This is a compliance matrix for an ultrasound system procurement at the Clement J. Zablocki VA Medical Center in Milwaukee, WI. The document outlines technical requirements for one (1) ultrasound system for the Surgical Services-ENT Clinic to perform fine needle aspirations of head and neck lesions/masses.

The required specifications include a minimum 23-inch screen size, 4 transducer ports, 40-minute minimum battery life, and multiple scanning modes including 2D, M-Mode, and various Doppler capabilities. Two specific transducers are required: ML6-15-D Matrix Linear Array Probe (4.5-15MHz) and L6-24-D Linear Array Probe (6.0-24MHz). The system must include DICOM compatibility, wireless connectivity with FIPS 140-2/3 compliance, and encrypted hard drive. Clinical training requirements include on-site training for 10 technologists and 10 physicians, with continuing education credits. The warranty must cover all parts and labor for one year following acceptance. The facility is offering a Phillips Sparq ultrasound system (Serial #US71410120, installed 7/15/2014) for potential trade-in credit.

View the file

Other files

File Type Posted
SF30 - Amendment 0004.pdf PDF
SEP 2024 Equipment Only Consolidation Schedule - Revised2.pdf PDF
Amendment 0004 - Continuation Page.pdf PDF
SF30 - Amendment 0003.pdf PDF
358B49001 - 695B48011.zip ZIP file
SEP 2024 EQ Consolidation - Schedule of Line Items w QA - ROUND 2.xlsx XLSX spreadsheet
SEP 2024 Equipment Only Consolidation Schedule - Revised.pdf PDF
Amendment 0003 - Continuation Page.pdf PDF
SF30 Amendment 0002.pdf PDF
644B48042 - XR US VAMC PHOENIX AZ_Compliance matrix updated.xlsx XLSX spreadsheet
Amendment 0002 - Continuation Page.pdf PDF
644B48043 - XR US VAMC PHOENIX AZ_Compliance matrix updated.xlsx XLSX spreadsheet
SEP 2024 EQ Consolidation - Schedule of Line Items w QA.xlsx XLSX spreadsheet
Amendment 0001 - Continuation Page.pdf PDF
SEP 2024 EQ Consolidation - Schedule of Line Items w QA - Updated.xlsx XLSX spreadsheet
573B45026 - 644B48042.zip ZIP file
SF30 Amendment 0001.docx DOCX document
644B48043 - 757B50001.zip ZIP file
358B49001 - 573B45024.zip ZIP file
Vendor Folders.zip ZIP file
691B40053 - 757B50001.zip ZIP file
516B40042 - 548B40010.zip ZIP file
499B49018 - 516B40041.zip ZIP file
548B40020 - 558B49028.zip ZIP file
654B30016 - 673B40053.zip ZIP file
673B40056 - 691B40052.zip ZIP file
SEP 2024 EQ Consolidation - RFO.pdf PDF
Memorandum for Training.pdf PDF
629B40355 - 654B30014.zip ZIP file
558B49030 - 596B40037.zip ZIP file
VA Directive 6550 - New Appendix A.pdf PDF
SEP 2024 Equipment Only Consolidation Schedule.pdf PDF
36A79725Q0002_1.docx DOCX document
SEP 2024 EQ Consolidation - Schedule of Line Items.xlsx XLSX spreadsheet
Vendor Workup Spreadsheet - 36A79725Q0002.xlsx XLSX spreadsheet
101EH2069 - 499B49015.zip ZIP file
596B40038 - 621B40006.zip ZIP file
Show all 37

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Compliance Matrix

UltrasonicCompliance Matrix Created:
V121/23/25
MILWAUKEE WI
695-B48007
OFFEROR
MANUFACTURER
MODEL
QUOTE NUMBER

VENDOR INSTRUCTIONS:

Please fill out the table above with the offeror name (if different than the device manufacturer), manufacturer name, model name/number, and the quote number. All required specifications are listed below. Please review the required specifications and provide supporting information about the proposed model. The "Vendor Response" column is limited to the following entries: "exceeds", "meets," "partially meets," and "does not meet." In the "Documentation" column, please indicate the location within the attached documentation (file name, page number) for information to support the vendor response. Please provide any additional information/explanations in the "Vendor Explanation" column. **NOTE THAT THIS SPREADSHEET HAS BEEN LOCKED TO ALLOW ONLY FOR FILLING OUT REQUIRED FIELDS (THE TABLE ABOVE AND COLUMNS E, F, & G). DO NOT MAKE ANY FORMATTING OR CONTENT CHANGES TO THE COMPLIANCE MATRIX.**

REQUIREMENT OVERVIEWVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
The Milwaukee surgical services ENT staff located at Clement J. Zablocki VA Medical Center require procurement of one (1) ultrasound to continue offering clinic based fine needle aspirations of head and neck lesions/masses.
Facility Name
Clement J. Zablocki VA Medical Center
Delivery Address
5000 W National Ave. Milwaukee, WI 53295
Requesting Department(s)
Surgical Services-ENT Clinic
Quantity
1
TECHNICAL REQUIREMENTSVENDOR RESPONSEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Unit physical specifications
a. Minimum screen size [in]
23
b. Minimum active screen size [in]
24
c. Minimum monitor rotation [degrees]
13.5 both directions
d. Minimum tilting [degrees]
4.5
e. Monitor height range from floor [in]
48 - 72
f. Minimum number of transducer ports
4
g. Minimum image storage [days or TB]
1
h. Minimum battery life [min]
40
i. Maximum equipment dimensions (HxWxD) [in]
50/23/35
j. Maximum equipment weight [lbs]
190
2. Scanning modes
☒ a. Two-dimensional (2D)
☒ b. Tissue Harmonic Imaging
☒ c. M-Mode
☒ d. Simultaneous M-Mode
☒ e. Pulsed Wave Doppler
☒ f. Continuous Wave Doppler
☒ g. Velocity Color Doppler
☒ h. Color Power Doppler
☒ i. Tissue Doppler Imaging
☒ j. Two-dimensional (2D)
3. Doppler displays
☒ a. Frequency
☒ b. Velocity
☒ c. Power (microvascular/directorial)
☒ d. Duplex
☒ e. Steering for color Doppler is required to be 3° increments or less
4. Control panel specifications
☒ a. Physical keyboard
☒ b. Touchscreen monitor
☒ c. Keyboard on touchscreen
☒ d. Black-and-white printer
☒ e. Articulating support arm to allow for vertical and horizontal adjustment of the monitors for viewing from anywhere in the room
5. Additional specifications
☒ a. Panoramic mode
☒ b. Multi-modality display
☒ c. Image annotation
☒ d. One-button image optimization
☒ e. One-button Doppler optimization
☒ f. Programmable protocols
☒ g. Built-in, customizable PACS-compatible form
☒ h. Ability to create patient worklist without order or network connection
☒ i. Specific applications and workflows for all imaging specialties, including interventional guidance
☒ j. Beam steering
☒ k. Ability to enter standby mode or sleep mode
☒ l. Needle enhancement
☒ m. Image editing – easy video editing and annotation
6. Analysis packages
☒ a. Procedural
☒ b. Musculoskeletal Clinical
☒ c. Small Parts
☒ d. Vascular – measurement and analysis of vessels
☒ e. Obstetric
☒ f. Breast
☒ g. Cardiovascular Clinical
☒ h. Thyroid productivity package
Vendors must include in their offers analysis packages that meet the criteria above. Please include all other analysis packages offered by your company in the optional section on the quotes.
7. Transducers
Description
Qty
Frequency Range [MHz]
a. ML6-15-D Matrix Linear Array Probe
1
4.5 – 15MHz
b. L6-24-D Linear Array Probe
1
6.0 – 24MHz
Vendors must include in their offers transducers that meet the criteria above. Please include all other transducers offered by your company in the optional section on the quotes.
8. Advanced features
☒ a. Image storage for any captured images
9. Security/Connectivity requirements
☒ a. OEM-supported operating system
☒ b. Latest DICOM print, store, commit, structured report (SR) and modality worklist
☒ c. Wireless connectivity to VA network – Compatible with 802.11b/g/n and FIPS 140-2 or FIPS-3 compliant
☒ d. Encrypted hard drive
☒ e. PACS compatibility – VA VistA Imaging
TRAINING REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. Clinical Training
☒ a. On-site clinical applications training for 10 technologists during go-live
☒ b. On-site follow-up clinical applications training for 10 technologists once technologists have hands-on experience with the system
☒ c. On-site clinical applications training for 10 physicians during go-live
☒ d. Technologists who complete the clinical applications training shall receive continuing education credits (CMEs).
☒ e. Vendors shall be responsible for accommodating different personnel shifts for clinical applications training during go-live.
2. Biomedical Technician Training
Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.
Technical training information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.
Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.
SERVICE REQUIREMENTSVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
1. VPN/Remote Access – The vendor shall provide any and all equipment service programs, such as remote diagnostics, during the warranty period. The vendor shall provide post-warranty remote diagnostic service program as an “Add Option” with the offer. The system shall provide vendor remote diagnostics via VPN. The vendor shall either utilize the VA national site-to-site VPN or work with the Office of Cyber and Information Security and the VAMC Information Systems Security Officer to establish a client-based VPN.
2. Service and Operator Manuals – The vendor shall provide the following documentation for the proposed systems:
a. Two (2) copies of operator instruction manuals (one (1) electronic and one (1) physical copy)
b. Two (2) copies of a service manuals (one (1) electronic and one (1) physical copy)
*Vendors can include the physical copy as a priced line item in their quote as applicable.
3. Minimum Warranty – The system and accessories shall be covered under the manufacturer’s warranty and shall include all parts and labor for one year following acceptance by the VAMC. This warranty must include PMs as required by the manufacturer. The manufacturer’s factory-trained field service representatives shall perform installation and maintenance during the warranty period.
Vendors are encouraged to include any offerings for service, warranty, and training that may exceed the minimum requirements, to include information on their service support structure during and after the warranty period. Vendors who do not include any added value offerings for service, warranty, and training will not be docked or excluded from consideration. However, any such offerings will be evaluated based on added value.
ADDED VALUEVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Specifications listed below are not required, but preferred. Vendors who do not include the below specifications in the submitted offer will not be docked or excluded from consideration. Specifications listed below will be evaluated based on added value.
☒ a. Additional year(s) of warranty
☒ b. Post-warranty remote diagnostic service program
☒ c. Version/platform long-range plan
☒ d. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7
☒ e. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device receives Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020)), and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) from worklist provider.

☒ f. Device compatible with Cerner CareAware MultiMedia (CAMM) Release 7:

Device sends Patient ID (0010, 0020), Issuer of Patient ID (0010, 0021), Other Patient IDs Sequence Patient ID (0010, 1002 (0010, 0020))and Other Patient IDs Sequence Issuer of Patient ID (0010, 1002 (0010, 0021)) to the storage location.

☒ g. List optionally purchasable fusion and navigation options
☒ h. Tracking and recording of needle movement during biopsy in 3D coordinates in real-time
☒ i. Availability of projected, future needle movement during biopsy in 3D coordinates
☒ j. 3D-generated model of prostate that can be exported to report, adjustable by additional non-imaging information, such as pathology-related data
OTHER INFORMATION/DOCUMENTATION REQUESTEDVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
Please reference the “Instructions to Offers” for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” for response format.
1. Completed pre-procurement assessment form (6550 Appendix A)
2. Completed Manufacture Disclosure Statement for Medical Device Security (MDS2) form
3. Federal Information Processing Standard (FIPS) 140-2 or (FIPS) 140-3 certification
4. Product brochures
5. Technical specification sheets, to include dimensions and weight of the system
6. Typical drawings (pdf version of the CAD drawings)
7. Technical training- Biomedical: information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type.
Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.
8. Support information to include your company’s support structure during and after the warranty period
On-line or telephonic applications support and availability (include third party coverage)
A listing of field service engineer locations and availability
A listing of part depots
TRADE-INVENDOR COMPLIANCEVENDOR DOCUMENTATIONVENDOR EXPLANATION
☒ a. In instances where sanitization of ePHI compromises the OS and/or application software, or requires the removal of internal storage media, the vendor accepts the equipment “as is” and can elect at their own discretion to contract with the original equipment manufacturer (OEM) to restore the system.
The following equipment is available for trade-in. Please reflect any credits provided for trade-in equipment in the proposal.
Station
695
Manufacturer
Phillips
Model
Sparq
EE/Asset Number
86396
Serial Number
US71410120
Current Condition
● Operational
Installation Date
07/15/2014
☒ a. VA has no ancillary equipment to offer.

File details come from the government source that posted it. Updated .