any questions or concerns.

NFPA 99 Chapter 1 Section # 5.1.1.4 An existing system that is not in strict compliance with the provisions of this code shall be permitted to be continued in use, if, the authority having jurisdiction has determined that such use does not constitute a distinct hazard to life.

Recommendations will be listed BLACK:

Observations will be listed BLUE:

Deficiencies will be listed RED:

#1.) On this inspection there were outlet / inlet stations with leaks, broken latches, or are in need for repair plus zone valve box gauges that are bad.:

SEE REPAIR LIST

#2.) During the inspection it was noted that several zone valves boxes or area alarms were not properly labeled:

SEE REPAIR LIST

#3.) Missing alarm signal or failed Signals

a. Medical Vacuum Pump (LAG PUMP IN USE) at both master alarm panels missing.

b. N2 & N2O Reserve In Use Signals Failed To Alarm at Boiler Room Location.

c. N2 Low Pressure Failed To Alarm at Boiler Room Location.

#4.) On this inspection there are notes on Medical Surgical Vacuum Pump Systems:

THIS MEDICAL VACCUM PUMP SYSTEM SHOULD BE REPLACEMENT SOON.

OVER 40 YEARS OLD AND OUTDATED PER NFPA 99 CODE 2012 EDITION

AND ALL NEWER EDITIONS

Annual Medical Gas and Vacuum Inspection

This review follows the recommendations of the National Fire Protection Association Standards for Health Care Facilities (NFPA 99 2012 Edition)

· Master Alarms Test: The master alarm system tests shall be performed for each of the medical gas and vacuum piping systems. See report for details. (5.1.12.3.5.2)

· Gas Supply Sources Test: The system apparatus shall be tested for proper function, including the changeover from primary to secondary supply, and the operation of the reserved. See report for details. (5.1.12.3.14.2)

· Medical Air Compressor Systems: Tests of the medical air compressor system shall include the purity test for air quality and the test of alarm sensors after calibration and setup per manufacturer's instructions, as well as lead-lag controls. See report for details. (5.1.12.3.14.3)

· Medical-Surgical Vacuum Systems: The proper functioning of the medical-surgical vacuum source systems shall be tested before it is put into service. See report for details. (5.1.12.3.14.4)

· Alarms Test: All warning systems for each medical gas and vacuum systems shall be tested to ensure that all components function properly. See report for details. (5.1.12.3.5)

· Valve Test: Valves installed in each medical gas and vacuum piping system shall be tested to verify proper operation and rooms or areas of control. See report for details. (5.1.12.3.4)

· Medical Gas Concentration Test: After purging each system with gas of system designation the following shall be performed: each pressure gas source and outlet shall be analyzed for concentration of gas by volume. Random sampling followed; all available areas were tested. See report for details. (5.1.12.3.11)

· Station Outlet / Inlet Data Passing Guidelines

· Oxygen 99 plus percent oxygen, Nitrous oxide 99 plus percent nitrous oxide, Nitrogen Less than 1 percent oxygen or 99 plus Percent nitrogen, medical air 19.5 percent to 23.5 percent oxygen, other gases Concentration as specified by their labeling ±1 percent, unless otherwise specified: Oxygen, nitrous oxide, and med air outlets shall deliver 3.5 SCFM with a pressure drop of no more than 5 psig (35 kPa), and static pressure of 48 psig (348 kPa). Nitrogen outlets shall deliver 5.0 SCFM with a pressure drop of no more than 5 psig and static pressure of 160 psig. Piping shall be sized such that 3 SCFM can be evacuated through any one station inlet without reducing vacuum pressure below 12 in Hg at an adjacent station inlet. Oxygen and air outlet serving critical care areas shall permit a transient flow rate of 6.0 SCFM for 3 seconds. NT = Not Tested Random Testing

This review was performed on the date noted and was found to be in compliance with the conditions listed above unless noted. The details of this report can be found on the attached pages.

AL & Chip MOON Al & Chip Moon

MGPHO CMGV #V-0002 ASSE 6030 #00021182

6010/6020/6030/6035/6040/6050

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT 22-24,
Zone Valves Floor & Location:
7th Floor (Outside Recovery Across From 7F-1089)

Recovery 7F-154, 7F-154B

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA

Comments:

Station Outlet/Inlet Data

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor
Room
OXY
AIR
VAC
Flows
Style
Comments

7F-154

Bed 1
2
1
1
P
Railway / Pin Index
Bed 2
2
1
1
NT
Railway / Pin Index
Bed 3
2
1
1
P
Railway / Pin Index
Bed 4
2
1
1
P
Railway / Pin Index
Bed 5
2
1
1
P
Railway / Pin Index
Bed 6
2
1
1
P
Railway / Pin Index
7F-108G
2
1
1
P
Railway / Pin Index

7F-154B

EYE PAIN

2
2
3
P
Railway / Pin Index

Alarm Panel Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
7th Floor (Inside Recovery By Bed 6)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
51
P

40 Low & 60 High

P
Allied / Digital Impact
Medical Air
50
P

40 Low & 60 High

P
Allied / Digital Impact
Vacuum
-22
P

-12 Low

P
Allied / Digital Impact

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Located By Bed 6 7F-154)

Recovery 7F-108, 7F-108G

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
NA
NA
Medical Air
P
Y
½”
NA
NA
Vacuum
P
Y
1”
NA
NA
Comments:
Not In Use
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Outside Recovery Next to 7F-156)

7F-148, 7F-150B

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
Medical Air
P
Y
½”
Vacuum
P
Y
3/4”
Comments:
NOT IN USE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

7F-148, 7F-148B

Room
OXY
AIR
VAC
Flows
Style
Comments
7F-148
1
1
1

Custom / Pin Index

7F-150B
2
2
2

Custom / Pin Index

NOT IN USE

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #8)

OR #8 7F-151

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y BAD
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y BAD
¾”
-16
NA
Comments:
BAD MED VACUUM AND WAGD GAUGES
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #8 7F-151

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-151

Boom #1 Stryker

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
NCC / DISS
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #7)

OR #7 7F-142

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y BAD
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
NA
Comments:
BAD VACUUM GAUGE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #7 7F-142

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-142

Boom #1

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
Chemetron / NCP /DISS
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #6)

OR #6 7F-140

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
BAD GAUGE
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
NA
Comments:
N2O BAD GAUGE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #6 7F-140

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-140

Booms
4
P
Pin Index
P
Pin Index
P
Pin Index
P
Pin Index
2
P
Pin Index
P
Chemetron / NCP /DISS
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #5)

OR #5 7F-134

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
N/A

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #5 7F-134

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-134

Boom #1

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
NCP / DISS

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #4)

OR #4 7F-128

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
N/A
Comments:
7F 124 NOT IN USE / NOT IN SERVICE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #4 7F-128

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

Boom #1

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
LEAK ON WAGD

BOOM #1

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index

Boom #3

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #3)

OR #3 7F-124

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
N/A

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #3 7F-124

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-124

Boom #1

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
Chemetron / NCP /DISS

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
NCP / DISS
Wall
1
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #2)

OR #2 7F-122

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #2 7F-122

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-122

.
2
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
LEAKING WAGD

0R#2

P
Chemetron / NCP /DISS

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
NCP / DISS

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Sterile Corridor Outside OR #1)

OR #1 7F-118

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
54
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-26
NA
Nitrous Oxide
P
Y
½”
48
99+
Nitrogen
P
Y
½”
160
0
Carbon Dioxide
N/A
N/A
N/A
N/A
N/A
WAGD
P
Y
¾”
-16
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor

OR #1 7F-118

Room
OXY
AIR
VAC
N2O
CO2
NIT
WAGD
Flows
Style
Comments

7F-118

Boom #1

P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
Chemetron / NCP /DISS

Boom #2

P
Series (B) / Pin Index
P
Series (B) / Pin Index
NOTE
Series (B) / Pin Index
LEAKING VACUUM UNIT (A)

OR #1

P
Series (B) / Pin Index
1
P
Series (B) / Pin Index
P
NCP / DISS
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
7th Floor (Inside Control 7F-155 NS)

OR #1-8

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Allied / Digital Impact
Medical Air
48
P

40 Low & 60 High

P
Allied / Digital Impact
Vacuum / WAGD
-26
P

-12 Low

P
Allied / Digital Impact
Nitrous Oxide
48
P

40 Low & 60 High

P
Allied / Digital Impact
Nitrogen
160
P

140 Low & 190 High

P
Allied / Digital Impact

MAINLINE AREA ALARM

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Next to Room 7A-135)

7A-151, 7A-153, 7A-154, 7A-159, 7A-160, 7A-162A, 7A-163, 7A-164, 7A-160

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
54
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1 1/4”
-26
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor
Room
OXY
AIR
VAC
Flows
Style
Comments
7A-151
1
1
1
P
Beacon / Pin Index
7A-153
1
1
1
NT
Beacon / Pin Index
7A-154
1
1
1
P
Beacon / Pin Index

7A-158

1
1
1
P
Beacon / Pin Index
7A-160
1
1
1
P
Pin Index / Ohio
7A-162
1
1
1
P
Beacon / Pin Index
7A-162A
1
1
1
P
Beacon / Pin Index
7A-163
1
1
1
P
Beacon / Pin Index
7A-164
1
1
1
P
Beacon / Pin Index
7A-165
1
1
1
P
Beacon / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
7th Floor (Nurse Station For Room Outpatient Surgery)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Pin Index / Beige
Medical Air
48
P

40 Low & 60 High

P
Pin Index / Beige
Vacuum
-26
P

-12 Low

P
Pin Index / Beige
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Next to Room 7A-105 )

7A106,7A108,7A112,7A113,7A115,7A116,7A121,7A122,7A123,7A124

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
51
99
Medical Air
P
Y
1”
48
21
Vacuum
P
Y
1 ¼”
21
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Next to Room 7A-105 )
Room
OXY
AIR
VAC
Flows
Style
Comments
7A-106
1
1
1
P
AMICO

PIN INDEX

7A-108
1
1
1
P
AMICO
7A-112
1
1
1
P
AMICO
7A-113
1
1
1
NT
AMICO
7A-115
1
1
1
P
AMICO
7A-116
1
1
1
NT
AMICO
7A-121
1
1
1
P
AMICO
7A-122
1
1
1
P
AMICO
7A-123
1
1
1
P
AMICO
7A-124
1
1
1
P
AMICO
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
7th Floor (IN NS)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
51
P

40 Low & 60 High

P
BEACON MEGA 3
Medical Air
48
P

40 Low & 60 High

P
BEACON MEGA 3
Vacuum
-22
P

-12 Low

P
BEACON MEGA 3

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Next to Room 7B-103 )

Bio Med Dept

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
F
Y
¾”
52
99
Medical Air
F
Y
¾”
48
21
Vacuum
F
Y
1”
-19
N/A

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
7th Floor (Next to Room 7B-103 )

Bio Med Dept

Room
OXY
AIR
VAC
Flows
Style
Comments

BIO-MED

#1

1
1
1
P
DISS

WITH HOSE DROP PIN INDEX

BIO-MED

#2

1
1
1
P
DISS

WITH HOSE DROP PIN INDEX

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6B-102) NORTH

6B-105, 6B-106, 6B-111, 6B-112, 6B-114, 6B-116, 6B-117, 6B-119, 6B-122, 6B-123, 6B-125, 6B-128, 6B-129

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
52
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor / NORTH

6B-105, 6B-106, 6B-111, 6B-112, 6B-114, 6B-116, 6B-117, 6B-119, 6B-122, 6B-123, 6B-125, 6B-128, 6B-129

Room
OXY
AIR
VAC
Flows
Style
Comments
6B-105
1
1
1
NT
Series 2 / Pin Index
6B-106
1
1
1
P
Series 2 / Pin Index
6B-111
4
3
4
P
Series 2 / Pin Index
6B-112
2
1
2
P
Series 2 / Pin Index
6B-114
4
2
4
P
Series 2 / Pin Index
6B-116
1
1
1
P
Series 2 / Pin Index
6B-117
1
1
1
P
Series 2 / Pin Index
6B-119
2
1
2
NT
Series 2 / Pin Index
6B-122
3
2
3
NT
Series 2 / Pin Index
6B-123
3
2
3
P
Series 2 / Pin Index
6B-125
4
2
4
9
Series 2 / Pin Index
6B-128
1
1
1
P
Series 2 / Pin Index
6B-129
1
1
1
P
Series 2 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
6th Floor (Nurse Station Room 6B-106) NORTH

6B-105, 6B-106, 6B-111, 6B-112, 6B-114, 6B-116, 6B-117, 6B-119, 6B-122, 6B-123, 6B-125, 6B-128, 6B-129

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
50
P

40 Low & 60 High

P
Allied / Impact
Medical Air
48
P

40 Low & 60 High

P
Allied / Impact
Vacuum
-23
P

-12 Low

P
Allied / Impact

MED AIR IN RED AS LOW PRESSURE (FF CODE)

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6A-165)

6A-165

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
52
NA
Medical Air
P
Y
¾”
48
NA
Vacuum
P
Y
¾”
-23
NA
Comments:
Not in use valve handles removed.

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6A-131) GI

6A-132, 6A-148, 6A-149, 6A-150, 6A-151, 6A-153, 6A-154, 6A-155, 6A-158

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
53
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor GI

6A-132, 6A-148, 6A-149, 6A-150, 6A-151, 6A-153, 6A-154, 6A-155, 6A-158

Room
OXY
AIR
VAC
Flows
Style
Comments
6A-132
1
1
3
P
Beacon / Pin Index
6A-148
2
2
2
P
Beacon / Pin Index
6A-149
2
2
2
P
Beacon / Pin Index
6A-150
2
2
2
P
Beacon / Pin Index
6A-151
2
2
2
P
Beacon / Pin Index
6A-153
4
1
4
P
Series 2 / Pin Index
6A-154
4
1
4
P
Series 2 / Pin Index
6A-155
4
1
4
P
Series 2 / Pin Index
6A-158
4
1
4
P
Series 2 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
6th Floor (Nurse Station Room 6A-151) GI

6A-132, 6A-148, 6A-149, 6A-148, 6A-151, 6A-153, 6A-154, 6A-155, 6A-158

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
54
P

40 Low & 60 High

P
Allied IMPAC
Medical Air
48
P

40 Low & 60 High

P
Allied IMPAC
Vacuum
-25
P

-12 Low

P
Allied IMPAC
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6A-103) EAST

6A-102 - 6A-130

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
53
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6A-103) EAST

6A-102 - 6A-125

Room
OXY
AIR
VAC
Flows
Style
Comments
6A-102
1
1
1
P
Diamondcare / Pin Index
6A-104
1
1
1
P
Diamondcare / Pin Index
6A-106
1
1
1
P
Diamondcare / Pin Index
6A-106 OXYGEN BROKE FRONT PLATE
6A-108
2
2
2
P
Diamondcare / Pin Index
6A-110
1
1
1
P
Diamondcare / Pin Index
6A-113
1
1
1
NT
Diamondcare / Pin Index
6A-114
1
1
1
P
Diamondcare / Pin Index
6A-116
1
1
1
NT
Diamondcare / Pin Index
6A-117
1
1
1
P
Diamondcare / Pin Index
6A-119
1
1
1
P
Diamondcare / Pin Index
6A-123
1
1
1
P
Diamondcare / Pin Index
6A-124
1
1
1
NT
Diamondcare / Pin Index
6A-126
1
1
1
P
Diamondcare / Pin Index
6A-126 OXYGEN BROKE FRONT /

MED VACUUM BAD FRONT

6A-127
1
1
1
P
Diamondcare / Pin Index
6A-129
1
1
1
P
Diamondcare / Pin Index
6A-130
1
1
1
P
Diamondcare / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
6th Floor (Nurses Station next to Room 6A-103) EAST

6A-102, 6A-125

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
54
P

40 Low & 60 High

P
Beacon / Mega 2
Medical Air
41
P

40 Low & 60 High

P
Beacon / Mega 2
Vacuum
-25
P

-12 Low

P
Beacon / Mega 2

MEDICAL AIR READING LOW AT 41 PSIG / GAUGE READING IS 49 PSIG

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor (Next to Room 6F-109)

6F-110 SPD Dispatch & Testing

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
Medical Air
P
Y
½”
Vacuum
P
Y
3/4”
Comments:
NOT IN SERVICE / UNDER CONSTRUUCTION /

VALVE AND CAP OFF / REVIEW CAP AS A SHARK BIT TYPE

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
6th Floor

6F-110 SPD Dispatch & Testing

Room
OXY
AIR
VAC
Flows
Style
Comments

NOT IN SERVICE / UNDERCONSTRUUCTION /

VALVE AND CAP OFF / REVIEW CAP AS A SHARK BITE TYPE

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-109F) SICU #3-10

5F-109Q-U, 5F-109A, 5F-109F, 5F-109G

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
53
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor SICU #3-10
Room
OXY
AIR
VAC
Flows
Style
Comments
5F-109Q
3
2
4
NT
Railway / Pin Index
5F-109R
3
2
4
P
Railway / Pin Index
5F-109S
3
2
4
P
Railway / Pin Index
5F-109T
3
2
4
NT
Railway / Pin Index
5F-109U
3
2
4
P
Railway / Pin Index
5F-109A
3
2
4
P
Railway / Pin Index
5F-109F
3
2
4
NT
Railway / Pin Index
5F-109G
3
2
4
P
Railway / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5F-109) SICU #3-10
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
51
P

40 Low & 60 High

P
Allied / Digital Impact
Medical Air
48
P

40 Low & 60 High

P
Allied / Digital Impact
Vacuum
-25
P

-12 Low

P
Allied / Digital Impact

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-109G) SICU

5F-109H, 5F-109I, 5F-109K, 5F-109L, 5F-109P, 5F-109N

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
53
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor SICU
Room
OXY
AIR
VAC
Flows
Style
Comments
5F-109H
6
4
8
P
Railway / Pin Index
5F-109I
3
2
4
P
Railway / Pin Index
5F-109K
3
2
4
NT
Railway / Pin Index
5F-109L
3
2
4
P
Railway / Pin Index
5F-109P
3
2
4
P
Railway / Pin Index
5F-109N
3
2
4
NT
Railway / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5F-109) SICU
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Allied / IMPAC
Medical Air
48
P

40 Low & 60 High

P
Allied / IMPAC
Vacuum
-25
P

-12 Low

P
Allied / IMPAC

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-117N) MICU #1-4 - 9 &1 0

5F-114, 5F-116, 5R-117A-C, 5F-117S, 5F-117M, 5F-117N

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
48
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor MICU #1-4 - 9 &1 0
Room
OXY
AIR
VAC
Flows
Style
Comments
5F-114
3
2
4
P
Railway / Pin Index
NON-PATIENT USE
5F-116
3
2
4
P
Railway / Pin Index
NON-PATIENT USE
5F-117A
3
2
4
P
Railway / Pin Index
5F-117B
3
2
4
NT
Railway / Pin Index
5F-117C
3
2
4
P
Railway / Pin Index
5F-117S
3
2
4
P
Railway / Pin Index
5F-117M
3
2
4
NT
Railway / Pin Index
5F-117N
3
2
4
P
Railway / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5F-114) MICU #1-4 - 9 &1 0
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
TRI TECH LCD
Medical Air
48
P

40 Low & 60 High

P
TRI TECH LCD
Vacuum
-25
P

-12 Low

P
TRI TECH LCD

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-117M) MICU #5-8

5F-117H-K

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
51
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor MICU #5-8
Room
OXY
AIR
VAC
Flows
Style
Comments
5F-117H
3
2
4
P
Railway / Pin Index
5F-117I
3
2
4
P
Railway / Pin Index
5F-117J
3
2
4
P
Railway / Pin Index
5F-117J OXYGEN (B) LEAKING
5F-117K
3
2
4
P
Railway / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5F-114) MICU #5-8
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Allied / Digital
Medical Air
48
P

40 Low & 60 High

P
Allied / Digital
Vacuum
-25
P

-12 Low

P
Allied / Digital

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-118) DIALYSIS

5F-118F, 5F-118G, 5F-118J

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
51
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor DIALYSIS

5F-118F, 5F-118G, 5F-118J

Room
OXY
AIR
VAC
Flows
Style
Comments
5F-118F
1
1
2
P
Diamond 3 / Pin Index
5F-118G
1
1
2
NT
Diamond 3 / Pin Index
5F-118J
1
1
2
P
Diamond 3 / Pin Index

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5F-127) RESP THERAPY

5F-124, 5F-130-132, 5F-137, 5F-146, 5F-127, 5F-144

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
51
99+
Medical Air
P
Y
½”
48
21.3
Vacuum
P
Y
1”
-26
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor RESP THERAPY

5F-124, 5F-130-132, 5F-137, 5F-146, 5F-127, 5F-144

Room
OXY
AIR
VAC
Flows
Style
Comments
5F-124
2
2
2
P
Series 3 / Pin Index
5F-130
1
1
1
P
Series 3 / Pin Index
5F-131
1
1
1
P
Series 3 / Pin Index
5F-132
1
1
1
P
Series 3 / Pin Index
5F-137
1
1
1
P
Series 3 / Pin Index
5F-144
4
2
4
P
400 Series / Chemetron
5F-146
2
2
2
P
Series 3 / Pin Index
5F-146 MED AIR BEHIND CABNIET LEAKING
5F-127
0
0
1
P
Series 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5A-179)

5A-164, 5A-172

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
NA
NA
Medical Air
P
Y
½”
NA
NA
Vacuum
P
Y
1”
NA
NA
Comments:
Not in-service valves turned off.
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5A-145)
Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
NA
NA
Medical Air
P
Y
½”
NA
NA
Vacuum
P
Y
1”
NA
NA
Comments:
Not in-service valves turned / NOT IN SERVICE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5A-104)

5A-104,108,112,117,118,121,124,129,132,133,136,140

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
52
99
Medical Air
P
Y
¾”
49
21
Vacuum
P
Y
1”
-22
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5A-104)
Room
OXY
AIR
VAC
Flows
Style
Comments
5A-104
2
2
4
P
BEACON Pin Index
5A-108
2
2
4
P
BEACON Pin Index
5A-112
2
2
4
P
BEACON Pin Index
5A-117
2
2
4
P
BEACON Pin Index
5A-118
2
2
4
P
BEACON Pin Index
5A-121
2
2
4
P
BEACON Pin Index
5A-124
2
2
4
P
BEACON Pin Index
5A-129
2
2
4
P
BEACON Pin Index
5A-132
2
2
4
P
BEACON Pin Index
5A-133
2
2
4
P
BEACON Pin Index
5A-136
2
2
4
P
BEACON Pin Index
5A-140
2
2
4
P
BEACON Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Next to Room 5A-104)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
BEACON MEGA 3
Medical Air
48
P

40 Low & 60 High

P
BEACON MEGA 3
Vacuum
-25
P

-12 Low

P
BEACON MEGA 3

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5B-107) NORTH

5B-103, 5B-104, 5B-132, 5B-134

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
50
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor NORTH
Room
OXY
AIR
VAC
Flows
Style
Comments
5B-103
1
1
1
P
Diamond 3 / Pin Index
5B-104
1
1
1
P
Diamond 3 / Pin Index
5B-132
2
2
2
NT
Diamond 3 / Pin Index
5B-134
1
1
1
P
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5B-109) NORTH
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / Mega 2
Medical Air
48
P

40 Low & 60 High

P
Beacon / Mega 2
Vacuum
-25
P

-12 Low

P
Beacon / Mega 2
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5B-127) NORTH

5B-123, 5B-124, 5B-125, 5B-126, 5B-127

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
53
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor NORTH
Room
OXY
AIR
VAC
Flows
Style
Comments
5B-123
1
1
1
P
Diamond 3 / Pin Index
5B-124
1
1
1
P
Diamond 3 / Pin Index
5B-125
1
1
1
P
Diamond 3 / Pin Index
5B-126
1
1
1
NT
Diamond 3 / Pin Index
5B-127
1
1
1
P
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5B-109) NORTH
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / Mega 2
Medical Air
48
P

40 Low & 60 High

P
Beacon / Mega 2
Vacuum
-25
P

-12 Low

P
Beacon / Mega 2

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor (Next to Room 5B-117) NORTH

5B-111, 5B-112, 5B-113, 5B-117

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
53
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
5th Floor NORTH

5B-111, 5B-112, 5B-113, 5B-115, 5B-117

Room
OXY
AIR
VAC
Flows
Style
Comments
5B-111
2
2
2
P
Diamond 3 / Pin Index
5B-112
2
2
2
NT
Diamond 3 / Pin Index
5B-113
1
1
1
NT
Diamond 3 / Pin Index
5B-115
2
2
2
P
Diamond 3 / Pin Index
5B-117
1
1
1
P
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
5th Floor (Nurse Station Room 5B-109) NORTH

5B-111, 5B-112, 5B-113, 5B-115, 5B-117

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / Mega 2
Medical Air
48
P

40 Low & 60 High

P
Beacon / Mega 2
Vacuum
-25
P

-12 Low

P
Beacon / Mega 2
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor (Next to Room 4B-142) UROLOGY

4B-143

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
48
99+
Medical Air
P
Y
½”
48
21.3

Vacuum

P
Y
¾”
-23
NA
Nitrous Oxide
P
Y
½’
48
NA
Comments:
No nitrous oxide outlets in this room
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor UROLOGY

4B-143

Room
OXY
AIR
VAC
N2O
Flows
Style
Comments
Bed 1
2
2
2
0
P
Diamond 3 / Pin Index
Bed 2
2
2
2
0
P
Diamond 3 / Pin Index

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor (Next to Room 4B-145) UROLOGY

4B-144

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
48
99+
Medical Air
P
Y
½”
48
21.3
Vacuum / WAGD
P
Y
¾”
-23
NA
Nitrous Oxide
P
Y
½’
48
99+

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor UROLOGY

4B-144

Room
OXY
AIR
VAC
N2O
WAGD
Flows
Style
Comments
4B-144
1
P
Ohio / DISS

Hose drop

P
Ohio / DISS
P
Ohio / DISS
P
Ohio / DISS
1
P
Tri-Tech / Pin Index HD
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
4th Floor (Nurse Station Room 4B-135) UROLOGY

4B-143, 4B-144

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Pin Index / Beige
Medical Air
48
P

40 Low & 60 High

P
Pin Index / Beige
Vacuum / WAGD
-25
P

-12 Low

P
Pin Index / Beige
Nitrous Oxide
48
P

40 Low & 60 High

P
Pin Index / Beige

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor (Next to Room 4A-148) PALLIATIVE CARE

A-129, 4A-131, 4A-133, 4A-134, 4A-138-142, 4A-145

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
51
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1 ¼”
-26
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor PALLIATIVE CARE

4A-129, 4A-131, 4A-133, 4A-134, 4A-138-142, 4A-145

Room
OXY
AIR
VAC
Flows
Style
Comments
4A-129
2
2
2
P
Series B / Pin Index
4A-131
1
1
1
P
Series B / Pin Index
4A-133
2
2
2
P
Series B / Pin Index
4A-134
1
1
1
P
Series B / Pin Index
4A-138
1
1
1
NT
Series B / Pin Index
4A-139
1
1
1
P
Series B / Pin Index
4A-141
1
1
1
P
Series B / Pin Index
4A-142
1
1
1
NT
Series B / Pin Index
4A-145
2
2
2
P
Series B / Pin Index
4A-148
1
1
1
P
Series B / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
4th Floor (Nurse Station Room 4A-148) PALLIATIVE CARE

4A-129, 4A-131, 4A-133, 4A-134, 4A-138-142, 4A-145

Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / Mega 1
Medical Air
48
P

40 Low & 60 High

P
Beacon / Mega 1
Vacuum
NA
F
F
Beacon / Mega 1

4th Floor (Nurse Station Room 4A-148) PALLIATIVE CARE

MEDICAL VACUUM ALARM MODULE / COVERED UP

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor (Next to Room 4A-106) CLC

4A-106-113, 4A-116-121, 4A-123

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
48
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
4th Floor CLC
Room
OXY
AIR
VAC
Flows
Style
Comments
4A-106
2
2
2
NT
Diamond 3 / Pin Index
4A-107
2
2
2
P
Diamond 3 / Pin Index
4A-108
1
1
1
NT
Diamond 3 / Pin Index
4A-109
1
1
1
NT
Diamond 3 / Pin Index
4A-110
1
1
1
NT
Diamond 3 / Pin Index
4A-111
1
1
1
P
Diamond 3 / Pin Index
4A-112
1
1
1
NT
Diamond 3 / Pin Index
4A-113
2
2
2
P
Diamond 3 / Pin Index
4A-116
1
1
1
P
Diamond 3 / Pin Index
4A-117
1
1
1
NT
Diamond 3 / Pin Index
4A-118
1
1
1
P
Diamond 3 / Pin Index
4A-119
1
1
1
NT

Diamond 3 / Pin Index

4A-120
1
1
1
P
Diamond 3 / Pin Index
4A-121
1
1
1
P
Diamond 3 / Pin Index
4A-123
2
0
2
NT
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
4th Floor (Nurse Station Room 4A-148) CLC
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beige
Medical Air
48
P

40 Low & 60 High

P
Beige
Vacuum
-23
P

-12 Low

P
Beige

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3Rd Floor (Next to Room 3A-139) EXAMS ENT SOUTH

3E106,3E107,3E108,3E109,3E110,3E111

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
3/4”
51
NA
Medical Air
P
Y
1”
48
NA
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3Rd Floor (Next to Room 3A-139) EXAMS ENT SOUTH
Room
OXY
AIR
VAC
Flows
Style
Comments
3E-106
1
1
1
P
P/B (B) PIN INDEX
3E-107
1
1
1
P
P/B (B) PIN INDEX
3E-108
1
1
1
P
P/B (B) PIN INDEX
3E-109
1
1
1
P
P/B (B) PIN INDEX
3E-110
1
1
1
P
P/B (B) PIN INDEX
3E-111
1
1
1
P
P/B (B) PIN INDEX
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
3Rd Floor (Inside 3E-117) EXAMS ENT SOUTH
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon MEGA 3

Medical Air

Vacuum

NO ALARMS FOR THE MED AIR AND MED VACUUM

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor (Next to Room 3A-102) EAST
Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1 ½”
NA
NA
Medical Air
P
Y
1 ½”
NA
NA
Vacuum
P
Y
1 ½”
NA
NA
Comments:
Not in use valve handles removed.
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor (Next to Room 3B-142) DENTAL

3B-106-140

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
52
99+
Medical Air
P
Y
1”
48
21.3
Vacuum
P
Y
1 1/4”
-23
NA
Nitrous Oxide
P
Y
¾”
48
99
Nitrogen
P
Y
¾”
168
0
Comments:
3rd Floor (Next to Room 3B-142) DENTAL

REMOVE ROOMS / 3B-104,3B-148,3B-149,3B-150

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor DENTAL
Room
OXY
AIR
VAC
N2O
NIT
Flows
Style
Comments
3B-106
1
0
1
0
0
P
Series B / Pin Index
3B-107
1
0
1
0
0
P
Series B / Pin Index
3B-108
1
1
1
1
1
P
Series B / Pin Index
3B-110
1
1
1
1
1
P
Series B / Pin Index
3B-112
1
1
1
1
1
P
Series B / Pin Index
3B-113
1
1
1
1
1
P
Series B / Pin Index
3B-114
1
0
1
1
1
P
Series B / Pin Index
3B-115
1
0
1
1
1
NT
Series B / Pin Index
3B-116
1
0
1
0
0
P
Series B / Pin Index
3B-119
1
1
1
1
1
P
Series B / Pin Index
Medical Air Is Not Live
3B-120
1
1
1
1
1
P
Series B / Pin Index
Medical Air Is Not Live
3B-121
1
1
1
1
1
NT
Series B / Pin Index
Medical Air Is Not Live
3B-122
1
1
1
1
1
NT

Series B / Pin Index

3B-123
1
1
1
1
1
P
Series B / Pin Index
3B-124
1
1
1
1
1
P
Series B / Pin Index
3B-125
1
1
1
1
1
P
Series B / Pin Index
3B-127
1
1
1
1
1
P
Series B / Pin Index
3B-127 OXYGEN LEAK
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor DENTAL
Room
OXY
AIR
VAC
N2O
NIT
Flows
Style
Comments
3B-129
1
1
1
1
1
NT
Series B / Pin Index
3B-131
1
1
1
1
1
P
Series B / Pin Index
3B-133
1
1
1
1
1
P
Series B / Pin Index
3B-135
1
1
1
1
1
P
Series B / Pin Index
3B-137
1
1
1
1
1
P
Series B / Pin Index
3B-138
1
1
1
1
1
P
Series B / Pin Index
Medical Air is not live
3B-139
1
1
1
1
1
NT
Series B / Pin Index
Medical Air is not live
3B-115
1
1
1
1
1
P
Series B / Pin Index
3B-140
1
1
1
1
1
P
Series B / Pin Index
Medical Air is not live
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
3rd Floor (FRONT CHECK IN Dental Clinic 3B-101)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / MEGA 1
Medical Air
48
P

40 Low & 60 High

P
Beacon / MEGA 1
Vacuum
-23
P

-12 Low

P
Beacon / MEGA 1
Nitrous Oxide
48
P

40 Low & 60 High

P
Beacon / MEGA 1
Nitrogen
159
P

140 Low & 190 High

P
Beacon / MEGA 1

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor (Next to Room 3B-148) CLINICAL RESEARCH

3B-148, 3B-149, 3B150

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
51
Medical Air
P
Y
½”
48
Vacuum
P
Y
¾”
-23

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
3rd Floor CLINICAL RESEARCH

3B-148, 3B-149, 3B150

Room
OXY
AIR
VAC
Flows
Style
Comments
3B-148
1
1
1
P
Series B / Pin Index
3B-149
1
1
1
P
Series B / Pin Index
3B-150
1
0
2
P
Series B / Pin Index

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor PT / WOUND CARE

2B-119, 2B-126

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
52
99+
Vacuum
P
Y
1 ¼”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor PT / WOUND CARE
Room
OXY
VAC
Flows
Style
Comments

2B-126

PT

1
1
P
Dimond 3 / Pin Index

2B-119

WOUND CARE

1
1
P
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
2nd Floor (TX ROOM (LEFT OF FRONT DOOR) PT / WOUND CARE
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / MEGA 1
Vacuum
-23
P

-12 Low

P
Beacon / MEGA 1
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2A-103) EAST INPATIENT REHAB

2A-103, 2A-104, 2A-106, 2A-108, 2A-110, 2A-113, 2A-114, 2A-115, 2A-116, 2A-118, 2A-122, 2A-123, 2A-124, 2A-127

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1 ½”
Medical Air
P
Y
1 ½”
Vacuum
P
Y
1 ½”
Comments:
NOT IN SERVICE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor EAST INPATIENT REHAB

2A-103, 2A-104, 2A-106, 2A-108, 2A-110, 2A-113, 2A-114, 2A-115, 2A-116, 2A-118, 2A-122, 2A-123, 2A-124, 2A-127

Room
OXY
AIR
VAC
Flows
Style
Comments
2A-103
2
2
2
NT
Dimond 3 / Pin Index
NOT IN SERVICE
2A-104
2
2
2
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-106
2
2
2
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-108
2
2
2
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-110
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-113
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-114
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-115
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-116
2
2
2
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-118
2
2
2
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-122
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-123
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-124
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
2A-127
1
1
1
NT
Diamond 3 / Pin Index
NOT IN SERVICE
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
2nd Floor (CHECK IN DESK Nurse Station 2A-103) EAST INPATIENT REHAB

2A-103, 2A-104, 2A-106, 2A-108, 2A-110, 2A-113, 2A-114, 2A-115, 2A-116, 2A-118, 2A-122, 2A-123, 2A-124, 2A-127

Gas
Pressure
Audible/Visual
Labeling
Make/Model

Oxygen

P
Allied / IMPACT

Medical Air

P
Allied / IMPACT

Vacuum

P
Allied / IMPACT

NOT IN SERVICE

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2A-139) ORTHO

2A-136 # A-E, 2A-142, 2A-143, 2A-144, 2A-153, 2A-154, 2A-155, 2A-156, 2A-157, 2E-118

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
¾”
48
99+
Medical Air
P
Y
¾”
48
21.3
Vacuum
P
Y
1”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2A-139) ORTHO
Room
OXY
AIR
VAC
Flows
Style
Comments
2A-136A
1
1
1
NT
Dimond 3 / Pin Index
UNDER CONTRUCTION
2A-136B
1
1
1
NT
Diamond 3 / Pin Index
UNDER CONTRUCTION
2A-136C
1
1
1
NT
Diamond 3 / Pin Index
UNDER CONTRUCTION
2A-136D
1
1
1
NT
Diamond 3 / Pin Index
UNDER CONTRUCTION
2A-136E
1
1
1
NT
Diamond 3 / Pin Index
UNDER CONTRUCTION
2A-142
1
1
1
P
Diamond 3 / Pin Index
2A-143
1
1
1
P
Diamond 3 / Pin Index
2A-144
1
1
1
P
Diamond 3 / Pin Index
2A-153
1
1
1
P
Diamond 3 / Pin Index
2A-154
1
1
1
P
Diamond 3 / Pin Index
2A-155
1
1
1
P
Diamond 3 / Pin Index
2A-156
1
1
1
P
Diamond 3 / Pin Index
2A-157
1
1
1
P
Diamond 3 / Pin Index
2E-118
1
1
1
P
Diamond 3 / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
2nd Floor (Nurse Station Across From 2A-159)
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
52
40/60
P
TRI TECH LCD
Medical Air
48
40/60
P
TRI TECH LCD
Vacuum
-22
-12
P
TRI TECH LCD

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (In Radiology Reception) RADIOLOGY

2F-109, (MRI 2F-111 & 2F-112), 2F-124A, 2F-124B, 2F-124D, 2F-126,

2F-127, 2F-130, 2F-135 CT, 2F-138, 2F-144, 2F-148, 2F-151, 2F-153B

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
1”
48
99+
Vacuum
P
Y
1 1/4”
-23
NA

Comments:

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (In Radiology Reception Area) RADIOLOGY
Room
OXY
AIR
VAC
Flows
Style
Comments
2F-109
1
0
1
NT
Ohio / DISS
2F-124A
1
0
1
P
Ohio / Pin Index
2F-124B
1
0
1
P
Ohio / Pin Index
2F-124D
1
0
1
P
Ohio / Pin Index
2F-126
2
0
2
P
Ohio / Pin Index
2F-127
1
1
1
P
Ohio / Pin Index
Medical Air Not In Use
2F-130
1
1
1
P
Ohio / Pin Index
Medical Air Not In Use
2F-135 CT
1
0
1
P
Ohio / Pin Index
2F-138
1
0
1
P
Ohio / Pin Index
2F-144
1
0
1
P
Ohio / Pin Index
2F-147
1
0
1
P
Ohio / Pin Index
2F-151
1
0
1
NT
Ohio / Pin Index
2F-153B
1
0
1
P
Ohio / Pin Index
2F-111 MRI
1
0
1
P
TRI TECH / Pin Index
2F-110 MRI
2
0
2
P
Ohio / Pin Index
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
2nd Floor (In Radiology Reception Area) RADIOLOGY
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / PB MEGA 1
Vacuum
-23
P

-12 Low

P
Beacon / PB MEGA 1
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2F-134) CT

2F-134 CT

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
48
99+
Medical Air
P
Y
½”
0
NA
Vacuum
P
Y
¾”
-23
NA
Comments:
Medical Air Is Not Live
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2F-134) CT
Room
OXY
AIR
VAC
Flows
Style
Comments
2F-134
2
2
2
P
Series B / Pin Index
Medical Air Is Not Live
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:
2nd Floor (Inside Control Room 2F-134) CT
Gas
Pressure
Audible/Visual
Labeling
Make/Model
Oxygen
53
P

40 Low & 60 High

P
Beacon / MEGA 2
Medical Air
NOT IN USE
NOT IN USE
NOT IN USE
Beacon / MEGA 2
Vacuum
-23
P

-12 Low

P
Beacon / MEGA 2

Zone Valve Box Data Sheet

Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2F-138) CT

2F-138 CT

Gas
Labeling
Gauges
Valve Size
Pressure
Percentage
Oxygen
P
Y
½”
52
99+
Medical Air
P
Y
½”
42
21
Vacuum
P
Y
¾”
-23
NA
Comments:
VERY LOW READING ON MED AIR ALARM / REPLACE OR REVIEW
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Zone Valves Floor & Location:
2nd Floor (Next to Room 2F-138) CT
Room
OXY
AIR
VAC
Flows
Style
Comments
2F-138
1
1
2
P
CHEMETRON CONNECT 2
WALL
2F-138
1
1
2
P
CHEMETRON DISS CEILING
NO HOSE DROPS
Facility:
VA Oklahoma City Health Care System
Location:
921 NE 13th Street - Oklahoma City, OK 73104
Date:
SEPT. 22-24, 2025
Alarm Panel Floor & Location:

This is the start of the file's text. The full file is on GovTribe.