This is a Schedule of Requirements and Quality Assurance (QA) document for a January 2026 consolidation procurement of high-tech medical equipment (HTME) with extended installation services for the Department of Veterans Affairs. The document lists 32 line items of medical imaging equipment distributed across multiple VA Healthcare Networks (VISNs) and facilities, including X-ray systems (XR), CT scanners, MRI systems, nuclear medicine equipment (NM SPECT, PET/CT), radiation therapy linear accelerators (RT LINAC), ultrasound/interventional radiology (URO, RF, CATH LAB), and other specialized imaging modalities. Each line item specifies the VISN location, facility station number, equipment purchase order (PO) number, equipment type, and quantity (all quantities are 1 unit per line item), with facilities spanning from Delaware to California.

The QA template section documents extensive vendor inquiries and government responses addressing technical specification clarifications and fairness of competition concerns across multiple equipment procurements. Key themes in the QA responses include: clarifications on power-assist table capabilities and motorization requirements; revisions to UPS (uninterruptible power supply) specifications to reduce duration from extended periods to 7-10 minutes for partial functionality; acceptance of alternative detector architectures and technologies (such as CZT-based systems instead of traditional NaI cameras, or fixed detectors instead of wireless) provided clinical performance is maintained; removal or repositioning of vendor-specific specifications to value-add sections to broaden competition; adjustments to dimensional requirements (table height ranges, detector weights, pixel sizes) and load capacities to allow multiple vendors to compete; and clarifications on integrated versus modular equipment designs. The government consistently balanced clinical requirements with fair competition principles, accepting alternative technologies and specifications when they maintained functional clinical capabilities while expanding the vendor pool eligible to bid.

View the file

Other files

File Type Posted
Amendment 0001 - Continuation Page.pdf PDF
605B60047 - 695B68003.zip ZIP file
36A79726Q0008 0001.docx DOCX document
521B62001 - 578B68008.zip ZIP file
578B58006 - 640B64047.zip ZIP file
JAN 2026 TKY Schedule of Requirements - Updated 3-11-26.xlsx XLSX spreadsheet
Vendor Workup Spreadsheet - 36A79726Q0008.xlsx XLSX spreadsheet
JAN 2026 TKY Consolidation Schedule.pdf PDF
NAC Turnkey Rubric - 2 Mar 2026.pdf PDF
JAN 2026 Extended Install and Turnkey Consolidation RFO.pdf PDF
36A79726Q0008_1.docx DOCX document
460B68004 - 556B61003.zip ZIP file
Memorandum for Training.pdf PDF
Vendor Folders.zip ZIP file
VA Directive 6550 - New Appendix A - published Jun. 2023.docx DOCX document
646B52002 - 695B68003.zip ZIP file
Show all 16

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Schedule of Line Items

HTME VISNStationEquip POTKY POEquipment TypeQTY
VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON, DE460B68004460B68005XR IR1
VISN 10: VA Healthcare System of Ohio506 ANN ARBOR, MI506B60010506B60013RT LINAC1
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001521B62002XR CT1
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B64088528B64082XR IR1
VISN 20: VA Northwest Network531 BOISE, ID531B63000531B63001XR RAD1
VISN 20: VA Northwest Network531 BOISE, ID531B63002531B63003XR RF1
VISN 7: VA Southeast Health Care Network534 CHARLESTON, SC534B62020534B62008XR CT1Updated Equipment PO
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B60019539B60020XR URO1
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001556B60002NM SPECT CT1
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B61003556B61004XR RAD1
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B58006578B68001XR MRI1
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B68002578B68007XR RAD1
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B68004578B68005XR RAD1
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B68008578B68009XR CT1
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047605B60049XR RAD1
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60011607B60012XR RF1
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60013607B60014XR CATH LAB1
VISN 5: VA Capitol Health Care Network613 MARTINSBURG, WV613B61003613B61004NM PET CT1
VISN 7: VA Southeast Health Care Network619 MONTGOMERY, AL619B62002619B63006XR RF1
VISN 21: VA Sierra Pacific Network640 PALO ALTO, CA640B64047640B64048XR RF1
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52002646B52003XR URO1
VISN 10: VA Healthcare System of Ohio655 SAGINAW, MI655B60002655B60003XR RF1
VISN 15: VA Heartland Network657 ST. LOUIS, MO657B60003657B60004RT LINAC1
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003663B63005NM SPECT CT1
VISN 20: VA Northwest Network663 SEATTLE, WA663B63004663B63006XR RAD1
VISN 22: VA Desert Pacific Healthcare Network664 SAN DIEGO, CA664B60033664B60032XR CT1
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000668B63001NM SPECT CT1
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048691B60049XR CT1
VISN 20: VA Northwest Network692 WHITE CITY, OR692B63002692B63001XR RAD1
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B68003695B68004XR CATH LAB1

QA Template

HTME VISNStationEquip POEquipment QuestionsGOVT Equipment Response
ALLALLALLALL QUESTIONS PERTAINING TOO DRAWINGS (AS-BUILTS, CAD, ETC.)AS DESCRIBED IN THE SOLICITATION RFO, WITH THE NEW PROCESS FOR EQUIPMENT WITH EXTENDED INSTALLATION INCLUDING TURNKEY; ALL DRAWINGS (AS-BUILTS, CAD, ETC.) WILL BE PROVIDED TO THE OFFER THAT IS RATED HIGHEST AFTER EVALUATION. ONCE NOTIFIED BY THE CO, THAT VENDOR AND STATION POC WILL WORK TOGETHER TO PROVIDE THOSE DRAWINGS, AS WELL AS WORK ON THE SITE VIST.
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B680032.s. Please remove the ability for the tabletop to move in cradle motion (side to side) or place in the "value add" section to make this more vendor inclusive.

2.w. Please remove "Dynamic Coronary Roadmap" as this is not vendor inclusive and is a lockout spec.

updated specs as requested
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B68003Is there injector needed in this quote?No, we have a Bayer/Medrad Mark 7 Arterion procured in September 2024.
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B68003Is there a need for additional 27” monitors to be mounted on the walls? If so, how many?No
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B68003What is the lab Number?We only have 1 electrophysiology lab in room 5535 building 111.
VISN 12: VA Great Lakes Health Care System695 MILWAUKEE, WI695B68003do you need an equipment boom?no, will be a separate future procurement.
VISN 20: VA Northwest Network692 WHITE CITY, OR692B630023.m. Please clarify, what are power assist table capabilities?

3.p. Please clarify the amount of rail coverage you require, as there are bridge and rail limitations -- unless this is being designed as a trauma room.

3.m. Table should be motorized in the z-axis to go up and down. Power assist can either be motorized in the xy plane or some other mechanism to assist the technologist with moving the table is said xy plane. All vendors don't have motorized capabilities in the xy plane so made the specification more generic to allow better competition for the technologist assist.

3.p. The facility wants the rails to cover the entire room to maximize tube placement in the space.

VISN 20: VA Northwest Network692 WHITE CITY, OR692B630023. O. Please remove need for UPS on Capital quote add if necessary to the turnkey.UPS has to be part of the equipment quote. It cannot be added to the turnkey quote.
VISN 20: VA Northwest Network692 WHITE CITY, OR692B630023m) Please clairfy 'power assist' requirement; is requiement asking for an elevating table or power assisted movement of OTC?
Added Value d) Please clarify if this is a third detector as well as total number of detectors required? Additionally, please move this requirement to section 1 System Digital Detector Configuration for clarity3m) Table should be motorized in the z-axis to go up and down. Power assist can either be motorized in the xy plane or some other mechanism to assist the technologist with moving the table is said xy plane. All vendors don't have motorized capabilities in the xy plane so made the specification more generic to allow better competition for the technologist assist.

Added Value d) Facility is requiring 2 detectors as part of the base proposal. There is active discussion on whether the facility will want a 2nd 14x17 detector so it was added included in the added value. This should be included as an option on your proposal.

VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Minimum reconstructed slice width set to 0.2mm. Request this be revised to 0.5mm to allow fair competition.Reconstructed slice width will be adjusted to 0.5mm or less for routine clinical imaging
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Table height range, would 33cm to 95 cm from floor be acceptable? This lower table height benefits veteransThis is acceptable and requirement will reflect minimum range
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Max table load capacity, request this be adjusted to 695 lbsThis will be adjusted to be at least 695lbs
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Photon counting detector, would you accept competitive technology that provides the same clinical value without the same marketing name? Request revision to Photon counting or like technology.This will be adjusted to photon-counting detector technology or equivalent advanced detector/imaging technology with comparable clinical capability
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Additional Specifications Line 42A Maintains full functionality of the entire system for a minimum of 60 minutes during a power outage; Please consider reducing the duration minutes to 10 -15 minutes which is typical for a standard systemUPS time will be reduced to a minimum of 10 minutes for emergency power.
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B60048Added Value Line 91A Needle Guidance system; Please clarify the procedures you plan to perform and whether you're looking for guidance display on screen only or projected directly onto the patient?On screen guidance is acceptable, projection directly onto the patient is not required. This is an added value function
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B600481.j. Please change table weight to 675 for fair competition.

2.k. Are you looking for a full UPS? This requires its own room and siting. Please reduce time 10 mins, this is enough time to finish the scan, save the date, and remove the patient from the scanner.

2.p. Please remove Photon Counting or place in value-add section, as this is a lock out spec for all vendors except for one.

3.e. Please reduce time 10 mins, this is enough time to finish the scan, save the date, and remove the patient from the scanner.

3.f.Please reduce time 10 mins, this is enough time to finish the scan, save the date, and remove the patient from the scanner.

TIV: For trade in system valuation, please provide number of scans on the tube.

1.j. To safely image Veterans across a broad range of body habitus, including bariatric patients, without unnecessary off-site referral or delay in care patient table weight capacity shall be at least 695 2.k. UPS time will be reduced to a minimum of 10 minutes for emergency power 2.p. This will be adjusted to photon-counting detector technology or equivalent advanced detector/imaging technology with comparable clinical capability 3.e./f. UPS time will be reduced to a minimum of 10 minutes for emergency power VISN 20: VA Northwest Network 668 SPOKANE, WA 668B63000 3.h., 3.aa., 6.e., and 6.j.Please remove or move to Value Adds to allow full and fair competition as only one vendor can meet these requirements.

5.i. Please remove as this is not something done in a NM department.

Both fully automated and semi-automated collimator configurations are acceptable. QC specs will not be moved to value add. Please respond with how QC is completed on the proposed system.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000Which cardiac software package would you prefer - Cedars, Invia or Emory?Cedars
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000The technical specifications reference a dual-detector, variable-angle, dual-head configuration.

Can the NAC confirm whether alternative detector architectures — such as a fixed 360-degree solid-state CZT SPECT system — are considered acceptable if they meet or exceed the clinical performance and diagnostic intent of the solicitation?

Alternative detector architectures will be considered as long as they meet the clinical performance and diagnostic intent of each of the individual technical specifications.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000The specifications list minimum quantities of interchangeable collimators (LEHR, MEGP, HEGP, pinhole) as well as automatic and semi-automatic collimator changers.

Are these requirements mandatory, or will the VA consider systems that utilize an integrated collimation design inherent to the detector architecture rather than removable collimators?

VA will consider a integrated collimation design.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000Several sections explicitly reference a dual-head configuration.

Is this language intended to prescribe a specific detector geometry, or is it intended to describe minimum clinical capability that could also be met by a full-ring CZT system without mechanically independent detector heads?

It is intended to describe minimum clinical capability.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000Certain performance parameters (e.g., sensitivity reported in cpm/µCi and detector crystal thickness in inches) appear aligned with traditional NaI-based Anger cameras.

Will the NAC accept alternative performance metrics appropriate to CZT-based systems, provided equivalent or superior clinical performance can be demonstrated?

Will accept alternative performance metrics for CZT based systems provided that they are equivalent to what is being asked.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000Should the technical requirements be interpreted as prescriptive (requiring a specific hardware configuration), or functional (allowing alternative technologies that meet the intended diagnostic workflows and outcomes)?Functional; however, it needs to be clear if the response how the response if it is not a direct answer to the tehcnical requirement, how it meets the intended functional requirement.
VISN 20: VA Northwest Network668 SPOKANE, WA668B63000Was the specification developed around a specific existing system or replacement baseline, and if so, is the intent to allow technologically advanced alternatives that differ in detector design but exceed performance expectations?Other detector designs will be considered if they meet or exceed the technical specifications for the relevant clinical function. It needs to be clearly stated in the responses how they meet or exceed that specification if they are not a direct response to the given technical spec.
VISN 22: VA Desert Pacific Healthcare Network664 SAN DIEGO, CA664B60033Numerous specs are specifically based off Siemens. Are you allowing for any fair competition? Should this just be sole sourced?We are allowing for fair competition for all dual source, photon counting CT scanners to provide the most advanced care for our Veterans.
VISN 22: VA Desert Pacific Healthcare Network664 SAN DIEGO, CA664B60033Security/Connectivity Line 75A Integration to Image Guidance system; Please clarify what capability you're referring to with Phillips ultrasound? Are you looking for basic biopsy guidance or more advanced navigation/fusion capabilities? what procedures would you plan to use it for?We are referring to using a image guidance system for advanced navigation and fusion capabilities. Procedures that we will use it for are interventional radiology.
VISN 22: VA Desert Pacific Healthcare Network664 SAN DIEGO, CA664B600331.a. Please confirm this number should be 192 Slices or what is being asked.

1.b. Please change to 80 Cm, only one vendor can meet this spec.

2.o. Remove or move to value add section for fair competition only one vendor can meet this spec.

3.g,h. Please change the requirement to partial UPS. Which will allow saved data and to get the patient off the table. Otherwise you will need a full ups and will need full sighting.

4.f. Please remove or place in value add section this is a lock out spec for all vendors except one.

TIV: For trade in system valuation, please provide number of scans on the tube.

1.a. The minimum number of slices is 2 x 144 1.b. Reducing this value impacts the benefits to our veteran population 2.o. Dual source imaging enables VA San Diego to provide the most advanced imaging for our Veterans 3.g.h The request is for partial, the requirement is for a UPS to support the CT scanner workstations only which will allow saved data and to get the patient off the table.

4.f We are requesting a dual source, photon counting CT scanner to provide the most advanced care for our Veterans.

VISN 20: VA Northwest Network663 SEATTLE, WA663B630043.m. Please clarify, what are power assist table capabilities?
Table should be motorized in the z-axis to go up and down. Power assist can either be motorized in the xy plane or some other mechanism to assist the technologist with moving the table is said xy plane. All vendors don't have motorized capabilities in the xy plane so made the specification more generic to allow better competition for the technologist assist.
VISN 20: VA Northwest Network663 SEATTLE, WA663B63004Would the VA consider revising the specification for 3 wireless detectors to accept fixed detectors in the wall stand and table? Fixed detectors can eliminate the risk of dropping the detector, the need to rotate the detectors in the bucky trays, and the need to monitor detector battery level.VA will consider fixed detectors. Will not be adding a 3rd detector right now.
VISN 20: VA Northwest Network663 SEATTLE, WA663B63004Respectfully request maximum pizel size of 150 be reduced to 148" to allow for open and fair competition.No. 148'' exceeds the specification of 150''
VISN 20: VA Northwest Network663 SEATTLE, WA663B630043m) Please clairfy 'power assist' requirement; is requiement asking for an elevating table?Table should be motorized in the z-axis to go up and down. Power assist can either be motorized in the xy plane or some other mechanism to assist the technologist with moving the table is said xy plane. All vendors don't have motorized capabilities in the xy plane so made the specification more generic to allow better competition for the technologist assist.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B630033h, 3aa, 6e, and 6l. Please remove or move to Value Adds to allow full and fair competition as only one vendor can meet these requirements.Both fully automated and semi-automated collimator configurations are acceptable. QC specs will not be moved to value add. Please respond with how QC is completed on the proposed system.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Which cardiac software package would you prefer - Cedars, Invia or Emory?Cedars
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Please be more specific on what advanced neurology applications you are interested in.Site does datascans and would like any packages that would optimize those.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003The technical specifications reference a dual-detector, variable-angle, dual-head configuration.

Can the NAC confirm whether alternative detector architectures — such as a fixed 360-degree solid-state CZT SPECT system — are considered acceptable if they meet or exceed the clinical performance and diagnostic intent of the solicitation?

Alternative detector architectures will be considered as long as they meet the clinical performance and diagnostic intent of each of the individual technical specifications.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003The specifications list minimum quantities of interchangeable collimators (LEHR, MEGP, HEGP, pinhole) as well as automatic and semi-automatic collimator changers.

Are these requirements mandatory, or will the VA consider systems that utilize an integrated collimation design inherent to the detector architecture rather than removable collimators?

VA will consider a integrated collimation design.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Several sections explicitly reference a dual-head configuration.

Is this language intended to prescribe a specific detector geometry, or is it intended to describe minimum clinical capability that could also be met by a full-ring CZT system without mechanically independent detector heads?

It is intended to describe minimum clinical capability.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Certain performance parameters (e.g., sensitivity reported in cpm/µCi and detector crystal thickness in inches) appear aligned with traditional NaI-based Anger cameras.

Will the NAC accept alternative performance metrics appropriate to CZT-based systems, provided equivalent or superior clinical performance can be demonstrated?

Will accept alternative performance metrics for CZT based systems provided that they are equivalent to what is being asked.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Should the technical requirements be interpreted as prescriptive (requiring a specific hardware configuration), or functional (allowing alternative technologies that meet the intended diagnostic workflows and outcomes)?Functional; however, it needs to be clear if the response how the response if it is not a direct answer to the tehcnical requirement, how it meets the intended functional requirement.
VISN 20: VA Northwest Network663 PUGET SOUND, WA663B63003Was the specification developed around a specific existing system or replacement baseline, and if so, is the intent to allow technologically advanced alternatives that differ in detector design but exceed performance expectations?Other detector designs will be considered if they meet or exceed the technical specifications for the relevant clinical function. It needs to be clearly stated in the responses how they meet or exceed that specification if they are not a direct response to the given technical spec.
VISN 15: VA Heartland Network657 ST. LOUIS, MO657B60003Given the advantages of a c-arm gantry, if Elekta has technologies that meet the facility requirements and meet or exceed the clinical requirements, can Elekta bid a C-arm gantry linac?
No - The requirement is for a LINAC without a C-Arm Gantry.
VISN 10: VA Healthcare System of Ohio655 SAGINAW, MI655B600023.v. Can the UPS requirement be reduced to 7 minutes full functionality?3u. Yes
VISN 10: VA Healthcare System of Ohio655 SAGINAW, MI655B600021.aa. Please clarify if wallstand is needed if the table can perform radiography work including chest exams.

1.cb. Please change detector size to 14x17 to allow full and fair competition.

3.c. Please clarify if an overhead tube is needed only for cross-table lateral work.

3.u. Please move barcode reader to value-add section to allow full and fair competition.

3.z. Please clarify if wallstand is needed if the table can perform radiography work including chest exams.

5.r. Please clarify if you want an embedded compression cone or if a manual compression cone is requested.

7.a. Please move bone suppression to the value-add section in order to allow full and fair competition.

1.aa. Yes 1.cb. Amended 3.c. Overhead tube desired with wall stand configuration 3.u. Amended 3.z. Yes, wall stand required 5.r. Embedded cone 7.a. Amended

VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52002What room is this going to? The specifications do not stipulate a room.This equipment is going to VA Pittsburgh University Drive Campus 3N105 OR# 8 Cystocopy
VISN 21: VA Sierra Pacific Network640 PALO ALTO, CA640B640472.c. Please change minimum SID range from 9" - 23" to allow for full and fair competition.
No. Multiple manufacturers can meet this clinical requirement.
VISN 21: VA Sierra Pacific Network640 PALO ALTO, CA640B640473. S. Please remove for need for UPS on Capital quote add if necessary to the tunkey.No. The VA cannot buy equipment on a turnkey quote due to funding appropriations.

VISN 21: VA Sierra Pacific Network 640 PALO ALTO, CA 640B64047 2c) Please update requiremtnt to 43"-71” SID range to broaden competition 3b) Please edit to allow for ceiling or table mounted lead shielding 3i) Plpease clairfy requiremed 'accessories' for ceiling mounitng aside from required dual monitors.

4d – Table width min 38”. Ours is 33.5”. A wider tabletop does not improve imaging, weight capacity, or patient safety — access, reach, and clearance matter more, and the 33.5" Excel Plus table is optimized for those clinical realities. (See if that’ll work, LOL) 4h) Please clarify if 'Ankle clamps' are they are the same as stirrups, used largely for OB/URO exams?.

7) Please remove, as studies on legacy unit should be sent to long term storage, as they will not be compatible with another vendor General: Will the station consider a remote style RF solution? 2c) No. Multiple manufacturers can meet this clinical requirement.

3b) No, a ceiling or table mounted lead shielding cannot be supported due to space constraints.

3i) 'accessories' is an equipment boom with at least 7 shelves.

4d) Changed in tech specs (not because the patient safety argument really worked, LOL).

4h) Removed from tech spec.

7) No. It is a clinical and VA requirement to store and maintain legacy patient data. We require the vendor to provide project management support & technical resources and required hardware to help migrate and maintain the studies from existing R/F system in some form of storage.

General: No.

VISN 5: VA Capitol Health Care Network613 MARTINSBURG, WV613B610031F. Can the requirement be lowered for fair competition?There is a clinical need for this specification and two or more vendors can meet it
VISN 5: VA Capitol Health Care Network613 MARTINSBURG, WV613B61003Remove term “rings” as that is vendor specific and replace with “Minimum PET detector axial FOV of 26 cm”. Please modified to allow for more competitive bids.There is a clinical need for this specification and two or more vendors can meet it
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B600132.t. Please place "Ability for tabletop to move in cradle motion (side to side)" in value add section as this is not industry standard.
No. The ability to cradle the table is essential for proper patient positioning if a transesophageal echocardiogram is necessary, particularly left lateral position. This is necessary for probe insertion, image acquisition, patient comfort, and airway protection.
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60013Do you want an equipment boom for EP?Yes-prefer Skytron
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60013Do you want Tilt for the Table?Requirement updated to include tilting table
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60013New Med Rad Injector Needed? If so, Pedestal or Table mounted?No, we prefer an ACIST that would be table mounted but able to be removed
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B60013What is the Lab number?C4156
VISN 12: VA Great Lakes Health Care System607 MADISON, WI607B600111.aa. Please change wireless detector for table to fixed.

2.l. Please clarify if wallstand is needed if the table can perform radiography work including chest exams.

2.q. Please change heat unit x-ray tube to 600 kHU to allow for full and fair competition.

3.c. Please clarify if an overhead tube is needed only for cross-table lateral work 5.c. Please change computer hard drive to 1TB to allow for full and fair competition.

6.a. Please move bone suppression to the value-add section to allow for full and fair competition.

1.aa. No, wireless required to reduce trip hazard in small room.

2.l. Spec revised. Wall stand is not needed, room too small.

2.q. Requirement changed to 600 kHU.

3.c. Yes, Overhead Tube for only cross table work, room too small.

5.c. Requirement changed to 1TB.

6.a. Requirement changed to value added section.

VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047LN # 15 XR RAD - 1.a.b.c. System Digital Detector Configuration:
Q: Would the VA consider revising the specification for 3 wireless detectors to accept fixed detectors in the wall stand and table? Fixed detectors can eliminate the risk of dropping the detector, the need to rotate the detectors in the bucky trays, and the need to monitor detector battery level.This would be considered.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047LN # 15 XR RAD - 1.a.b.c. System Digital Detector Configuration:
Q: Would the VA consider revising the specification for 3 wireless detectors to accept fixed detectors in the wall stand and table? Fixed detectors can eliminate the risk of dropping the detector, the need to rotate the detectors in the bucky trays, and the need to monitor detector battery level.This would be considered.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Line 15 XR RAD 2.i. Minimum image plate height from floor:
Q: Please change minimum image plate height from floor to 13 in to allow full and fair competition.When doing standing feet or ankle scans we need it as low as possible to prevent falls from a step stool. We cannot increase this height beyond 11.5 in.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Line 15 XR RAD 3.t. ceiling and floor rail
Q: Can you please clarify the request for a ceiling and floor rail?Some systems have a floor mounted system and as such would require a floor rail. For room standardization, we require a floor rail in addition to a ceiling rail.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 50e/51f Table holder for wireless/Mobile Holder for wireless; Which method for charging is preferred; mobile charging holder or in table charger holder?Table charging holder
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 54 I Automatic tube crane with protocol based movement (option) ; Please clarify if you would like this as an option and should the software for automation also be an option (which was requested in Software Requirements Line 84 F?)Yes, we would like both as an option.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications; Line L Free Floating manual tabletop; Please clarify is the "ask" is for a free floating table or a motor driven table?Free floating table top
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications; Line 58M Motor-driven tabletop; If the tabletop requested above is free floating, what is the need for the motorization in the X and Y axis? Can this specification be removed?Yes the requirement for motorization in the X, Y axis can be removed. We need it motorized in the z-axis (up and down).
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 61P Removeable/adjustable patient hand grips; Duplicate specification, listed on Line 67VThe removeable/adjustable patient hand grips are for the wall bucky for patient use and a requirement. The ones listed in 67V are for the technician to move the table top but these are a preferred option and not a requirement and can be removed
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 62 Q ; Please clarify if this requirement is needed since Bone Suppression Software is requested on Line 80BThese are distinct requirements and both required. The dual energy is not only for bone supression.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 64 S Maintains full functionality of the entire system for a minimum of 6 minutes during a power outage; Please clarify why a Full UPS is needed for an Xray room and if the facility has the space, electrical and HVAC to support this system? Please consider removing this specification.Spec removed. Workstation UPS is a requirement and would not require specfic space, electrical or HVAC concerns.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Additional Specifications Line 65T Ceiling and Floor rail; Why is the floor rail needed for an overhead tube crane system? Please consider removing this specification.Some systems have a floor mounted system and as such would require a floor rail. For room standardization, we require a floor rail in addition to a ceiling rail.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B60047Software Requirements Line 84 F Auto Positioning Software ; Line 54 I this is listed as optional , please clarify the "ask"If the Automatic tube crane with protocol based movement is offered, then we require the auto positioning software.
VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B600471.cc. Please change wireless detector weight to 9 lbs to allow for full and fair competition.

2.i. Please change image plate height from floor to 11.5 inches to allow for full and fair competition.

3.t. Please clarify the floor railing need.

1.cc the weight can be increased to 7lbs to match the other wireless detector required but cannot be increased to 9 lbs for a 14x17in detector.

2.i. This requirement can be changed to 11.5 in.

3.t. Some systems have a floor mounted system and as such would require a floor rail. For room standardization, we require a floor rail in addition to a ceiling rail.

VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B600473t) Please specify if this requirement is for a rail mounted moveable wallstandYes
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B68008Max system weight set to 5500 lbs, can this be expendad to 5600 lbs?Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680082.k. Please clarify if a partial or full UPS is desired? Are you connected to back up facility power? If you're looking for a full UPS is there siting to accommodate? If you're looking for partial only, please adjust from 10 minutes to "allow for saving exam, and getting patient off the system."

3.f. Please clarify if a partial or full UPS for workstation is desired? Are you connected to back up facility power? If you're looking for a full UPS is there siting to accommodate? If you're looking for partial only, please adjust from 10 minutes to "allow for saving exam, and getting patient off the system" 2.k Partial UPS. Yes connected to back up facility power. No, we are nto looking for full UPS. For partial, at minimum 10 minutes.

3.f Partial UPS. Yes connected to back up facility power. No, we are nto looking for full UPS. For partial, at minimum 10 minutes.

VISN 12: VA Great Lakes Health Care System578 HINES, IL578B68008Regarding trade in, has there been any tube replacements? Please provide the date the new tube was installed if applicable.Yes, April 10, 2024.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680043.t. Can the UPS requirement be reduced to 7 minutes full functionality?Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680042.f. Maximum pixel size [µm] - for chest x-rays, 140um is optimal, please adjust.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680043.r. Linear tomography - please remove to increase competition or move to ADDED VALUE.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680044.d. Minimum processing workstation monitor size [in] - please change 23.8" to 23".Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680042.g. Minimum patient table load capacity (static) [lbs] - please change 880 lbs to 770 lbs top increase competition, unless you have 800 lb patientsSpec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680042.h. Minimum patient table load capacity (dynamic) [lbs]- please change 705lbs to 683 lbs.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680023.t. Can the UPS requirement be reduced to 7 minutes full functionality?Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680022.f. Maximum pixel size [µm] - for chest x-rays, 140um is optimal, please adjust.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680023.r. Linear tomography - please remove to increase competition or move to ADDED VALUE.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680024.d. Minimum processing workstation monitor size [in] - please change 23.8" to 23".Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680022.g. Minimum patient table load capacity (static) [lbs] - please change 880 lbs to 770 lbs top increase competition, unless you have 800 lb patientsSpec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B680022.h. Minimum patient table load capacity (dynamic) [lbs]- please change 705lbs to 683 lbs.Spec has been revised.
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B580068.f What is expected for the Advanced Visualization package?There is no 8.f requirement for Advanced Visualization
VISN 12: VA Great Lakes Health Care System578 HINES, IL578B580061.f. Please change channel specifications to reflect the maximum number of channels per Field of View (FOV) without table movement each generating an independent partial image; not simply the number of elements that can be connected.

The 204 Channel specification is only used by one vendor and does not reflect the true number of RF channels in a system.

3.f. Please clarify if a partial or full UPS is desired? Are you connected to back up facility power? If you're looking for a full UPS is there siting to accommodate? If you're looking for partial only, please adjust from 10 minutes to "allow for saving exam, and getting patient off the system."

5.m. Is an endorectal coil required or just the ability/preference to do high-quality prostate imaging?

1.f 64 minimum channel spec is standard

3.f Partial UPS is sufficient. Must maintain functionality for minimum of 10 minutes. Spec revsied.

5.m An endorecal coil is not required. The bility to do high-quality prostate imaging is required.

VISN 12: VA Great Lakes Health Care System 556 NORTH CHICAGO, IL 556B61003 2.f. Please change the minimum image plate height from floor to 11.5 inches to allow for full and fair competition.

2.p. Please change minimum SID range to 25 to allow for full and fair competition.

The plate height will not be altered since increasing the minimum from 10 in to 11.5 in would tighten the requirements.

2p. This is not a technical requirement of this matrix.

VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B610032.c. Please confirm generator voltage range [kV] at 220Range has been updated to 50kV to 220kV
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B610032.d. Maximum spatial resolution [lp/mm] - Please change to 3.57Requirement has been revised.
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B600013.u. Please clarify if a partial or full UPS is desired? Are you connected to back up facility power?

If you require a full UPS is there siting to accommodate?

Alternatively, if you're looking for partial only UPS, please adjust from 10 minutes to "allow for saving exam, and getting patient off the system."

3.h. Please remove or move to Value Adds in order to allow for full and fair competition. Only one vendor can meet this requirement.

3.cc. Please remove or move to Value Adds in order to allow for full and fair competition. Only one vendor can meet this requirement.

5.c. What 3rd party cardiac SW is preferred… Cedars or 4DM?

5.i. Please remove as this is not typically done in an NM department 6.e. Please remove or move to Value Adds in order to allow for full and fair competition. Only one vendor can meet this requirement.

6.j. Please remove or move to Value Adds in order to allow for full and fair competition. Only one vendor can meet this requirement.

3u. Full UPS - requirement has been revised

3h. Requirement has been revised

3cc. Feature moved to Value-Add section.

5c. 4DM

5i. No, half time imaging is a standard pratice at this facility.

6e. Requirement has been revised

6j. No, this is standard of care nuclear imaging processing at this facility.

VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001Which cardiac software package would you prefer - Cedars, Invia or Emory?4DM
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001The technical specifications reference a dual-detector, variable-angle, dual-head configuration.

Can the NAC confirm whether alternative detector architectures — such as a fixed 360-degree solid-state CZT SPECT system — are considered acceptable if they meet or exceed the clinical performance and diagnostic intent of the solicitation?

Acceptable if clinical functionality is not impacted
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001The specifications list minimum quantities of interchangeable collimators (LEHR, MEGP, HEGP, pinhole) as well as automatic and semi-automatic collimator changers.

Are these requirements mandatory, or will the VA consider systems that utilize an integrated collimation design inherent to the detector architecture rather than removable collimators?

Not mandatory if clinical functionality is not impacted.
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001Several sections explicitly reference a dual-head configuration.

Is this language intended to prescribe a specific detector geometry, or is it intended to describe minimum clinical capability that could also be met by a full-ring CZT system without mechanically independent detector heads?

It is intended to describe minimum clinical capability
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001Certain performance parameters (e.g., sensitivity reported in cpm/µCi and detector crystal thickness in inches) appear aligned with traditional NaI-based Anger cameras.

Will the NAC accept alternative performance metrics appropriate to CZT-based systems, provided equivalent or superior clinical performance can be demonstrated?

Yes, as long as it is comparable to traditional systems.
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001Should the technical requirements be interpreted as prescriptive (requiring a specific hardware configuration), or functional (allowing alternative technologies that meet the intended diagnostic workflows and outcomes)?Functional.
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B60001Was the specification developed around a specific existing system or replacement baseline, and if so, is the intent to allow technologically advanced alternatives that differ in detector design but exceed performance expectations?There is a specific intended function for this system to compliment existing services. "Technologically advanced alternatives" could be acceptable if required clinical capabilities are met.
VISN 7: VA Southeast Health Care Network534 CHARLESTON, SC534B62006 --> PO changed to 534B620202.j. Please clarify if a partial or full UPS is desired? Are you connected to back up facility power? If you require a full UPS is there siting to accommodate?

Alternatively, if you're looking for partial only UPS, please adjust from 10 minutes to "allow for saving exam, and getting patient off the system."

We at least need a UPS that maintains functionality of the workstation to properly shutdown equipment during a power outage – Full system UPS has been added to “Added Value” section and may be considered if space and budget allow
VISN 7: VA Southeast Health Care Network534 CHARLESTON, SC534B62006 --> PO changed to 534B62020Line 21 0.5 mm is vendor specific. Request is 0.625m for fair competitionYes, Tech Specs have been updated
VISN 7: VA Southeast Health Care Network535 CHARLESTON, SC534B62006 --> PO changed to 534B62020Line 52 foot and table slickers. Request utilization of table pads vs slicker for fair competitionYes, Tech Specs have been updated
VISN 7: VA Southeast Health Care Network536 CHARLESTON, SC534B62006 --> PO changed to 534B62020Line 47 Rear Gantry Display. Request use rear gantry controls vs. display. Language correction requestYes, Tech Specs have been updated
VISN 7: VA Southeast Health Care Network537 CHARLESTON, SC534B62006 --> PO changed to 534B62020Line 29 Request clarification on 140 msec Maximum non-Segmented reconstruction timeThis is not required, so line has been removed.
VISN 20: VA Northwest Network531 BOISE, ID531B630025.d. Please change minimum table width to 31 in. to allow for full and fair competition.
Change to min table width of 31 in is approved.
VISN 20: VA Northwest Network531 BOISE, ID531B630023u) Is a TIMMS unit required or does the quoted solution simply need to be compatible?Quoted solution simply needs to be compatible with TIMMS.
VISN 20: VA Northwest Network531 BOISE, ID531B630001.bc. Please change maximum wireless detector weight to 7lbs to allow for full and fair competition.

1.bd. Please change IP67 rating to IPX4 rating to allow for full and fair competition.

2.k.,2.l.,2.m. Please clarify that a table is not needed as these specifications refer to table.

1.bc. Change to 7 lbs is approved. 1bd, Change to IPX4 is approved. 2. Confirmed table is not needed. Please disregard the specs.
VISN 20: VA Northwest Network531 BOISE, ID531B630003. N. Please remove need for UPS on Capital quote add if necessary to the turnkey.The UPS needs to remain on the Capital quote as it is equipment
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B640881.b. Please clarify image intensifier. This is outdated technology.Minimum flat-panel detector field-of-view (FOV): 48 cmA flat-panel detector is required.
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B640881.i. Please clarify 120in x 120in requirement.This restriction ensures the system physically fits within the room layout, leaves adequate clearance for safe patient and staff movement, and complies with facility space planning constraints. The largest horizontal area (width x length) occupied by the R/F system—when fully assembled and installed—must not exceed a rectangle of 120 inches by 120 inches (that is, 10 feet by 10 feet). This includes all main components such as the gantry, C-arm, patient table, control panels, and any fixed system accessories that form part of the assembled unit.
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B64088General: Will any additional upper body shields be required?No
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B64088General: Will any surgical lights be required?No
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B64088General: Will a contrast media injector be required?No, a contrast injector is not included in the procurement, however the current contrast injector will be interfaced to the selected equipment
VISN 2: VA Healthcare Network Upstate New York528 SYRACUSE, NY528B64088General: Will metal artifact reduction capability for Cone Beam CT be required?No
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Please confirm the minimum detector width z-axis [mm]. There is currently no conventional clinical CT Scanner that provides 192 mm (19.2cm) z-axis coverage.We confirm it is 40mm. Tech Specs have been updated.
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Please confirm the minimum reconstructed slice width [mm].We confirm it is 40mm. Tech Specs have been updated.
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Does CT Fluoroscopy need to be ceiling mounted or on a pedestal?If referring to the contrast injector or monitor, both should be ceiling mounted.
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Is there a identified contrast injector vendor make/model?Bayer Stellant D
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Please confirm if CT Perfusion listed under Advanced Applications is for Body Perfusion, Brain Perfusion, or both.We confirm it is for both. Techs Specs have been updated.
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001In Section C. Training Requirements, Line E indicates Technologists shall receive continuing education credits (CME’s). Please confirm if this is supposed to be CEU’s, instead of CME’s.We confirm this is for CEUs. Tech Specs have been updated.
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B620011.c. Please change minimum detector width to 40mm for fair competition/ industry standard.

1.d. Please update scan field of view FOV) to 50cm for fair competition.

2.k. Full UPS - Please confirm a Full UPS is required, and is there siting to accommodate a full UPS?

Alternatively, please modify tech specs to partial UPS, adjusting to 10mins to "allow for saving exam and getting patient off the system."

1.c. Updated to 40mm 1.d. Updated to 50cm 2.k. A full UPS is not required; Tech Specs have been updated with suggested language
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B62001Line 23 matrix states 192mm z axis detector. Requesting a minimum of 80mm for more accurate and fair competition. No known 192mm in ITN website.Tech Specs have been updated on minimum z axis detector width to 40mm.
VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON, DE460B68004What Model is the current IVUS system?Philips IntraSight 5