36C10G20Q0010-023.docx

DOCX document Posted

Attached to
NX EQ Patient Temperature Management Systems Brand Name or Equal Solicitation Amendment 0001
Solicitation number
36C10G20Q0010
Issued by

About this file

36C10G20Q0010 ATTACHMENT B - PATIENT TEMPERATURE MANAGEMENT SYSTEM - PRODUCT DESCRIPTION.docx

View the file

Other files

File Type Posted
36C10G20Q0010-0001007.docx DOCX document
36C10G20Q0010-022.xlsx XLSX spreadsheet
36C10G20Q0010-026.docx DOCX document
36C10G20Q0010-025.docx DOCX document
36C10G20Q0010-021.docx DOCX document
36C10G20Q0010-024.docx DOCX document
36C10G20Q0010-027.xlsx XLSX spreadsheet

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ATTACHMENT B - PRODUCT DESCRIPTION

NON-EXPENDABLE EQUIPMENT

PATIENT TEMPERATURE MANAGEMENT SYSTEM

Prepared By:

Strategic Acquisition Center Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400 Fredericksburg, VA 22408

1. SCOPE OF WORK:

The Veterans Health Administration (VHA) Equipment Life Cycle Management (ELCM) Program identified Patient Temperature Management System, which are also called Circulating-Liquid Patient Warming Cooling Units, as a candidate for a VA-wide (otherwise referred to as “national”) single award Requirements contract.

The Patient Temperature Management Systems in this requirement circulate warm or cold liquid through channels to raise, lower, or maintain a patient's body temperature. The system allows clinicians to control and stabilize a patient’s body temperature by programing it to heat or cool fluid in the disposable large pads or blankets. They are used to warm hypothermic patients or to cool feverish patients and are frequently used to induce mild hypothermia during neurosurgery or in patients who have experienced a stroke or cardiac arrest. Induced hypothermia lowers a patient's metabolism and has been shown to have neuroprotective effects. Preventing heat loss and hypothermia during other surgical procedures helps reduce the risk of pressure damage and wound infection due to decreased circulation.

The contractor must demonstrate the ability to meet all requirements for the solicitation. The objective is to provide Patient Temperature Management Systems to be used by clinicians throughout the VA medical centers and facilities. The contractor will be required to deliver Patient Temperature Management Systems and accessories to VHA facilities throughout the United States. The period of performance is for one (1) 12-month base period with two (2) twelve (12) month option periods.

2. REQUIREMENT

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires the quoter indicate that each product being offered as an “equal” product to the Blanketrol III® Model 233 which is manufactured by Gentherm Medical Company. For each equal product, the offeror must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The vendor must also clearly identify the item by brand name (if any) and make/model number. Finally, the vendor must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer, and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

CONTRACT LINE ITEMS
MANUFACTURER
PART NUMBER
ITEM DESCRIPTION
0001
Gentherm
Model 233
Blanketrol III Hyper-Hypothermia Temperature Management System
0002
Gentherm
286R
Hose Extension, reusable 9 feet
0003
Gentherm
4872MS
Interconnect Cable
0004
Gentherm
233-CAP
Blanketrol III Interconnectivity Kit

The Department of Veterans Affairs (VA) is seeking vendors who can provide Blanketrol III® Model 233, Patient Temperature Management System as listed above or equal commodities which meet all of the following salient characteristics:

The following SC’s apply to CLIN 0001 only:

SC #
SALIENT CHARACTERISTIC
METHOD OF EVALUATION
SC 1
Patient Temperature Set Range of 86°F to 104°F (30°C - 40°C)
Literature Review
SC 2
Must have the option to display the temperature in either Celsius or Fahrenheit
Literature Review
SC 3
Must have high temperature fluid and low temperature fluid visual and audible alarm
Literature Review
SC4
Must have both a visual and audible alarm to alert of low fluid levels
Literature Review
SC 5
Must have a minimum fluid temperature range of of 39.2°F to 104°F (4°C - 42°C)
Literature Review
SC 6
Must have a minimum reservoir capacity of 3.5 liters
Literature Review
SC 7
Must have manual, automatic and monitoring operation modes
Literature Review
SC 8
Must have the capability to monitor patient temperature
Literature Review