This is a Performance Work Statement (PWS) for Diagnostic Medical Physics Services supporting the Veterans Health Administration (VHA) at multiple facilities in North Florida and South Georgia.
The contractor must provide comprehensive diagnostic medical physics support at seven VA locations: Malcom Randall VA Medical Center (Gainesville), Gainesville Primary Care VA Clinic, Lake City VA Medical Center, Jacksonville OPC, Jacksonville North OPC, Tallahassee OPC, and The Villages OPC. Key service requirements include annual quality assurance testing and certification of all diagnostic x-ray and imaging equipment (x-ray, nuclear medicine cameras, PET/CT cameras, ultrasound units, and MRIs) in compliance with American College of Radiology (ACR) and Mammography Quality Standards Act (MQSA) requirements. The contractor must provide written reports signed by a diagnostic medical physicist within five working days of inspection completion, including testing methods, instruments used, measurements, and identified deficiencies. Additional requirements include annual review of the Diagnostic Imaging Service's Quality Assurance program, equipment recertification following repairs or modifications, dose rate surveys in unrestricted areas, shielding design approvals and surveys for new construction, consultation on dose estimate calculations, and emergency response within 24 hours. Services are provided during normal business hours (Monday-Friday, excluding federal holidays) with the government providing working space.
Personnel must be certified by the American Board of Radiology, American Board of Medical Physics, or Canadian College of Physicists in Medicine, or meet alternative qualification standards including a graduate degree in physics or related discipline, formal graduate-level coursework in biological sciences, and three years of documented experience in the modality being inspected. The Quality Assurance Surveillance Plan establishes performance standards with Acceptable Quality Levels ranging from 98–99 percent, monitored through random inspections and customer feedback, with Contract Discrepancy Reports issued for non-compliance.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Updated NFSG Radiology Physics Equipment List 2026.pdf | ||
| 36C24827Q0077 0001.docx | DOCX document | |
| 36C24827Q0077.docx | DOCX document | |
| NFSG Radiology Physics Equipment List.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Performance Work Statement
Diagnostic Medical Physics Services
1.0. Introduction.
1.1. The Government requires Diagnostic Medical Physics Services in support of the Malcom Randall VA Medical Center, Gainesville Primary Care VA Clinic, Lake City Medical Center, Jacksonville OPC, Jacksonville North OPC, Tallahassee OPC and the Villages OPC, Diagnostic Imaging Service (DIS).
2.0. Description / Scope / Objective.
2.1 The Contractor must provide all labor, material, supplies, equipment, and qualified personnel to provide on-site diagnostic medical physics support or services for the Veterans Health Administration (VHA) as defined in this Performance Work Statement.
3.0 Applicable Documents.
3.1. The following laws, regulations, policies, and procedures in effect on date of contract issuance and all subsequent changes or updates apply:
FAR Federal Acquisition Regulation VAAR Veterans Administration Acquisition Regulation JCAHO Joint Commission on Accreditation of Healthcare Organizations
Requirments VHA Handbook VHA Directive 1105.01, Management of Radioactive Materials VHA Handbook VHA Handbook 1105.02, Nuclear Medicine and Radiation Safety
Service VHA Handbook VHA Handbook 1105.03, Mammography Program Procedures and
Standards VHA Handbook VHA Handbook 1105.04, Fluoroscopy Safety VHA Handbook VHA Handbook 1105.05, MRI Safety ACR American College of Radiology Guidance
3.2. Example: Glossary. Acronyms used in this PWS are listed below for easy reference:
COR Contracting Officer’s Representative FAR Federal Acquisition Regulation VHA Veterans Health Administration NF/SG North Florida/South Georgia Veteran Healthcare Systems RSO Radiation Safety Officer DIS Diagnostic Imaging Service
4.0. Performance Requirements.
4.1. The Contractor shall provide imaging equipment (x-ray equipment, nuclear medicine cameras, PET/CT cameras, ultrasound units, and MRIs) inspections to ensure compliance with the current American College of Radiology (ACR) and Mammography Quality Standards Act (MQSA) requirements. All work shall be performed in accordance with any and all Federal, State, Local, and Industry: laws, codes, standards, procedures, and/or regulations for the proper intended performance for
Diagnostic Medical Physics services. The primary objective is to provide services and deliverables through performance of physics services that may be needed for Government imaging equipment.
4.2. The contractor shall provide annual review of Diagnostic Imaging Service (DIS) overall Imaging Quality Assurance program for diagnostic imaging equipment.
4.3. The contractor shall provide annual quality assurance testing survey and certification of all diagnostic x- ray and imaging equipment.
4.4. The contractor shall provide recertification of equipment for performance when required due to repair or modification, i.e., x-ray tube replacement, upgrades, etc.
4.5. The contractor shall provide written reports signed by the diagnostic medical physicist documenting the testing, including a summary of methods, instruments used, measurements and deficiencies identified, for all diagnostic x-ray and imaging equipment within five (5) working days after completion of the inspections. The Contractor shall provide two (2) written copies of the physicist report to the RSO and the Chief of Biomedical Services within five days of completion of the work. The report may be in electronic format.
4.6. The contractor shall report deficiencies or non-conformances which represent unsafe to the RSO immediately. Verbal reports must be provided after testing so that DIS will know that equipment can be immediately put into use.
4.7. The contractor shall provide dose rate surveys in unrestricted areas adjacent to the radiation areas are to be included during testing and in the final report.
4.8. The contractor shall provide shielding design approvals for new construction and renovations.
4.9. The contractor shall provide shielding surveys for new construction, modifications and replacement of equipment.
4.10 The contractor shall annual quality assurance testing of mammography equipment.
4.11. The contractor shall provide consultation and final reports for dose estimates calculations for patient, staff and fetal exposures and sentinel event investigations.
4.12. The contractor shall provide emergency response time on equipment within 24 hours.
4.13. The contractor shall provide Quality Assurance Testing, all Quality Assurance Tests shall meet or exceed current and possible future changes in NRC, ACR, MQSA, NCRP, FDA and TJC requirements and recommendations. All Quality Assurance Tests shall meet or exceed the requirements in the following VHA directives and handbooks and future directives and handbooks.
5.0. Performance Location and Hours.
5.1. Diagnostic Medical Physics Services are to be provided at imaging facilities throughout North Florida/South Veterans Health System. The Contractor shall provide necessary audits, reviews, spot surveys and equipment testing in all areas where diagnostic imaging equipment is located including, but not limited to:
- Malcom Randall VA Medical Center; 1601 S.W. Archer Road; Gainesville, FL 32608-1197
- Gainesville Primary Care VA Clinic; 5469 S.W. 34th St.; Gainesville, FL 32608
- Lake City VA Medical Center; 619 S. Marion Ave.; Lake City, FL 32025-5808
- Jacksonville OPC; 1536 North Jefferson St.; Jacksonville, FL 32209
- Jacksonville North OPC; 145 Heron Bay Rd.; Jacksonville, FL 32218-2670
- Tallahassee OPC; 2181 Orange Avenue; Tallahassee, FL 32311
- The Villages OPC; 8900 SE 165th Mulberry Ln.; The Villages, FL 32162
5.1.2. Working space will be provided by the Government.
5.2. Regularly scheduled work shall be performed during normal business hours, Monday-Friday, excluding Federal holidays.
6.0. Deliverables.
6.1. Annually evaluation reports to the Government.
6.2. Satisfaction surveys and reports follow-up with the Government.
7.0. Contractor Personnel and Management.
7.1. Contractor Personnel. The Contractor shall be responsible for employing technically qualified personnel to perform the work specified in this PWS. The Contractor shall maintain the personnel, organization and administrative control necessary to ensure that the work delivered meets the contract specifications and requirements.
7.2. Government Badges. Not required. VA Visitor badges will be used in conjunction with Contractor-supplied badges when on VA property.
7.3. Contractor Badges. As required by Contractor
7.4. All Contractor personnel attending meetings, answering Government telephones, and working in other situations where their Contractor status is not obvious to third parties are required to identify themselves as such to avoid creating an impression in the minds of members of the public or Congress that they are Government officials. They must also ensure that all documents or reports produced by Contractor personnel are suitably marked as Contractor products or that Contractor participation is appropriately disclosed.
7.5. Safety/Security Requirements. The Contractor shall comply with all applicable federal, state and local laws and ensure performance is secure while protecting material, equipment, and supplies from damage and loss. Government security personnel shall have the express right to inspect for security violations at any time during the term of the contract. The C&A requirements do not apply; A Security Accreditation Package is not required.
7.6. All information generated in performance of this contract becomes property of the Government and must be surrendered upon request and/or completion of the contract. The Contractor must protect against the destruction and/or misplacement of all information produced under this contract, unless directed under local policy to destroy. This includes both hard and soft copies of data.
7.7. The Contractor shall not release any information (including photographs, files, public announcements, statements, denials, or confirmations) on any part of the subject matter of this contract or any phase of any program hereunder without the prior written approval of the COR.
7.7.1. Contractor’s patient’s records will be maintained in accordance with all applicable regulations.
7.8. Standards of conduct
7.8.1. Conduct of Personnel. Contractor personnel shall conduct themselves in a professional manner (i.e.
timeliness, communication; spoken and written etc.). The Contracting Officer may require the Contractor to remove from the job site any employee working under this contract for reasons of suspected misconduct, a suspected security breach, or suspected to be under the influence of alcohol, drugs, or any other incapacitating agent. Contractor employees shall be subject to dismissal from the premises upon determination by the Contracting Officer that such action is necessary in the interests of the Government.
The hospital director or designated representative has the authority to bar individuals from the Government facility. The removal from the job site or dismissal from the premises shall not relieve the Contractor of the requirement to provide sufficient personnel to perform the services as required by this performance work statement. The Government shall not reimburse the Contractor for travel and other expenses associated with the removal of personnel.
7.8.2. Contractor personnel are not authorized to carry or possess personal weapons to include, but not limited to, firearms and knives with a blade length in excess of three inches, while assigned under this contract when working in a Government Facility,.
7.8.3. Working Attire and Appearance. Contractor personnel shall present a professional appearance commensurate with standards delineated for Government civilian personnel acting in similar capacities.
7.8.4 The Government reserves the right to review Contractor personnel qualifications. Upon request, resumes shall be provided to the Contracting Officer prior to clearance approval, for review and acceptance by the COR. The intent is verification purpose and not for the Government to become the hiring authority.
8. TRAVEL: NOT APPLICABLE
9. NON-PERSONAL SERVICE STATEMENT
9.1. Contractor employees performing services under this order will be controlled, directed, and supervised at all times by management personnel of the contractor. Contractor management will ensure that employees properly comply with the performance work standards outlined in the PWS. Contractor employees will perform their duties independent of, and without the supervision of, any Government official. The tasks, duties, and responsibilities set forth in the contract may not be interpreted or implemented in any manner that results in any contractor employee creating or modifying Federal policy, obligating the appropriated funds of the United States Government, overseeing the work of Federal employees, providing direct personal services to any Federal employee, or otherwise violating the prohibitions set forth in Parts 7.5 and 37.1 of the Federal Acquisition Regulation (FAR).
10. PERSONNEL QUALIFICATIONS
10.1. The Contractor shall be responsible for employing technically qualified personnel to perform the work specified in this PWS. The Contractor shall maintain the personnel, organization, and administrative control necessary to ensure that the work delivered meets the contract specifications and requirements.
The work history of each contractor employee must contain experience directly related to the task and functions he/she is intended to perform under this contract. The Government reserves the right, during the life of the resulting contract, to request work histories on any contractor employee for the purposes of verifying compliance with the above requirements; additionally, the Government reserves the right to review resumes of contractor personnel proposed to be assigned. Personnel assigned to, or utilized by, the Contractor in performance of work shall be fully capable of performing the requirements contained in the PWS in an efficient, reliable, and professional manner.
10.2. Specific Personnel Qualifications Requirements – Personnel assigned to or utilized by the contractor in the performance of this contract shall, as a minimum:
Be certified by the American Board of Radiology, American Board of Medical Physics, or the Canadian College of Physicists in Medicine. For mammography, a qualified medical physicist can have state licensure to perform mammography inspections in lieu of board certification. For diagnostic computed tomography (CT), nuclear medicine, PET, or MRI, a qualified medical physicist can meet the following requirements in lieu of board certification:
- A graduate degree in physics, medical physics, biophysics, radiologic physics, medical health physics, or a closely related science or engineering discipline from an accredited college or university
- Formal graduate-level coursework in the biological sciences with at least one course in biology or radiation biology and one course in anatomy, physiology, or a similar topic related to the practice of medical physics
- Physicist must document the 3 years of experience for the modality being inspected11.3. Be authorized to work in the United States.
11. PERFORMANCE STANDARDS AND QUALITY MEASUREMENT
11.1. Performance standards define desired services. The Government performs surveillance to determine if the Contractor exceeds, meets, or does not meet these standards. The Government shall use these standards to determine Contractor performance and shall compare Contractor performance to the Acceptable Quality Level (AQL). The Quality Assurance Surveillance Plan (QASP) method of surveillance will be by Clinical Staff report of negative compliance.
11.2 The QASP and its performance objectives are as follows:
PWS
Para.
Performance Objective Performance
Standard
Acceptable Quality Level
Surveillance Method
Compliance
4.2.
Annual review of Diagnostic Imaging Service (DIS) overall Imaging Quality Assurance program for diagnostic imaging equipment.
Annual review coordinated with the Government..
99% of the time.
Review held annually and coordinated with the Government.
Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the PCO will be notified, the performance record will be documented, and PCO may pursue contract action.
4.5
Written reports signed by the diagnostic medical physicist documenting the testing, including a summary of methods, instruments used, measurements and deficiencies identified, for all diagnostic x-ray and imaging equipment within five (5) working days after completion of the inspections
Written report within five (5) working days after completion of the inspections.
98% of the time.
Random inspection conducted by the COR and/or customer comment.
Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the PCO will be notified, the performance record will be documented, and PCO may pursue contract action.
4.6
Report deficiencies or non-conformances which represent unsafe to the RSO immediately.
Access immediate reporting time.
99% of the time.
Random inspection conducted by the COR and/or customer comment.
Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the PCO will be notified, the performance record will be documented, and PCO may pursue contract action.
4.12
Provide emergency response time on equipment within 24 hours.
Access response within twenty-four
(24) hours.
99% of the time.
Random inspection conducted by the COR and/or customer comment.
Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the PCO will be notified, the performance record will be documented, and PCO may pursue contract action.
6.0
All deliverables received. Annually evaluation reports to the Government.
Satisfaction surveys and reports follow-up with the Government.
99% of the time.
Deliverable received in a timely manner.
Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the PCO will be notified, the performance record will be documented, and PCO may pursue contract action.