About this file

This document is a periodic inspection checklist for a Guldmann lifter, conforming to DS/EN ISO 10535:2006 standard for a 12-month inspection. The checklist is divided into three main sections: electrical connections, lifter components, and lifting hanger.

The electrical connections section requires checking cable integrity, securing, isolation, and charger functionality. The lifter section involves comprehensive checks including emergency stop and lowering functions, full lifting and leg spreading cycles with rated load, welding inspection, wheel condition and brakes, and fastener assessment. The lifting hanger section focuses on attachment safety mechanisms, sling attachment rubber finger condition, and potential damage to plastic covers. The form provides spaces to record inspection location, serial number, date, and inspector, with a "SAMPLE" watermark indicating this is a template document for standard inspection procedures.

View the file

Other files for this federal contract opportunity

Other files attached to J065--NTX Mobile Lift PM Services_Base plus 4 Option Years, newest first.
File Type Posted
SF30 36C25726Q0041 0001.docx DOCX document
36C25726Q0041 0001_1.docx DOCX document
Service Contract Act WD_2015-5257_Smith.pdf PDF
Service Contract Act WD_2015-5273_Fannin.pdf PDF
Attachment B_PM Checklist.pdf PDF
S02 36C25726Q0041.docx DOCX document
Service Contract Act WD_2015-5227_Dallas.pdf PDF
Service Contract Act WD_2015-5255_Grayson.pdf PDF
Attachment A_Patient Safety Alert AL 14-04.pdf PDF
Service Contract Act WD_2015-5227_Hunt.pdf PDF
36C25726Q0041_3.docx DOCX document
Service Contract Act WD_2015-5231_Tarrant.pdf PDF
Service Contract Act WD_2015-5839_Hood.pdf PDF
Service Contract Act WD_2015-5227_Denton.pdf PDF
Service Contract Act WD_2015-5227_Collin.pdf PDF
Show all 15

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

G ul dm an n

G B

/U S

-0 1/

CO

NT

RO

L

Checklist

G ul dm an n G

B -2

5/

/1

Periodic Inspection according to DS/EN ISO 10535:2006 12-month inspection of lifter

Checklist

1. Electrical connections Check that all cables are intact Check that all cables are secured with a cable relief Check for rupture on cables and on isolation Check that the charger I fully functional

2. Lifter Check connection between hook and hanger Check the emergency stop Check the emergency lowering Do a full lifting cycle with rated load – must be effortless and without jarring sound Do a full leg spreading cycle with rated load – must be effortless and without jarring sound Check that all welding are free of cracks Check that the wheels are free of filth and hair – must rotate effortless Check the wheel brakes – must lock the wheel Visual control of all fasteners – loose fasteners with thread locking adhesive must not be tightened, but replaced with new thread locking adhesive applied

Place a sticker with a new date for next inspection

3. Lifting Hanger Check hanger attachment safety mechanism for correct function Check sling attachment rubber finger for correct function and wear Check for damage to plastic covers that could compro-mise use of the hanger

Comments:

Place of Inspection:

Serial number:

Inspection date:

Inspection by:

SAMPLE

SAM

PLE

Text1:
Text3:
Text4:
Text5:
Text6:
Text2:
Text7:
Text8:
Text9:
Text10:
CheckBox1: Off
CheckBox2: Off
CheckBox3: Off
CheckBox4: Off
CheckBox5: Off
CheckBox6: Off
CheckBox7: Off
CheckBox8: Off
CheckBox9: Off
CheckBox10: Off
CheckBox11: Off
CheckBox12: Off
CheckBox13: Off
CheckBox14: Off
CheckBox15: Off
CheckBox16: Off
CheckBox17: Off

File details come from the government source that posted it. Updated .