Attachment D - Minimum Technical Requirements _revised.docx

DOCX document 13 KB Posted

Attached to
6515--Anesthesia Units
Solicitation number
36C24124Q0652
Issued by

About this file

This document is an attachment titled "Minimum Technical Requirements" that defines product specifications for anesthesia units sought by the Department of Veterans Affairs (VA) Veterans Health Administration (VHA) under a pre-solicitation request for information (RFI).

The key technical requirements include the anesthesia units must be FDA-approved, have safety systems to prevent dangerous gas connections, include hypoxic mixture fail-safe controls, support CPAP/volume/pressure ventilation modes, have respirator frequency adjustable between 3-100 breaths per minute, have 1-180 L/min ventilator inspiratory gas flow, and interface with VA health records systems. The RFI seeks information from potential suppliers, including company details, small business status, and ability to meet the technical requirements. Responses are due by August 27, 2024, and an industry day is scheduled for August 23, 2024. The anticipated NAICS code is 339112 Surgical and Medical Instrument Manufacturing, and the Product Service Code is 6515 Medical and Surgical Instruments, Equipment, and Supplies.

View the file

Other files

File Type Posted
Attachment C - Industry Day Slide Presentation.pdf PDF
Attachment B - Industry Day Questions _Answers.docx DOCX document
36C24124Q0652 0003.docx DOCX document
36C24124Q0652 0002.docx DOCX document
36C24124Q0652 0001.docx DOCX document
Attachment A - Minimum Technical Requirements.docx DOCX document
36C24124Q0652.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Minimum Technical Requirements MTR 1 – The anesthesia units must be approved by the Food & Drug Administration (FDA).

MTR 2 – The anesthesia units’ pipeline connections must have a diameter-indexed safety system (DISS) or other equivalent means to prevent connection of dangerous gasses.

MTR 3 – The anesthesia units must have a hypoxic mixture fail-safe system control to prevent the unit from being set to dispense a hypoxic mixture. The N20 and O2 flow must be interlocked in ratio proportions with a minimum value of 25% to produce a breathing mixture.

MTR 4 – The anesthesia units’ ventilation modes should support continuous positive airway pressure (CPAP) volume and pressure control.

MTR 5 – The anesthesia units’ respirator frequency must be able to be set within a range of 3-100 breaths per minute.

MTR 6 – The anesthesia unit’s ventilator inspiratory gas flow range must be between 1-180 L/min.

MTR 7 – The anesthesia unit’s monitors must include a continuing pressure alarm capable of sensing high gas pressure and include minimum selectable threshold levels in the patient’s breathing circuit.

MTR 8 – The anesthesia units must be able to interface with VA VISTA, CPRC, CERNER, and anesthesia health records keeping systems.

MTR 9 – The anesthesia unit packages must be severable such that separate items can be added to complete bundles (to include freight, warranty, and technical support).