36C25626Q0852_Draft PWS.pdf

PDF 137 KB Posted

Attached to
6525--RFI - NeuroSignal Robotic Transcranial Doppler (raTCD)
Solicitation number
36C25626Q0852
Issued by

About this file

This is a Performance Work Statement (PWS) for the acquisition and implementation of a Robotic Transcranial Doppler (TCD) system at the Michael E. DeBakey VA Medical Center (MEDVAMC). The contractor must procure one robotic TCD system with a medical-grade console, adjustable robotic headset, two 2 MHz transducers, mobile cart, cables, accessories, battery backup/UPS, and printer. The system must include software licenses for automated vessel search, probe actuation, continuous signal optimization, spectral Doppler/M-mode visualization, microembolic signal (MES) detection, waveform trending, and secure data export. The platform must support bilateral monitoring with autonomous probe reacquisition after patient movement, hands-free continuous monitoring capability, and integration with VA-approved hardware and network environments. Data export must be available in secure formats (PDF, CSV, image files) for CPRS clinical documentation, with HL7 integration if supported or secure file-based transfers to VA servers; all patient data must remain on VA-controlled systems with no offsite or cloud storage authorized.

The contractor shall complete delivery within 45–60 days after purchase order issuance, installation within 30 days of equipment receipt, and training within 30 days of installation. Installation and configuration must be coordinated with VA IT and Biomedical Engineering and verified through acceptance testing demonstrating successful bilateral insonation, autonomous reacquisition, MES detection functionality, and report generation to designated VA directories. The contractor must provide at least one structured training session covering device orientation, robotic headset application, protocols for vasospasm surveillance and emboli detection, artifact reduction, alarm configuration, and report generation, along with training materials and competency checklists. Support and warranty requirements include a 12-month minimum device and software warranty from installation date, remote technical support, software updates during warranty, onsite service with 24-hour response for critical issues, and preventive maintenance aligned to Biomedical Engineering guidance. All contractor personnel requiring access to VA systems must complete background checks, and the contractor must comply with VA Handbook 6500, HIPAA requirements, and coordinate with the Information Security Officer for cybersecurity review before installation proceeds.

View the file

Other files

File Type Posted
36C25626Q0852_NOI SS.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Performance Work Statement (PWS) – Robotic Transcranial Doppler (TCD) System

Title: Acquisition and Implementation of Robotic Transcranial Doppler (TCD) System (NeuroSignal)

1. Purpose

The purpose of this PWS is to outline the requirements for acquisition, installation, configuration, and implementation of a Robotic Transcranial Doppler (TCD) system to support advanced cerebrovascular monitoring, including detection and trending of vasospasm, evaluation of cerebral autoregulation, real-time assessment of hemodynamics, and microembolic signal (MES) surveillance across neurocritical care, stroke, and perioperative settings at the Michael E. DeBakey VA Medical Center

(MEDVAMC).

2. Scope

This PWS covers:

• Procurement of one (1) Robotic TCD system, including console, robotic headset(s), 2 MHz transducers, fixation mechanisms, mobile cart, cables, and required accessories.

• Software licenses/modules necessary for automated vessel search, probe actuation, continuous signal optimization, spectral Doppler/M-mode visualization, emboli detection, waveform trending, and data export/reporting.

• Integration with VA-approved hardware and network environments, including user authentication and secure storage on VA-controlled systems.

• Installation and configuration of the device, software, and network connectivity.

• Clinical training for physicians, technologists, and nursing staff.

• Initial technical support, warranty services, and implementation assistance.

3. Objectives

• Enable continuous, high-quality neuromonitoring for patients at risk of vasospasm, hypoperfusion, embolic phenomena, or autoregulatory failure.

• Standardize TCD acquisition quality and reduce operator dependency through robotic automation and autonomous signal optimization.

• Improve workflow efficiency and expand monitoring duration without sustained technologist presence.

• Integrate TCD data with clinical documentation workflows and case conferences (stroke, neuro-ICU).

• Ensure secure implementation within VA IT and cybersecurity standards and maintain all patient data on VA-controlled systems.

• Support quality improvement initiatives and research protocols that require uninterrupted, reproducible monitoring.

4. Requirements

4.1 System Specifications (Device & Software)

The contractor shall provide a Robotic TCD platform with the following capabilities:

• Robotic automation: Automated vessel search, probe alignment, and continuous signal optimization for bilateral monitoring (e.g., MCA) with minimal operator intervention.

• Autonomous reacquisition: Automatic probe adjustment/reacquisition after patient movement, with audible/visual alerts for signal loss and recovery.

• Continuous monitoring stability: Hands-free monitoring suitable for prolonged trending in neuro-ICU environments.

• Ultrasound modalities: Spectral Doppler and M-mode visualization, real-time hemodynamic measurements and beat-to-beat waveform capture.

• Advanced analytics:

o Microembolic signal (MES) detection and logging.

o Trend analysis and configurable alarms for vasospasm risk (user-defined thresholds).

o Optional autoregulation indices and physiologic correlation (if available from vendor).

• User workflows: Protocol-driven studies (vasospasm surveillance, emboli detection, autoregulation), configurable presets, and role-based user access.

• Data management:

o Export to secure formats (PDF, CSV, image files) for inclusion in CPRS/clinical documentation.

o HL7 or equivalent integration path if supported; otherwise secure file-based transfers to VA servers.

o Local storage on VA-approved workstation or network share; no offsite/cloud storage.

4.2 Hardware Requirements

The contractor shall provide:

• One (1) Robotic TCD console with medical-grade display, keyboard/mouse, and integrated control software.

• One (1) Robotic headset (adjustable for adult head sizes) with fixation and comfort accessories; optional pediatric accommodations if available.

• Two (2) 2 MHz transducers (primary use) and any required connectors/cables.

• Mobile cart (lockable) with cable management and footprint suitable for ICU rooms.

• Battery backup/UPS compatible with VA standards to protect against brief power interruptions.

• Printer (or VA-approved print workflow) for bedside summary reports as needed.

• Any necessary interface modules to capture or reference physiologic timing signals (if supported), provided they meet VA cybersecurity and biomedical standards.

4.3 Installation and Configuration

The contractor shall:

• Coordinate with VA IT and Biomedical Engineering for installation planning, network configuration, and device inspections.

• Perform onsite or remote installation of the system and software.

• Configure user roles, study protocols, and data export destinations on VA-controlled storage.

• Verify functionality via acceptance testing, including:

o Successful bilateral insonation and stable continuous monitoring on test subjects or approved phantom.

o Demonstration of automated search/alignment and autonomous reacquisition after controlled head movement.

o Generation and export of test reports to designated VA directories.

• Installation shall not proceed without VA IT and ISO approval.

4.4 Data Handling and Security

• All VA patient data shall remain on VA-controlled systems; no offsite or vendor cloud retention is authorized under this PWS.

• Contractor shall comply with VA Directive and Handbook 6500 series and HIPAA requirements.

• Contractor personnel shall complete required background checks if access to VA systems or facilities is needed.

• No cloud-based storage or remote data hosting is authorized unless expressly approved by VA in writing.

4.5 Training

The contractor shall provide clinical training for relevant staff, including:

• Device orientation and robotic headset application.

• Protocols for vasospasm surveillance, emboli detection, and autoregulation workflows.

• Best practices for study setup, artifact reduction, alarm configuration, and report generation.

• Training materials (manuals and digital documentation) and at least one (1) structured session plus Q&A support.

4.6 Support and Warranty

The contractor shall provide:

• One-year device and software warranty (minimum).

• Remote technical support and software updates during the warranty period.

• Onsite service when required; target response within 24 hours for critical issues (or per vendor standard terms).

• Preventive maintenance schedule aligned to Biomedical Engineering guidance.

5. Performance Standards

Performance shall be evaluated based on:

Standard Acceptable Performance

Delivery & Installation

Completed within agreed timeframe after PO issuance; all hardware/software installed and networked per VA requirements.

Functionality Successful bilateral insonation; continuous monitoring with autonomous reacquisition; MES detection features operational;

export/report functions verified.

Data Quality Stable, reproducible Doppler signals suitable for trending and clinical decision-making; minimal artifact during standard ICU conditions.

Training Completion of structured session(s); delivery of training materials;

users demonstrate proficiency in protocol setup and report generation.

Support Response to technical issues within 24 hours; escalation pathway defined; updates applied without disruption to clinical care.

Security Compliance

Full adherence to VA IT and security policies (VA Handbook 6500, HIPAA); no offsite data storage; access controls implemented. [1]

6. Deliverables

The contractor shall provide:

• Robotic TCD system hardware (console, headset, transducers, cart, accessories).

• Software licenses/modules and activation keys.

• Installation documentation, configuration notes, and acceptance test results.

• Training materials and schedule; attendance log and competency checklist.

• Warranty documentation and service contact information.

• Security compliance documentation required for VA IT/ISO review.

7. Period of Performance

• Delivery: Within 45–60 days ARO (or per vendor terms).

https://dvagov-my.sharepoint.com/personal/kasra_rahbar_va_gov/_layouts/15/Doc.aspx?sourcedoc=%7BCD55D172-71C8-46A3-85FA-4DE61FAC6E3F%7D&file=PWS%20for%20Neura%20Signal%20Doppler.docx&action=default&mobileredirect=true

• Installation: Within 30 days of equipment receipt.

• Training: Within 30 days of installation.

• Warranty: 12 months from installation date.

8. Security Requirements

The contractor shall:

• Comply with VA Handbook 6500 and associated directives.

• Ensure no offsite storage or external hosting of VA data.

• Coordinate with Information Security Officer (ISO) for cybersecurity review.

• Provide documentation required for VA IT approval, including network diagrams, software versioning, and patch/update procedures.

• Participate in security review or site visit if required.

(End of PWS)