36C25020B0030-006.pdf
PDF Posted
- Solicitation number
- 36C25020B0030
About this file
This pre-solicitation notice announces a forthcoming solicitation for a New MRI Site Prep project at the Dayton VA Medical Center in Dayton, Ohio. The solicitation will require preparation of the site for a new building to house MRI equipment, including all labor, supervision, materials, and equipment, to be completed according to drawings and specifications within 365 calendar days of notice to proceed. The project value is estimated between $2 million and $5 million.
The solicitation will be issued pursuant to 38 U.S.C. 8127 and is a 100% set-aside for service-disabled veteran-owned small businesses. Interested SDVOSBs must be small businesses under the relevant NAICS code of 236220 and registered in verification websites including VIP.vetbiz.gov and SAM.gov. The SDVOSB prime contractor or subcontractor must perform a minimum of 15% of the construction work. Bid, payment and performance bonds will be required. The solicitation and specifications will be posted on March 20, 2020 at FBO.gov, with proposals due approximately 30 days later. A one-time site visit will be scheduled. Amendments will be posted at FBO.gov and questions should be directed to the specified contracting officer by email.
36C25020B0030 Project Drawings (Part 3 of 4).pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25020B0030-0002001.docx | DOCX document | |
| 36C25020B0030-0001000.docx | DOCX document | |
| 36C25020B0030-008.pdf | ||
| 36C25020B0030-004.pdf | ||
| 36C25020B0030-007.pdf | ||
| 36C25020B0030-005.pdf | ||
| 36C25020B0030-001.docx | DOCX document | |
| 36C25020B0030-002.pdf | ||
| 36C25020B0030-003.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
P.5 O.7P.81P.9
A110
EQUIPMENT
A111
MRI ROOM
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
1/4" / 1'-0"
ROOF SCREEN WALL ROOF SCREEN WALL
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
6.29.7 8.8 7.9 6.46.97.3
A111
MRI ROOM
A112
CONTROL ROOM
A113
TOILET
A104
PREP AREA
A103
ENTRANCE
A102
ENTRANCE HALL
AE-501
AE-502
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
MILLWORK REMOVED IN MRI ROOM FOR CLARITY.
REFER TO FLOOR PLAN AND INTERIOR ELEVATIONS
CUBICLE
CURTAIN,
OFOI
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
8.87.96.9 7.3
TOP OF ROOF (HIGH SIDE)
16' - 3"
A105
SUB WAITING
A112
CONTROL ROOM
A101
LOBBY/WAITING
A111
MRI ROOM
STRUCTURAL
UNI-STRUT
SUPPORT
SYSTEM
PATIENT LIFT
PATIENT RAIL
SYSTEM
SUPPORT RAIL
TIE-ROD BY PATIENT
RAIL MANUFACTURER
METAL
DETECTOR
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o t
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :5
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
BUILDING SECTIONS
AE-500
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-500
2 BUILDING SECTION - CROSS
SCALE: 1/4" = 1'-0"AE-500
1 BUILDING SECTION - LONGITUDINAL
SCALE: 1/4" = 1'-0"AE-500
3 BUILDING SECTION - PATIENT LIFT SYSTEM
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
BID & CONSTRUCTION 12/19/2019
CAM REVIEW 02/21/2020
EXISTING FIRST FLOOR
0"
P.5 O.7P.81P.9Q.4
AE-501
AE-500
AE-510
A103
ENTRANCE
A101
LOBBY/WAITING
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
A100
VESTIBULE
GYP. BD.
SOFFIT POCKET
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
6.2
6.4 6.9 7.3
A100
VESTIBULE
AE-501
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
A101
LOBBY/WAITING
NEW STOREFRONT SYSTEM
EXTEND STUD AND EXTERIOR
SHEATHING TO DECK
METAL PANEL
COLUMN COVER
PERFORATEDD METAL
PANEL SCREEN WALL
EXISTING ANGLED
MANUFACTURED STONE
VENEER WALL
CANOPY SKYLIGHT
AE-510
SIM
AE-510
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :0
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
BUILDING SECTIONS
AE-501
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-501
1 LOBBY SECTION
SCALE: 1/4" = 1'-0"AE-501
2 BUILDING SECTION - VESTIBULE
Revisions Date
95% CD SUBMISSION 10/07/2019
EXISTING FIRST FLOOR
0"
5/8" GYP. BD. ON 2X4 WOOD
STUD
1 3/4" RF SHIELD WALL
2" ISOLATION
SCHEDULED CEILING REFER
TO CEILING PLAN
RF CEILING
SCHEDULED WALL, REFER
TO FLOOR PLAN
A111
MRI ROOM
A110
EQUIPMENT
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
9.7 8.8
AE-500
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
AE-511
AE-511
MILLWORK REMOVED FOR CLARITY. REFER TO
FLOOR PLAN AND INTERIOR ELEVATIONS
5/8" GYP. BD. ON 2X4 WOOD
STUD
1 3/4" RF SHIELD WALL
2" ISOLATION
SCHEDULED CEILING REFER
TO CEILING PLAN
RF CEILING
EXTERIOR WALL
AE-511
AE-511
C O
O R
D
IN
A T
E W
IT
H
M R
I M
A N
U F
A C
T U
R E
R
AE-511
RADIATION SHIELDING
(GAUSS LINE SHIELDING)
STRUCTURAL STEEL FRAMING,
PROVIDE 1HR FIREPROOFING (TYP.)
STRUCTURAL STEEL
FRAMING, PROVIDE 1HR
FIREPROOFING (TYP.)
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :0
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
BUILDING SECTIONS
AE-502
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/2" = 1'-0"AE-502
4 RF - Wall
SCALE: 1/2" = 1'-0"AE-502
1 ENLARGED MRI ROOM SECTION
Revisions Date
95% CD SUBMISSION 10/07/2019
EXISTING FIRST FLOOR
0"
P.81
1 3/8" STONE VENEER ON 2" RIGID
INSULATION ON 3/4" GLASS MAT SHEATHING
5/8" GYP. BD. ON 6" METAL STUD
WITH MIN. R-13 BATT INSULATION
TOP OF ROOF (HIGH SIDE)
16' - 3"
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
AE-510
AE-510
AE-510
FOUNDATION AND FOOTER, REFER
TO STRUCTURAL DRAWINGS
PERIMETER FOUNDATION DRAIN,
REFER TO CIVIL DRAWINGS
CANOPY ROOFING SYSTEM
PERFORATED METAL PANEL ROOF SCREEN WALL
METAL PANEL
TOP OF MANUF. STONE
18' - 0"
EXISTING FIRST FLOOR
0"
P.81P.9
AE-511
AE-511
R .O
TOP OF ROOF (HIGH SIDE)
16' - 3"
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
GYP. BD. SOFFIT @14'- 0 1/2" A.F.F.
TPO ROOFING MEMBRANE OVER
TAPERED POLYISOCYANURATE
INSULATION ON METAL DECK
METAL PANEL
AE-510
1 3/8" STONE VENEER ON 2" RIGID
INSULATION ON 3/4" GLASS MAT SHEATHING
5/8" GYP. BD. ON 6" METAL STUD
WITH MIN. R-13 BATT INSULATION
WINDOW SYSTEM
PERFORATED METAL PANEL ROOF SCREEN WALL
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :0
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
WALL SECTIONS
AE-505
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 3/4" = 1'-0"AE-505
1 Wall Section - Typical
SCALE: 3/4" = 1'-0"AE-505
2 Wall Section @ Window
FOUNDATION AND FOOTER, REFER
TO STRUCTURAL DRAWINGS
PERIMETER FOUNDATION DRAIN,
REFER TO CIVIL DRAWINGS
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
P.81
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
45.00°
ALUMINUM PERFORATED,
CORRUGATED METAL PANELS WITH
ISOLATION TAPE (MT-2)
REFER TO STRUCTURAL DRAWINGS FOR
TUBE STEEL AND ANCHORING. REFER TO
ROOF PLAN FOR POST LOCATIONS
TUBE STEEL KICKER AND POST,
REFER TO STRUCTURAL DRAWINGS
6"
24' - 0"
AE-510
STRUCTURAL STEEL FRAMING
WITH 1HR FIREPROOFING (TYP.)
REFER TO STRUCT. DWGS.
SCHEDULED CEILING
SKYLIGHT ON 12" CURB PROVIDED
BY SKYLIGHT MANUFACTURER
MOISTURE RESISTANT GYP. BD. ON 3 5/8"
METAL STUDS FRAMED OPENING
AE-510
SCHEDULED CEILING
1HR SPRAY-ON
FIREPROOFING (TYP.)
1'-6"
STRUCTURAL STEEL
FRAMING, SEE
STRUCT. DWGS.
COMPOSITE METAL
WALL SYSTEM
SHIM AS REQUIRED
3 5/8" COLD-FORMED
FRAMING
/2
ALUMINUM EXTRUSION
JOINT SYSTEM BY
PANEL MFG.
P.9
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
1'-0"
/2
/2
COMPOSITE METAL
WALL SYSTEM
METAL FASCIA
TPO ROOF MEMBRANE
3/4" PLYWOOD
COLD-FORMED METAL
FRAMING
8" 5/8"
TAPERED INSULATION
STRUCTURAL STEEL
FRAMING, SEE
STRUCT. DWGS.
TPO ROOF MEMBRANE
ROOF DECK, SEE
STRUCT. DWGS.
5/8" EXTERIOR
GYPSUM BOARD
COLD-FORMED METAL
FRAMING
COLD- FORMED
ANCHOR - GRAVITY &
LATERAL
COLD- FORMED
ANCHOR - LATERAL
BENT STEEL PLATE -
SEE STRUCT. DWGS.
VAPOR PERMEABLE
BARRIER
TOP OF MANUF. STONE
18' - 0"
P.81
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
THIN CALCIUM
SILICATE UNIT
STAINLESS STEEL
FLASHING
WATER-RESISTIVE
BARRIER
TPO ROOF MEMBRANE
TAPERED INSULATION
ROOF DECK, SEE
STRUCT. DWGS.
5/8" EXTERIOR
GYPSUM BOARD
BAR JOIST, SEE
STRUCT. DWGS.
2 5/8" 2"3/4" 6"
1'-6"
STRUCTURAL STEEL
FRAMING WITH 1HR
FIREPROOFING (TYP.),
SEE STRUCT. DWGS.
/2
/2
4" RIGID INSULATION
METAL COPING
BLOCKING AS
REQUIRED
ROOF SCREEN PANEL
TPO ROOF MEMBRANE
5/8" ROOFING COVER
BOARD
6" COLD-FORMED
METAL FRAMING
TAPERED INSULATION
CRICKET - SEE ROOF
PLAN
STRUCTURAL STEEL
FRAMING, SEE
STRUCT. DWGS.
TPO ROOF MEMBRANE
TPO ROOF MEMBRANE
SEALANT AND BACKER
ROD
TPO ROOF MEMBRANE
RIGID INSULATIONRIGID INSULATION
ROOF VAPOR BARRIER
PANEL STARTER CLIP
P.81
T.O.S. SCREENWALL
24'-0"
ROOF SCREEN PANEL
STRUCTURAL STEEL
FRAMING, SEE
STRUCT. DWGS.
SHIM AS REQUIRED
CONCEALED
FASTENER
SUPPORT Z
J CHANNEL
S.S. RIVET PAINTED TO
MATCH PANEL
5/8" 3 5/8" 5/8"
2"
/2
'J' BEAD
BLOCKING AS
REQUIRED
EXTERIOR GYPSUM
SHEATHING
CLOSURE TRIM
SKYLIGHT SILL
EXTRUSION
INSULATED LAMINATED
GLASS
GASKET
SEALANT AND BACKER
ROD
WEEP HOLES AT EACH
RAFTER & HIP
SKYLIGHT SILL
EXTRUSION
SKYLIGHT BASE PLATE
0.032" ALUMINUM
APRON
FLEXIBLE FLASHING
ROOF VAPOR BARRIER
TPO ROOF MEMBRANE
RIGID INSULATION
RIGID INSULATION
COVER BOARD
TPO ROOF MEMBRANE
RIGID INSULATION
ROOF VAPOR BARRIER
COLD-FORMED METAL
FRAMING
EXTERIOR GYPSUM
SHEATHING
ROOF DECKING
R13 INSULATION
S K
Y L
IG
H
T M
A N
U F
A C
T U
R E
R
(C
O O
R D
IN
A
T E
W
IT
H
6.8
TOP OF STEEL (LOW SIDE)
15' - 3 1/2"
WATER-RESISTIVE
BARRIER
EXISTING CORE &
SHELL INSULATING
GLASS
WINDOW HEAD
FRAMING MEMBER
5/8" EXTERIOR
GYPSUM BOARD
COLD-FORMED METAL
FRAMING
TPO ROOF MEMBRANE
TAPERED INSULATION
ROOF VAPOR BARRIER
ROOF DECK, SEE
STRUCT. DWGS.
FLEXIBLE FLASHING
5 1/8"
/2
SEALANT AND BACKER
ROD
STAINLESS STEEL
FLASHING
FLEXIBLE FLASHING
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :0
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
DETAILS
AE-510
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1" = 1'-0"AE-510
1 ROOF TOP SCREEN WALL
SCALE: 1" = 1'-0"AE-510
2 SKYLIGHT SECTION
SCALE: 3" = 1'-0"AE-510
6 COLUMN COVER
SCALE: 1 1/2" = 1'-0"AE-510
8 CANOPY METAL PANEL
SCALE: 1 1/2" = 1'-0"AE-510
4 CANOPY ROOF @ WALL
SCALE: 1 1/2" = 1'-0"AE-510
5 SCREEN WALL TOP
SCALE: 3" = 1'-0"AE-510
3 SKYLIGHT CURB
SCALE: 1 1/2" = 1'-0"AE-510
7 STOREFRONT HEAD @ CANOPY
Revisions Date
95% CD SUBMISSION 10/07/2019
EXISTING FIRST FLOOR
0"
P.81
CONCRETE FOUNDATION,
REFER TO STRUCTURAL
DRAINGS
CONCRETE SLAB ON
GRADE, REFER TO
STRUCTURAL DRAWINGS
MIN. R13 BATT
INSULATION
/8
1 3/8"
1 3/8"
2" 5/8" 6" 5/8"
1'-6"
1'-0"
INSULATED SPANDREL
GLASS UNIT
WINDOW SILL FRAMING
MEMBER & SILL
RECEPTOR
BLOCKING AS
REQUIRED
5/8" GYPSUM BOARD
6" COLD-FORMED
METAL FRAMING
1 5/8" LIGHT GAUGE
METAL FRAMING
WATER-RESISTIVE
BARRIER
EXTERIOR GYPSUM
SHEATHING
THIN CALCIUM
SILICATE UNIT
RIGID INSULATION
ALUMINUM STARTER
CHANNEL
STAINLESS STEEL
FLASHING
SEALANT AND BACKER
ROD W/ WEEP VENTS
@ 24"O.C. TO ALIGN W/
CHANNEL WEEPS
FLEXIBLE FLASHING
SEALANT AND BACKER
ROD
STAINLESS STEEL
FLASHING
PLASTIC ISOLATING
SHIM AS REQUIRED
ALUMINUM
INGTERMEDIATE
CHANNEL
(2'X2') 2" PERIMETER
INSULATION
PERIMETER FOUNDATION
DRAIN, REFER TO CIVIL
DRAWINGS
POROUS FILL
FLUID APPLIED
WATERPROOFING
P.81
SCHEDULED CEILING
MIN. R13 BATT
INSULATION
WINDOW HEAD
9'-0"
STAINLESS STEEL
FLASHING
WINDOW HEAD
FRAMING MEMBER
INSULATED GLAZING
UNIT
WINDOW HORIZONTAL
FRAMING MEMBER
INSULATED SPANDREL
GLASS UNIT
/8
/8
5/8" GYPSUM BOARD
6" COLD-FORMED
METAL FRAMING
SEALANT AND BACKER
ROD
SEALANT AND BACKER
ROD - BEHIND
CHANNEL WEEPS
ALUMINUM STARTER
CHANNEL
BLOCKING AS
REQUIRED
THIN CALCIUM
SILICATE UNIT
RIGID INSULATION
WATER-RESISTIVE
BARRIER
2" COLD-FORMED "Z"
EXTERIOR GYPSUM
SHEATHING
PLASTIC ISOLATING
SHIM AS REQUIRED
SEALANT AND BACKER
ROD
1 3/8"
1 3/8"
2" 5/8" 6" 5/8"
1'-0"
1'-6"
COLD-FORMED BOX
HEADER, SEE STRUCT.
DWGS.
MULLION EXTENSION
'J' BEAD
5/8" GYPSUM BOARD
1 5/8" LIGHT GAUGE
METAL FRAMING
EXISTING FIRST FLOOR
0"
8.8
RECESSED RF FLOORING, REFER
TO RF FLOORING DETAIL
PARENT WALL, REFER TO
FLOOR PLAN FOR WALL TYPE
RF WALL, REFER TO RF WALL DETAIL 2" ISOLATION
SCHEDULED BASE
R F
W A
L L P
A N
E L
RECESSED REINFORCED SLAB ON
GRADE, REFER TO STRUCTURAL
DRAWINGS
CONCRETE SLAB ON
GRADE, REFER TO
STRUCTURAL DRAWINGS
SCHEDULED BASE
-1 7/8"
RECESSED FLOOR RF FLOOR PANEL 1
FINISH FLOOR, REFER TO FINISH
SCHEDULE
EXISTING FIRST FLOOR
0"
9.7
WALL FOOTING, REFER TO
STRUCTURAL DRAWINGS
RECESSED RF FLOORING,
REFER TO RF FLOORING DETAIL
RF SHIELDED WALL; REFER TO RF WALL
DETAIL
1" FILLER
2" ISOLATION SPACE
SCHEDULED BASE
CONCRETE SIDEWALK;
REFER TO CIVIL DRAWINGS
-1 7/8"
RECESSED FLOORRF FLOOR PANEL
FINISH FLOOR, REFER TO FINISH
SCHEDULER
F W
A L L P
A N
E L
/8
THIN CALCIUM
SILICATE UNIT
ALUMINUM
INGTERMEDIATE
CHANNEL
SHIM AS REQUIRED
SEALANT AND BACKER
ROD
RIGID INSULATION
2" COLD-FORMED "Z"
FLEXIBLE FLASHING
ALUMINUM STARTER
CHANNEL
STAINLESS STEEL
FLASHING
WATER-RESISTIVE
BARRIER
3/4" PLYWOOD
6" COLD-FORMED
METAL FRAMING
6"3/4"2"
1 1/4"
1 3/8"
1'-6"
PERIMETER FOUNDATION
DRAIN
(2'X2') 2" PERIMETER
INSULATION
FLUID APPLIED
WATERPROOFING
POROUS FILL
CONCRETE SLAB ON GRADE
BRONZE SPRAYED STRUCTURAL
AND ELECTRICAL SEAM CLAMP
1/4 - 20 FLAT HEAD SELF THREADING SCREW
POLYETHYLENE UNDERLAYMENT
(2 LAYERS MIN. 4MIL EACH)
3/4" RF FLOOR PANEL WITH 3OZ
COPPER ON TOP SIDE
1/8" SUB -FLOOR
1/8" VINYL FILLER TILE
1/8" HARDBOARD LEVELING LAYER
COORDINATE OVERALL FLOOR DEPTH WITH RF
PANEL MANUFACTURER AND MRI MANUFACTURER
FINISH FLOOR, REFER TO FINISH SCHEDULE
MRI EQUIPMENT BASEPLATE AND
HARDBOARD LEVELING LAYER, COORDINATE
WITH MRI EQUIPMENT MANUFACTURER
8.8
R F
W A
L L P
A N
E L
SOLID SURFACE SILL,
COORDINATE WITH RF VENDOR
RF WINDOW
CONTROL ROOM COUNTER WITH
BACKSPASH, REFER TO INTERIOR
ELEVATIONS
2" ISOLATION SPACE
PARENT WALL, REFER TO
FLOOR PLAN FOR WALL TYPE
RF WALL, REFER TO RF WALL
WALL DETAIL
MRI ROOM COUNTER WITH
BACKSPASH; REFER TO
INTERIOR ELEVATIONS
SOLID SURFACE SILL,
COORDINATE WITH RF VENDOR
5/8" 3 5/8"
2"1 3/4"
3 1/2" 5/8"
SHIM AS REQ'D. BY RF
MANUFACTURER
8.8
R F
W A
L L P
A N
E L
MRI ROOM LAY-IN CEILING, REFER
TO REFLECTED CEILING PLAN AND
FINISH SCHEDULE
RF WINDOW
CONTROL ROOM LAY-IN
CEILING, REFER TO
REFLECTED CEILING PLAN AND
FINISH SCHEDULE
CLIPS & FASTENERS PER SHIELDING
MANUFACTURER'S RECOMMENDATION.
TYPICAL. DO NOT PENETRATE THROUGH
SHIELDING.
RF CEILING PANEL BY RF
SHIELDING MANUFACTURER
RADIATION SHIELDING (GAUSS LINE
SHIELDING). REFER TO SIEMENS SHIELDING
REPORT FOR ADDITIONAL INFORMATION
SCHEDULED CEILING. REFER TO
FINISH SCHEDULE
SUPPORT SYSTEM TO STRUCTURE
ABOVE PER MANUFACTURER'S
RECOMMENDATION
EXTERIOR WALL CONSTRUCTION, REFER
TO WALL SECTIONS
C O
O R
D
IN
A T
E W
IT
H
M R
I M
A N
U F
RF SHIELD WALL PANEL
5/8" GYP. BD. ON 2X4 WOOD STUD
2" ISOLATION
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :0
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
DETAILS
AE-511
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1 1/2" = 1'-0"AE-511
1 WINDOW - DETAIL @ SILL
SCALE: 1 1/2" = 1'-0"AE-511
2 WINDOW - DETAIL @ HEAD
SCALE: 1 1/2" = 1'-0"AE-511
3 ENLARGED RF FLOOR/WALL SHIELD SECTION
SCALE: 1 1/2" = 1'-0"AE-511
6 ENLARGED RF FLOOR SHIELD SECTION
SCALE: 6" = 1'-0"AE-511
8 RF FLOORING
SCALE: 1 1/2" = 1'-0"AE-511
4 ENLARGED RF WINDOW SILL SECTION
SCALE: 1 1/2" = 1'-0"AE-511
5 ENLARGED RF WINDOW HEAD SECTION
SCALE: 1 1/2" = 1'-0"AE-511
7 ENLARGED RF CEILING DETAIL
Revisions Date
95% CD SUBMISSION 10/07/2019
4'
- 0
1/ 4"
/4
1"
NOTES:
1. SEE RCP FOR CEILING HEIGHTS.
2. CEILING TRACKS TO BE LOCATED 3"
OFF OF WALL, UNLESS OTHERWISE
NOTED
STEEL L 1"x1"x1/8"x4'-0"
LONG W/ LEGS TURNED
DOWN 3/4" ANCHOR TO
MAIN
TEES W/ #8 SHEET METAL
SCREWS
SPACER
FASTEN 30" O.C. MAX.
CEILING GRID
CUBICLE TRACK
CEILING TILE
NOTE:
PATIENT LIFT TRACK TO BE SUPPORTED ABOVE CEILING BY
ENGINEERED MEDICAL SUPPORT SYSTEM, COORDINATE WITH
ALL TRADES FOR LOCATIONS AND MOUNTING INSTRUCTIONS.
CL STRAIGHT SECTION
TRACK, RCP FOR LOCATIONS
SCHEDULED CEILING
NOTE: SEE RCP FOR CEILING HEIGHTS
M
IN
M
IN
4 7/8"
ACOUSTICAL
CEILING PANEL
ACOUSTICAL
CEILING PANEL
ALIGN WITH EDGE
OF WALL BEYOND
1/2" MAX
3 1/8"
1 7/8"
METAL CONTROL JOINT
DOUBLE STUD @ DOORS
GYP. BOARD, SEE
PARTITION TYPES
METAL STUD
1/2" FIRE RESISTIVE GYP.
BD.
(FULL HEIGHT OF PARTITION)
NOTE:
CONTROL JOINT TO OCCUR AT CORNERS OF ALL GYPSUM
BOARD OPENINGS AND 30'-0" MAX. AT ALL GYPSUM BOARD
PARTITIONS TO STRUCTURE ABOVE.
CONTINUOUS
SEALANT ON
ALL SIDES
P L A
N
S E
E
PLAN
SECTION
CONTINUOUS SEALANT
ALL SIDES
CONTINUOUS SEALANT
ALL SIDES
METAL STUD FRAMING
METAL STUD FRAMING
FIRE EXTINGUISHER
CABINET (F.E.C.)
SEMI RECESSED
CABINET
3x3x3/8 STEEL POST AND
1/4" BASE PLATE FROM
FLOOR TO DECK ABOVE
WOOD BLOCKING
SCHEDULED PARTITION
NEW GRAB BAR
CL
P.81
7.9
10 1/4" 1'-1 3/8"
G
C
G
(2)
LOCKER
UNITS
(L-1)
STRUCTURAL STEEL WITH
1HR FIREPROOFING (TYP.)
P.81
8.8
/4
2'-11 1/8" 10 1/4"
CG-1
G
G C
G
OVERFLOW
DOWNSPOUT WITH
REMOVABLE
SCREEN
ROOF DRAIN PIPING STRUCTURAL STEEL WITH
1HR FIREPROOFING (TYP.)
P.81
1'-3"
CG-1
OVERFLOW
DOWNSPOUT WITH
REMOVABLE
SCREEN
G
G
G
C
ROOF DRAIN PIPING
P.81
9.7
G
G
/4
1'-4 1/4"
CG-1
STRUCTURAL STEEL WITH
1HR FIREPROOFING (TYP.)
O.7
7.9
/4
8" 8"
C
G
G
GG
STRUCTURAL STEEL WITH
1HR FIREPROOFING (TYP.)
O.7
8.8
G
10 1/4"
2'-3 7/8"
/4
2'-2"
10 1/4" 1 7/8"
STRUCTURAL STEEL WTH
1HR FIREPROOFING (TYP.)
O.7
9.7
/4
1'-0 7/8"
11 5/8"
MANUFACTURED
STONE VENEER WALL
STRUCTURAL STEEL WITH
1HR FIREPROOFING (TYP.)
P.81
THIN CALCIUM
SILICATE UNIT -
FINISHED END UNIT
RIGID INSULATION
2" COLD-FORMED "Z"
WATER-RESISTIVE
BARRIER
3/4" PLYWOOD
6" COLD FORMED
STRUCTURAL FRAMING
AT 16" O.C MAX.
INSULATED SPANDREL
GLASS UNIT
WINDOW JAMB
FRAMING MEMBER
/8
/4
/4
/8
5/8" GYPSUM BOARD
BLOCKING AS
REQUIRED -
LIGHT GAUGE
FRAMING -
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :1
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
INTERIOR DETAILS AND SECTIONS
AE-700
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1 1/2" = 1'-0"AE-700
1 CURTAIN TRACK ISOMETRIC
SCALE: 3" = 1'-0"AE-700
2 CEILING_UNISTRUT
SCALE: 1 1/2" = 1'-0"AE-700
3 CEILING_GYPSUM BOARD
SCALE: 3" = 1'-0"AE-700
4 GYPSUM BOARD CONTROL JOINT
SCALE: 1" = 1'-0"AE-700
5 FEC CABINET_SEMI RECESSED
SCALE: 1 1/2" = 1'-0"AE-700
6 GRAB BAR SUPPORT
SCALE: 1" = 1'-0"AE-700
8 PLAN DETAIL
SCALE: 1" = 1'-0"AE-700
9 PLAN DETAIL
SCALE: 1" = 1'-0"AE-700
7 PLAN DETAIL
SCALE: 1" = 1'-0"AE-700
10 PLAN DETAIL
SCALE: 1" = 1'-0"AE-700
12 FIRST FLOOR PLAN - Callout 5
SCALE: 1" = 1'-0"AE-700
13 FIRST FLOOR PLAN - Callout 6
SCALE: 1" = 1'-0"AE-700
14 FIRST FLOOR PLAN - Callout 7
SCALE: 1 1/2" = 1'-0"AE-700
11 WINDOW JAMB DETAIL
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
PLASTIC LAMINATE ON
MDF
BASE CABINET
PLASTIC
LAMINATE
ADJUST. SHELF
DRAWER
/2
/8
/8
/8
/8
3" SCHEDULED BASE
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
SOLID SURFACE MATERIAL
COUNTERTOP AND
BACKSPLASH
25"
1/2" 25"
/2
/2
" SOLID SURFACE MATERIAL
COUNTERTOP AND
BACKSPLASH
METAL SUPPORT
BRACKET
PENCIL DRAWER
1/2"
1"
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
SCHEDULED BASE
PLASTIC
LAMINATE
FASCIA
PANEL
METAL SUPPORT
BRACKET
SCHEDULED BASE
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
GROMMET,
COORD.
LOCATIONS WITH
OWNER
25"
/2
SOLID SURFACE MATERIAL
COUNTERTOP AND
BACKSPLASH
1/2"
PLASTIC LAMINATE ON
MDF
BASE CABINET
DRAWER
SCHEDULED BASE
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
/8
/8
/8
/8
/8
/8
/2
3"
DRAWER
DRAWER
SOLID SURFACE MATERIAL
COUNTERTOP AND
BACKSPLASH
25"
1/2"
PLASTIC LAMINATE ON
MDF
BASE CABINET
SINK
/2
/8
/8
/8
/8
SCHEDULED BASE
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
3"
SOLID SURFACE
MATERIAL
COUNTERTOP AND
BACKSPLASH WITH
INTEGRAL SINK
25"
1/2"
/2
1/2" SOLID SURFACE
MATERIAL
1 1/2"
EASED
EDGE
SUBSTRAT
E
/2
1/2" SOLID SURFACE
MATERIAL
EASED
EDGE
DOOR/DRAW
ER MAY
OCCUR
1"
SUBSTRAT
E
SUBSTRAT
E
/2
1/2" SOLID SURFACE
MATERIAL
EASED
EDGE
DOOR/DRAW
ER MAY
OCCUR
1"
3/4"
1 1/2" DEPTH 1" DEPTH 3/4" DEPTH
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
UNDER CABINET LIGHTING
MAY OCCUR; REFER TO
ELECTRICAL DWGS.
PLASTIC LAMINATE ON MDF
ADJUST. SHELVES
14"
PROVIDE COMMON BLOCKING,
FURRING AND HANGING DEVICES
CEILING. SEE REFLECTED
CEILING PLANS FOR HEIGHTS.
GYP. BD. SOFFIT, REFER TO
INTERIOR ELEVATIONS
1"
PROVIDE COMMON BLOCKING,
FURRINGAND HANGING DEVICES
PROVIDE COMMON BLOCKING,
FURRING AND HANGING
DEVICES
SCHEDULED BASE
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
PLASTIC LAMINATE ON MDF
ROLLING SHELVES
PROVIDE COMMON BLOCKING,
FURRING AND HANGING
DEVICES
24"
CEILING. SEE REFLECTED
CEILING PLANS FOR HEIGHTS.
GYP. BD. SOFFIT, REFER TO
INTERIOR ELEVATIONS
1"
U .O
.N
/1
MIN.
8 1/4"
U.O.N.
24 1/4"
NOTE: LOOSE
BACKSPLASH TO
OCCUR ONLY AT
WET LOCATIONS
SOLID SURFACE
TOP W/ INTEGRAL
SINK BOWL, NO
DRIP EDGE
PLYWOOD
SUBSTRATE
REMOVABLE HIGH
PRESSURE
DECORATIVE
LAMINATE PANEL
CONCEALED
COUNTERTOP
SUPPORT
BRACKETS, TYP.
SCHEDULED WALL
BASE, TYP.
MAX.
6".
PROVIDE COMMON BLOCKING,
FURRING AND HANGING
DEVICES
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
PROVIDE COMMON BLOCKING,
FURRING AND HANGING
DEVICES
36"
FULL LENGTH CABINET
DOORS
CEILING. SEE REFLECTED
CEILING PLANS FOR HEIGHTS.
GYP. BD. SOFFIT, REFER TO
INTERIOR ELEVATIONS
1"
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
PROVIDE COMMON
BLOCKING, FURRING AND
HANGING DEVICES
CEILING. SEE REFLECTED
CEILING PLANS FOR HEIGHTS.
PLASTIC LAMINATE ON MDF
ADJUST. SHELVES
SCHEDULED BASE
24"
R F
S H
IE
L D
O
P E
N
IN
G
C O
O R
D
W
IT
H R
F V
E N
D O
R
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
COODINATE TOP OF
CABINET WITH RF & MRI
MANUFACTURER
1. ALL BASE CABINETS, WALL CABINETS, AND TALL CABINETS SHALL BE
PLASTIC LAMINATE WITH FINISHED ENDS WHERE REQUIRED, UNLESS
OTHERWISE NOTED.
2. ALL WALL CABINETS SHALL HAVE PLASTIC LAMINATE FACE RAIL TO
CONCEAL UNDERCABINET LIGHT FIXTURE UNLESS OTHERWISE NOTED.
3. ALL COUNTERS AND SPLASHES SHALL BE PLASTIC LAMINATE WITH
FINISHED ENDS, UNLESS OTHERWISE NOTED.
4. ALL SPLASHES SHALL BE 4" HIGH, UNLESS OTHERWISE NOTED. RETURN ALL
SPLASHES AT SIDE WALLS.
5. ALL SHELVING INDICATED SHALL BE ADJUSTABLE, PLASTIC LAMINATE WITH
FINISHED EDGES, UNLESS OTHERWISE NOTED.
6. ALL FILLER PANELS, COUNTER SUPPORT PANELS, BACK PANELS, END
PANELS, AND CLOSURE PANELS SHALL BE PLASTIC LAMINATE, UNLESS
OTHERWISE NOTED.
7. ALL BACK PANELS AT KNEE SPACES SHALL BE REMOVABLE.
8. ALL DOORS AND DRAWERS MARKED WITH "L" INDICATED SHALL HAVE
LOCKS. SEE INTERIOR ELEVATIONS.
9. ALL FILE DRAWERS (FD) SHALL HAVE "SUSPENSION FILE" RACKS. SEE
INTERIOR ELEVATIONS.
GYP. BD. SOFFIT, REFER TO
INTERIOR ELEVATIONS
PLASTIC LAMINATE ON ALL
EXPOSED SURFACES
EQUIPMENT (N.I.C.)
COORDINATE WITH END USER
PROVIDE COMMON BLOCKING,
FURRING AND HANGING
DEVICES
OUTLET, SEE ELECTRICAL
DWGS.
14"
CEILING. SEE REFLECTED
CEILING PLANS FOR HEIGHTS.
1"
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :1
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
MILLWORK SECTIONS
AE-701
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 3/4" = 1'-0"AE-701
3 BASE CABINET w/ DRAWER
SCALE: 3/4" = 1'-0"AE-701
2 COUNTERTOP with DRAWER
SCALE: 3/4" = 1'-0"AE-701
1 COUNTERTOP
SCALE: 3/4" = 1'-0"AE-701
5 DRAWER (3)
SCALE: 3/4" = 1'-0"AE-701
4 SINK BASE
SCALE: 3" = 1'-0"AE-701
20 06_EASED EDGE 1 1/2"_SSM
SCALE: 3/4" = 1'-0"AE-701
7 WALL with TASK
SCALE: 3/4" = 1'-0"AE-701
8 TALL CABINET
SCALE: 1" = 1'-0"AE-701
19 SINK BASE W/ REMOVABLE PANEL
SCALE: 3/4" = 1'-0"AE-701
9 TALL CABINET NO SHELF
SCALE: 3/4" = 1'-0"AE-701
10 FILTER CABINET WITH CABINET BELOW
MILLWORK GENERAL NOTES
SCALE: 3/4" = 1'-0"AE-701
6 OPEN CABINET
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
RF VIEW WINDOW
AE-701
SCHEDULED
BASE, TYP.
END PANEL
SS-2
PL-12
AE-701
METAL SUPPORT
BRACKET
2'-0"
AE-701
1'-3" 1'-3"
AE-500
FECSCHEDULED
BASE, TYP.
RF SHIELDED DOOR
IRWC-1
IRWC-1
B
COUNTER
TOP
F L U
S H
V
A L V
E
ROBE
HOOK
(RH)
FRAMED
MIRROR
(FM)
SHOWER
CURTAIN
ROD
(SCR)
TOWEL
BAR
(TB)
ACCESSORIES
PAPER
TOWEL
DISPENSER
(PTD)
GRAB BAR
(GB), SEE
DET.6/AE700
TOILET
TISSUE
DISPENSER
(TTD)
SANITARY
NAPKIN
DISPOSAL
(SND)
URINAL
PLUMBING
LAVATORY TOILET DUPLEX RECEPT.,
DATA, INTERCOM
OR TELEPHONE
OUTLET
WALL SWITCH
OR DUPLEX
OUTLET
AT COUNTER
TOP
WALL SWITCH,
T-STAT,
PUSH PLATE
ELECTRICAL
TOILET
SEAT
COVER
DISPENSER
(SCD)
NOTE:
MOUNTING HEIGHTS ARE AS INDICATED UNLESS OTHERWISE NOTED.
NOT ALL ACCESSORIES / FIXTURES MAY BE USED.
BABY
CHANGING
STATION
(BCS)
FIRE
EXTINGUISHER
CABINET
(FEC), SEE
DET. 5/AE-700
A D
A
A D
A
S T
A N
D A
R D
C O
O R
D w
" 6
A
D A
S T
A N
D A
R D
" 4
CL
(39"- 41")
40"
(3
CL
MAX
12"
FINISHED WALL
CL
18"
36" MIN.
42"
C O
O R
D
W F
L U
S H
V A
L V
E
FLUSH
CONTROL
8"
24"
M
IN
M
A X
M
IN
24" TOILET TISSUE
DISPENSER
FINISHED
WALL
7"
SANITARY NAPKIN
DISPOSAL
GRAB BAR
GRAB BAR
M A
X
GLOVE
DISPENSER
SPECIMEN
PASS - THRU
CABINET
(SPC)
M
A X
M
IN
MOP BASIN
SERVICE FAUCET
& CONTROLS
MOP BROOM HOLDER
UTILITY SHELF
MOP AND BROOM
HOLDER WITH SHELF
WATER
COOLER
STANDING
A D
A
S T
A N
D A
R D
CLEARANCE
60" MIN.
CLEARANCE
56" MIN.
MOP AND BROOM
HOLDER WITH SHELF
(MBH)
FINISHED
WALL
18"
CL
COMBINATION
TOWEL
DISP. AND
WASTE
RECEPT.
(CTD)
S T
A N
D A
R D
M
A X
A
D A
TOILET
SCHEDULED BASE, TYP.
OPEN TO
BEYOND
4'-4 1/8"
MED-GAS
ZONE VALVE,
REFER TO
PLUMB. DWGS.
METAL
DETECTOR
IRWC-1
IRWC-1
AE-701
AE-701
SCHEDULED BASE, TYP.
OMNICELL (OFOI)
SS-2
PL-12
PL-12
OPEN SHELF
WITH TASK
LIGHT
1'-3" 2'-0" 1'-3"
1'-3" 2'-0" 1'-3"
L L
L
L
L
L
L
L
AE-500
L = LOCKABLE
GYP. BD. SOFFIT
OPENING FOR PATIENT LIFT,
COORDINATE WITH PATIENT
LIFT MANUFACTURER
MED GAS ALARM
PANEL, REFER TO
PLUMB. DWGS.
METAL
DETECTOR, OFCI
MFD (OFOI)
IRWC-1
SCHEDULED BASE, TYP.
SS-2 SS-2
HTD-1
TR-2
TR-1
CT-1
CT-3
TR-3
PT-2/A
CT-1
TR-2
PT-2/A
TR-1
GB-2
GB-3
TTD-1
PT-2/A
CT-1
TR-2
TR-1
GB-1
GB-3
FM-1
SF-1
2'-0"
1'-6" 8"
CT-3 TR-3
TR-1
RF VIEW WINDOW
AE-701
AE-701
SCHEDULED BASE, TYP.END
PANEL
SS-2
SUPPORT BRACKET,
TYP.
AE-500
B
FILTER PANEL CABINET
SCHEDULED BASE, TYP.
4'-4"
SS-2
SUPPORT BRACKET,
TYP.
AE-701
OUTLINE OF IMAGE. OWNER
PROVIDED CONTRACTOR INSTALLED.
COORDINATE WITH COR
PL-12
PL-12
OUTLINE OF
10'-0"W X 8'-0"H
KNOCK OUT PANEL
SCHEDULED
BASE, TYP.
AE-701
AE-701
SCHEDULED BASE, TYP.
SS-2
PL-12
PL-12
PL-12
AE-701
3'-0 1/8" 3'-0" 3'-0" 3'-0" 3'-0" 3'-0"
PL-12PL-12PL-12PL-12PL-12
3 EQ. SPACES
7'-3"
GYP. BD. SOFFIT
AE-701
AE-701
AE-501
2 EQ.
42"
2 EQ.
42"
SCHEDULED BASE, TYP.
SS-2
PL-12
PL-12
IRWC-1
SCHEDULED BASE, TYP.
SS-2
IRWC-1
OPEN TO BEYOND
SCHEDULED BASE, TYP.
TV (OFOI)
IRWC-1
OUTLINE CASE WORK
REMOVED FOR CLARITY
ISOCENTER
OF MAGNET
ISOCENTER
OF MAGNET
IRWC-1
SCHEDULED BASE
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :1
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
INTERIOR ELEVATIONS
AE-800
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-800
3 CONTROL ROOM / CORRIDOR - SOUTH
MOUNTING HEIGHTS LEGEND
SCALE: 1/4" = 1'-0"AE-800
2 CONTROL ROOM - EAST
SCALE: 1/4" = 1'-0"AE-800
1 CONTROL ROOM / CORRIDOR - NORTH
SCALE: 1/4" = 1'-0"AE-800
4 CONTROL ROOM - WEST
SCALE: 1/4" = 1'-0"AE-800
6 TOILET - NORTH
SCALE: 1/4" = 1'-0"AE-800
7 TOILET - EAST
SCALE: 1/4" = 1'-0"AE-800
8 TOILET - SOUTH
SCALE: 1/4" = 1'-0"AE-800
9 TOILET - WEST
SCALE: 1/4" = 1'-0"AE-800
10 MRI - NORTH
SCALE: 1/4" = 1'-0"AE-800
11 MRI - EAST
SCALE: 1/4" = 1'-0"AE-800
12 MRI - SOUTH
SCALE: 1/4" = 1'-0"AE-800
13 MRI - WEST
SCALE: 1/4" = 1'-0"AE-800
16 PREP AREA - WEST
SCALE: 1/4" = 1'-0"AE-800
15 PREP AREA - SOUTH
SCALE: 1/4" = 1'-0"AE-800
14 PREP AREA - EAST
SCALE: 1/4" = 1'-0"AE-800
5 CORRIDOR - EAST
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
SCHEDULED WALL BASE, TYP.
PATIENT LIFT
PATIENT RAIL
SYSTEM
TV (OFOI)
IRWC-1
SCHEDULED BASE, TYP.
PATIENT LIFT
IRWC-1
SCHEDULED WALL BASE, TYP.
IRWC-1
AE-500
SCHEDULED WALL BASE, TYP.
AE-500
IRWC-1
AE-501
SCHEDULED WALL
BASE, TYP.
STOREFRONT
WINDOW SYSTEM
OPEN TO
BEYOND
SCHEDULED
WALL BASE, TYP.
AE-500
EQ EQ
CL
LINEAR WALL DIFFUSER,
REFER TO MECH. DWGS.
AE-501
SCHEDULED WALL BASE, TYP.
3'-10" CL
TV (OFOI)
LINEAR WALL DIFFUSER,
REFER TO MECH. DWGS.
(2)
LOCKER
UNITS
(L-1)
SCHEDULED BASE, TYP.
SPANDREL PANEL
BEHIND
AE-701
AE-701
2 EQ.
42"
2 EQ.
42"
SCHEDULED BASE, TYP.
SS-2
PL-12
PL-12
GYP. BD. SOFFIT
(2)
LOCKER
UNITS
(L-1)
SCHEDULED BASE, TYP.
SPANDREL PANEL
BEHIND
AE-701
AE-701
2 EQ.
42"
2 EQ.
42"
SCHEDULE BASE, TYP.
SS-2
PL-12
PL-12
GYP. BD. SOFFIT
STOREFRONT
WINDOW SYSTEM
MANUFACTURED
STONE VENEER
WALL SYSTEM
SCHEDULED BASE, TYP.
TV (OFOI)
IRWC-1
SCHEDULED BASE, TYP.
TV (OFOI)
IRWC-1
AE-501
OPEN TO
BEYOND
IRWC-1
SCHEDULED BASE
OUTLINE OF NEW
MANUFACTURED STONE
VENEER WAL SYSTEM
EXISTING MANUFACTURED
STONE SYSTEM
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :2
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
INTERIOR ELEVATIONS
AE-801
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-801
1 SUB WAITING - EAST
SCALE: 1/4" = 1'-0"AE-801
2 SUB WAITING - SOUTH
SCALE: 1/4" = 1'-0"AE-801
3 SUB WAITING - WEST
SCALE: 1/4" = 1'-0"AE-801
4 SUB WAITING - NORTH
SCALE: 1/4" = 1'-0"AE-801
5 LOBBY/WAITING - EAST
SCALE: 1/4" = 1'-0"AE-801
6 LOBBY/WAITING - SOUTH
SCALE: 1/4" = 1'-0"AE-801
7 LOBBY/WAITING - WEST
SCALE: 1/4" = 1'-0"AE-801
11 DRESSING 107 - EAST
SCALE: 1/4" = 1'-0"AE-801
12 DRESSING 107 - WEST
SCALE: 1/4" = 1'-0"AE-801
9 DRESSING 106 - EAST
SCALE: 1/4" = 1'-0"AE-801
10 DRESSING 106 - WEST
SCALE: 1/4" = 1'-0"AE-801
15 VESTIBULE - WEST
SCALE: 1/4" = 1'-0"AE-801
8 DRESSING 106 - SOUTH
SCALE: 1/4" = 1'-0"AE-801
13 DRESSING 107 - NORTH
SCALE: 1/4" = 1'-0"AE-801
14 ENTRANCE - EAST
SCALE: 1/4" = 1'-0"AE-801
16 VESTIBULE - NORTH
Revisions Date
95% CD SUBMISSION 10/07/2019
P.5
O.7
6.25.2
P.81
9.78.87.96.4 6.9
P.9
Q.4
AE-500
AE-505
10' - 0"
AE-505
SIM.
AE-700
AE-700
7.3
13' - 0"
AE-501
AE-500
AE-510
AE-501
8'-6"
CL
ISOCENTER OF MAGNET
CL
EQ EQ
2'-6"
5'-0" 2'-0"
.0
30.00°
15.00°
.0
13'-5" 8'-10"
/8
ELLIPTICAL DISPLAY
LIGHTING8'-0"
QUENCH VENT; COORDINATE
WITH MRI VENDOR AND
MECHANICAL DAWINGS
"9' - 0"
9' - 0"
9' - 0"
9' - 0"
9' - 0"
9' - 0"
OPEN TO
ABOVE
2'-0"
E Q
E Q
EQ EQ
EQ EQ
E Q
E Q
EQ EQ
6'-6" 6'-6" 6'-6" 6'-6" 6'-6" 6'-6"
/4
E Q
E Q
EQ EQ
1'-10"
MATCH CELING HEIGHT
AND DIRECTION
EXISTING CEILING
SYSTEM TO REMAIN
MATCH CELING HEIGHT
AND DIRECTION
PATIENT LIFT RAIL
SYSTEM
/8
/8
4'-2" 4'-1"
2'-1" 4'-2"
R 25'-0"
14' - 0"
EXISTING CEILING
SYSTEM TO REMAIN
6'-4 1/2" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8"
EXPOSED TO
STRUCTURE ABOVE
E Q
E Q
7' - 2" 7' - 2"
7' - 2"
7' - 2"
7' - 2"
9' - 0"
8' - 6"
EXISTING SOFFIT TO
REMAIN
EXISTING SOFFIT TO
REMAIN
OPENING FOR
FILTER CABINET
5'-0"
/8
/8
/8
6'-5 3/4" 8'-3" 20'-3" 22'-3" 28'-8"
4'-5 1/8"
EQ EQ
E Q
E Q
AE-500
E Q
'A
E Q
'A
ALIGN CEILING GRID
" 9'-1" 9'-1"
14' - 0"
13' - 0"
13' - 0"
CUBICLE CURTAIN TRACK
RETURN AIR DIFFUSER; REFER TO MECH. DWGS.
NEW ACOUSTICAL LAY-IN CEILING
2x4 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
RECESSED LED DOWN LIGHT;
REFER TO ELECT. DWGS.
SUSPENDED DIRECT/INDIRECT LINEAR
FIXTURE; REFER TO ELECT. DWGS.
GYPSUM BOARD CEILING OR SOFFIT
SUPPLY AIR DIFFUSER; REFER TO MECH. DWGS.
EXISTING ACOUSTICAL LAY-IN CEILING
2x2 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
CEILING HEIGHT SYMBOL
EXIT LIGHT; REFER TO ELECT. DWGS.
7'-10"
LINEAR DIFFUSER
1.
2.
3.
4.
5.
6.
7.
8.
CENTERLINES OF ROOM SHALL BE STARTING POINT OF CEILING GRID LAYOUT,
UNLESS OTHERWISE NOTED.
SPRINKLER HEADS, INCANDESCENT DOWNLIGHTS, DIFFUSERS, ETC. SHALL BE
CENTERED IN CEILING PANELS IN BOTH DIRECTIONS, UNLESS OTHERWISE
INDICATED.
DIFFUSERS AND RETURN GRILLES NOT SHOWN IN ROOMS WITHOUT CEILINGS.
ALL CEILING HEIGHTS SHALL BE 9'-0" A.F.F. UNLESS OTHERWISE NOTED.
REFER TO ELECTRICAL DRAWINGS FOR LOCATION OF EXIT LIGHTS.
ALL SOFFITS AND VERTICAL SURFACES CONNECTING CEILINGS AND/OR
SOFFITS AT DIFFERENT ELEVATIONS SHALL BE CONSTRUCTED OF METAL
STUDS/FURRING AND GYPSUM WALLBOARD, UNLESS OTHERWISE INDICATED.
P&M DENOTES PATCH TO MATCH EXISTING CONSTRUCTION AND FINISHES. ALL
PATCHING OF EXISTING MATERIAL SHALL BE DONE WITH MATERIALS AND
WORKMANSHIP MATCHING ADJACENT SURFACES.
NO ATTEMPT HAS BEEN MADE TO INDICATE ALL AREAS REQUIRING CUTTING
AND PATCHING DUE TO MECHANICAL AND/OR ELECTRICAL WORK.
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :2
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
REFLECTED CEILING PLANS
AE-900
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-900
1 FIRST FLOOR REFLECTED CEILING PLAN
REFLECTED CEILING PLAN LEGEND
REFLECTED CEILING PLAN GENERAL NOTES
Revisions Date
65% DD SUBMISSION 07/29/2019
95% CD SUBMISSION 10/07/2019
O.7
6.25.2
P.81
9.78.87.96.4 6.9
P.9
Q.4
AE-500
AE-505
AE-505
SIM.
7.3
AE-501
AE-500
AE-501
10' - 0"
AE-700
AE-700
13' - 0"
8'-6"
CL
ISOCENTER OF MAGNET
CL
EQ EQ
2'-6"
5'-0" 2'-0"
.0
30.00°
15.00°
.0
13'-5" 8'-10"
/8
ELLIPTICAL DISPLAY
LIGHTING8'-0"
QUENCH VENT; COORDINATE
WITH MRI VENDOR AND
MECHANICAL DAWINGS
"9' - 0"
9' - 0"
9' - 0"
9' - 0"
9' - 0"
9' - 0"
OPEN TO
ABOVE
2'-0"
E Q
E Q
EQ EQ
EQ EQ
E Q
EQ EQ
6'-6" 6'-6" 6'-6" 6'-6" 6'-6" 6'-6"
/4
E Q
E Q
EQ EQ
1'-10"
MATCH CELING HEIGHT
AND DIRECTION
EXISTING CEILING
SYSTEM TO REMAIN
MATCH CELING HEIGHT
AND DIRECTION
/8
/8
4'-2" 4'-1"
2'-1" 4'-2"
R 25'-0"
14' - 0"
EXISTING CEILING
SYSTEM TO REMAIN
6'-4 1/2" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8"
EXPOSED TO
STRUCTURE ABOVE
E Q
E Q
7' - 2"
7' - 2"
7' - 2"
7' - 2"
7' - 2"
9' - 0"
8' - 6"
EXISTING SOFFIT TO
REMAIN
EXISTING SOFFIT TO
REMAIN
ALIGN CEILING GRID
OPENING FOR
FILTER CABINET
5'-0"
/8
/8
/8
6'-5 3/4" 8'-3" 20'-3" 22'-3" 28'-8"
4'-5 1/8"
EQ EQ
E Q
AE-500
13' - 0"
13' - 0"
CUBICLE CURTAIN TRACK
RETURN AIR DIFFUSER; REFER TO MECH. DWGS.
NEW ACOUSTICAL LAY-IN CEILING
2x4 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
RECESSED LED DOWN LIGHT;
REFER TO ELECT. DWGS.
SUSPENDED DIRECT/INDIRECT LINEAR
FIXTURE; REFER TO ELECT. DWGS.
GYPSUM BOARD CEILING OR SOFFIT
SUPPLY AIR DIFFUSER; REFER TO MECH. DWGS.
EXISTING ACOUSTICAL LAY-IN CEILING
2x2 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
CEILING HEIGHT SYMBOL
EXIT LIGHT; REFER TO ELECT. DWGS.
7'-10"
LINEAR DIFFUSER
1. SEE SHEET AE-900 FOR GENERAL REFLECTED CEILING PLAN NOTES
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :2
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
REFLECTED CEILING PLAN - DEDUCT
ALTERNATE #1
AE-910
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-910
1 FIRST FLOOR REFLECTED CEILING PLAN DEDUCT #1
REFLECTED CEILING PLAN LEGEND
REFLECTED CEILING PLAN GENERAL NOTES
Revisions Date
R
O.7
6.25.2
P.81
9.78.87.96.4 6.9
P.9
Q.4
AE-500
AE-505
AE-505
SIM.
7.3
AE-501
AE-500
AE-510
AE-501
10' - 0"
AE-700
AE-700
13' - 0"
E Q
E Q
8'-6"
CL
ISOCENTER OF MAGNET
CL
EQ EQ
/8
PATIENT LIFT
SYSTEM
DEDUCT ALTERNATE #3 (SHORT CANOPY)
EXISTING CEILING
SYSTEM TO REMAIN
MATCH CELING HEIGHT
AND DIRECTION
EXISTING CEILING
SYSTEM TO REMAIN
MATCH CELING HEIGHT
AND DIRECTION
/8
/8
/8
6'-5 3/4" 8'-3" 20'-3" 22'-3" 28'-8"
AE-500
ALIGN CEILING GRID
7' - 2"
7' - 2" 7' - 2"
9' - 0" 9' - 0" 9' - 0"
13' - 0"
10' - 0"
9' - 0"
9' - 0"
9' - 0"
6'-4 1/2" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8" 6'-5 1/8"
14' - 0"
13' - 0"
CUBICLE CURTAIN TRACK
RETURN AIR DIFFUSER; REFER TO MECH. DWGS.
NEW ACOUSTICAL LAY-IN CEILING
2x4 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
RECESSED LED DOWN LIGHT;
REFER TO ELECT. DWGS.
SUSPENDED DIRECT/INDIRECT LINEAR
FIXTURE; REFER TO ELECT. DWGS.
GYPSUM BOARD CEILING OR SOFFIT
SUPPLY AIR DIFFUSER; REFER TO MECH. DWGS.
EXISTING ACOUSTICAL LAY-IN CEILING
2x2 RECESSED LED FIXTURE; REFER
TO ELECT. DWGS.
CEILING HEIGHT SYMBOL
EXIT LIGHT; REFER TO ELECT. DWGS.
7'-10"
LINEAR DIFFUSER
1. SEE SHEET AE-900 FOR GENERAL REFLECTED CEILING PLAN NOTES
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :2
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
REFLECTED CEILING PLAN - DEDUCT
ALTERNATE #3
AE-911
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
SCALE: 1/4" = 1'-0"AE-911
1 FIRST FLOOR REFLECTED CEILING PLAN DEDUCT #3
REFLECTED CEILING PLAN LEGEND
REFLECTED CEILING PLAN GENERAL NOTES
Revisions Date
95% CD SUBMISSION 10/07/2019
FLOOR
CPT CARPET TILE
CT CERAMIC TILE
LVT LUXURY VINLY TILE
WSF WELDED VINYL SHEET FLOORING
ROOM FINISH LEGEND
WALL
P PAINT
WP IMPACT RESISTANT WALLCOVERING
BASE
RB RUBBER COVE BASE
WSF WELDED SHEET VINYL, INTEGRAL COVE BASE
CEILINGS
ACT ACOUSTICAL CEILING TILE
P PAINT
GENERAL NOTES
1. SEE REFLECTED CEILING PLAN FOR CEILING HEIGHTS.
2. SEE DRAWINGS FOR LOCATION AND ELEVATIONS OF
ARCHITECTURAL WOODWORK.
3. GYPSUM BOARD SOFFITS: UNLESS NOTED, HORIZONTAL SURFACES
SHALL BE SAME AS ANY CONNECTING WALL SURFACE. SEE
REFLECTED CEILING PLAN FOR SOFFIT HEIGHT.
4. PAINT ALL NEW METAL DOORS AND DOOR FRAMES (INCLUDING
VISION PANELS IN DOORS) UNLESS NOTED. SEE PAINT
SPECIFICATIONS FOR FINISH.
5. P&M DENOTES PATCH AND MATCH TO EXISTING ADJACENT
MATERIALS, FINISHES AND COLORS
SUBSTRATE
C CONCRETE
CBB CEMENTITIOUS BACKER BOARD
GWB GYPSUM BOARD
MRGB MOISTURE RESISTANT GYPSUM BOARD
ARCHITECTURAL WOODWORK
PL PLASTIC LAMINATE
SS SOLID SURFACE MATERIAL
SCHEDULE NOTES
1. SEE DRAWINGS FOR LOCATION OF CORNER GUARDS, CRASH RAILS,
AND BUMPER GUARDS.
2. REFER TO FLOOR PATTERN PLANS FOR FLOOR PATTERN
INFORMATION ON DRAWINGS.
3. SEE DRAWINGS FOR WALL ELEVATIONS
EXISTING OR
SCHEDULED
PARTITION
BLOCKING AS
REQUIRED
CONTINUOUS
ALUMINUM
RETAINER
VINYL CRASH RAIL
EDGE OF WALL,
DOOR FRAME, OR
CORNER GUARD
ELEVATION SECTION
2"
S E
E E
L E
V A
T
IO
N S
V A
R
IE
SATTACH PER
MANUFACTURER'S
STANDARDS
METAL STUD
FRAMING, SEE
PARTITION TYPE
VINYL EXTRUSION
BACK-UP SHEET
VINYL ACRYLIC
CORNER GUARD
CONT. ALUM.
RETAINER2
2"
ATTACH PER
MANUFACTURER'S
STANDARDS
METAL STUD
FRAMING, SEE
PARTITION TYPE
VINYL ACRYLIC
CORNER GUARD
3"
" CONT. ALUM.
RETAINER
PARTITION TYPE 'C'
NON-RATED SMOKE PARTITION
(1) LAYER OF 5/8" GYP. BD. ON EACH SIDE
OF 3 5/8" METAL STUDS @ 16" O.C.
PARTITION TYPE 'D'
NON-RATED
(1) LAYER OF 5/8" GYP. BD. ON EACH
SIDE OF 3 5/8" METAL STUD @ 16"O.C.
PARTITION AT FLOOR PARTITION AT FLOOR
PARTITION AT CEILING
STRUCTURE ABOVE
DEFLECTION JOINT
CONTINUOUS
RUNNER TRACK
CONTINUOUS
ACOUSTICAL
SEALANT
CEILING SYSTEM
CONT. RUNNER
TRACK
CORRIDOR SIDE
INFILL VOID AREAS
W/ SOUND
BLANKET
CONTINUOUS
ACOUSTICAL
SEALANT
CONTINUOUS
RUNNER TRACK
CEILING
STRAP BRIDGING
STRUCTURE ABOVE
EXTEND GYP.
BD. 6" ABOVE
CEILING
DEFLECTION JOINT
CONT. RUNNER TRACK
STRUCTURE ABOVE
SMOKESTOP ALL VOID
AREAS TIGHT TO
UNDERSIDE OF
STRUCTURE W/ SAFING
14 GA. STL PLATE
AS REQ.
EXISTING CONSTR.
OR NEW MASONRY
CONSTR. (SEE PLANS
FOR SPECIFIC
CONDITIONS)
SOUND ATTENUATION
BLANKET
HEAD OF PARTITION
METAL DECK PERPENDICULAR TO FLUTES
HEAD OF PARTITION
METAL DECK PARALLEL TO FLUTES
HEAD OF PARTITION
CONCRETE SLAB
HEAD OF PARTITION
METAL DECK PERPENDICULAR TO FLUTES
HEAD OF PARTITION
METAL DECK PARALLEL TO FLUTES
4 7/8" 4 7/8"
3 1/2" SOUND
ATTENUATION
BLANKET
CONTINUOUS
ACOUSTICAL
SEALANT EACH SIDE
STRUCTURE ABOVE
3 1/2" SOUND
ATTENUATION
BLANKET
CONTINUOUS
RUNNER TRACK
CONTINUOUS
RUNNER
TRACK DEFLECTION JOINT
STRUCTURE ABOVE
CONTINUOUS
RUNNER TRACK
DEFLECTION JOINT
3 1/2" SOUND
ATTENUATION
BLANKET
DEFLECTION JOINT
14 GA. STEEL
PLATE AS
REQUIRED
CONTINUOUS
ACOUSTICAL SEALANT
EACH SIDE
STRUCTURE ABOVE
CONTINUOUS
RUNNER TRACK
3 1/2" SOUND
ATTENUATION
BLANKET
CONTINUOUS
RUNNER TRACK
3 1/2" SOUND
ATTENUATION
BLANKET
EXTEND GYP. BD. 6"
ABOVE CEILING, THIS
SIDE
3 1/2" SOUND
ATTENUATION
BLANKET
CONTINUOUS
ACOUSTICAL SEALANT
EACH SIDE
3 1/2" SOUND
ATTENUATION
BLANKET
3 1/2" SOUND
ATTENUATION
BLANKET
4 1/4"
EXISTING CONSTR.
OR NEW MASONRY
CONSTR. (SEE PLANS
FOR SPECIFIC
CONDITIONS)
PARTITION TYPE 'G'
NON-RATED
(1) LAYER OF 5/8" GYP. BD. ON ONE SIDE OF
3 5/8" METAL STUD FURRING @ 16" O.C.
PARTITION AT FLOOR
PARTITION AT CEILING
HEAD OF PARTITION
4 7/8"
PARTITION TYPE 'A'
ONE HOUR RATED (U.L. #U419)
(1) LAYER OF 5/8" FIRE RESISTIVE GYP. BD. ON
EACH SIDE OF 3 5/8" METAL STUDS @ 16" O.C.
HEAD OF PARTITION
METAL DECK PERPENDICULAR TO FLUTES
HEAD OF PARTITION
METAL DECK PARALLEL TO FLUTES
PARTITION AT FLOOR
HEAD OF PARTITION
CONCRETE SLAB
STRUCTURE ABOVE
CONTINUOUS
RUNNER TRACK
FIRESTOP JOINT
BETWEEN TOP OF FIRE-
RATED PARTITION AND
STRUCTURE ABOVE W/
U.L. OR W.H.I. PPROVED
(1) HR. JOINT FIRESTOP
SYSTEM.
SOUND
ATTENUATION
BLANKET
CONTINUOUS
ACOUSTICAL SEALANT
EACH SIDE
CONTINUOUS
ACOUSTICAL
SEALANT
DEFLECTION JOINT
CONTINUOUS
RUNNER TRACK
STRUCTURE ABOVE
STRUCTURE ABOVE
DEFLECTION JOINT
DEFLECTION JOINT
SOUND
ATTENUATION
BLANKET
SOUND
ATTENUATION
BLANKET
CONTINUOUS
RUNNER TRACK
CONTINUOUS RUNNER
TRACK
FIRESTOP JOINT
BETWEEN TOP OF FIRE-
RATED PARTITION AND
STRUCTURE ABOVE W/
U.L. OR W.H.I. PPROVED
(1) HR. JOINT FIRESTOP
SYSTEM.
EXISTING CEILING
EXISTING FLOOR
(SLAB)
METAL TRACK
FOR 3 5/8" STUD
OR 2X TOP PLATE
METAL RUNNER
FOR 3 5/8" METAL
STUD OR 2X SILL
PLATE
5/8" EXTERIOR
GRADE PLYWOOD
OR 5/8" GYP. BD.
WITH LEVEL 4 FINISH
PUBLIC SIDE CONSTRUCTION SIDE
MAX GAP OF 1/4"
BETWEEN EXISTING
CEILING AND TOP OF
WALL
TEMPORARY CONSTRUCTION
BARRIER DETAIL
EXISTING CEILING
EXISTING FLOOR (SLAB)
METAL TRACK FOR 8"
STUD OR 2X TOP PLATE
METAL RUNNER FOR 8" METAL
STUD OR 2X SILL PLATE
5/8" GYP. BD. WITH
LEVEL 4 FINISH
PUBLIC SIDE CONSTRUCTION SIDE
MAX GAP OF 1/4"
BETWEEN EXISTING
CEILING AND TOP OF
WALL
SPRAY APPLIED VAPOR
BARRIER ON EXTERIOR
GRADE PLYWOOD
TEMPORARY CONSTRUCTION BARRIER
WHERE EXPOSED TO ELEMENTS
4 7/8"
TEMPORARY
CONSTRUCTION BARRIER
HEAD OF PARTITION
METAL DECK PARALLEL TO FLUTES
PARTITION AT FLOOR
SOUND ATTENUATION
BLANKET CONTINUOUS ACOUSTICAL
SEALANT EACH SIDE
STRUCTURE ABOVE
DEFLECTION JOINT
SOUND ATTENUATION
BLANKET
CONTINUOUS RUNNER TRACK
CONTINUOUS RUNNER TRACK
FIRESTOP JOINT BETWEEN TOP
OF FIRE-RATED PARTITION AND
STRUCTURE ABOVE W/ U.L. OR
W.H.I. PPROVED (1) HR. JOINT
FIRESTOP SYSTEM.
ONE HOUR RATED (U.L. #U419)
(1) LAYER OF 5/8" FIRE RESISTIVE GYP. BD. ON
EACH SIDE OF 3 5/8" METAL STUDS @ 16" O.C.
PUBLIC SIDE CONSTRUCTION SIDE
5/8" GYP. BD. WITH
LEVEL 4 FINISH
MICHAEL DWAYNE
LIEZERT
S
TATE OF OHIO
R E
G
ISTERED ARCHIT
E C
T9939
Project Number
Office of Construction and Facilities Management
Building Number
Drawing Number
VA FORM 08-6231
CONSULTANTS: ARCHITECTS/ENGINEERS: ARCHITECTURAL PROJECT #
VA Building th re e in c h e s o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t o n e in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t o n e e ig h th in c h o n e fo o
F
E
D
C
B
A
F
E
D
C
B
A
987654321
987654321
Project Title
Location
Date Checked Drawn
Drawing Title
Approved: Project Director
7850 Freeway Circle
440.243.2000 t
Cleveland, OH 44130
440.243.3305 f
KEY PLAN
Dwg. of
FULLY SPRINKLERED
PHILIP P. KIRK
CHIEF ENGINEER
:2 :2
A M
C U s e rs jc h o D o c u m e n ts R
N E W
M R I S
IT
E
P R E P C e n tr a l_ jc h o .r v t
PARTITION TYPES, FINISH SCHEDULE,
LEGEND AND DETAILS
AE-1000
BH Designer
552-CSI-176
NEW MRI SITE PREP
02/21/2020
4100 WEST THIRD STREET, DAYTON, OHIO 45428
18106
BID & CONSTRUCTION
ROOM FINISH SCHEDULE
ROOM FLOOR BASE WALLS CEILING ARCH.
WOODWORK NOTESNUMBER NAME SUBSTR. FINISH SUBSTR. FINISH SUBSTR. NORTH SOUTH EAST WEST SUBSTR. FINISH
A100 VESTIBULE C CPT-8 GWB RB-1 GWB - P-2/A - P-2/A* MRGB PT-32/A - *WITHSTONE VENEER
A101 LOBBY/WAITING C LVT-3/LVT-8* GWB RB-1 GWB - P-16/A P-2/A P-2/A - ACT-1 - REWORK EXISTING FLOORING AND CELING TILE WITH NEW TILE AND PATTERN
A102 ENTRANCE HALL C LVT-8 GWB RB-1 GWB - P-2/A P-2/A P-2/A - ACT-1 - *SEE FINISH FLOOR PLAN
A103 ENTRANCE C LVT-8 GWB RB-1 GWB P-2/A - P-2/A P-16/A - ACT-1 -
A104 PREP AREA C LVT-8 GWB RB-1 GWB P-2/A P-2/A* P-2/A P-16/A - ACT-1 SS-2/PL12 *WITH/ IRWC1A WAINSCOT
A105 SUB WAITING C LVT-3/LVT-8* GWB RB-1 GWB P-2/A P-16/A P-2/A P-16/A - ACT-1 - *SEE FINISH FLOOR PLAN
A106 DRESSING C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-16/A P-2/A - ACT-1 SS-2/PL12 WINDOW SILL: SS-2
A107 DRESSING C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-16/A P-2/A - ACT-1 SS-2/PL12 WINDOW SILL: SS-2
A108 OFFICE C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-16/A P-2/A - ACT-1 - WINDOW SILL: SS-2
A109 CORRIDOR C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-2/A - - ACT-1 -
A110 EQUIPMENT C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-2/A P-2/A - ACT-1 -
A111 MRI ROOM C WSF-10 GWB WSF-10 GWB P-2/A P-16/A P-2/A P-16/A - ACT-1 SS-2/PL12
A112 CONTROL ROOM C LVT-8 GWB RB-1 GWB P-2/A P-2/A P-2/A P-16/A - ACT-1 SS-2/PL12
A113 TOILET C CT-1 CBB -- CBB *CT-1 *CT-1 *CT-1 ct-3 - ACT-1 - *WITH PT2/A ABOVE
E113 EX. CORRIDOR C LVT-8 GWB RB-1 ETR P-2/A P-2/A P-2/A P-2/A - GWB* - *PT2/A: SOFFITS
E131 EX.
This is the start of the file's text. The full file is on GovTribe.