671A4-22-172 Med Air Final Design Drawings - Revs No.1 - 01-10-2025.pdf
PDF 3 MB Posted
- Solicitation number
- 36C25725B0001
About this file
This is a set of construction drawings for a project to remove and replace the medical air vacuum system at the Kerrville VA Medical Center in Kerrville, TX (Project #671A4-22-172). The drawings detail the installation of new medical air compressors and vacuum pumps in the basement mechanical room of Building 11, including demolition of existing equipment and installation of new equipment with associated electrical, plumbing and medical gas connections.
The drawings include architectural, plumbing, and electrical plans showing equipment locations, piping routes, and electrical connections. Key specifications include a new duplex oil-less scroll medical air compressor system (24 CFM at 120 PSI) with 200 gallon receiver tank and a new duplex dry claw medical vacuum pump system (65 SCFM at 19" Hg) with 200 gallon receiver. The electrical work involves connecting the new equipment to Panel PBYEA with 30A/3P circuit breakers. The project requires infection control measures including sticky mats, zipper entries, and ICRA barriers during construction. The facility is fully sprinklered and the drawings are marked as 100% bid documents dated January 11, 2022 with revisions dated January 10, 2025.
View the file
Other files
Show all 23
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DocuSign Envelope ID: D2FF90DF-8975-4E1C-9A25-4C043CFA3793
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
KERRVILLE VA MEDICAL CENTER
3600 MEMORIAL BLVD.
KERRVILLE, TX 78028-5758
VA PROJECT #: 671A4-22-172
PROJECT SITE MAP
PROJECT IMAGE
PROJECT AREA PROJECT LOCATION MAP DRAWING LIST
GENERAL
G-001 COVER SHEET
G-100 GENERAL INFORMATION
GL100 ENLARGED BASEMENT LIFE SAFETY PLAN
PLUMBING
PG001 PLUMBING GENERAL INFORMATION
PG400 MECHANICAL ROOM MEDICAL GAS PLANS
P-601 MEDICAL GAS SCHEDULES AND EXISTING EQUIPMENT PHOTOGRAPHS
ELECTRICAL
E-001 ELECTRICAL GENERAL NOTES, SYMBOL LEGENDS, AND ABBREVIATIONS
E-401 ELECTRICAL ENLARGEMENTS - DEMOLITION AND NEW WORK
E-601 ELECTRICAL SCHEDULES
FULLY SPRINKLERED
100% BID DOCUMENTS
CONSULTANTS: PROJECT MANAGER
Headquarters:
ACG Project Number 21-040 Office of
Construction
Drawing Title
COVER SHEET
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
VA PROJECT NUMBER
671A4-22-172
Building Number
Apogee Consulting Group, P.A. and Facilities # 11
Engineers Architects
1151 Kildaire Farm Road, Suite 120 Cary, North Carolina 27511 www.acg-pa.com 919-858-7420
Management
U.S. Department
Location
KERRVILLE VA MEDICAL CENTER
3600 MEMORIAL BLVD.
Approved
Date
Checked
Drawn
Drawing Number
G-001 # Revisions: Date
1/11/2022 © COPYRIGHT Apogee Consulting Group, P.A. of Veterans Affairs KERRVILLE, TX 78028-5758 JANUARY 11, 2022 TM AT Dwg
VA FORM 08-6231
KERRVILLE VA
/1
/2
:2
:0
A
M
C U s e rs
\n s to ltz \D o c u m e n ts
\A -0
-V
A K e rr v ill e -M e d A ir V a c S y s te m -T
X n s to ltz .r v t
1 3
2 3
3 3
1 3
1 3 http://www.acg-pa.com/
VA FORM 08-6231 1
/1 /2
:2
:2
U s e rs
\n s to ltz \D o c u m e n ts
\A -0
-V
A K e rr v ill e -M e d A ir V a c S y s te m -T
X n s to ltz .r v t
PROJECT MANAGER
ACG Project Number
21-040
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
Engineers Architects www.acg-pa.com 919-858-7420 © COPYRIGHT Apogee Consulting Group, P.A.
Approved Drawing Number
G-100 Dwg 2 of 3
Date Checked
JANUARY 11, 2022 TM
AT
CODE STANDARDS AND EXECUTIVE ORDERS
PURPOSE: THE PUBLIC BUILDINGS AMENDMENT ACT OF 1988, PUBLIC LAW (PUB. L.) 100-678 REQUIRES FEDERAL AGENCIES TO FOLLOW NATIONAL RECOGNIZED "MODEL" BUILDING CODES. THE FEDERAL PARTICIPATION IN THE DEVELOPMENT AND USE OF VOLUNTARY STANDARDS, OFFICE OF MANAGEMENT AND BUDGET (OMB) CIRCULAR A-119, REQUIRES ALL EXECUTIVE AGENCIES TO RELY ON VOLUNTARY STANDARDS, BOTH DOMESTIC AND INTERNATIONAL, WHENEVER FEASIBLE, AND TO
PARTICIPATE IN VOLUNTARY STANDARD BODIES.
GENERAL: VA HAS ADOPTED THE LATEST EDITION OF THE FOLLOWING CODES AND STANDARDS AS A MINIMUM FOR ALL PROJECTS PERFORMED IN THE MODERNIZATION, ALTERATION, ADDITION, OR IMPROVEMENT OF ITS REAL PROPERTY AND THE CONSTRUCTION OF NEW STRUCTURES. VA DESIGN MANUALS AND MASTER SPECIFICATIONS SPECIFY OTHER CODES AND
STANDARDS THAT VA FOLLOWS ON ITS PROJECTS:
• VA DIRECTIVES, DESIGN MANUALS, MASTER SPECIFICATIONS, VA NATIONAL CAD STANDARD APPLICATION GUIDE, AND OTHER GUIDANCE ON THE TECHNICAL INFORMATION LIBRARY (TIL) (HTTP://WWW.CFM.VA.GOV/TIL/).
• INTERNATIONAL BUILDING CODE (IBC) 2021 (ONLY WHEN SPECIFICALLY REFERENCED IN VA DESIGN DOCUMENTS, SEE NOTES
BELOW)
• NFPA 101 LIFE SAFETY CODE 2021 (SEE NOTES BELOW)
• NFPA NATIONAL FIRE CODES WITH THE EXCEPTION OF NFPA 5000 AND NFPA 900
• OCCUPATIONAL, SAFETY AND HEALTH ADMINISTRATION (OSHA) STANDARDS.
• VA SEISMIC DESIGN REQUIREMENTS, H-18-8
• NFPA 70 NATIONAL ELECTRICAL CODE 2020
• INTERNATIONAL PLUMBING CODE (IPC) 2018
• ASME CODE FOR PRESSURE PIPING
• ARCHITECTURAL BARRIERS ACT ACCESSIBILITY STANDARDS (ABAAS) INCLUDING VA SUPPLEMENT, BARRIER FREE DESIGN
STANDARD (PG-18-13)
• THE PROVISIONS FOR CONSTRUCTION AND SAFETY SIGNS. STATED IN THE GENERAL REQUIREMENTS SECTION 01010 OF THE
VA MASTER CONSTRUCTION SPECIFICATION.
LOCAL CODES: AS AN AGENCY OF THE FEDERAL GOVERNMENT, VA IS NOT SUBJECT TO LOCAL IMPOSITION OF CODE ENFORCEMENT PROCEDURES (DRAWING REVIEWS, BUILDING PERMITS, INSPECTIONS, FEES, ETC.). VA MUST FUNCTION AS THE AUTHORITY HAVING JURISDICTION (AHJ) AND THUS HAS THE RESPONSIBILITY TO GUARD PUBLIC HEALTH AND SAFETY THROUGH ENFORCING ITS ADOPTED CODES. HOWEVER, LOCAL AUTHORITIES SHOULD BE NOTIFIED ABOUT PLANNED PROJECTS AND GIVEN OPPORTUNITY TO REVIEW DRAWINGS PROVIDED THAT VA DOES NOT PAY FOR REVIEW OR INSPECTION
FEES.
NOTES:
1. NFPA 101 PRIMARILY ADDRESSES LIFE SAFETY AND FIRE PROTECTION FEATURES WHILE THE IBC ADDRESSES A WIDE RANGE OF CONSIDERATIONS, INCLUDING, BUT NOT LIMITED TO, STRUCTURAL STRENGTH, SEISMIC STABILITY, SANITATION, ADEQUATE LIGHT AND VENTILATION, AND ENERGY CONSERVATION. VA BUILDINGS MUST MEET THE REQUIREMENTS OF NFPA 101 AND DOCUMENTS REFERENCED BY NFPA 101 IN ORDER TO COMPLY WITH THE ACCREDITATION REQUIREMENTS OF THE JOINT COMMISSION. THEREFORE, DESIGNS SHALL COMPLY WITH THE REQUIREMENTS OF THE LATEST EDITION OF NFPA 101 AND DOCUMENTS REFERENCED THEREIN. DESIGN FEATURES NOT ADDRESSED BY NFPA 101 OR DOCUMENTS REFERENCED THEREIN SHALL COMPLY WITH THE REQUIREMENTS OF THE LATEST EDITION OF THE IBC OR AS OTHERWISE ADDRESSED ABOVE IN THIS PROGRAM GUIDE. FOR DESIGN FEATURES THAT ARE ADDRESSED BY BOTH THE IBC AS WELL AS NFPA 101 OR A DOCUMENT REFERENCED BY NFPA 101, THE REQUIREMENTS OF NFPA 101 OR THE DOCUMENT REFERENCED BY NFPA 101 SHALL BE USED EXCLUSIVELY (THIS APPLIES EVEN IF THE IBC REQUIREMENTS ARE DIFFERENT).
2. CONFLICTS BETWEEN NATIONALLY RECOGNIZED CODES AND STANDARDS AND VA REQUIREMENTS – SHOULD A CONFLICT EXIST BETWEEN VA REQUIREMENTS AND VA ADOPTED NATIONALLY RECOGNIZED CODES AND STANDARDS, THE CONFLICT SHALL BE BROUGHT TO THE ATTENTION OF VA. THE RESOLUTION OF THE CONFLICT SHALL BE MADE BY THE AUTHORITY
HAVING JURISDICTION FOR VA TO ENSURE A CONSISTENCY SYSTEM WIDE.
Drawing Title Project Title VA PROJECT NUMBER
CONSULTANTS: Office of GENERAL INFORMATION REMOVE/REPLACE MEDICAL 671A4-22-172
Construction AIR VACUUM SYSTEM Building Number and Facilities # 11 Management Location
KERRVILLE VA MEDICAL CENTER
U.S. Department 3600 MEMORIAL BLVD.
1/11/2022 of Veterans Affairs KERRVILLE, TX 78028-5758
# Revisions: Date http://www.cfm.va.gov/TIL/)
A. SEE SHEETS A001 FOR SYMBOLS AND
ABBREVIATIONS.
LIFE SAFETY LEGEND
SHEET NOTES:
EGRESS PATH
8 9 10 11 12 13 14 15
XXX
XXX
EXIT
IDENTIFICATION
EGRESS CLEAR
WIDTH
EGRESS PATH TAG
EGRESS PATH TAG - COMMON PATH
EGRESS IDENTIFICATION
EGRESS CAPACITY
AREA OF REFUGE
B. REFER TO MECHANICAL DRAWINGS FOR
INFORMATION
C. REFER TO ELECTRICAL DRAWINGS FOR
INFORMATION CONCERNING POWER, LIGHTING
AND COMMUNICATION SYSTEM(S)
D. CONTRACTOR TO PROVIDE ALL ADDITIONAL
FRAMING NECESSARY FOR ALL OPENINGS AND
SUPPLEMENTAL FRAMING ABOVE PARTITIONS.
E. ALL FIRE RATED ASSEMBLIES SHALL BE
INSTALLED IN STRICT ACCORDANCE WITH TESTED
ASSEMBLIES INDICATED.
F. EXTEND FIRE RATED PARTITIONS, BARRIERS, AND
OTHER SEPARATIONS TO BOTTOM OF
UP
FEB
FEC
FBC
CA
KB
DE
HAZARDOUS LOCATION
FIRE EXTINGUISHER BRACKET
FIRE EXTINGUISHER CABINET
FIRE BLANKET CABINET
EMERGENCY RESPONSE REFERENCE
CONTROLLED ACCESS - ALWAYS
UNLOCKED IN DIRECTION OF EGRESS
KNOX BOX
DELAYED EGRESS W/ ALARM
ROOF/FLOOR DECK ABOVE (OR AS DIRECTED BY
UL ASSEMBLY) AND TO EXTERIOR WALL. SEAL
JOINT BETWEEN EDGES OF PARTITION WITH FIRE
RATED SEALANT AND/OR INTUMESCENT
ASSEMBLY.
G. ALL PENETRATIONS OF FIRE-RATED ASSEMBLIES
SHALL BE FIRE-SEALED IN ACCORDANCE WITH
APPROVED MANUFACTURER'S DETAIL FOR
LOCATION, TYPE OF CONSTRUCTION,
PENETRATING ITEM AND RATING REQUIRED.
H. SMOKE RESISTANT CONSTRUCTION SHALL MEET
THE REQUIREMENTS OF ALL APPLICABLE CODES.
AREA SHALL BE SEPARATED FROM THE
REMAINDER OF THE BUILDING BE CONSTRUCTION
CAPABLE OF RESISTING THE PASSAGE OF
SMOKE. THE PARTITIONS SHALL EXTEND FROM
THE FLOOR TO THE UNDERSIDE OF THE FLOOR
OR ROOF ASSEMBLY ABOVE. DOORS SHALL BE
SELF OR AUTOMATIC CLOSING. DOORS SHALL
EGRESS LEGEND NOT HAVE AIR TRANSFER OPENINGS AND SHALL
NOT BE UNDERCUT IN EXCESS OF CLEARANCE
PERMITTED WITH ACCORDANCE TO NFPA 80.
MECHANICAL
ROOM
028A
PIPE
BASEMENT
TOTAL SQUARE
FOOTAGE:
OCCUPANCY:
EGRESS
REQUIRED:
TRAVEL
DISTANCE
MECHANICAL ROOM 028 3,409 GSF
HEALTHCARE - OUTPATIENT
- 240 SF / PERSON
EXISTING MECHANICAL SPACE. NO
SMOKE MODIFICATION OF THE
SPACE OR EGRESS BEING
PERFORMED
150'-0" MAXIMUM DISTANCE.
I. STORAGE OF ANY FLAMMABLE OR COMBUSTIBLE
LIQUIDS IS UNDERSTOOD TO NOT BE WITHIN THE
BUILDING. STORAGE OF ANY MATERIAL IS TO BE
IN COMPLIANCE WITH ALL APPLICABLE CODES
AND AUTHORITIES HAVING JURISDICTION.
J. FACILITY SAFETY PERSON TO VERIFY IF EXISTING
EGRESS FROM MECHANICAL SPACES MEET THE
REQUIREMENTS OF NFPA 101.
CODE INFORMATION
PROJECT NAME: REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
PROJECT ADDRESS: KERRVILLE VA MEDICAL CENTER
3600 MEMORIAL BLVD.
KERRVILLE, TX 78028-5758
OWNER OR AGENT: KERRVILLE VA MEDICAL CENTER
PROPOSED USE: Relevant Code Goes Here
INCREASE
(per NFPA 101-19.2.6.2.1)
COMMON PATH
OF TRAVEL
(NFPA 101 -
7.13.1(2) )
REQUIRED # OF
A EXITS
(per NFPA 101, 19.2.4)
DEAD END
CORRIDOR
(NFPA 101-19.2.5.2)
INCREASE TRAVEL DISTANCE 50FT IF
BUILDING IS SPRINKLED
IN EXISTING BUILDING, COMMON
PATH OF TRAVEL NOT TO EXCEED
150'
- 2 EXITS MIN. PER FLOOR REQ'D
- 2 EXISTING EXITS, NOT BEING
MODIFIED
EXISTING DEADEND CORRIDOR
EXCEEDING 30' SHALL BE
PERMITTED TO CONTINUE IN USE IF
IT IS UNPRACTICAL AND
UNFEASIBLE TO ALTER FROM.
OWNED BY: FEDERAL
CODE ENFORCEMENT JURISDICTION: FEDERAL
APPLICABLE CODES:
20XX International Building Code 20XX Electrical Code 20XX ADA Code 20XX Plumbing Code Relevant Code Goes Here
GENERAL CODE INFORMATION:
CONSTRUCTION TYPE:Relevant Code Goes Here
PRIMARY OCCUPANCY: Relevant Code Goes Here
SECONDARY OCCUPANCY: N/A
OCCUPANCY CLASSIFICATION
(NFPA 2018 101 SECTION 4.1):
HEALTHCARE OCCUPANCY BUILDING, SPRINKLERED
CONSTRUCTION TYPE I (442)
B
MECH.
MEANS OF EGRESS:
STAIR ENCLOSURE: 2-HR (NFPA 101.7.1.3.2.1)
MIN. EXIT WIDTH: 32" CLEAR (NFPA 101.2.1.2.2)
OCCUPANT LOAD: (NFPA 101.7.3)
HEALTHCARE INPATIENT: 1:240
STORAGE: 1:300
CAPACITY FACTORS: (NFPA 101.7.3.3):
STAIRS: 0.3
OTHER: 0.2
SERVICE
ELEVATOR
LOBBY
ELECT.
CLOSET
SERVICE
ELEVATOR
NO. 3
KEY PLAN
SCALE: NO SCALE
D
EMS
EQUIPMENT
010A
STORAGE
010B
IDF 010D
010D
CRAWL
SPACE
E.C.
MAIN
LOBBY
PASSEN GER
NO. 2
PASSENGER
NO. 1
MUSEUM
028A
WEST
NORTH
AREA OF WORK
EAST
E
CORRIDOR BASEMENT
WALL RATING LEGEND
1 HR - FIRE SEPARATION
2 HR - FIRE SEPARATION
ENLARGED BASEMENT LIFE SAFETY PLAN
0 4 8 16
S S S S
SMOKE RESISTIVE HAZARDOUS AREA
Scale: 1/8" = 1'-0" FULLY SPRINKLERED
100% BID DOCUMENTS
CONSULTANTS: PROJECT MANAGER
Headquarters:
ACG Project Number 21-040 Office of
Construction
Drawing Title
ENLARGED BASEMENT LIFE SAFETY
PLAN
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
VA PROJECT NUMBER
671A4-22-172
Apogee Consulting Group, P.A. and Facilities # 11
Engineers Architects
1151 Kildaire Farm Road, Suite 120 Cary, North Carolina 27511
Management
U.S. Department
Location
KERRVILLE VA MEDICAL CENTER
Date Checked Drawn
GL100
# Revisions:
1/11/2022 www.acg-pa.com 919-858-7420 of Veterans Affairs KERRVILLE, TX 78028-5758 JANUARY 11, 2022 TM AT Dwg
/1
/2
:2
:2
U s e rs
\n s to ltz \D o c u m e n ts
\A -0
-V
A K e rr v ill e -M e d A ir V a c S y s te m -T
X n s to ltz .r v t
U P
T R
A V
E L D
IS
T
A N
C E
EXTERIOR
EXIT
CLR CAPACITY
XX" ###
WORK SHALL CONFORM TO APPLICABLE LOCAL, STATE, AND NATIONAL CODES AND APPLICABLE VA STANDARDS.
INSTALL EQUIPMENT IN ACCORDANCE WITH MANUFACTURER'S WRITTEN RECOMMENDATIONS.
OBTAIN AND PAY FOR PERMITS, FEES, AND INSPECTIONS REQUIRED FOR THIS WORK.
PROVIDE NEW MATERIALS, EQUIPMENT AND PRODUCTS INCORPORATED IN THE WORK, OF A SUITABLE GRADE FOR THE
PURPOSES INTENDED.
EQUIPMENT OR MATERIAL DEVIATIONS FROM INDICATED CONTRACT REQUIREMENTS ARE SUBJECT TO THE APPROVAL OF THE COR. SUBMIT PROPOSED EQUIPMENT DEVIATIONS THAT ARE SIMILAR BOTH IN QUALITY AND CAPACITY TO THAT
EQUIPMENT SPECIFIED, AS SOLELY DETERMINED BY THE COR.
INSTALL EQUIPMENT AS INDICATED, ALLOWING FOR SUFFICIENT ACCESS AND CLEARANCE SPACE FOR EQUIPMENT
MAINTENANCE, REPAIRS, AND REPLACEMENT.
SUPPORT PIPING SYSTEM AS REQUIRED BY APPLICABLE LOCAL, STATE, AND NATIONAL CODES ALONG WITH
SPECIFICATIONS, VA STANDARDS AND MANUFACTURER'S RECOMMENDATIONS.
FIRESTOP AND SEAL PIPING PENETRATIONS THROUGH WALLS AND FLOORS EQUAL TO THE RATING OF WALLS AND FLOORS.
TEST PLUMBING SYSTEM IN ACCORDANCE WITH APPLICABLE LOCAL, STATE AND NATIONAL CODES AND APPLICABLE VA
STANDARDS.
COORDINATE PLUMBING PIPING WITH WORK OF OTHER DISCIPLINES.
PROVIDE SHUT OFF VALVES AT BRANCH PIPING TO FIXTURES.
COORDINATE PLUMBING WORK WITH ALL OTHER TRADES PRIOR TO INSTALLATION. COORDINATE ROUTING OF PIPING WITH
EXISTING CONDITIONS.
PROVIDE ALL NECESSARY OFFSETS, TEES, ELBOWS, ETC. FOR A COMPLETE AND COORDINATED WORKING PLUMBING
SYSTEM. DRAWING PLANS, SCHEMATICS, AND DIAGRAMS INDICATE GENERAL LOCATION AND ARRANGEMENT OF PIPING
SYSTEMS.
CONTRACTOR SHALL COORDINATE ANY PLUMBING SYSTEM REQUIRED SHUTDOWN WITH THE COR 1 WEEK IN ADVANCE.
PATCH, PAINT AND CLEAN SURFACES SUSTAINING DAMAGE ASSOCIATED WITH PLUMBING WORK OF THIS CONTRACT, UNLESS NOTED OTHERWISE.
DO NOT INSTALL PIPES ABOVE TELECOM OR ELECTRICAL EQUIPMENT.
PROVIDE RECERTIFICATION OF NEW INSTALLED MEDICAL GAS EQUIPMENT. TEST SYSTEMS FOR OPERATION, ALARMS AND
CAPACITY AND PROVIDE FINAL COMMISSIONING REPORT.
PIPE SLOPE ARROW
3-WAY CONTROL VALVE
PRESSURE REGULATOR (GAS)
HOSE BIBB / SILLCOCK
PRESSURE GAUGE
F FLOW SWITCH
PUMP
WYE STRAINER W/DRAIN VALVE
M WATER METER
FLOW ARROW
CIRCUIT SETTER
BACKFLOW PREVENTER
THERMOMETER
WYE STRAINER
OR FLOOR DRAIN - ROUND OR SQUARE
PIPE ANCHOR
PIPE GUIDES
HANGER ROD
INSULATION (VAPOR
BARRIER TYPE IS
REQUIRED FOR
LOW TEMPERATURE
PIPE)
PROVIDE HIGH
COMPRESSIVE
STRENGTH
INSULATION (9 PSF
MIN. DENSITY) UNDER
INSULATION SHIELD 1
INSULATION SHIELD AT
HANGER
WELD
SADDLE
ADJUSTABLE CLEVIS HANGER TYPE 1 - SEE
SPECIFICATIONS
ADJUSTABLE CLEVIS HANGER TYPE 43 - SEE
SPECIFICATIONS
PROVIDE INSULATION SHIELD &
INSERT FOR ALL PIPING (8" MIN.)
1/2" DIA. HANGER RODS WITH 36"
MAX. SPACING ON EACH CHANNEL
NOTES:
SEE SPECIFICATIONS FOR
DETAILED HANGER
1" MAX.
BAND
1-5/8" 12 GAUGE CHANNEL
OR 2" x 2" x 1/4" ANGLE
REQUIREMENTS. SIDE VIEW TRAPEZE HANGER
FOR UP TO 1000 LB. UNIFORM
LOAD
MAXIMUM PIPE/TUBING SUPPORT SPACING
NOM. SIZE (IN.) THRU 3/4 1 1 1/4 1 1/2 2 2 1/2 3 4 5 6 8 10 12
PIPE (FT.) 7 7 7 9 10 11 12 14 16 17 19 22 23
TUBING (FT.) 5 FT 6 7 8 8 9 10 12 13 14 16 - -
NOTE: FOR TRAPEZE HANGER TAKE SPACING OF SMALLEST SIZE ON TRAPEZE
PIPE HANGER DETAIL
Scale: No Scale
1 Revision #1
01/10/2025
PG001
Date Dwg 1 of 3
ANGLE GATE VALVE
SAFETY RELIEF VALVE
TWO POSITION CONTROL VALVE
AV
AUTOMATIC AIR VENT
P PRESSURE SWITCH
T TEMPERATURE SWITCH
PIPE UNION
FLEXIBLE PIPE CONNECTOR
PIPE TEE
PLAN 90° ELBOW
CROSS AND RISER
TEE OFF BOTTOM
TEE OFF TOP
DROP AND TURN
DROP AND RUN
TEE DOWN
TEE UP
PIPE TURNING UP
PIPE TURNING DOWN
PIPE CAP
WCO WALL CLEANOUT
CO SUSPENDED CLEANOUT
FLOOR CLEANOUT
U.S. Department of Veterans Affairs
PROVIDE STICKY MAT ON ENTIRE FLOOR AREA OF SERVICE ELEVATOR LOBBY BETWEEN SERVICE ELEVATOR #3 AND THE
DOOR TO THE MECHANICAL ROOM WHERE NEW EQUIPMENT IS BEING INSTALLED.
PROVIDE ZIPPER ENTRY AT ACCESS POINT TO CONSTRUCTION AREA, AT DOOR TO THE MECHANICAL ROOM, IF THE WORK
REQUIRED TO INSTALL THE EQUIPMENT WILL GENERATE ANY DUST.
PROVIDE ICRA BARRIERS AS NECESSARY TO PERFORM THE WORK PER THE MECHANICAL DRAWINGS. THE USE OF
TEMPORARY, REUSABLE TYPE BARRIERS (EX. STARC WALL SYSTEMS) IS ACCEPTABLE. ENSURE THAT ACCESS THROUGH
THE BARRIERS IS PROVIDED IF THE TEMPORARY BARRIERS CUTS OFF ACCESS FROM THE MECHANICAL ROOM OR ANY
OTHER SPACE TO THE MEANS OF EGRESS.
VERIFY EXISTING CONDITIONS PRIOR TO BEGINNING WORK. NOTIFY COR REGARDING CONFLICTS BEFORE PROCEEDING WITH
WORK.
REMOVED ITEMS REMAIN PROPERTY OF GOVERNMENT UNTIL COR DETERMINES THAT THEY ARE NOT REQUIRED. REMOVE
FROM SITE ITEMS NOT BEING RETAINED BY GOVERNMENT OR REINSTALLED AS PART OF THIS WORK.
DO NOT DISTURB ITEMS NOT INDICATED FOR DEMOLITION.
VERIFY SIZE AND LOCATION OF EXISTING UTILITIES PRIOR TO START OF DEMOLITION. RELOCATE, REMOVE AND ADJUST
MECHANICAL AND ELECTRICAL ITEMS AS NECESSARY TO COORDINATE WITH NEW WORK.
AFTER DEMOLITION IS COMPLETE, PATCH AND REPAIR FINISHES TO MATCH EXISTING ADJACENT CONDITIONS.
COORDINATE DEMOLITION WITH COR. SUBMIT SHUTDOWN PLAN AND SCHEDULE TO COR FOR REVIEW AND APPROVAL.
PROTECT EXISTING UTILITIES TO REMAIN FROM DAMAGE DURING DEMOLITION. REPAIR UTILITIES AND SERVICES DAMAGED AT
NO ADDITIONAL COST.
TEMPORARILY MOVE OR TAKE EQUIPMENT OUT OF SERVICE AS NECESSARY TO COMPLETE WORK.
COORDINATE WITH NEW PLUMBING WORK AND WORK OF OTHER TRADES.
TEST EXISTING SYSTEMS FOR OPERATION, ALARMS, AND CAPACITY THEN PROVIDE REPORT TO COR PRIOR TO START OF
DEMOLITION.
ISOLATE SUPPLY EQUIPMENT FROM DISTRIBUTION SYSTEMS BY SHUTTING OFF SOURCE VALVES BEFORE BEGINNING
DEMOLITION.
PRESSURE REDUCING VALVE (WATER)
PLUG VALVE
GLOBE VALVE
GATE VALVE
CONTROL VALVE
CHECK VALVE
BUTTERFLY VALVE
BALL VALVE
ANGLE GLOBE VALVE
ECCENTRIC REDUCER
CONCENTRIC REDUCER
CONSULTANTS:
Drawn
NAS
Checked
RBK
Date
JANUARY 11, 2022
Approved Location
KERRVILLE VA MEDICAL CENTER
3600 MEMORIAL BLVD.
KERRVILLE, TX 78028-5758
Building Number
# 11
VA PROJECT NUMBER
671A4-22-172
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
Drawing Title
PLUMBING GENERAL INFORMATION
/1
/2
:1
:2
U s e rs
\s m is e v ic h \D o c u m e n ts
\M F
P -0
-V
A K e rr v ill e -M e d
A ir V a c S y s te m -T
X s m is e v ic h .r v t
PLUMBING SYMBOLS & ABBREVIATIONS
ANNOTATION ABBREVIATIONS
A/E ARCHITECT/ENGINEER F FAHRENHEIT O2 OXYGEN
AC ABOVE CEILING FA FROM ABOVE OC ON CENTER
AD AREA DRAIN FB FROM BELOW OD OUTSIDE DIAMETER
AFF ABOVE FINISHED FLOOR FBO FURNISHED BY OTHERS OFD OVERFLOW DRAIN
AG AIR GAP FCO FLOOR CLEANOUT OR OPERATING ROOM
AP ACCESS PANEL FCW FILTERED COLD WATER OVFL OVERFLOW
AS AUTOMATIC SPRINKLER FD FLOOR DRAIN
AFG ABOVE FINISHED GRADE FDC FIRE DEPARTMENT (HOSE) PA PASCAL
ASD ADJUSTABLE SPEED DRIVE CONNECTION PC PLUMBING CONTRACTOR
ASHRAE AMERICAN SOCIETY HEATING, FM FLOW METER PD PRESSURE DROP
REFRIGERATING, AIR CONDITIONING FOP FUEL OIL PUMP PDI PLUMBING AND DRAINAGE INSTITUTE
ENGINEERS FOR FUEL OIL RETURN PG PRESSURE GAUGE
ASME AMERICAN SOCIETY OF MECHANICAL FOS FUEL OIL SUPPLY PP PLUMBING PUMP
ENGINEERS FOV FUEL OIL VENT PPM PARTS PER MILLION
ASPE AMERICAN SOCIETY PLUMBING FS FLOOR SINK PRS PRESSURE REDUCING STATION
ENGINEERS FU FIXTURE UNITS PRV PRESSURE REDUCING VALVE
ASR AUTOMATIC SPRINKLER RISER PSI POUNDS PER SQUARE INCH
AV ACID VENT G GALLON PSIA POUNDS PER SQUARE INCH -
AW ACID WASTE GC GENERAL CONTRACTOR ABSOLUTE
GCO GRADE CLEANOUTS PSIG POUNDS PER SQUARE INCH - GAUGE
BAS BUILDING AUTOMATION SYSTEM GD GARBAGE DISPOSAL PTRV PRESSURE TEMPERATURE RELIEF
BF BELOW FLOOR GPD GALLONS PER DAY VALVE
BFP BACKFLOW PREVENTER GPH GALLONS PER HOUR PW POTABLE WATER
BG BELOW GRADE GPM GALLONS PER MINUTE
BH BOOSTER HEATER GPR GAS PRESSURE REGULATOR RD ROOF DRAIN
BHP BREAK HORSEPOWER GRS GAS REGULATOR STATION RDL ROOF DRAIN LEADER
BJ BETWEEN JOISTS GT GREASE TRAP RL ROOF LEADER
BOP BOTTOM OF PIPE GVTR GAS VENT THROUGH ROOF RO REVERSE OSMOSIS
BSP BLACK STEEL PIPE GWH GAS FIRED WATER HEATER RWL RAIN WATER LEADER
BT BATHTUB
BTUH BRITISH THERMAL UNITS PER HOUR H&CW HOT AND COLD WATER SAN SANITARY SEWER
HB HOSE BIBB (INTERIOR) SCFM STANDARD CUBIC FOOT/MIN
C CELSIUS HD HUB DRAIN SCW SOFTENED COLD WATER
CF COMBINATION FIXTURE HEX HEAT EXCHANGER SDMH STORM DRAIN MANHOLE
CGA COMPRESSED GAS ASSOCIATION HP HORSEPOWER SP SUMP PUMP
CI CAST IRON HS HAND SINK SPR SPRINKLER LINE
CO CLEANOUT HST HOT WATER STORAGE TANK SQFT SQUARE FOOT
COND CONDENSATE (DOMESTIC) SS STAINLESS STEEL
CP CONDENSATE PUMP HWB HOT WATER BOILER ST STORAGE TANK
CS CLINICAL SINK HWCP HOT WATER CIRCULATING PUMP SW STORM WATER
CV CONTROL VALVE HWP HOT WATER PUMP
HYD HYDRANT TCV TEMPERATURE CONTROL VALVE
DA DENTAL COMPRESSED AIR TD TRENCH DRAIN
DCW DOMESTIC COLD WATER ICW INDUSTRIAL COLD WATER TDH TOTAL DYNAMIC HEAD
DHW DOMESTIC HOT WATER INV INVERT TDL TEMPERATURE DIFFERENTIAL
DHWR DOMESTIC HOT WATER RETURN IPC INTERNATIONAL PLUMBING CODE TEMP TEMPERATURE
DHWS DOMESTIC HOT WATER SUPPLY IRW IRRIGATION WATER TMV THERMOSTATIC MIXING VALVE
DF DRINKING FOUNTAIN IW INDIRECT WASTE TP TRAP PRIMER
DI DEIONIZED WATER IWH INSTANTANEOUS WATER HEATER TSTAT THERMOSTAT
DN DOWN IWR INDUSTRIAL WATER RETURN TWR TEMPERED WATER RETURN
DOE DEPARTMENT OF ENERGY IWS INDUSTRIAL WATER SUPPLY TWS TEMPERED WATER SUPPLY
DS DOWNSPOUT TYP TYPICAL
DV DENTAL VACUUM KW KILOWATT
DW DISHWASHER KWHR KILOWATT-HOUR UPC UNIFORM PLUMBING CODE
DWG DRAWING
DWH DOMESTIC WATER HEATER L/S LITER PER SECOND V SANITARY VENT
DWR DRINKING WATER RETURN LA LABORATORY AIR VAC VACUUM
DWS DRINKING WATER SUPPLY LAV LAVATORY VB VACUUM BREAKER
DWV DRAIN WASTE VENT LBS/HR POUNDS PER HOUR VCO VACUUM CLEANER OUTLET
LCW LABORATORY COLD WATER VP VACUUM PUMP
EC ELECTRICAL CONTRACTOR LHW LABORATORY HOT WATER VS VENT STACK
EL ELEVATION LNG LIQUID NATURAL GAS VTR VENT THROUGH ROOF
EMCS ENERGY MONOSERRAT AND CENTRAL LOX LIQUID OXYGEN
SYSTEM LV LABORATORY VACUUM W WASTE
EPA ENVIRONMENTAL PROTECTION LW LOW WATER WC WATER CLOSET
AGENCY WCO WALL CLEANOUT
EPACT ENERGY POLICY ACT M METER WG WATER GAUGE
ESC ESCUTCHEON MA MEDICAL AIR WH WATER HEATER
ESH EMERGENCY SHOWER MAV MANUAL AIR VENT WHA WATER HAMMER ARRESTOR
ET EXPANSION TANK MBH THOUSANDS OF BTU PER HOUR WHD WALL HYDRANT
EWC ELECTRIC WATER COOLER MC MECHANICAL CONTRACTOR WL WATER LINE
EWH ELECTRIC WATER HEATER MED MEDICAL WM WATER METER
EWS EYEWASH STATION MER MECHANICAL EQUIPMENT ROOM WPD WATER PRESSURE DROP
EX EXISTING MH MANHOLE WS WASTE STACK
MOU MEMORANDUM OF UNDERSTANDING
MSB MOP SINK BASIN YCO YARD CLEANOUT
MV MEDICAL VACUUM YH YARD HYDRANT
N2 NITROGEN
N2O NITROUS OXIDE
NC NORMALLY CLOSED
NG NATURAL GAS
NIC NOT IN CONTRACT
NO NORMALLY OPEN
NOM NOMINAL
NPW NON POTABLE WATER
NTS NOT TO SCALE
PLUMBING SYMBOLS
PIPING SYSTEM
AW AW ACID WASTE
CA CA COMPRESSED AIR
CD CD CONDENSATE DRAIN
CO2 CO2 CARBON DIOXIDE
NG NG NATURAL GAS
GW GW GREASE WASTE
MA MA MEDICAL AIR
N2 N2 NITROGEN
N2O N2O NITROUS OXIDE
OFD OFD OVERFLOW STORM DRAIN
OW OW OIL WASTE
O2 O2 OXYGEN
PD PD PUMP DISCHARGE
PG PG PROPANE GAS
ST ST STORM
VAC VAC VACUUM
WAGD WAGD WASTE ANESTHETIC GAS DISPOSAL
SAN SAN SANITARY WASTE
AV
OV
V
AV ACID VENT
OV OIL VENT
V SANITARY VENT
DCW DOMESTIC COLD WATER
DI DI DE-IONIZED WATER
NPCW NPCW NONPOTABLE COLD WATER
RO RO REVERSE OSMOSIS WATER
SCW SCW SOFTENED COLD WATER
HW 140
DHW DOMESTIC HOT WATER
DHW 140 DOMESTIC HOT WATER (OTHER TEMP)
DCWR DOMESTIC CW RECIRCULATION
DHWR DOMESTIC HW RECIRCULATION
ANNOTATION SYMBOLS
# NEW WORK KEYNOTE
# DEMOLITION KEYNOTE
NEW BOLD TEXT INDICATES NEW ITEM
EXISTING ITALICIZED TEXT INDICATES EXISTING ITEM
DETAIL MODULE NUMBER
DETAIL OR SECTION MARK
SHOWN ON DRAWING
CONNECT TO EXISTING
LIMIT OF DEMOLITION
P-### NEW PLUMBING EQUIPMENT MARK
P-### EXISTING PLUMBING EQUIPMENT MARK
PROJECT MANAGER
ACG Project Number
21-040
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
10 11 12 13 14
CRAWL SPACE
UNEXCAVATED
11 12
TO BUILDING
DISTRIBUTION SYSTEM
13 FROM BUILDING
1.
1" MA
2" MV
EX-11-VP-1
A. COORDINATE ANY LOSS OF MEDICAL GAS
SERVICE WITH COR AT LEAST 1 WEEK IN
ADVANCE. PROVISION OF TEMPORARY MEDICAL
GAS SUPPLY EQUIPMENT DURING EQUIPMENT
CHANGEOVER IS NOT IN CONTRACT.
B. PROVIDE STICKY MAT ON ENTIRE FLOOR AREA
OF SERVICE ELEVATOR LOBBY BETWEEN
SERVICE ELEVATOR #3 AND THE DOOR TO THE
MECHANICAL ROOM WHERE NEW EQUIPMENT IS
BEING INSTALLED.
C. PROVIDE ZIPPER ENTRY AT ACCESS POINT TO
CONSTRUCTION AREA, AT DOOR TO THE
MECHANICAL ROOM, IF THE WORK REQUIRED TO
STAIRWELL EXIT TO
OUTSIDE AT FIRST
FLOOR LEVEL
EX-11-AC-1
OUTLINE OF EXISTING
HOUSKEEPING PAD
4" MV EXHAUST
INSTALL THE EQUIPMENT WILL GENERATE ANY
DUST.
D. PROVIDE ICRA BARRIERS AS NECESSARY TO
PERFORM THE WORK PER THE MECHANICAL
DRAWINGS. THE USE OF TEMPORARY, REUSABLE
TYPE BARRIERS (EX. STARC WALL SYSTEMS) IS
ACCEPTABLE. ENSURE THAT ACCESS THROUGH
THE BARRIERS IS PROVIDED IF THE TEMPORARY
BARRIERS CUTS OFF ACCESS FROM THE
MECHANICAL ROOM OR ANY OTHER SPACE TO
THE MEANS OF EGRESS.
E. CONTRACTOR IS TO PROVIDE FLOORING
PROTECTION FROM ENTRY DOOR TO SERVICE
ELEVATOR OF THE FIRST FLOOR AND FROM THE
SERVICE ELEVATOR TO THE MECHANICAL ROOM
DOOR AT THE BASEMENT LEVEL. PLYWOOD
FLOOR COVERING IS SUGGESTED, BUT THE
CONTRACTOR CAN UTILIZE WHATEVER FLOOR
PROTECTION MATERIAL THEY FEELS WILL
ACCOMPLISH THE JOB. CONTRACTOR SHALL BE
RESPONSIBLE FOR AND DAMAGE TO THE
FLOORING, ADJACENT SURFACES, ETC. CAUSED
BY THE PERFORMANCE OF THE CONTRACTOR’S
WORK OR BY THEIR PERSONNEL.
F. FOR PHOTOGRAPHS OF EXISTING CONDITIONS
SEE PG601.
MECH. ROOM
4" MV EXHAUST UP
THROUGH ROOF
EXHAUST TO
OUTLET
PIPE BASEMENT
FROM INTAKE
# NEW WORK KEY NOTES:
1 NEW MEDICAL AIR COMPRESSOR SYSTEM.
CONNECT TO EXISTING SUPPLY PIPING AND AIR
INTAKE.
2 NEW MEDICAL VACUUM PUMP. CONNECT TO
EXISTING SUPPLY PIPING AND VACUUM EXHAUST.
3 RECONNECT NEW EQUIPMENT TO EXISTING
MASTER ALARM AT JUNCTION BOX.
4 PROVIDE SUPPORT FOR NEW AND EXISTING
ELECT.
CLOSET
NO. 3
ELEVATOR
LOBBY
SERVICE ELEVATOR FOR
ACCESS WITH EQUIPMENT
BASEMENT MECHANICAL ROOM MEDICAL GAS DEMOLITION PLAN
Scale: 1/4" = 1'-0"
0 2 4 8 MEDGAS PIPING ASSOCIATED WITH AIR
COMPRESSOR AND VACUUM PUMP IN MECHANICAL
ROOM. SUPPORT FOR THIS PIPING SHALL CONFORM
WITH SPECIFICATIONS.
5 PROVIDE STICKY MAT AT ENTRANCE TO
CONSTRUCTION AREA PER FACILITY ICRA
REQUIREMENTS.
6 PROVIDE NEW MEDICAL AIR INTAKE PIPING.
1 PROVIDE SUPPORTS FOR MEDICAL PIPING INSIDE
MECHANICAL ROOM AND OUTSIDE. SUPPORTS FOR
E.C.
MAIN
LOBBY
NO. 2
NO. 1
11 12 TO BUILDING
13 FROM BUILDING
1" MA
2" MV
THIS PIPING SHALL CONFORM WITH
SPECIFICATIONS.
7 TURN DOWN AND SCREEN OUTSIDE INTAKE PIPING.
E UNEXCAVATED
11-AC-1
OUTLINE OF EXISTING
HOUSKEEPING PAD
11-VP-1
# DEMOLITION KEY NOTES:
F
BASEMENT MECHANICAL ROOM INFECTION CONTROL PLAN
Scale: 1/8" = 1'-0"
2 1/2" MA
4" MV EXHAUST
1 EXISTING MEDICAL AIR COMPRESSORS, RECEIVER,
AND DESICCANT COOLER TO BE REMOVED.
DEMOLISH ASSOCIATED PIPING TO POINTS
INDICATED.
2 EXISTING MEDICAL VACUUM PUMP TO BE REMOVED.
DEMOLISH ASSOCIATED PIPING TO POINTS
INDICATED.
3 REMOVE EXISTING DEW POINT SENSOR ON MEDICAL
AIR SUPPLY MAIN AND TURN OVER TO VA.
4 DEMOLISH MEDICAL AIR INTAKE PIPING COMPLETE
INCLUDING PIPING INSTALLED OUTSIDE.
TO OUTLET
FROM 3RD
FLOOR ROOF
WEST
NORTH MECHANICAL
EAST
BASEMENT MECHANICAL ROOM MEDICAL GAS PLAN
Scale: 1/4" = 1'-0"
0 2 4 8
S S S S SMOKE RESISTIVE HAZARDOUS AREA
Drawing Title
Project Title
100% BID DOCUMENTS
VA PROJECT NUMBER
Office of Construction
MECHANICAL ROOM MEDICAL GAS
PLANS
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
671A4-22-172 and Facilities # 11
1 Revision #1
01/10/2025
Management
U.S. Department
Location
KERRVILLE VA MEDICAL CENTER
Date Checked Drawn
PG400
Date of Veterans Affairs KERRVILLE, TX 78028-5758 JANUARY 11, 2022 RBK NAS Dwg
CONSULTANTS:
/1
/2
:1
:2
U s e rs
\s m is e v ic h \D o c u m e n ts
\M F
P -0
-V
A K e rr v ill e -M e d
A ir V a c S y s te m -T
X s m is e v ic h .r v t
U
M
V
E
X H
A U
S
M
V
E
X H
A U
S
I D
A I
PROJECT MANAGER
ACG Project Number
21-040
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
DocuSign Envelope ID: 23E13BA4-D0F9-42A1-BCB8-E4C79E81E576
VACUUM PUMP SCHEDULE
MARK
LOCATION
AREA AND/OR BLDG
SERVED
SYSTEM AND/OR
TYPE UNIT
RECEIVER SIZE VACUUM PRESSURE
CFM DELIVERY @ VACUUM
PRESSURE (SI)
ELECTRIC POWER
WEIGHT
(LBS)
REMARKS
NO. PUMPS
POWER EACH
PHASE VOLT FLA
GAL [L] in Hg [kPa] SCFM STANDARD [L/s] HP [Kw]
11-VP-1 BASEMENT... BUILDING 11 MEDICAL VACUUM DUPLEX DRY CLAW 200 [ 760 ] 19 [ 64 ] 65 [ 31 ] 2 7.5 5.6 [ 3 ] 460 21 2190 1, 2, 3, 4, 5, 6, 7, 8, 9
NOTES:
1. CAPACITIES ARE AT SITE CONDITION.
2. BASIS OF DESIGN IS AMICO MODEL V-CCD-D-200P-SS-N-075-MS.
3. MODULAR UNITS THAT CAN BE DISASSEMBLED AND REASSEMBLED ON SITE.
4. PROVIDE ALL ACCESSORIES FOR FULLY FUNCTIONAL SYSTEM.
5. PROVIDE ALL ALARMS PER NFPA AND CONNECT TO EXISTING ALARM SYSTEM.
6. PROVIDE DRAIN PIPING WITH VALVE AND MAIN ISOLATION VALVE FOR THE UNIT.
7. UNIT SHALL OPERATE BETWEEN 19" Hg AND 25" Hg WITH CAPACITY AT LISTED SCFM.
8. LISTED CAPACITY IS WITH ONE PUMP IN OPERATION.
9. PROVIDE GAUGES, SENSORS, AND SAFETY DEVICES.
AIR COMPRESSOR SCHEDULE
MARK
LOCATION
AREA AND/OR
BLDG SERVED
SYSTEM AND/OR
TYPE
NFPA AIRFLOW CAPACITY
NFPA DISCHARGE
PRESSURE
RECEIVER SIZE
RECEIVER DIMENSIONS, DIA
X LENGTH
ELECTRIC POWER [60 Hz]
WEIGHT
(LBS)
REMARKS
NO. MOTORS
POWER EA.
PHASE
VOLT
FLA
CFM [L/s] PSIG [kPa] GAL [L] IN [mm] HP [kW]
11-AC-1
BASEMENT
MECHANICAL
BUILDING 11 MEDICAL AIR
DUPLEX OIL-LESS
SCROLL
24 [ 94 ] 120 [ 830 ] 200 [ 760 ] 36X84 [900X2100] 2 7.5 [ 6 ] 3 460 22 2330 1, 2, 3, 4, 5, 6, 7, 8, 9
NOTES:
1. CAPACITIES ARE AT SITE CONDITION.
2. BASIS OF DESIGN IS AMICO MODEL A-SCD-D-200P-SS-N-075-TS
3. MODULAR UNIT THAT CAN BE DISASSEMBLED AND REASSEMBLED ON SITE.
4. PROVIDE ALL ACCESSORIES FOR FULLY FUNCTIONAL SYSTEM.
5. PROVIDE ALL ALARMS PER NFPA AND CONNECT TO EXISTING ALARM SYSTEM.
6. PROVIDE DRAIN PIPING WITH ISOLATION VALVE AND MAIN ISOLATION VALVE FOR THE UNIT.
7. PROVIDE DESSICANT AIR DRYER WITH ALL ACCESSORIES.
8. LISTED CAPACITY IS WITH ONE PUMP IN OPERATION.
9. PROVIDE GAUGES, SENSORS, AND SAFETY DEVICES.
MEDICAL AIR COMPRESSOR VIEW 1
Scale: No Scale
EXISTING AIR AND VACUUM SYSTEMS
Scale: No Scale
VACUUM PUMPS
Scale: No Scale
MEDICAL AIR COMPRESSOR VIEW 3
Scale: No Scale
MEDICAL AIR COMPRESSOR VIEW 2
Scale: No Scale
EXISTING DESSICANT SYSTEM
Scale: No Scale
VACUUM TANK AND CONTROL PANEL
Scale: No Scale
MEDICAL AIR INTAKE
Scale: No Scale
P-601 # Revisions: Date
JANUARY 11, 2022 Dwg of Veterans Affairs
Drawn
NAS
Checked
RBK
Location
KERRVILLE VA MEDICAL CENTER
3600 MEMORIAL BLVD.
KERRVILLE, TX 78028-5758
Building Number
# 11
919-858-7420 www.acg-pa.com Engineers Architects © COPYRIGHT Apogee Consulting Group, P.A.
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
VA PROJECT NUMBER
671A4-22-172
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
Drawing Title
MEDICAL GAS SCHEDULES AND
EXISTING EQUIPMENT PHOTOGRAPHS
Office of Construction and Facilities Management
ACG Project Number
21-040 PROJECT MANAGER
1/11/2022
CONSULTANTS:
/7
/2
:5
:0
U s e rs
\n s to ltz \D o c u m e n ts
\M F
P -0
-V
A K e rr v ill e -M e d
A ir V a c
S y s te m -T
X n s to ltz .r v t
DocuSign Envelope ID: FCFB9153-3C99-48D2-8B65-8D3C2A627379
ELECTRICAL ABBREVIATIONS LIST
1P
1 POLE (2P, 3P, 4P, ETC.)
CONST
CONSTRUCTION
GAL
GALLON
MTR
MOTOR, MOTORIZED
SPKR
SPEAKER
1PH 1 PHASE CONT CONTINUATION OR CONTINUOUS GALV GALVANIZED MTS MANUAL TRANSFER SWITCH SP SPARE 2/C TWO-CONDUCTOR (3/C, 4/C, ETC.) CONTR CONTRACTOR GC GENERAL CONTRACTOR SR SURFACE RACEWAY 2W TWO-WIRE (3W, 4W, ETC.) COORD COORDINATE GEN GENERATOR NA NOT APPLICABLE SS STAINLESS STEEL 3PH 3 PHASE COR CONTRACTING OFFICER'S GFI GROUND FAULT CIRCUIT INTERRUPTER NC NORMALLY CLOSED SSW SELECTOR SWITCH
A AMPERE REPRESENTATIVE GND GROUND NEC NATIONAL ELECTRICAL CODE STD STANDARD
AAP ALARM ANNUNCIATOR PANEL CPT CONTROL POWER TRANSFORMER GRS GALVANIZED RIGID STEEL (CONDUIT) NEMA NATIONAL ELECTRICAL SURF SURFACE MOUNTED AC ABOVE COUNTER CRI COLOR RENDERING INDEX GYP BD GYPSUM BOARD MANUFACTURER'S ASSOCIATION SW SWITCH
AIR CONDITIONER CT CURRENT TRANSFORMER NFDS NON-FUSED SAFETY DISCONNECT SWBD SWITCHBOARD ALTERNATING CURRENT CTV CABLE TELEVISION HID HIGH-INTENSITY DISCHARGE SWITCH SYM SYMMETRICAL
ACC ACCESSIBLE CTR CENTER HOA HAND-OFF-AUTOMATIC NIC NOT IN CONTRACT SYS SYSTEM
ACLG ABOVE CEILING CU COPPER HP HORSEPOWER NL NIGHT LIGHT
ADO AUTOMATIC DOOR OPENER CU FT CUBIC FEET NO NORMALLY OPEN TC TIMECLOCK
A/E ARCHITECT/ENGINEER HT HEIGHT NTS NOT TO SCALE TEL TELEPHONE
AF AMP FRAME D DEMOLITION HTR HEATER TEL/DATA TELEPHONE/DATA
AFC AVAILABLE FAULT CURRENT (KILOAMPS) dB DECIBEL HV HIGH VOLTAGE OC ON CENTER TERM TERMINAL AFF ABOVE FINISHED FLOOR DB DIRECT BURIAL HVAC HEATING, VENTILATING AND AIR OD OUTSIDE DIAMETER TP TWISTED PAIR AFG ABOVE FINISHED GRADE DC DIRECT CURRENT CONDITIONING OH OVERHEAD TPS TWISTED PAIR SHIELDED AFI ARC FAULT CIRCUIT INTERRUPTER DEG C DEGREES CELSIUS OL OVERLOAD TR TAMPER RESISTANT AHJ AUTHORITY HAVING JURISDICTION DEG F DEGREES FAHRENHEIT IES ILLUMINATION ENGINEERING SOCIETY T-STAT THERMOSTAT AHU AIR HANDLING UNIT DEMO DEMOLISH/DEMOLITION IG ISOLATED GROUND PA PUBLIC ADDRESS TTB TELEPHONE TERMINAL BOARD AIC AMPERE INTERRUPTING CAPACITY DEPT DEPARTMENT IMC INTERMEDIATE METAL CONDUIT PB PULL BOX OR PUSHBUTTON TTC TELEPHONE TERMINAL CABINET AL ALUMINUM DIA DIAMETER IR INFRARED PBPU PREFABRICATED BEDSIDE PATIENT TV TELEVISION
ALT ALTERNATE DISC DISCONNECT I/W INTERLOCK WITH UNIT TYP TYPICAL
AMP AMPERE DIST DISTRIBUTION IWH INSTANTANEOUS WATER HEATER PCB POLYCHLORINATED BIPHENYL
APPROX APPROXIMATELY DN DOWN PC PHOTOELECTRIC CELL UC UNDER COUNTER
ARCH ARCHITECT, ARCHITECTURAL DS SAFETY DISCONNECT SWITCH J-BOX JUNCTION BOX PED PEDESTAL UE UNDERGROUND ELECTRICAL
AT AMP TRIP DT DOUBLE THROW PEND PENDANT UG UNDERGROUND
ATS AUTOMATIC TRANSFER SWITCH DWG DRAWING KV KILOVOLT PF POWER FACTOR UH UNIT HEATER
AUTO AUTOMATIC KVA KILOVOLT-AMPERE PH PHASE UL UNDERWRITER'S LABORATORY
AUX AUXILIARY EC ELECTRICAL CONTRACTOR KVAR KILOVOLT-AMPERE REACTIVE PIV POST INDICATING VALVE UT UNDERGROUND TELEPHONE AV AUDIO VISUAL EG EQUIPMENT GROUND KW KILOWATT PNL PANEL UTP UNSHIELDED TWISTED PAIR AWG AMERICAN WIRE GAUGE ELEC ELECTRIC, ELECTRICAL KWH KILOWATT HOUR POE POWER-OVER-ETHERNET UTIL UTILITY
ELEV ELEVATOR PR PAIR UV UNIT VENTILATOR OR
BAS BUILDING AUTOMATION SYSTEM EM EMERGENCY LED LIGHT-EMITTING DIODE PRI PRIMARY ULTRAVIOLET BATT BATTERY EMI ELECTROMAGNETIC INTERFERENCE LF LINEAR FEET/FOOT PROJ PROJECTION
BC BARE COPPER EMS ENERGY MANAGEMENT SYSTEM LM LUMEN PRV POWER ROOF VENTILATOR V VOLT BD BOARD EMT ELECTRICAL METALLIC TUBING LRA LOCKED ROTOR AMPS PT POTENTIAL TRANSFORMER VA VOLT-AMPERES BFF BELOW FINISHED FLOOR EPO EMERGENCY POWER OFF LT LIGHT PVC POLYVINYL CHLORIDE (CONDUIT) VERT VERTICAL BLDG BUILDING EQUIP EQUIPMENT LTG LIGHTING PWR POWER VFD VARIABLE FREQUENCY DRIVE
BMS BUILDING MANAGEMENT SYSTEM ESMT EASEMENT LTNG LIGHTNING VOL VOLUME
BYP BYPASS ETR EXISTING TO REMAIN LV LOW VOLTAGE QTY QUANTITY
EWC ELECTRIC WATER COOLER W WATT
C CONDUIT EWH ELECTRIC WALL HEATER MATV MASTER ANTENNA TELEVISION SYSTEM R RELOCATE/RELOCATED W/ WITH CAB CABINET ELECTRIC WATER HEATER MAX MAXIMUM RCP REFLECTED CEILING PLAN WH WATER HEATER
CALC CALCULATED EX EXISTING MC MECHANICAL CONTRACTOR RCPT RECEPTACLE W/O WITHOUT
CAP CAPACITY EXP EXPLOSION PROOF METAL-CLAD (CONDUIT) REQD REQUIRED WP WEATHERPROOF
CAT CATALOG, CATEGORY MCA MINIMUM CIRCUIT AMPACITY RGS RIGID GALVANIZED STEEL
CATV COMMUNITY ANTENNA TELEVISION FA FIRE ALARM MCB MAIN CIRCUIT BREAKER RM ROOM XFMR TRANSFORMER CB CIRCUIT BREAKER FAAP FIRE ALARM ANNUNCIATOR PANEL MCC MOTOR CONTROL CENTER RMS ROOT MEAN SQUARE XFR TRANSFER CCR CONTROL CONTACTOR FACP FIRE ALARM CONTROL PANEL MDP MAIN DISTRIBUTION PANEL RTU ROOF TOP UNIT
CCT CORRELATED COLOR TEMPERATURE FC FOOTCANDLE MECH MECHANICAL
CCTV CLOSED CIRCUIT TELEVISION FCU FAN COIL UNIT MFR MANUFACTURER SC SURFACE CONDUIT ANGLE cd CANDELA FIXT FIXTURE MFS MAIN FUSED DISCONNECT SWITCH SEC SECONDARY @ AT
CD CONSTRUCTION DOCUMENTS FLR FLOOR MH MANHOLE SF SQUARE FEET/FOOT
DELTA
CF CONTRACTOR-FURNISHED MIC MICROPHONE SHT SHEET ' FEET
CKT CIRCUIT FU FUSE MIN MINIMUM SI INTERNATIONAL SYSTEM OF UNITS " INCHES
CLG CEILING FUDS FUSED SAFETY DISCONNECT MISC MISCELLANEOUS SIM SIMILAR # NUMBER
CO CONTRACTING OFFICER SWITCH MLO MAIN LUGS ONLY S/N SOLID NEUTRAL Φ PHASE
COAX COAXIAL CABLE FVR FULL VOLTAGE REVERSING MOA MULTIOUTLET ASSEMBLY SPEC SPECIFICATION CL CENTER LINE COMM COMMUNICATION FVNR FULL VOLTAGE NON-REVERSING MSBD MAIN SWITCHBOARD PL PLATE
CONN CONNECTION MT MOUNT
100% CONSTRUCTION DOCUMENTS
E-001
Date Dwg
EXAMPLE 1: WALL MOUNTED EXIT SIGN TYPE 'E1' WITH SINGLE FACE
AND DIRECTIONAL ARROW.
'LPN'-1,3,5
HOME RUN TO BRANCH CIRCUIT PANEL: THE PANEL DESIGNATION AND CIRCUIT
DESIGNATION ARE SHOWN ADJACENT TO THE HOME RUN ARROW.
CIRCUIT BREAKER SIZES (AMPS/NUMBER OF POLES) ARE SHOWN IN THE PANEL
SCHEDULE WITH THE CORRESPONDING PANEL AND CIRCUIT DESIGNATION.
EXAMPLE: HOME RUN TO PANELBOARD 'LPN'; CIRCUITS 1, 3, 5.
DEMOLITION KEY NOTE: SEE THE SHEET KEY NOTES ON THAT SHEET FOR THE
NOTE NUMBER INDICATED IN THE SQUARE.
E1
T1
EXAMPLE 2: CEILING MOUNTED EXIT SIGN TYPE 'E2' WITH DOUBLE
FACE AND DIRECTIONAL ARROWS.
d
SYMBOL NOTATIONS: UPPER CASE LETTERS ADJACENT TO SYMBOLS INDICATE
A UNIT TYPE. SEE APPROPRIATE SCHEDULE OR SPECIFICATIONS.
c
THE LIGHTING FIXTURE TYPE IS INDICATED BY AN UPPER CASE LETTER OR AN UPPER
CASE LETTER AND NUMBER.
THE SWITCH DESIGNATION (IF NEEDED) IS INDICATED BY A LOWER CASE LETTER.
EXAMPLE: LIGHTING FIXTURE TYPE 'A1' IS CONTROLLED BY SWITCH 'b'.
EXIT SIGNS: STEM INDICATES WALL MOUNTING. NO STEM INDICATES CEILING
MOUNTING. SHADED AREA INDICATES ILLUMINATED FACE(S). ARROW INDICATES
DIRECTIONAL ARROW ON ILLUMINATED FACE(S).
DEVICES: THE SWITCH DESIGNATION IS INDICATED BY A LOWER CASE LETTER.
EXAMPLE: SPLIT DUPLEX RECEPTACLE; ONE RECEPTACLE OUTLET IS
CONTROLLED BY SWITCH 'c'.
THE CONTROL DEVICE DESIGNATION IS INDICATED BY A LOWER CASE LETTER.
E2
EXAMPLE: SINGLE POLE SWITCH 'd' TO CONTROL LIGHTING
FIXTURES INDICATED BY 'd'.
TRANSFORMERS: THE TRANSFORMER TYPE IS INDICATED BY A NUMBER
FOLLOWING THE UPPER CASE LETTER 'T'. SEE THE TRANSFORMER SCHEDULE OR
THE SINGLE LINE DIAGRAM FOR THE TRANSFORMER DESCRIPTION AND
REQUIREMENTS.
EXAMPLE: TRANSFORMER TYPE 'T1'.
'LPN'
PANELBOARDS: PANELBOARD DOORS MAY BE SHOWN TO INDICATE OPENING
SIDE OF RECESSED PANELBOARDS. SEE PANELBOARD IDENTIFICATION FOR
DESIGNATION CODES.
SHEET KEY NOTE: SEE THE SHEET KEY NOTES ON THAT SHEET FOR THE NOTE
NUMBER INDICATED IN THE HEXAGON.
A1 b of Veterans Affairs
06/30/23
USING OHM'S LAW AND THE KNOWN RESISTIVITY CONSTANT OF CONDUCTOR METAL (CIRCULAR MIL-OHM PER
FOOT) FOR COPPER CONDUCTORS, CALCULATE THE REQUIRED WIRE SIZE NECESSARY SO AS NOT TO EXCEED THE
RECOMMENDED VOLTAGE DROP OF 3% FOR BRANCH CIRCUIT CONDUCTORS.
ALL ELECTRICAL DEVICES, FIXTURES, EQUIPMENT AND FEEDERS SHALL BE INSTALLED IN ACCORDANCE WITH THE
PROJECT SPECIFICATIONS, THE MANUFACTURER'S RECOMMENDED PROCEDURES, ALL APPLICABLE LOCAL AND STATE
CODES, AMERICANS WITH DISABILITIES ACT AND WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE AND VA
STANDARDS.
PROVIDE ADDITIONAL SUPPORT FOR DEVICES, FIXTURES, EQUIPMENT AND FEEDERS WHERE THE BUILDING
CONSTRUCTION IS NOT SUITABLE FOR DIRECT MOUNTING.
FIRESTOP, DRAFTSTOP, SMOKESTOP AND/OR PROTECT THE ANNULAR SPACE AROUND ALL PENETRATIONS THROUGH
WALLS, PARTITIONS, FLOORS, CEILING, AND ROOFS IN ACCORDANCE WITH THE REQUIREMENTS OF THE NATIONAL
ELECTRICAL CODE, UL LISTING REQUIREMENT AND THE APPLICABLE BUILDING CODES.
VERIFY CEILING SYSTEMS AND PROVIDE MOUNTING ACCESSORIES, TRIMS AND ALL REQUIRED MOUNTING HARDWARE TO
SUIT THE PARTICULAR INSTALLATION.
PROTECT EXISTING UNDERGROUND AND BUILDING INTERIOR UTILITIES DURING CONSTRUCTION.
BRANCH CIRCUIT CONDUCTORS SHALL BE 12 AWG COPPER MINIMUM.
COORDINATE ANY AND ALL WORK WITH ALL OTHER TRADES PRIOR TO INSTALLATION SO AS TO AVOID CONFLICT DURING
CONSTRUCTION.
ALL PANELS SHALL HAVE TYPED, COMPLETED DIRECTORIES INDICATING EQUIPMENT SERVED AND ROOM NUMBER (AS
INDICATED ON FINAL BUILDING ROOM SIGNAGE) OF EQUIPMENT LOCATION, OR SPARE, OR SPACE.
MANUFACTURER'S NAME AND MODEL NUMBER ARE GIVEN FOR DESCRIPTIVE PURPOSES, TO INDICATE A QUALITY
STANDARD, AND ARE NOT INTENDED TO LIMIT PRODUCTS TO A PARTICULAR MANUFACTURER. PRODUCTS DEEMED EQUAL
AND APPROVED BY THE DESIGNER WILL BE ACCEPTED. ALL PRODUCTS MUST COMPLY WITH "BUY AMERICAN ACT".
ALL FEEDERS AND CIRCUITRY SHALL BE TORQUED PER THE PANEL, BREAKER, AND/OR PARTICULAR EQUIPMENT
MANUFACTURER'S SPECIFICATIONS.
CIRCUITRY TO SWITCHES, RECEPTACLES, AND ALL OTHER DEVICES SHALL BE TERMINATED ON THE DEVICE'S SCREW
TERMINALS.
MOUNTING HEIGHTS INDICATED ARE TO CENTER OF DEVICE, OUTLET, FIXTURE, OR EQUIPMENT UNLESS NOTED
OTHERWISE.
ALL WIRE TERMINATIONS SHALL BE RATED FOR 75 DEGREES C.
ALL CONDUCTORS SHALL HAVE THHN/THWN INSULATION, UNLESS OTHERWISE NOTED.
ALL CONDUIT SHALL BE RGS OR EMT UNLESS OTHERWISE NOTED. FMC CONDUIT SHALL BE USED ON VIBRATING
EQUIPMENT (MAXIMUM LENGTH OF 6 FEET). PVC SHALL BE USED FOR UNDERGROUND CONDUIT OR WHERE CONCRETE- ENCASED. WHERE VIBRATING EQUIPMENT IS LOCATED OUTSIDE, LFMC CONDUIT SHALL BE USED (MAXIMUM SIX FEET) FOR
FINAL CONNECTIONS.
ALL ELECTRICAL MATERIALS, DEVICES, APPLIANCES AND EQUIPMENT SHALL BE LABEL LISTED BY AN APPROVED THIRD
PARTY TESTING AGENCY.
FOR ALL LED FIXTURES, THE FIXTURE MUST BE CAPABLE OF SELF-RESETTING TO THE SWITCHED/CONTROLLED STATE
DURING ANY FLUCTUATION IN POWER SUPPLY WHERE AUTOMATIC PROTECTIVE MEASURES DISABLE THE LED LAMPS.
PROVIDE A LETTER OR STATEMENT FROM THE MANUFACTURER, OR OTHER ACCEPTABLE PROOF, THAT ALL LED FIXTURES, WITH OR WITHOUT BROWNOUT PROTECTION, WILL RETURN TO THE SWITCHED/CONTROLLED STATE AUTOMATICALLY.
PROVIDE STATEMENT WITH THE FIXTURE SUBMITTALS.
IN CASE OF CONFLICTS OR DISCREPANCIES WITHIN OR AMONG THE CONTRACT DRAWINGS, THE BETTER QUALITY, MORE
STRINGENT REQUIREMENTS OR GREATER QUANTITY OF WORK , AS DETERMINED BY THE GOVERNMENT, SHALL BE
PROVIDED.
COMPRESSION COUPLINGS SHALL BE USED. NO SET SCREW FITTINGS ARE ALLOWED.
MINIMUM CONDUIT SIZE TO BE USED IS 3/4".
FEEDER TAGS APPENDED WITH 'VD' ARE UPSIZED IN AMPACITY DUE TO VOLTAGE DROP.
SHORT CIRCUIT CURRENT CALCULATIONS SHALL BE PERFORMED ON THE ELECTRICAL SYSTEM USING A BOLTED
THREE PHASE FAULT CONDITION FOR CALCULATIONS TO ACHIEVE MAXIMUM THERMAL AND MECHANICAL STRESS TO
THE SYSTEM FOR CALCULATIONS PURPOSES. EXISTING SHORT CIRCUIT VALUES WILL BE USED OUTSIDE THE WORK
AREA, FOR THE STARTING POINT OF THE CALCULATIONS.
JWB
Checked
MJJ
Date
JANUARY 11, 2022
Approved Location
KERRVILLE VA MEDICAL CENTER
3600 VETERANS DRIVE, KERRVILLE, TX
28028
Building Number
# 11
VA PROJECT NUMBER
671-A4-22-172
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
Drawing Title
ELECTRICAL GENERAL NOTES,
SYMBOL LEGENDS, AND
ABBREVIATIONS
/7
/2
:2
:2
U s e rs
\jw b e e z le y \D o c u m e n ts
\E -0
-V
A K e rr v ill e -M e d
A ir V a c S y s te m -T
X jw b e e z le y a c g -p a .c o m
.r v t
LIGHTING SYMBOL LEGEND
SYMBOL
HEIGHT
AFF * DESCRIPTION
--- 2' X 2' RECESSED ARCHITECTURAL TROFFER
--- 2' X 4' RECESSED ARCHITECTURAL TROFFER
--- 2' X 4' RECESSED FLAT LENS TROFFER
7'-0"
CENTER
2' LINEAR - WALL MOUNTED
7'-0"
CENTER
4' LINEAR - WALL MOUNTED
--- 4' LINEAR STRIP - SUSPENDED
--- 6" DIAMETER RECESSED DOWNLIGHT
7'-5"
BOTTOM
EXIT SIGN - WALL MOUNTED
--- EXIT SIGN - CEILING MOUNTED
7'-6"
CENTER
EMERGENCY LIGHT BATTERY
OPERATED - WALL MOUNTED
6'-0"
CENTER
WALL SCONCE
48"
CENTER
SINGLE POLE WALL SWITCH
SWITCH MODIFIERS:
OS: OCCUPANCY SENSOR, LOW VOLTAGE,
DUAL TECHNOLOGY
VS: VACANCY SENSOR, LOW VOLTAGE,
DUAL TECHNOLOGY
DR: DUAL RELAY
VSD: VACANCY SENSOR, LINE VOLTAGE, DIMMER
OSD: OCCUPANCY SENSOR, LINE VOLTAGE, DIMMER
LV: LOW VOLTAGE
LVD: LOW VOLTAGE DIMMER
D: LINE VOLTAGE DIMMER
T: DIGITAL TIMER
K: KEYED
3: 3 WAY
4: 4 WAY
OS ---
OCCUPANCY SENSOR, LOW VOLTAGE,
DUAL TECHNOLOGY - CEILING MOUNTED
OS --- OCCUPANCY SENSOR, LOW VOLTAGE, DUAL
TECHNOLOGY - WALL/CORNER MOUNTED
RC --- LOW VOLTAGE LIGHTING CONTROL DEVICE
RCD ---
LOW VOLTAGE LIGHTING CONTROL
DEVICE WITH 0-10V DIMMING
LS ---
LIGHT LEVEL SENSOR, SINGLE ZONE
DIMMING CLOSED LOOP - CEILING MOUNTED
* - ANY HEIGHTS INDICATED IN PLANS SHALL SUPERSEDE THOSE LISTED HERE.
ELECTRICAL SYMBOL LEGEND
SYMBOL HEIGHT DESCRIPTION
18" 120V SIMPLEX RECEPTACLE
18" 120V DUPLEX RECEPTACLE
18"
120V DUPLEX RECEPTACLE EMERGENCY OUTLET
AND COVER PLATE SHALL BE RED IN COLOR
18" 120V QUADRUPLEX RECEPTACLE
18" 120V QUADRUPLEX RECEPTACLE EMERGENCY
OUTLET AND COVER PLATE SHALL BE RED IN COLOR
18" 208V DUPLEX RECEPTACLE
18" SPECIAL RECEPTACLE
120V DUPLEX RECEPTACLE FLOOR BOX
120V DUPLEX RECEPTACLE EMERGENCY FLOOR BOX
OUTLET AND COVER PLATE SHALL BE RED IN COLOR
120V QUADRUPLEX RECEPTACLE FLOOR BOX
120V QUADUPLEX RECEPTACLE EMERGENCY
FLOOR BOX OUTLET AND COVER PLATE SHALL BE
RED IN COLOR
RECEPTACLE MODIFIERS:
AC: 3" ABOVE COUNTERTOP/BACKSPLASH
GFI: GROUND FAULT CURRENT INTERRUPTER
WP: WEATHERPROOF ENCLOSURE
DED: DEDICATED CIRCUIT
J AS
NOTED
JUNCTION BOX, WALL MOUNTED, PURPOSE AS NOTED
J
CLNG
JUNCTION BOX, CEILING MOUNTED, PURPOSE AS
NOTED
J
JUNCTION BOX, FLOOR MOUNTED, PURPOSE AS
NOTED
E
AS
NOTED
EQUIPMENT CONNECTION
72" TOP DISCONNECT SWITCH NON-FUSED
72" TOP DISCONNECT SWITCH FUSED
48" SINGLE POLE WALL SWITCH: 48" AFF STANDARD UNO
SWITCH MODIFIERS:
M: MOTOR-RATED
ALL HEIGHTS ARE TO CENTER OF DEVICE UNLESS NOTED OTHERWISE.
TOP: HEIGHT TO TOP OF DEVICE
ANY HEIGHTS INDICATED IN PLANS SHALL SUPERSEDE THOSE LISTED HERE.
PROJECT MANAGER
ACG Project Number
21-040
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511 mailto:System-TX_jwbeezley@acg-pa.com.rvt
11 12 13 14
A. COORDINATE DISCONNECTION OF EXISTING
EQUIPMENT AND CONNECTION TO NEW
EQUIPMENT WITH MECHANICAL CONTRACTOR.
B. EQUIPMENT CONNECTIONS SHALL BE MADE
ACCORDING TO MANUFACTURER'S
INSTRUCTIONS AND RECOMMENDATIONS.
C. PROVIDE EQUIPMENT CONNECTIONS TO
INCLUDE DISCONNECTING MEANS,
CONDUCTORS, CONDUIT, & OCPD ACCORDING
TO THE EQUIPMENT CONDUCTOR SCHEDULE
ON SHEET-'E-601'.
D. EXISTING CONNECTIONS TO MEDICAL-GAS
ALARM PANEL TO REMAIN FOR USE WITH NEW
EQUIPMENT. SEE EQUIPMENT CONDUCTOR
SCHEDULE NOTES.
EX-11-AC-1 1
EX-11-VP-1 2
E. DURING THE COURSE OF CONSTRUCTION,
TEMPORARY EQUIPMENT (UNITS AND/OR
CYLINDERS) FOR MAINTAINING AIR-VACUUM
OPERATION MAY BE REQUIRED. THE VA WILL
FURNISH AND INSTALL TEMPORARY SYSTEMS;
HOWEVER, COORDINATE TEMPORARY
EQUIPMENT WORK WITH VA PERSONNEL AS
NEEDED.
# NEW WORK KEY NOTES:
EXISTING 3 4
PANEL 'PBYEA'
1 CONNECT NEW PACKAGED MEDICAL AIR
COMPRESSOR EQUIPMENT. SEE
MECHANICAL EQUIPMENT CONDUCTOR
SCHEDULE AND ASSOCIATED PANEL
SCHEDULE.
2 CONNECT NEW PACKAGED MEDICAL
VACUUM PUMP EQUIPMENT. SEE
MECHANICAL EQUIPMENT CONDUCTOR
SCHEDULE AND ASSOCIATED PANEL
SCHEDULE.
3 INSTALL A NEW 30A/3P CIRCUIT BREAKER
TO FEED THE NEW VACUUM PLANT
EQUIPMENT IN PANEL-'PBYEA',
CIRUITS-1,3,5.
4 INSTALL A NEW 30A/3P CIRCUIT BREAKER
TO FEED THE NEW AIR COMPRESSOR
EQUIPMENT IN PANEL-'PBYEA',
CIRCUITS-19,21,23.
BASEMENT MECHANICAL ROOM ELECTRICAL DEMOLITION PLAN
# DEMOLITION KEY NOTES:
11 12 13 14
1 DISCONNECT AND REMOVE EXISTING
ELECTRICAL CONNECTION TO PACKAGED
MEDICAL AIR COMPRESSOR EQUIPMENT.
REMOVE EXISTING DISCONNECTING
MEANS AS WELL AS ASSOCIATED
CONDUITS AND CONDUCTORS, BACK TO
SOURCE OF ORIGIN.
2 DISCONNECT AND REMOVE EXISTING
ELECTRICAL CONNECTION TO PACKAGED
MEDICAL VACUUM EQUIPMENT. REMOVE
EXISTING DISCONNECTING MEANS AS
WELL AS ASSOCIATED CONDUITS AND
1 11-AC-1
11-VP-1 2
CONDUCTORS BACK TO SOURCE OF
ORIGIN.
3 REMOVE THE EXISTING 40A/3P CIRCUIT
BREAKER FEEDING THE EXISTING VACUUM
PLANT EQUIPMENT IN PANEL -'PBYEA',
CIRCUITS-1,3,5.
4 REMOVE THE EXISTING 30A/3P CIRCUIT
BREAKER FEEDING THE EXISTING AIR
COMPRESSOR EQUIPMENT IN
PANEL-'PBYEA', CIRCUITS-19,21,23.
EXISTING 3 4
PANEL 'PBYEA'
NORTH MECHANICAL
WEST EAST
BASEMENT MECHANICAL ROOM ELECTRICAL PLAN
S S S S
SMOKE RESISTIVE HAZARDOUS AREA
ELECTRICAL ENLARGEMENTS -
DEMOLITION AND NEW WORK
100% CONSTRUCTION DOCUMENTS
Location
KERRVILLE VA MEDICAL CENTER
3600 VETERANS DRIVE, KERRVILLE, TX
E-401
28028 JANUARY 11, 2022 Dwg
VA PROJECT NUMBER
671-A4-22-172
U.S. Department of Veterans Affairs
Drawn
JWB
Checked
MJJ
Building Number
# 11
919-858-7420 www.acg-pa.com Engineers Architects © COPYRIGHT Apogee Consulting Group, P.A.
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
Office of Construction and Facilities Management
ACG Project Number
21-040 PROJECT MANAGER
/7
/2
:2
:2
U s e rs
\jw b e e z le y \D o c u m e n ts
\E -0
-V
A K e rr v ill e -M e d
A ir V a c S y s te m -T
X jw b e e z le y a c g -p a .c o m
.r v
MICHAEL J.
JENSEN
Exp. 06/30/23
ELECTRICAL
Branch Panel: PBYEA Location: MECHANICAL ROOM-028 Volts: 277/480 A.I.C. Rating: 14,000 AMPS SYMMETRICAL
Supply From: SWITCHBOARD-'1YEC'... Phases: 3 Mains Type: MLO
Mounting: RECESSED Wires: 4 Mains Rating: 225 A
Enclosure: NEMA 1 MCB Rating: 225 A
Notes:
EQUIPMENT BRANCH PANEL
CKT
Circuit Description
Trip
Poles
Poles
Trip
Circuit Description
CKT
VACUUM SYSTEM
30 A
6.10 5.54
40 A
<CONNECTED LOAD>
3 6.10 5.54 4
5 6.10 5.54 6
SPD COMP - 'OFF'
20 A
0.00 0.00
SPARE - 'OFF'
9 0.00 0.00 10
11 0.00 0.00 12
XFMR OLD DRYER
3.88 6.65
60 A
A/C #5
15 3.88 6.65 16
17 3.88 6.65 18
MEDICAL AIR
6.10 5.54
BAILER
21 6.10 5.54 22
23 6.10 5.54 24
DOMESTIC WATER BOOSTER PUMP
2.77 4.43
2 30 A SPD CONTROL
27 2.77 4.43 28
29 2.77 -- 1 -- SPACE 30
31 SPACE -- 1 -- 0.00
70 A
33 SPACE -- 1 -- 0.00 34
35 SPACE -- 1 -- 0.00 36
SPS AIR COMP
2.22 0.00
39 2.22 0.00 40
41 2.22 0.00 42
Total Load: 43.23 kVA 43.23 kVA 38.80 kVA
Total Amps: 159 A 159 A 140 A
Legend:
Load Classification Connected Load Demand Factor Estimated Demand Panel Totals
Spare 125.26 kVA 125.00% 156.58 kVA
Total Conn. Load: 125.26 kVA Total Est. Demand: 156.58 kVA Total Conn.: 151 A
Total Est. Demand: 188 A
Notes:
ELECTRICAL SCHEDULES
FULLY SPRINKLERED
100% CONSTRUCTION DOCUMENTS
VA PROJECT NUMBER
671-A4-22-172
Building Number
# 11
Location
KERRVILLE VA MEDICAL CENTER
3600 VETERANS DRIVE, KERRVILLE, TX
E-601 # Revisions: Date
28028 JANUARY 11, 2022 Dwg of Veterans Affairs
Drawn
JWB
Checked
MJJ
919-858-7420 www.acg-pa.com Engineers Architects
© COPYRIGHT Apogee Consulting Group, P.A.
Headquarters:
Apogee Consulting Group, P.A.
1151 Kildaire Farm Road, Suite 120
Cary, North Carolina 27511
Project Title
REMOVE/REPLACE MEDICAL
AIR VACUUM SYSTEM
Office of Construction and Facilities Management
ACG Project Number
21-040 PROJECT MANAGER
/7
/2
:2
:2
U s e rs
\jw b e e z le y \D o c u m e n ts
\E -0
-V
A K e rr v ill e -M e d
A ir V a c S y s te m -T
X jw b e e z le y a c g -p a .c o m
.r v
MICHAEL J.
JENSEN
Exp. 06/30/23
ELECTRICAL
File details come from the government source that posted it. Updated .