This file is a Schedule of Line Items and Questions & Answers document for a Special February 2026 Equipment Only Consolidation solicitation (36A79726Q0014) issued by the Department of Veterans Affairs National Acquisition Center under the PACT Act Leases program. The consolidation encompasses medical imaging equipment for 13 VA Health Systems across multiple VISNs (Veterans Integrated Service Networks), with line items totaling approximately 45 pieces of equipment distributed across Charleston SC, North Chicago IL, Fayetteville NC, Hampton VA, Nashville TN, New Orleans LA, Killeen TX, Temple TX, and Walla Walla WA.

Equipment requirements include various radiological modalities: X-ray (XR) systems including radiography (RAD), computed tomography (CT), magnetic resonance imaging (MRI), ultrasound (US), fluoroscopy (RF), and mammography (MAMMO), with quantities ranging from one to two units per location. Specific technical specifications address gantry apertures, detector configurations, table specifications, workstation requirements, image processing capabilities, and power backup systems (UPS). The Q&A section documents extensive vendor feedback requesting specification modifications to increase competition, with government responses either accepting changes (highlighted in green on amended compliance matrices) or maintaining original requirements. Significant changes include adjustments to gantry aperture dimensions (reduced from 80cm to 70-72cm on CT systems), table weight capacities, pixel size specifications, wireless detector configurations, and UPS functionality durations. The deadline for vendor proposals is April 14, 2026, with technical specifications and compliance matrices dated after March 25, 2026 required for submission. Amendment 0003 addresses all vendor questions and updates affected technical specifications, with vendors instructed to use only the most current compliance matrix versions.

View the file

Other files

File Type Posted
36A79726Q0014 0003.docx DOCX document
Zipped Files Updated.zip ZIP file
36A79726Q0014 0002.docx DOCX document
SPECIAL FEB 2026 EQ Consolidation Schedule of Line Items Updated.xlsx XLSX spreadsheet
687B63002.zip ZIP file
36A79726Q0014 0002.docx DOCX document
Vendor Folders.zip ZIP file
SPECIAL FEB 2026 Equipment Only Consolidation Schedule.pdf PDF
36A79726Q0014_2.docx DOCX document
SPECIAL FEB 2026 EQ Consolidation - Schedule of Line Items.xlsx XLSX spreadsheet
Memorandum for Training.pdf PDF
534B62010 - 626B64017.zip ZIP file
Vendor Workup Spreadsheet - 36A79726Q0014.xlsx XLSX spreadsheet
626B64018 _ 687B63002.zip ZIP file
VA Directive 6550 - New Appendix A - published Jun. 2023.pdf PDF
Show all 15

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

VISN Names

HTME VISN
VISN 1: VA New England Healthcare System
VISN 2: VA Healthcare Network Upstate New York
VISN 4: VA Stars & Stripes Healthcare Network
VISN 5: VA Capitol Health Care Network
VISN 6: VA Mid-Atlantic Health Care Network
VISN 7: VA Southeast Health Care Network
VISN 8: VA Sunshine Healthcare Network
VISN 9: VA Mid South Healthcare Network
VISN 10: VA Healthcare System of Ohio
VISN 12: VA Great Lakes Health Care System
VISN 15: VA Heartland Network
VISN 16: South Central VA Health Care Network
VISN 17: VA Heart of Texas Health Care Network
VISN 19: VA Rocky Mountain Network
VISN 20: VA Northwest Network
VISN 21: VA Sierra Pacific Network
VISN 22: VA Desert Pacific Healthcare Network
VISN 23: VA Midwest Health Care Network

Schedule of Line Items

HTME VISNStationEQ POEquipment TypeQTY
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62012XR RAD2
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62011XR CT1
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62010XR MRI1
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR US1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69014XR RAD1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69012XR RAD1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69013XR MRI1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69011XR CT1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69006XR RF1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69009NM BONE DEN1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69010XR MAMMO1
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69000XR MRI1
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMO1
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69001XR RF2
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69004NM BONE DEN0
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US0
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69003XR CT1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64017XR CT1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64016XR CT1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64018XR RAD1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64019XR RAD1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMO1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64022XR RAD1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64021NM BONE DEN1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64023XR RAD1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64024XR MRI1
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CT1
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60018XR MRI1
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD1
VISN 17: VA Heart of Texas Health Care Network674 KILLEEN TX674B60240XR MRI1
VISN 17: VA Heart of Texas Health Care Network674 KILLEEN TX674B60239XR MAMMO1
VISN 17: VA Heart of Texas Health Care Network674 KILLEEN TX674B60238XR MAMMO1
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60236XR RF1
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60233XR RAD1
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60232XR RAD1
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60231XR CT1
VISN 20: VA Northwest Network687 WALLA WALLA WA687B63002XR RAD2
VISN 20: VA Northwest Network687 WALLA WALLA WA687B63000XR MRI1
VISN 20: VA Northwest Network687 WALLA WALLA WA687B63001XR CT1

Q&A

HTME VISNStationEQ POEquipment TypeEquipment QuestionsGovt Equipment Response
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62010XR MRI2.c. please change table weight limit to 500lbs
2.d. please change minimum table height to 70cm2c. Can change to 500lbs

2d. Minimum table height not on tech specs

VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62011XR CTPlease clarify if both the TeraRecon server and post processing server are needed for 3D visualization?Both are not needed
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62011XR CTPlease confirm if the number of slices acquired simultaneously is 512.Corrected to 128
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62011XR CTCan the 80 cm gantry aperture be revised to 70 cm to allow for additional competition?No change from 80cm
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62011XR CTCan minimum reconstructed slice thickness be adjusted to 0.625? (We can technically reprocess to .5 and lower, but at acquisition we are .625 so this may allow for more competition).
Does siting allow for full system UPS as listed in tech specs?1d. Changed to 0.625

2i. At least need UPS that maintains functionality of the workstation to properly shutdown equipment during a power outage – Full system UPS has been added to “Added Value” section and may be considered if space and budget allow VISN 7: VA Southeast Health Care Network 534 CHARLESTON SC 534B62012 XR RAD 2f. Please change minimum tube height from floor when used with wall stand to 12 inches for full and fair competition. 3m. Please remove or clarify if a wireless foot pedal is an acceptable substitution. 2f. Changed to 12” 3m. Wireless foot pedal acceptable or change verbiage to read “controls accessible on each side of the table” VISN 12: VA Great Lakes Health Care System 556 NORTH CHICAGO IL 556B61006 XR US Can the minimum monitor size be updated to 19 inches to increase competition?

Can the minimum width be expanded to 23 inches to increase competition?

Please confirm the type of navigation software required.

Yes

Yes

Navigation software specific to markers/overlays

VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould it be advantageous to have minimum monitor rotation be 180 degrees? (Section B. Technical Requirements, Unit physical specifications, line c.)Advantageous but not required
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould it be advantageous to have minimum tilting be at least 90 degrees? (Section B. Technical Requirements, Unit physical specifications, line d.)Not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USHow is it that the “Monitor height range from floor [in]” is listed as a singular number of “48 in” rather than an actual range? Would a range of 40-70 in be sufficient? (Section B. Technical Requirements, Unit physical specifications, line f.)Requirement has been revised
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USHow can the “Control panel height range from floor [in]” be listed as a single number of 36 inches rather than an actual range? Would a range of 30-40 in be sufficient? (Section B. Technical Requirements, Unit physical specifications, line h.)Requirement has been revised
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould it be beneficial to have “Minimum control panel rotation” be at least 180 degrees? (Section B. Technical Requirements, Unit physical specifications, line g.)No
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USCan the “Minimum image storage [days or TB]” be 1 day or 1 TB? (Section B. Technical Requirements, Unit physical specifications, line j.)No, minimum storage must be 7 days
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould they like the potential or option of 3D capability? (Section B. Technical Requirements, Scanning Modes , line b.)No
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould they ever use Pulsed Wave Doppler, Color Power Doppler, Continuous Wave Doppler, or Tissue Harmonic Imaging? (Section B. Technical Requirements, Scanning Modes, lines h., k., i., and e.)Color Power Doppler
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USFor Control panel specifications, why don’t they require or desire a touchscreen monitor and a keyboard on touchscreen ?(Section B. Technical Requirements, Control panel specifications, lines b. and c.)Prefer physical buttons for simpler operation
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould one-button image optimization and one-button doppler optimization benefit them and help their efficiency? (Section B. Technical Requirements, Additional specifications, lines d. and e.)Not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWould they ever use shear wave and strain-based elastography? (Section B. Technical Requirements, Additional specifications, line j.)Not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USIs a 4 way laproscopic transducer absolutely necessary or is it an offering of only one particular company that can be viewed as a whole separate device, given that it’s a robotic transducer? (Section B. Technical Requirements, Transducers, line a.)Not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USThe “T style fingertip transducer” mentioned in Section B. Technical Requirements under the “Transducers” section on line b. is something only offered by FujiFilm. Is this transducer an absolute necessity to complete the procedures and scans that this system will be used for?Not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USWouldn’t a Hockey Stick probe with a larger frequency range of perhaps 25-7 MHz be desirable, given how frequently this probe may get used on this particular system? (Section B. Technical Requirements, Transducers, line d.)Desirable, but not a requirement
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO IL556B61006XR USCan one of the “Clinical Training” requirements be “Technologists who complete the clinical applications training shall receive continuing medical education credits (CMEs)?” This would be a great value add for their sonographers/technicians. (Section C. Training Requirements, Clinical Training, Line d.)Not a requirement
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B690011XR CTPlease confirm the number of processing workstations.Updated to quantity of 1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69006XR RFSystem digital detector configuration:

Can the wall detector be changed to fixed to minimize drops since there will be another wireless detector that can be moved throughout the room?

Please change the table detector to integrated fixed as this will be the dynamic flat panel detector used for both rad and fluoro, the 3rd detector can be wireless for maximum use in room Can you please change maximum spatial resolution [lp/mm] to 3.4 to accommodate other vendors?

Table Requirements:

Can you please change Maximum lowest table height from floor [in] to 25.6 to accommodate other vendors?

Can you please change Minimum patient table load capacity (static) [lbs] to 660 to accommodate other vendors?

Please remove bariatric specific table to accommodate other vendors.Tech Spec updated to reflect the following:17x 17 Fixed wall, 17x 17 fixed table, additional 14x 17 wireless. spacial resolution and table requirements remained unchanged, multiple vendors can meet.
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69006XR RF1ab. Please change wireless detector size to 14x17 to allow for full and fair compeittion.

1cb. Please change wireless detector size to 14x17 to allow for full and fair competition.

1cb. Please clarify that 3 detectors are needed.

3t. Please remove this specification as it only pertains to fixed radiograhy.Tech Spec updated to reflect the following:17x 17 Fixed wall, 17x 17 fixed table, additional 14x 17 wireless. 3t specification removed.
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69010XR MAMMOWill they be doing biopsies at this location?No
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69010XR MAMMODo they plan on reading images on site (at Jacksonville)? Will they want to share that functionality with Durham or Fayetteville?Yes, share functionality with Fayetteville
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69010XR MAMMOCan you provide the physical address for this location?4386 Western Blvd, Jacksonville, NC 28546
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69010XR MAMMOOn our system priors can be reviewed at the acquisition workstation. Does the technologist want to review priors at the acquisition workstation or want a second/separate monitor to review priors on? How are the radiologist reviewing, is a separate station required for them?Yes, we want to include a separate monitor and workstation
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69011XR CTDoes siting allow for full system UPS as listed in tech specs?Yes, The current building design includes a 150sq ft Imaging UPS room
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69012XR RAD2d. Change maximum pixel size to 140 for full and fair competition. 2f. Change static patient table load capacity to 770lbs for full and fair competition. 2g. Change dynamic patient table load capacity to 680lbs for full and fair competition.Tech Spec updated to reflect pixel size change. Table requirement remain unchanged, multiple vendors are capable of meeting
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69012XR RADSystem digital detector configuration:

Please confirm if there is a total of 2 wireless detectors (one in wall bucky and one in table bucky only) or 3 wireless detectors within the room?

Unit physical specifications:

Please clarify if the Maximum time for image display [s] at 3 seconds is preview or full image. If full image, please change to 7 seconds to accommodate all vendors.

Can the minimum patient table load capacity dynamic be changed to 660 to accommodate other vendors?

Additional Specifications:

Can the Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity be removed to accommodate other vendors.3 Wireless detectors. Tech spec updated to reflect preview display time. Table requirements remain unchanged, multiple vendors can meet.
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69013XR MRI2.c. please change table weight limit to 500lbs
2.d. please change minimum table height to 70cmTech spec changed to update table height. Weight limit remains unchanged, multiple vendors can meet.
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69014XR RAD2d. Change maximum pixel size to 140 for full and fair competition. 2f. Change static patient table load capacity to 770lbs for full and fair competition. 2g. Change dynamic patient table load capacity to 680lbs for full and fair competition.Tech Spec updated to reflect pixel size change. Table requirement remain unchanged, multiple vendors are capable of meeting
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69000XR MRI1.d. minimum FOV 500cm. Please clarify minimum or maximum FOV specification and units (mm vs cm). Typical maximum FOV is 50cm
1.f. please change channel specifications to reflect the Maximum number of channels per Field of View without table movement each generating an independent partial image, not simply the number of elements that can be connected. The 204 Channel specification is only used by one vendor and does not reflect the true number of RF channels in a system.1.d. changed minimum to 45 cm

1.f no change

VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69000XR MRIThe Schedule of Line Items and the Matrix and Tech Spec documents are labeled as "XR US". In review, the documents reads "MRI". Please confirm Equipment Type for this line.This modality is MRI
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69001XR RFTechnical Requirements Section 1. Please confirm the amount of wireless detectors. In the matrix it requests 4 detectors of 17x17 configuration.1.c.b was updated to 14x17

total wireless detectors are 4 (1 wall, 1 table and 2 additional) VISN 6: VA Mid-Atlantic Health Care Network 590 HAMPTON VA 590B69001 XR RF 1aa. Please change wall detector to fixed to allow full and fair competition.

1ba. Please change table detector to fixed to allow full and fair competition.

1ca. Please clarify the number of detectors being request per each unit. If four are being requested per unit, please lower to 3 detectors to allow full and fair competition.

1cb. Please change wireless detector size to 14x17 to allow full and fair competition.

6a/6b. PLease change monitor size to 21inches to allow full and fair competition.

7a. Please move bone suppression to the value-add section to allow full and fair competition. 1aa. No change 1ba. No change 1ca. No change 1.cb - changed 6a/6b -changed 7a. No change

VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR USPlease confirm this is the correct obligation number. This is the same obligation number used on the September 2025 Equipment only consolidation.Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR USCan the minimum monitor size be updated to 19 inches to increase competition?

Can the minimum image storage but changed to 500gb to increase competition?

Image Fusion Software. Please confirm what procedures will be performed utilizing Fusion.

If Fusion is required, will you be utilizing a 3rd party Fusion solution or desiring an all-inclusive Fusion system?

If utilitizing a 3rd party, which navigation system is currently in use?

If Fusion is required, will it be needed for TP, TR, of both?

If Fusion is required, will it be required to be freehand or stabilized?

If Fusion is required, will you need endfire, sidefire, or both?

If Fusion is required, please identify probe requirements and quantities.

If TP is required, please add Bi-plane probe.

Can Phantom for quality control testing be moved to added value?

Can Automatic point shear wave elastography, linear strain elastography, utrasound derived fatt fraction (UDFF) be moved to added value?

Can AI powered semi automated measurement tool be moved to added value?

Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR USPlease clarify if there is a need for prostate Fusion? If no prostate Fusion is needed, the following specs should be removed from Section 8. Advanced features:

Fusion-Targeted Biopsy System

Capable of accepting MRI (both 1.5T and 3.0T) and ultrasound images from other modalities as input and displaying these images on a screen

Prostate gland volume and edge boundaries computation capabilities

Tracking and recording of needle movement during biopsy in 3D coordinates in real-time

Availability of projected, future needle movement during biopsy in 3D coordinates

Ability to overlay previous biopsy images on top of real-time prostate images

3D-generated model of prostate that can be exported to report, adjustable by additional non-imaging information, such as pathology-related data

Scanning modes

Navigation Software – note: no needle guide is requested for Navigation SW on benchmark.

Image Fusion Software Under 6. Analysis packages

Other analysis packages: FusionRemoved from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR USCan you verify the quantity of Endocavitary Transducers? 4 or 8?
Can you verify if 8.f. 3D-generated model of prostate that can be exported to report, adjustable by additional non-imaging information, such as pathology-related data is required?Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US1.j. Can maximum depth be opened up to 49.02" for increased competition?Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US5.n. Can the phantom requirement be placed in the Value Add section?Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US8. Is Fusion clinically necessary?Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US2.c. Image Navigation software & Fusion Software – is this clinically necessary on both systems? If not clinically necessary, please move to added value.Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69002XR US8 Is Fusion-targeted Biopsy System and all the features in this section required on both systems? If not clinically necessary, please move to added Value.Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69003XR CTCan the 80 cm gantry aperture be revised to 70 cm to allow for additional competition?No change
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69003XR CTa. Minimum number of slices acquired simultraneously: 64
Q:Please change 1a to reflect 256 slices acquired simultaneously to be in line with the 160 mm detector and 0.625 mm slice thickness specifications (160/0.625 = 256).changed to 256
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69003XR CTLine 22-Minimum gantry aperture Can this be changed to 72cm for free and fair competition?No change
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69003XR CTLine 23-Minimum detector width Can this be changed to 40mm for free and fair competition?No change
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69004NM BONE DENThe Schedule of Line Items list the Equipment Type as "NM BONE DEN"; however, the Matrix and Tech Spec documents are labeled as "XR US". In review, the documents reads as "Bone Density". Please confirm the Equipment Type for this line.Removed from consolidation
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMO. Technical requirements
•1.a. Generator Voltage Range [KV] – Request consideration of revising the specified generator voltage range to include 23–49 kV to broaden vendor participation and enhance competition.

• 1.d. Number of generator phases – Please clarify the intended meaning and requirement for “generator phases,” as this term may be interpreted in multiple ways. The number specified may be specific to a vendor’s design approach. If this specification is not mandatory, please confirm whether it may be addressed as a value-added feature. 1.a. changed 1.d changed to 3 phases

VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMO. Workstation Requirements
•2. c. Minimum number of review workstations – Please clarify if the request is to quote a new Radiologist Review Station2c. changed to minimum 2 for Radiologist Review Station
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMOAdditional Specifications

AI Enhancement capabilities – Please clarify the intended scope of “AI enhancement capabilities,” as this term can encompass a broad range of functionalities. If specific capabilities are not required as part of this solicitation, please confirm whether this may be moved to value-added offerings.

Biopsy Software for Multi-room capabilities – Please clarify what is meant by Biopsy Software for Multi-room capabilities. If this is not needed, please confirm whether this may be moved to value added offerings. AI Enhancement capabilities - reading assistance for radiologists Biopsy Software - licensing available to have the biopsy software avaialble in multiple room workstations

VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMOPlease update equipment from XR US to XR Mammo.Listed as XR Mammo
VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMOLabeled as US but appears to be Mammo, please confirm

What VABB Device adapter is needed? (Eviva, Brevera, Mammetome, Bard?) Are you looking for a Dedicated Breast Radiologist Reading Workstation?

Do you need a positioning Chair for the patient procedure for the Breast Biopsy? ' Are you interested in also having a Needle Localization kit? This is a mammo.

No adapter requested.

Yes 2 dedicated workstations.

Yes positioning chair.

Yes to needle kit

VISN 6: VA Mid-Atlantic Health Care Network590 HAMPTON VA590B69005XR MAMMOThe Schedule of Line Items and the Matrix and Tech Spec documents are labeled as "XR US". In review, the documents reads "Mammography". Please confirm Equipment Type for this line.This is mammo.
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64016XR CTCan the 75 cm gantry aperture be revised to 70 cm to allow for additional competition?Yes, Specification has been updated
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64016XR CTLine 23: b. Minimum gantry aperture (cm)- 75: Can the minimum be lowered to 72 cm for free and fair competition?Yes, Specification has been updated
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64016XR CT2K.) Full functionality for 30 minutes of the modality/system is not necessary and extremely cost prohibitive for a CBOC. Please change to full functionality of 6.1 minutes and partial functionality for 10 minutes.
3F.) Do you really need to maintain full functionality for 20 minutes for a workstation?2K.) Full functionality changed from 30 minutes to 10 minutes to allow for proper shut down of device during outage

3F.) Changed from 20 minutes to 10 minutes to allow for proper shut down of device during outage

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64017XR CTCan the 75 cm gantry aperture be revised to 70 cm to allow for additional competition?Yes, Specification has been updated
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64017XR CT2K.) Full functionality for 30 minutes of the modality/system is not necessary and extremely cost prohibitive for a CBOC. Please change to full functionality of 6.1 minutes and partial functionality for 10 minutes.
3F.) Do you really need to maintain full functionality for 20 minutes for a workstation?2K.) Full functionality changed from 30 minutes to 10 minutes to allow for proper shut down of device during outage

3F.) Changed from 20 minutes to 10 minutes to allow for proper shut down of device during outage

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64017XR CTLine 23: b. Minimum gantry aperture (cm)- 75: Can the minimum be lowered to 72 cm for free and fair competition?Yes, Specification has been updated
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64017XR CTTraining Requirements, 1d) Please specify which OS support is required; Is this a question about Windows 10 / Windows 11 compatibility and support?Windows 11 is preferred, but if that is not possible, we require OEM support for any OS that might be end-of-life within the VA
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64018XR RAD2i. Please change image plate height from floor to 11.5 inches to allow full and fair competition.
2m. Pleasd change minimum table height from floor to 20 inches to allow full and fair competition.2i. Tech spec updated to 12 in.

2m. Tech spec updated to 21 in.

VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64018 XR RAD 2h. Please change minimum tube height from floor when used with wall stand to 12 inches for full and fair competition. 2i. Please change minimum image plate height from floor to 12 inches for full and fair competition. 2l. Change dynamic patient table load capacity to 680lbs for full and fair competition. 2m. Please change minimum table height from floor to 21 inches for full and fair competition. 3n. Please remove or clarify if a wireless foot pedal is an acceptable substitution. 2h. Tech spec updated to 12 in.

2i. Tech spec updated to 12 in.

2l. Tech spec changed to 680 lbs.

2m. Tech spec updated to 21 in.

3n. Wireless foot pedal is acceptable VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64019 XR RAD Can the maximum wireless wall detector weight with battery installed (lbs) and the maximum wireless table detector weight with battery installed (lbs) be changed to 9.5lbs to accommodate other vendors?

Can the minimum tube height from floor when used with wall stand be changes to 12” to accommodate other vendors? Can the minimum image plate height from floor when used with wall stand be changes to 12” to accommodate other vendors?

Can the minimum patient table load capacity dynamic be changed to 660 to accommodate other vendors?

Can the minimum table height from floor (in) be changed to 20” to accommodate other vendors?

Can the maximum system weight (lbs) be changed to 5000lbs to accommodate other vendors?

Can the UPS – maintains full functionality of the entire modality / system for a minimum of 10 minutes during a power outage – be removed in both section 3 and 4 as this is not necessary for the entire system in radiography? 1. Yes and yes

2. Tech spec updated to 12 in.

3. Tech spec updated to 12 in.

4. Tech spec changed to 660 lbs.

5. Tech spec updated to 21 in.

6. Tech spec changed to 5,000 lbs

7. UPS needed for both sections to allow for proper shut down of the system in the event of a power outage

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64019XR RAD2i. Please change image plate height from floor to 11.5 inches to allow full and fair competition.
2m. Pleasd change minimum table height from floor to 20 inches to allow full and fair competition.2i. Tech spec updated to 12 in.

2m. Tech spec updated to 21 in.

VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64019 XR RAD 2h. Please change minimum tube height from floor when used with wall stand to 12 inches for full and fair competition. 2i. Please change minimum image plate height from floor to 12 inches for full and fair competition. 2l. Change dynamic patient table load capacity to 680lbs for full and fair competition. 2m. Please change minimum table height from floor to 21 inches for full and fair competition. 3n. Please remove or clarify if a wireless foot pedal is an acceptable substitution. 2h. Tech spec updated to 12 in.

2i. Tech spec updated to 12 in.

2l. Tech spec changed to 680 lbs.

2m. Tech spec updated to 21 in.

3n. Wireless foot pedal is acceptable

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOWould you like the 3D imaging to be high resolution?Yes
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOWould you like a synthetic 2D image (C-View/Intelligent 2D) to be created from the 3D imaging?Yes
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOWill you require a Biopsy positioning chair? Do you need a breast biopsy system such as an ATEC or prefer a breast biopsy system with imaging like our Brevera?No
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOThe specs asks for iCAD, Hologic does not sell iCAD so please confirm you are requesting density software?Yes; density software is required
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOPlease advise if you want 2D CAD, 3D CAD, or 2D & 3D CAD?2D and 3D CAD
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOWould you like us to include needle loc?Yes
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMOWill you need an ACR phantom with case?Yes
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64020XR MAMMO3C.) Is full functionality for 10 minutes really necessary for the UPS for mammography? This is not standard practice for the VA. If a Full UPS is needed, do you have the required space/siting/HVAC etc. to support one?.
3d.) Is a vacuum assisted device needed for the guided biopsy system?3C.) Partial functionality would meet the needs of the service

3D.) No VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64022 XR RAD System Digital Detector Configuration Can the maximum wireless wall detector weight with battery installed (lbs) and the maximum wireless table detector weight with battery installed (lbs) be changed to 9.5lbs to accommodate other vendors?

Unit physical Specifications:

Can the minimum tube height from floor when used with wall stand be changes to 12” to accommodate other vendors? Can the minimum image plate height from floor when used with wall stand be changes to 12” to accommodate other vendors?

Can the minimum patient table load capacity dynamic be changed to 660 to accommodate other vendors?

Can the minimum table height from floor (in) be changed to 20” to accommodate other vendors?

Can the maximum system weight (lbs) be changed to 5000lbs to accommodate other vendors?

Workstation Requirements:

Can the UPS – maintains full functionality of the entire modality / system for a minimum of 10 minutes during a power outage – be removed in both section 3 and 4 as this is not necessary for the entire system in radiography? 1. Yes and yes

2. Tech spec updated to 12 in.

3. Tech spec updated to 12 in.

4. Tech spec changed to 660 lbs.

5. Tech spec updated to 21 in.

6. Tech spec changed to 5,000 lbs

7. UPS needed for both sections to allow for proper shut down of the system in the event of a power outage

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64022XR RAD2i. Please change image plate height from floor to 11.5 inches to allow full and fair competition.
2m. Pleasd change minimum table height from floor to 20 inches to allow full and fair competition.2i. Tech spec updated to 12 in.

2m. Tech spec updated to 21 in.

VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64022 XR RAD 2h. Please change minimum tube height from floor when used with wall stand to 12 inches for full and fair competition. 2i. Please change minimum image plate height from floor to 12 inches for full and fair competition. 2l. Change dynamic patient table load capacity to 680lbs for full and fair competition. 2m. Please change minimum table height from floor to 21 inches for full and fair competition. 3n. Please remove or clarify if a wireless foot pedal is an acceptable substitution. 2h. Tech spec updated to 12 in.

2i. Tech spec updated to 12 in.

2l. Tech spec changed to 680 lbs.

2m. Tech spec updated to 21 in.

3n. Wireless foot pedal is acceptable VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64023 XR RAD System Digital Detector Configuration Can the maximum wireless wall detector weight with battery installed (lbs) and the maximum wireless table detector weight with battery installed (lbs) be changed to 9.5lbs to accommodate other vendors?

Unit physical Specifications:

Can the minimum tube height from floor when used with wall stand be changes to 12” to accommodate other vendors? Can the minimum image plate height from floor when used with wall stand be changes to 12” to accommodate other vendors?

Can the minimum patient table load capacity dynamic be changed to 660 to accommodate other vendors?

Can the minimum table height from floor (in) be changed to 20” to accommodate other vendors?

Can the maximum system weight (lbs) be changed to 5000lbs to accommodate other vendors?

Workstation Requirements:

Can the UPS – maintains full functionality of the entire modality / system for a minimum of 10 minutes during a power outage – be removed in both section 3 and 4 as this is not necessary for the entire system in radiography? 1. Yes and yes

2. Tech spec updated to 12 in.

3. Tech spec updated to 12 in.

4. Tech spec changed to 660 lbs.

5. Tech spec updated to 21 in.

6. Tech spec changed to 5,000 lbs

7. UPS needed for both sections to allow for proper shut down of the system in the event of a power outage VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64023 XR RAD System Digital Detector Configuration Can the maximum wireless wall detector weight with battery installed (lbs) and the maximum wireless table detector weight with battery installed (lbs) be changed to 9.5lbs to accommodate other vendors?

Unit physical Specifications:

Can the minimum tube height from floor when used with wall stand be changes to 12” to accommodate other vendors? Can the minimum image plate height from floor when used with wall stand be changes to 12” to accommodate other vendors?

Can the minimum patient table load capacity dynamic be changed to 660 to accommodate other vendors?

Can the minimum table height from floor (in) be changed to 20” to accommodate other vendors?

Can the maximum system weight (lbs) be changed to 5000lbs to accommodate other vendors?

Workstation Requirements:

Can the UPS – maintains full functionality of the entire modality / system for a minimum of 10 minutes during a power outage – be removed in both section 3 and 4 as this is not necessary for the entire system in radiography? 1. Yes and yes

2. Tech spec updated to 12 in.

3. Tech spec updated to 12 in.

4. Tech spec changed to 660 lbs.

5. Tech spec updated to 21 in.

6. Tech spec changed to 5,000 lbs

7. UPS needed for both sections to allow for proper shut down of the system in the event of a power outage

VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64023XR RAD2i. Please change image plate height from floor to 11.5 inches to allow full and fair competition.
2m. Pleasd change minimum table height from floor to 20 inches to allow full and fair competition.2i. Tech spec updated to 12 in.

2m. Tech spec updated to 21 in.

VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64023 XR RAD 2h. Please change minimum tube height from floor when used with wall stand to 12 inches for full and fair competition. 2i. Please change minimum image plate height from floor to 12 inches for full and fair competition. 2l. Change dynamic patient table load capacity to 680lbs for full and fair competition. 2m. Please change minimum table height from floor to 21 inches for full and fair competition. 3n. Please remove or clarify if a wireless foot pedal is an acceptable substitution. 2h. Tech spec updated to 12 in.

2i. Tech spec updated to 12 in.

2l. Tech spec changed to 680 lbs.

2m. Tech spec updated to 21 in.

3n. Wireless foot pedal is acceptable VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64024 XR MRI Row 28: Max system dimensions: 85 x 85 x 105: please consider increasing to at least 100 x 100 x 105 to allow for competition.

Row 29: Max system weight is 4000 lbs: please consider increasing to at least 6,000lb to allow for competition.

Row 40: UPS – maintains full functionality of the entire modality / system for a minimum of 20 minutes during a power outage: please consider reducing the duration to 10 minutes for fair competition Row 50: Ceiling-mounted injector: Will a pedestal mounted injector be acceptable? 28. Tech spec updated

29. Tech spec updated to 8,000 lbs

40. Tech spec updated

50. Pedestal mounted injector will meet the needs of the facility VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN 626B64024 XR MRI 1.g. maximum system dimensions. Please change to 100x100x100 1.h. maximum system weight. Please change to 8000lbs 3.d. Would you like a FULL UPS OR would you like a Partial? If a Full UPS is needed, do you have the required space/siting/HVAC etc. to support one?. Please change runtime to 7 minutes 3.n. Please change to pedestal mounted which is the standard for MR injectors 5.m. Is an endorectal coil required or simply the ability to do high- quality prostate MR? 1.g. Tech spec updated to 100 x 100 x 105 1.h. Tech spec updated 3.d. Full UPS is required and we have required space 3.n. Tech spec updated 5.m. No, endorectal coil is not needed. We do prostate MRI without it

VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RADTechnical Requirements Line 25A. Please change the detector in the Table to wireless to allow additional competitionUpdated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.f Respectfully request to change the maximum pixel size to 150 from 148 to increase competition.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.h Respectfully request minimum tube height from floor to 16" to the middle of the collimation from 1" for fair competition. There is no known vendor that supports tube height to 1" from floor.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.j Please clarify if this is referring to the wallstand.Deleted
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.l Respectfully request minimum patient table load capacity (dynamic) to 650lbs from 700lbs for fair competition.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.m Respectfully request minimum table height from floor to 21" from 20" to increase competition.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.o Respectfully clarify maximum system dimension details. This would be room dependent. Please provide minimum room dimension.Currently, the room design has not been finalized and will be built to accommodate the unit so we cannot provide a minimum.
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD3.l Respectfully clarify if UPS is for proper shutdown of workstation safely. UPS for the processing workstation that maintains functionality for a minimum of 5 minutes during the power outage. Are you requiring a full UPS to run the system and take x-rays during a power outage?Just for ensuring patient safety and proper shutdown procedures. Tech spec has been updated
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RADIf Yes, Is there enough room in the x-ray room to install a UPS?Building has not been created yet, but they intend to build the facility based on equipment so it should be, yes.
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD4.h Respectfully clarify if UPS is for proper shutdown of workstation safely to store images.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2i. Please change image plate height from floor to 11.5 inches to allow full and fair competition.Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2.e. Can the maximum spatial resolution be increased to 3.7lp/mm to increase competitionUpdated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD2. o. Please clarify maximum system dimensions, is this for table?Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RAD3. h. Is a six-way manual table top movement acceptable? An electronic tabletop is usually required in RF room notUpdated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60017XR RADValue Add-What size detector is the site looking for?Fixed detector 17 in by 17 in
VISN 16: South Central VA Health Care Network629 NEW ORLEANS, LA629B60017XR RADLine 139: Please verify Serieal number of the system being traded in.1000118 is the SN of the system within Maximo
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60018XR MRIAdditional Specifications Line 44O States "Cryogens"(provide and install) but on Line 47R States "low/no helium", please clarify what type of magnet is preferredNo preference - but attempts should be made by chosen vendor to reduce helium in their designs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60018XR MRICoils Line 63L Please clarify the need for an Endorectal coilRemoved
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60018XR MRI3d. Would you like a FULL UPS (e.g., 150-200 kVa OR just the Partial? If a Full UPS is needed, do you have the required space/siting/HVAC etc. to support one?Full UPS. Building will be designed around specifications.
VISN 16: South Central VA Health Care Network629 NEW ORLEANS, LA629B60018XR MRILine 49 asks for 2 workstations. Does this mean we need to include a radiologist reading workstation or is this another workstation for technologists. Or does this mean 2 monitors?2 workstations should be included
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTRequirement Overview Line 13 Please clarify total number of workstations being requested including acquisition workstations2 workstations should be included
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTechnical Requirements Line 22A Please change the minimum number of slices to 128 slices simultaneously to allow additional competitionUpdated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTechnical Requirements Line 33F Is this clinical intent to perform coronary CTA?Mainly calcium score, minimal coronary CTA expected
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTechnical Requirements Line 34G Is this clinical intent to perform coronary CTA?Mainly calcium score, minimal coronary CTA expected
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTechnical Requirements Line 35H Is this clinical intent to perform coronary CTA?Mainly calcium score, minimal coronary CTA expected
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTechnical Requirements Line 37J Please clarify if a FULL UPS is required?Full UPS
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTTrade in information needed Line 87-93 Please provide the following trade in information for the Toshiba scanner; Date of last tube installation, Scan seconds on the last tube, scan seconds on the gantry, Software versionUnable to provide, stated in specifications
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CT1)The number of slices required is listed at 160. Normally for any vendor the number of slices is listed as a multiple of 64 for newer systems from any vendor. Did they mean to list this as 128 slices or 256 slices?
2)The Z-axis detector width, is also another somewhat odd number at 35mm. Normally the z-axis detector width is 40, 80, or 160mm. Can they please validate this number?
3)Instead of 2 AW workstations we believe the better solution is an AW server, can that be listed as an option vs the two workstation configuration?1) Edited to 128

2) 35 is minimum value, 40 or greater would be fine

3) alternate solutions ok, as long as two computers/monitors allow for 2 staff to work

VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTLine 23 a. Minimum number of slices acquired simultaneously – 160 - Can the minimum be lowered to 128 slices for free and fair competition?Updated tech specs
VISN 16: South Central VA Health Care Network629 NEW ORLEANS LA629B60019XR CTLine 25 d. Minium gantry aperture (cm) – 75 - Can the minimum be lowered to 72 cm for free and fair competition?No, this is needed for patient population
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60231XR CT1e. Would you please adjust the desired slice thickness to 0.625mm to allow for more competition?

2k. Would you like a FULL UPS (e.g., 150-200 kva) OR would you like a Partial? If a Full UPS is needed, do you have the required space/siting/HVAC etc. to support one?

2jj. would you like a standard injector OR would you like a syringeless injector that allows for much less contrast to be used and wasted per day??

3c and d: Since there are other CT systems planned for either this outpatient center and/or the Central Texas VAMC Network, would you consider a server-based post-processing solution that would be installed in your VM environment?

4a and 4b. Will you be doing both cardiac and neuro perfusions? If so, would the inclusion of a CT shuttle solution be desired?

Ec, Would an applications software delivery model that allows for continuous software enhancements during the life of the agreement as the vendor releases them be desired OR would you just prefer to quote discrete software as mentioned in the solicitation?? 1e - No 2k - removing System UPS requirement 2jj - Standard injector only 4a/4b - No Ec - if the software enhancements are covered under the purchase of the system and not as an add-on, yes.

VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60232XR RAD1a,c & 1b,c) Please increase max weight to 9lbs to maximize competition
2a) Will an 80kw generator be acceptable to the station?1a,c/1b,c - Yes

2a - Yes

VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60232XR RAD2.h Respectfully request minimum tube height from floor to 16" to the middle of the collimation from 1" for fair competition. There is no known vendor that supports tube height to 1" from floor.2h -Will increase to 15
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE TX674B60232XR RAD2.k Respectfully request minimum table height from floor to 21" from 15" to increase competition.2k - increased to 22