P01_Attachment B Steris Sterilizer Preventive Maintenance Schedule.pdf

PDF 81 KB Posted

Attached to
J065--Actionable - 554 (S) | New Requirement | Biomed Steris Sterilizer Maintenance (VA-26-00024834)
Solicitation number
36C25926Q0115
Issued by

About this file

Attachment B is a Preventive Maintenance Schedule that details equipment maintenance requirements for various medical and sterilization devices. The document outlines specific maintenance frequencies and Original Equipment Manufacturer (OEM) requirements for nine pieces of equipment, including washers, sterilizers, conveyors, and ultrasonics equipment.

Maintenance frequencies range from daily to semi-annual, with each equipment type having unique inspection and cleaning protocols. For example, the Vision 1327 Washer requires daily debris screen cleaning, weekly arm inspection, monthly gasket inspection, and quarterly full OEM preventive maintenance. Similarly, the 5052 Washer/Disinfector requires quarterly maintenance, with inspections 1 and 3 including required parts replacement. Other equipment like the AMSCO HC-1500 Steam Sterilizer and V-PRO Max Low-Temp Sterilizer have daily, weekly, and annual maintenance tasks to ensure optimal performance and equipment longevity.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment B – Preventive Maintenance Schedule (OEM-Aligned) Equipment PM Frequency OEM Requirements

Vision 1327 Washer Daily / Weekly / Monthly / Quarterly

Daily debris screen cleaning; Weekly arm inspection; Monthly gasket inspection; Quarterly full

OEM PM.

5052 Washer/Disinfector Quarterly (4×/year) OEM Quarterly PM;

Inspections 1 & 3 include required parts replacement.

AMSCO HC‑1500 Steam Sterilizer

Daily / Weekly / Annual Daily drain strainer cleaning; Weekly leak test;

Annual chamber cleaning and OEM PM.

SCS‑M‑F1 Conveyor Daily / Quarterly / Annual Daily roller & guide cleaning; Quarterly drip pan cleaning; Annual lift-gate verification.

400 Series Model 16 Sterilizer

Daily / Weekly / Annual Same OEM routine as HC‑1500: daily drain cleaning, weekly leak test, annual OEM PM.

Caviwave Ultrasonics Semi‑Annual (2×/year) Semi‑annual OEM PM including transducer testing, calibration, and parts replacement.

V‑PRO Max Low‑Temp Sterilizer

Daily / Weekly / Monthly Daily printer check; Weekly chamber & touchscreen cleaning; Monthly leak test.

InnoWave Pro Sonic Irrigator

Daily / Weekly / Semi‑Annual

Daily inspections; Weekly deep cleaning; Semi‑annual OEM service.

Attachment B – Preventive Maintenance Schedule (OEM-Aligned)

File details come from the government source that posted it. Updated .