May 2022 Consolidation - Schedule of Line Items with QAs - Updated 8-24-22.xlsx
XLSX spreadsheet 50 KB Posted
- Attached to
- 6525--May 2022 Consolidation
- Solicitation number
- 36A79722Q0013
View the file
Other files
Show all 39
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Schedule of Line Items
| HTME VISN | Station | Requisition Number in Action | Equipment Type | Qty | Benchmark (BM) or Technical Specification (TS) | |
| VISN 19: VA Rocky Mountain Network | 436 FT HARRISON, MT | 436B20008 | XR US | 1 | TS | |
| VISN 19: VA Rocky Mountain Network | 436 FT HARRISON, MT | 436B20009 | XR US | 1 | TS | |
| VISN 19: VA Rocky Mountain Network | 436 FT HARRISON, MT | 436B20010 | XR US | 1 | TS | |
| VISN 19: VA Rocky Mountain Network | 436 FT HARRISON, MT | 436B20011 | XR US | 2 | TS | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26001 | NM BONE DEN | 1 | TS | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26002 | XR RAD | 1 | TS | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26003 | XR RAD | 1 | TS | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26005 | XR US | 1 | BM | |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 NORTHFIELD, NJ | 460B28012 | XR CT | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 NORTHFIELD, NJ | 460B28013 | XR RAD | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B28019 | XR US | 5 | TS | |
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29003 | XR RAD | 1 | TS | |
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29004 | XR RAD | 1 | TS | |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B29005 | XR US | 1 | BM | |
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29006 | XR US | 1 | BM | |
| VISN 23: VA Midwest Health Care Network | 636 OMAHA, NE | 499B29007 | XR PORT C-ARM | 2 | TS | |
| VISN 23: VA Midwest Health Care Network | 636 OMAHA, NE | 499B29008 | XR US | 1 | BM | |
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29010 | XR US | 1 | BM | |
| VISN 23: VA Midwest Health Care Network | 636 OMAHA, NE | 499B29011 | XR US | 1 | BM | |
| VISN 4: VA Stars & Stripes Healthcare Network | 503 ALTOONA, PA | 503B20024 | NM NUC | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | XR RAD | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20003 | NM BONE DEN | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | XR RAD | 1 | TS | |
| VISN 7: VA Southeast Health Care Network | 509 AUGUSTA, GA | 509B22009 | XR US | 2 | TS | |
| VISN 7: VA Southeast Health Care Network | 521 BIRMINGHAM, AL | 521B22010 | XR US | 2 | BM | |
| VISN 7: VA Southeast Health Care Network | 521 BIRMINGHAM, AL | 521B22011 | XR US | 1 | BM | |
| VISN 2: VA Healthcare Network Upstate New York | 528 BUFFALO, NY | 528B20037 | XR US | 1 | BM | |
| VISN 2: VA Healthcare Network Upstate New York | 528 SYRACUSE, NY | 528B24270 | XR US | 1 | TS | |
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B27046 | XR PORT RAD | 1 | BM | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B24002 | XR US | 1 | TS | |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B22028 | XR US | 6 | TS | |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B22031 | XR US | 4 | TS | |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B22032 | XR PORT C-ARM | 2 | TS | |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B22043 | XR US | 1 | TS | |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B22047 | XR US | 1 | TS | |
| VISN 12: VA Great Lakes Health Care System | 537 CHICAGO, IL | 537B28006 | NM SPECT CT | 1 | TS | |
| VISN 12: VA Great Lakes Health Care System | 537 CHICAGO, IL | 537B28007 | XR US | 1 | TS | |
| VISN 10: VA Healthcare System of Ohio | 539 CINCINNATI, OH | 539B20032 | XR US | 2 | BM | |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B20044 | XR US | 2 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22025 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22043 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22044 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22045 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22046 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22047 | XR PORT C-ARM | 1 | TS | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22051 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22052 | XR PORT RAD | 1 | TS | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22055 | XR US | 1 | BM | |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22056 | XR US | 1 | BM | |
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | XR RAD | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20021 | XR RAD | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20052 | XR RAD | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20065 | XR RAD | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20066 | XR RAD | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20067 | XR RAD | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 562 ERIE, PA | 562B20027 | XR CT | 1 | BM | |
| VISN 16: South Central VA Health Care Network | 564 FAYETTEVILLE, AR | 564B10043 | NM BONE DEN | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 564 FAYETTEVILLE, AR | 564B20006 | XR US | 1 | BM | |
| VISN 16: South Central VA Health Care Network | 564 FAYETTEVILLE, AR | 564B20010 | XR US | 1 | BM | |
| VISN 16: South Central VA Health Care Network | 564 FAYETTEVILLE, AR | 564B20013 | XR PORT C-ARM | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 573 GAINESVILLE, FL | 573B25014 | XR C-ARM EP | 1 | TS | |
| VISN 12: VA Great Lakes Health Care System | 578 HINES, IL | 578B28009 | XR PORT C-ARM | 1 | BM | |
| VISN 16: South Central VA Health Care Network | 580 HOUSTON, TX | 580B20048 | XR PORT C-ARM | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 580 HOUSTON, TX | 580B20049 | XR US | 3 | BM | |
| VISN 16: South Central VA Health Care Network | 580 HOUSTON, TX | 580B20052 | XR PORT C-ARM | 1 | BM | |
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20034 | XR CATH LAB | 1 | TS | |
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20036 | XR US | 2 | TS | |
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20046 | XR PORT RAD | 6 | TS | |
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20056 | XR US | 3 | TS | |
| VISN 15: VA Heartland Network | 589 WICHITA, KS | 589B21003 | XR C-ARM | 3 | TS | |
| VISN 15: VA Heartland Network | 589 WICHITA, KS | 589B21004 | XR C-ARM | 1 | TS | |
| VISN 15: VA Heartland Network | 589 WICHITA, KS | 589B21005 | XR C-ARM | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 595 LEBANON, PA | 595B19029 | XR US | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 595 LEBANON, PA | 595B29011 | NM NUC | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 598 N. LITTLE ROCK, AR | 598B20017 | XR RF | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 598 N. LITTLE ROCK, AR | 598B20026 | XR US | 1 | TS | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20017 | NM PET CT | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20018 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20023 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20025 | XR PORT C-ARM | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20026 | XR RAD | 1 | TS | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20027 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20010 | XR PORT C-ARM | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20011 | XR PORT C-ARM | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20013 | XR PORT C-ARM | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20014 | XR PORT C-ARM | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20021 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20022 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20024 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20025 | XR US | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 605 LOMA LINDA, CA | 605B20026 | XR US | 1 | BM | |
| VISN 12: VA Great Lakes Health Care System | 607 MADISON, WI | 607B20003 | XR US | 1 | BM | |
| VISN 12: VA Great Lakes Health Care System | 607 MADISON, WI | 607B20004 | XR US | 1 | BM | |
| VISN 12: VA Great Lakes Health Care System | 607 MADISON, WI | 607B20006 | XR CATH LAB | 1 | BM | |
| VISN 9: VA Mid South Healthcare Network | 621 MTN HOME, TN | 621B20036 | XR US | 1 | BM | |
| VISN 9: VA Mid South Healthcare Network | 621 MTN HOME, TN | 621B20037 | XR US | 1 | BM | |
| VISN 9: VA Mid South Healthcare Network | 621 MTN HOME, TN | 621B20038 | XR US | 1 | BM | |
| VISN 9: VA Mid South Healthcare Network | 621 MTN HOME, TN | 621B20039 | XR US | 1 | BM | |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B20018 | XR US | 10 | TS | |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B20027 | XR CT | 1 | TS | |
| VISN 2: VA Healthcare Network Upstate New York | 632 NORTHPORT, NY | 632B21010 | XR PORT C-ARM | 1 | BM | |
| VISN 19: VA Rocky Mountain Network | 635 OKLAHOMA CTY, OK | 635B20007 | XR MRI LINAC | 1 | TS | |
| VISN 23: VA Midwest Health Care Network | 636 OMAHA, NE | 636B20077 | XR CATH LAB | 1 | TS | |
| VISN 22: VA Desert Pacific Healthcare Network | 644 PHOENIX, AZ | 644B28050 | XR US | 2 | TS | |
| VISN 22: VA Desert Pacific Healthcare Network | 644 PHOENIX, AZ | 644B28054 | XR US | 2 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B22013 | XR US | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B22014 | XR US | 1 | TS | |
| VISN 20: VA Northwest Network | 648 PORTLAND, OR | 648B12486 | XR RAD | 1 | TS | |
| VISN 20: VA Northwest Network | 648 PORTLAND, OR | 648B12487 | XR RAD | 1 | TS | |
| VISN 20: VA Northwest Network | 648 PORTLAND, OR | 648B22537 | XR CT | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B29002 | XR MAMMO | 2 | TS | |
| VISN 21: VA Sierra Pacific Network | 654 RENO, NV | 654B20018 | XR US | 1 | BM | |
| VISN 23: VA Midwest Health Care Network | 656 ST. CLOUD, MN | 656B29003 | XR MRI | 1 | TS | |
| VISN 15: VA Heartland Network | 657 MARION, IL | 657B22006 | XR RF | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 658 SALEM, VA | 658B29003 | XR US | 1 | TS | |
| VISN 6: VA Mid-Atlantic Health Care Network | 658 SALEM, VA | 658B29004 | XR US | 3 | TS | |
| VISN 19: VA Rocky Mountain Network | 660 SALT LKE CTY, UT | 660B20008 | XR US | 1 | TS | |
| VISN 21: VA Sierra Pacific Network | 662 SAN FRAN, CA | 662B20015 | XR US | 1 | TS | |
| VISN 21: VA Sierra Pacific Network | 662 SAN FRAN, CA | 662B20016 | XR US | 1 | TS | |
| VISN 20: VA Northwest Network | 663 TACOMA, WA | 663B23007 | XR PORT C-ARM | 1 | TS | |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663B23013 | XR US | 1 | TS | |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663B23014 | XR US | 1 | TS | |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663B23015 | XR US | 1 | TS | |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663B23017 | XR US | 2 | TS | |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663B23018 | XR US | 2 | TS | |
| VISN 22: VA Desert Pacific Healthcare Network | 664 SAN DIEGO, CA | 664B20017 | XR RAD | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 667 SHREVEPORT, LA | 667B29003 | XR RAD | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 667 SHREVEPORT, LA | 667B29004 | XR RF | 1 | TS | |
| VISN 16: South Central VA Health Care Network | 667 SHREVEPORT, LA | 667B29005 | XR RF | 1 | TS | |
| VISN 20: VA Northwest Network | 668 SPOKANE, WA | 668B23000 | XR US | 1 | TS | |
| VISN 8: VA Sunshine Healthcare Network | 672 SAN JUAN, PR | 672FC2002 | XR US | 12 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22012 | XR US | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22014 | XR US | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22027 | XR US | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22028 | XR RAD | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22038 | XR PORT RAD | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22039 | XR C-ARM | 1 | TS | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22041 | XR C-ARM | 1 | TS | |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B22042 | XR RAD | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20045 | XR US | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20077 | NM BONE DEN | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20079 | XR US | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20080 | XR US | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20097 | XR US | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B20098 | XR US | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 AUSTIN, TX | 674B20104 | XR MRI | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 674 AUSTIN, TX | 674B20105 | NM BONE DEN | 1 | TS | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27006 | XR PORT C-ARM | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27007 | XR CT | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27008 | XR US | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27009 | XR CT | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27010 | XR RAD | 1 | BM | |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B27011 | XR RAD | 1 | BM | |
| VISN 22: VA Desert Pacific Healthcare Network | 678 TUCSON, AZ | 678B22034 | XR US | 2 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 693 WILKES-BARRE, PA | 693B20004 | NM SPECT | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 693 WILKES-BARRE, PA | 693B20007 | XR US | 1 | TS | |
| VISN 4: VA Stars & Stripes Healthcare Network | 693 WILKES-BARRE, PA | 693B20008 | NM SPECT | 1 | BM | |
| VISN 4: VA Stars & Stripes Healthcare Network | 693 WILKES-BARRE, PA | 693B20011 | XR MRI | 1 | BM | remove from solicitation- this is an MRI upgrade. |
| VISN 12: VA Great Lakes Health Care System | 695 MILWAUKEE, WI | 695B28005 | XR US | 1 | BM | |
| VISN 12: VA Great Lakes Health Care System | 695 MILWAUKEE, WI | 695B28006 | XR CATH LAB | 1 | BM | |
| VISN 17: VA Heart of Texas Health Care Network | 740 HARLINGEN, TX | 740B20013 | NM BONE DEN | 1 | TS | |
| VISN 17: VA Heart of Texas Health Care Network | 740 HARLINGEN, TX | 740B20014 | NM BONE DEN | 1 | TS |
Questions_Answers
| VISN | Station | Requisition Number | QUESTIONS | GOVT Response |
| VISN 19: VA Rocky Mountain Network | 436 FT Harrison | 436B20008 | Request 8. Fusion-Targetd Biopsy System be removed or added to Value Add as this is typically not required for Cardiology. |
Request 8.b. Prostate Glan volume and edge boundaries computation capabilities be removed or added to Value Add as this is not typically required for Cardiology.
Request 8.c. Tracking and recording of needle movement during biopsy in 3D coordinates be removed or added to Value Add as this is typically not required for Cardiology.
Request 8.d. Availability of projected, future needle movement during biopsy in 3D coordinates be removed or added to Value Add as this is typically not required for Cardiology.
Request 8.e. Semi-robotic arm, physically attached to biopsy system, with at least 3 degrees of freedom used to stabilize needle and ultrasound probe movements be removed or added to Added Value as this is typically not required for Cardiology. Request 8. Removed Request 8.b. Removed Request 8.c. Removed Request 8.d. Removed Request 8.e. Removed
| VISN 19: VA Rocky Mountain Network | 436 FT Harrison | 436B20009 | Request 8.g. Semi-robotic arm, physically attached to biopsy system, with at least 3 degrees of freedom used to stabilize needle and ultrasound probe movements be removed or added to Added Value as this is typically not required for Radiology. | Request 8g. Removed | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26001 | 1i. Please change to 778lbs to allow full and fair competition. | ||
| 2k. Please move offline results and analysis email and fax to optional/value add to allow full and fair competition. | 1i. Changed to 800lbs. |
2k. Moved to Value Add
| VISN 21: VA Sierra Pacific Network | VA HONOLULU | 459B26002 | 1) Benchmark states - Minimum table height from floor [in] -- 53 -- This is probably just an oversite (in vs. cm), but 53 inches does not seem right for a minimum table height? | Removed Requirement |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26002 | 2i. Please updated minimum patient load capacity to 705lb (dynamic) to allow full and fair competition. | 2i. Changed to 705lbs |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26002 | 2 Unit Physical Specifictions h. Minimum patient table load capacity (static) [lbs] 800. Should this be Maximum vs. Minimum? Can the limit be changed from 800 to 705 lbs? | 2h. Changed to 705 lbs |
| VISN 21: VA Sierra Pacific Network | 459 VAMC HONOLULU | 459B26002 | 2.h Minimum patient table load capacity(static) [Ibs] 800lbs --> Please change from 800lbs to 771lbs to ensure maximum competition. |
2.i Minimum patient table load capacity(Dynamic) 800lbs --> Please change from 800lbs to 683lb to ensure maximum competition.
2.k Minimum table width 35" --> Please change from 35" to 32" to ensure maximum competition. Only one vendor meet 35" and it appears to be a lockout spec eliminating other vendors from competing.
3.n Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy.
3.r Please remove "Dual Energy" or modify to equivalent function of bone suppression in SW to ensure maximum competition.
3.s Please remove "Digital tomography" for full competition 2h. Changed to 705 lbs 2i. Changed to 683 lbs 2k. Changed to 32”
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26002 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. | |
| VISN 21: VA Sierra Pacific Network | VA HONOLULU | 459B26003 | 1) Benchmark states - Minimum table height from floor [in] -- 53 -- This is probably just an oversite (in vs. cm), but 53 inches does not seem right for a minimum table height? | Removed Requirement | |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26003 | 2 Unit Physical Specifictions h. Minimum patient table load capacity (static) [lbs] 800. Should this be Maximum vs. Minimum? Can the limit be changed from 800 to 705 lbs? | Tech specs does not specify 800lbs. | |
| VISN 21: VA Sierra Pacific Network | 459 VAMC HONOLULU | 459B26003 | 3.l Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. | ||
| 3.p Please remove "Dual Energy" or modify to equivalent function of bone suppression in SW to ensure maximum competition. | 3i. Requirement removed |
3p. Requirement removed
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B26003 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 NORTHFIELD, NJ | 460B28012 | 2c. please clarify if you would like dose after scan to align with industry standard. |
2g. please clarify if you are in fact looking for a full UPS or a Partial UPS?
3a. please change monitor size to 21.0 x 20.5 to allow full and fair competition.
3d. please change monitor size to 21.0 x 20.5 to allow full and fair competition. 2c – Yes
2g – Full UPS
3a – Approved to change, new Specifications attached
3d – Approved to change, new Specifications attached VISN 23: VA Midwest Health Care Network 636 IOWA CITY, IA 499B29003 • 1.19 Would customer consider a 17x17 inch Fixed Detector in the Wall stand vs a wireless detector with the benefit of less damage, avoiding drops, etc.? (**A wireless digital detector would be proposed in the overall system configuration)
• 1.23 Would customer consider a 17x17 inch wireless detector vs. a 17x17 inch fixed detector in the patient table?
• 1.23 Would customer consider a 14x17 inch wireless detector vs. a fixed 17x17 detector in patient table? (*A 17x17 inch detector would be proposed at wall stand for improved anatomical coverage.)
| • | 1.28 Would customer consider a Digital Xray system with no 10x12 detector if all other specifications were met or exceeded? |
| 1.19 – yes, if wall bucky is moveable (slides from one end of the room to the other end of the room along the table, and wall bucky pivots/swings clockwise & counter clockwise to perform x-table lateral hips) |
1.23 – Yes (wireless vs fixed)
1.23 – Yes, 14x17 easier for patients to hold for patella view, 17x17 nice for more coverage on table abdomen views
1.28 - Yes
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29003 | 5a. Would you like stitching on the wall stand and table? | |
| 4d. Please make touchscreen optional to allow full and fair competition. | 5a – Yes |
4d - Yes; touch screen requirement removed from tech spec
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29003 | Is there full leg full spine or stiching required? | Yes, both legs and full spine are needed | |
| VISN 23: VA Midwest Health Care Network | 499 VAMC IOWA CITY | 499B29003 | 2.l Minimum tube central ray height from floor when used with wall stand : 11" --> Please change from11" to 14" to ensure maximum competition. | ||
| 3.n Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. | 2.l – the closer the tube & wall bucky can get to the floor is better for the lateral foot view so patient doesn’t have to step up/stand higher on a platform |
3.n. - Motor-driven X-Y table top movement IS a spec for general radiography rooms AND fluoroscopy.
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29003 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. |
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29004 | • | 1.19 Would customer consider a 17x17 inch Fixed Detector in the Wall stand vs a wireless detector with the benefit of less damage, avoiding drops, etc.? (**A wireless digital detector would be proposed in the overall system configuration) |
• 1.23 Would customer consider a 17x17 inch wireless detector vs. a 17x17 inch fixed detector in the patient table?
• 1.23 Would customer consider a 14x17 inch wireless detector vs. a fixed 17x17 detector in patient table? (*A 17x17 inch detector would be proposed at wall stand for improved anatomical coverage.)
| • | 1.28 Would customer consider a Digital Xray system with no 10x12 detector if all other specifications were met or exceeded? |
| 1.19 – yes, if wall bucky is moveable (slides from one end of the room to the other end of the room along the table, and wall bucky pivots/swings clockwise & counter clockwise to perform x-table lateral hips) |
1.23 – Yes (wireless vs fixed)
1.23 – Yes, 14x17 easier for patients to hold for patella view, 17x17 nice for more coverage on table abdomen views
1.28 - Yes
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29004 | 5a. Would you like stitching on the wall stand and table? | |
| 4d. Please make touchscreen optional to allow full and fair competition. | 5a – Yes |
4d - Yes; touch screen requirement removed from tech spec
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29004 | Is there full leg full spine or stiching required? | Yes, both legs and full spine are needed | |
| VISN 23: VA Midwest Health Care Network | 499 VAMC IOWA CITY | 499B29004 | 2.l Minimum tube central ray height from floor when used with wall stand : 11" --> Please change from11" to 14" to ensure maximum competition. | ||
| 3.n Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. | 2.l – the closer the tube & wall bucky can get to the floor is better for the lateral foot view so patient doesn’t have to step up/stand higher on a platform |
3.n. - Motor-driven X-Y table top movement IS a spec for general radiography rooms AND fluoroscopy.
| VISN 23: VA Midwest Health Care Network | 636 IOWA CITY, IA | 499B29004 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | 1. | Please clarify if you would like a fixed detector in the wall stand or a wireless detector in the wall stand. If a wireless detector is needed, do you want one detector to share between the table and wall bucky or two wireless detectors? In line 19, you ask for a fixed wall detector, but in line 21 you ask for the weight of a wireless wall detector. | |
| 2. | Please clarify on line 36 under unit physical specs “j” what you mean by minimum tilt of imaging surface? What is the image surface you are referring to? A table, tube, detector? Line 39 is specific to the wall stand. (-20/+90 table tilt is typically for a fluoroscopic table) | ||||
| 3. | Please clarify if you want a free floating tabletop or a motorized tabletop. Lines 51 and 52 contradict each other. Line 52 (motorized tabletops are traditionally for fluoroscopic tables) | ||||
| Requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet -specs J – changed to 0, do not need table to tilt Spec 3.l – changed, only need free floating manual tabletop | |||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | 1)CAN THE PIXEL MINIMUM SIZE BE CHANGED FROM 140 MICRONS TO 150 MICRONS? |
2)CAN THE MINIMUM TOUCHSCREEN MONITOR SIZE CHANGE FROM 20" TO 19"?
3)CAN YOU FURTHER EXPLAIN ITEM 3(i) IN GREATER DETAIL?
4)CAN YOU FURTHER EXPLAIN ITEM 3(m) IN GREATER DETAIL?
5)CAN THE TABLE WEIGHT LOAD MAX OF 700LBS BE MOVED TO 649LBS?
| 1. Pixel size changed to 150 (industry standard) 2.We will keep current requirements for the monitor size the same 3i- is viewed as the ability to adjust the selected protocol from the tube head (kv,mAs) 4. Protocol adjustment via tube head or other in room monitors 5.We will keep current requirements for the table weight the same | |||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | Specifications call for a fixed detector in the wall stand and a iwreless detector in the table. Please confrm that you'd rather have 2 wireless detectors as opposed to one wireless detector that can be moved between the two locations. | We are requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | 2 - Unit physical specifications - f. Maximum pixel size [μm] 140. Can this be changed to 150 (industry standard)? | -This has been changed to 150 spec sheet updated | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | 2 - Unit physical specifications - f. Maximum pixel size [μm] 140. Can this be changed to 150 (industry standard)? | ||
| 5. Software Requirements a. Stitching Does this need to be performed on Full Leg Full Spine? | -This has been changed to 150 spec sheet updated | ||||
| VISN 17: VA Heart of Texas Health Care Network | 504 VAMC AMARILLO | 504B20002 | 2.e Minimum spatial resolution is 3.57lp/mm when required pixel size of 140um. |
2.h Minimum tube height from floor when used with wall stand : 12" --> Please change from12" to 14" to ensure maximum competition.
| 3.l Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. | -This has been changed to 150 spec sheet updated Tube height from floor- 12” was selected to ensure lowest possible height from the floor 3i – changed on spec sheet | ||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20002 | 1.a & 1.b: Requesting fixed 14”x 17” detector but states wireless in 1.c. Can you please confirm if it's fixed or wireless and only one shared between wall and table? | -Requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20003 | 1c. Please change Minimum scan time at 1% for lateral spine to 62 seconds (for 15.8x12cm) for standard mode to allow full and fair competition. |
1e. Please change minimum scan time at 1% CV for femur to 10 seconds to allow full and fair competition.
| 1h. Please change minimum distance between table and carm to 18 inches to allow full and fair competition. | The spec sheet has been updated | ||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | 1.a & 1.b: Requesting fixed 14”x 17” detector but states wireless in 1.c. Can you please confirm if it's fixed or wireless and only one shared between wall and table? | -Requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | 1. | Please clarify if a fixed detector in the wall stand or a wireless detector in the wall stand is required. If a wireless detector is needed, is one detector desired, to share between the table and wall bucky or two wireless detectors? In line 19, a fixed wall detector is indicated, but in line 21 the weight of a wireless wall detector is stated. | |
| 2. | Please clarify on line 36 under unit physical specs “j” what is meant by "minimum tilt of imaging surface"? What is the image surface that is refered to -- table, tube, detector? Line 39 is specific to the wall stand. (-20/+90 table tilt is typically for a fluoroscopic table) | ||||
| 3. | Please clarify if a free floating tabletop or a motorized tabletop is desired? Lines 51 and 52 contradict each other. Line 52 (motorized tabletops are traditionally for fluoroscopic tables) | ||||
| 1. Requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet 2.-specs J – changed to 0, do not need table to tilt 3.Spec 3.l – changed, only need free floating manual tabletop | |||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | Specifications call for a fixed detector in the wall stand and a iwreless detector in the table. Please confrm that you'd rather have 2 wireless detectors as opposed to one wireless detector that can be moved between the two locations. | Requesting Fixed detector in the Wall stand and wireless for the table bucky/corrected spec sheet | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | 2 - Unit physical specifications - f. Maximum pixel size [μm] 140. Can this be changed to 150 (industry standard)? | ||
| 5. Software Requirements a. Stitching Does this need to be performed on Full Leg Full Spine? | 2. changed to 150 spec sheet updated 5. yes | ||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B20004 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. | |
| VISN 7: VA Southeast Health Care Network | 509 Augusta | 509B22009 | Request 2.f. Monitor Height range from floor (in) 10 be revised to 55 in. |
Request 2.k. Maximum equipment weight of 230 lbs. be revised to 265 lbs.
Request 2.j. Minimum battery life (min) 60 be moved to value add.
Request 6.l. Ability to enter standby mode or sleep mode be moved to Vale Add Section.
Request Section 9 Advanced Features b to h be removed as these are not typical procedures for a Cardiology System. The monitor height from the floor was changed from 10in to 55in The equipment weight from 230lb to 265lb The battery life and standby/sleep mode was moved to the added value section Removed the “standard planning templates for prostate biopsies” and “prostate gland volume and edge boundaries computation capabilities” from the Advanced features section
| VISN 20: VA Northwest Network | 531 Boise | 531B24002 | Request 1.m. Maximum equipment weight of 100 lbs. be revised to 135 lbs. | Tech Specs Updated | |
| VISN 7: VA Southeast Health Care Network | 534 Charleston | 534B22031 | Request 2.c. Navigation Software be removed or added to Value Add. | ||
| Request 2.d. Image Fusion Software be removed or added to Value Add. | Moved to Value add | ||||
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22052 | 1c - To ensure maximum competition, please change to 0.1 - 320. There are no portable x-ray exams that should ever require 500 mAs given the efficiency (DQE and TQF) of today's detectors. |
1f - To maximize competition, please change to +/- 270 degrees.
1l - Please remove spec. Unit battery life is incredibly subjective based on type of images taken, age of battery, distance and incline driven, kV, mA, repeat rate and many other factors.
3p - Please clarify what functionality is being requested 1c – TS updated with change 1f – TS updated with change 1l – removed from TS 3p – removed from TS
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544B22052 | 2. | What type of grid do you want to include with the detector, physcial and/or virtual? | |
| 2. | Do you want a spare battery and charger with the detector? | ||||
| 4. | Do you want software to capture the rejected images | 2 – virtual is preferred, but physical is still acceptable. TS updated |
2 – yes; TS updated 4 – no
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | 1. Can you please confirm the drawing is correct for the new RAD room for the VA Pueblo, CO Clinic? | |
| 2. Is this Clinic owned by the VA, or leased space ? | 1. I’m not sure what drawing they are referring to, a drawing was not provided. |
2. Leased space
| VISN 19: VA Rocky Mountain Network | VA PUEBLO | 554B23018 | 1) Benchmark requires: Minimum tilt of imaging surface [deg] of 315. Is that degrees? What imaging surface is referred to here? | Clarification – NO requirement for table tilt of 315 degrees; should read Zero (0) Wall Stand bucky tilt 110 degrees tilt ( -20 degrees to 90 degrees ) | |
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | Please confirm which component the 315 degree tilt requirement is for? | Clarification – NO requirement for table tilt of 315 degrees; should read Zero (0) Wall Stand bucky tilt 110 degrees tilt ( -20 degrees to 90 degrees ) | |
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | 1.b. Please clarify if you are looking for a 14x17 or a 17x17 detector for the table. | ||
| 2.f. Please clarify what you are looking to accomplish clinically with a 315 degree tilt | 1. Minimum detector size for both table and wall stand 17 inch x 17 inch |
2. Clarification – NO requirement for table tilt of 315 degrees; should read Zero (0) Wall Stand bucky tilt 110 degrees tilt ( -20 degrees to 90 degrees )
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | 2 - Unit physical specifications - f. Maximum pixel size [μm] 140. Can this be changed to 150 (industry standard)? | |
| 2 Unit Physical Specifictions h. Minimum patient table load capacity (static) [lbs] 880. Should this be Maximum vs. Minimum? Can the limit be changed from 880 to 705 lbs? | 1. | Should read “minimum” pixel size [um] of 100. | ||
| 2. | Minimum patient table load capacity is changed to (static) [lbs] = 700 lbs |
Minimum patient table load capacity is changed to (dynamic) [lbs] = 650 lbs VISN 19: VA Rocky Mountain Network 554 VAMC PUEBLO 554B23018 2.d Minimum spatial resolution 5.0lp/mm --> Please change to 3.57lp/mm to ensure maximum competition. 5.0lp/mm resolution means 100um pixel size and 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 micron pixel pitch for chest images that mostly common exams in DR. It means 100 micron pixel pitch resolution has not clinical benefit at chest X-Ray. It appears to be a lockout spec eliminating other vendors from competing.
2.e Please modify from 100 micron pixel pitch detector to allow 140 micron pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 micron pixel pitch for chest images that mostly common exams in DR. It means 100 micron pixel pitch resolution has not clinical benefit at chest X-Ray. It appears to be a lockout spec eliminating other vendors from competing.
2.f Please explain clearly what requirement of "Minimum tilt of imaging surface 315 degrees" means.
2.g Minimum patient table load capacity(static) [Ibs] 880lbs --> Please change from 880lbs to 771lbs to ensure maximum competition.
2.h Minimum patient table load capacity(Dynamic) 700lbs --> Please change from 700lbs to 683lb to ensure maximum competition.
2.i Minimum table height from floor [in] 19 --> Please change from 19" to 21.5" for fair competition. It appears to be a lockout spec eliminating other vendors from competing.
| 3.h Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. | 1. | Changed minimum spatial resolution [lp/mm] from 5.0 to 3.5 lp/mm |
| 2. | Should read “minimum” pixel size [um] of 100 | |
| 3. | Clarification – NO requirement for table tilt of 315 degrees; should read Zero (0), Wall Stand bucky tilt 110 degrees tilt ( -20 degrees to 90 degrees ) | |
| 4. | Minimum patient table load capacity changed to (static) [lbs] = 700 lbs | |
| 5. | Minimum patient table load capacity changed to (dynamic) [lbs] = 650 lbs | |
| 6. | Minimum table height changed to = 22 inches |
Removed motor-driven tabletop movement in the x and y axis
| VISN 19: VA Rocky Mountain Network | 554 PUEBLO, CO | 554B23018 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20021 | 5.a. Do you want stitching on the wall and table? | Wall stitching only. Updated tech spec |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20021 | 2 - Unit physical specifications - f. Maximum pixel size [μm] 140. Can this be changed to 150 (industry standard)? |
2 Unit Physical Specifictions h. Minimum patient table load capacity (static) [lbs] 800. Should this be Maximum vs. Minimum? Can the limit be changed from 800 to 705 lbs?
| Maximum pixel size [μm] can this be 150 instead of 140? g. Minimum patient table load capacity (static) [lbs] can this be changed from 800 to 705. | Updated tech spec. Changed to 150 micron pixel. Changed to 700 lb table limit. | ||||
| VISN 6: VA Mid-Atlantic Health Care Network | 558 VAMC DURHAM | 558B20021 | 2.g Minimum patient table load capacity(static) [Ibs] 800lbs --> Please change from 800lbs to 771lbs to ensure maximum competition. | Updated tech spec. Changed to 700 lb table limit. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20021 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20052 | There are 4 Req # (RAD) Not on May Schedule of Line items updated 07/13/22 | They are listed on the schedule of line items. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20052 | 5.a. Do you want stitching on the wall and table? | Wall stitching only. Updated tech spec | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20065 | There are 4 Req # (RAD) Not on May Schedule of Line items updated 07/13/22 | They are listed on the schedule of line items. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20065 | 5.a. Do you want stitching on the wall and table? | Wall stitching only. Updated tech spec | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20066 | 5.a. Do you want stitching on the wall and table? | Wall stitching only. Updated tech spec | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20066 | There are 4 Req # (RAD) Not on May Schedule of Line items updated 07/13/22 | They are listed on the schedule of line items. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20067 | There are 4 Req # (RAD) Not on May Schedule of Line items updated 07/13/22 | They are listed on the schedule of line items. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B20067 | 5.a. Do you want stitching on the wall and table? | Wall stitching only. Updated tech spec | |
| VISN 16: South Central VA Health Care Network | 564 FAYETTEVILLE, AR | 564B10043 | 1d. Please change maximum scan time for forearm to 24 seconds to allow full and fair competition. |
1k. Please change Maximum system dimensions to 113x52x49 to allow full and fair competition.
| 1l. Please change maximum system weight to 780lbs to allow full and fair competition. | Changed as requested; updated tech spec | |||
| VISN 8: VA Sunshine Healthcare Network | 573 GAINESVILLE, FL | 573B25014 | 1.k. Please change to +/- 5.5 inches to allow full competition |
1. w. Please change to 19 inches to allow full competition
| 2. j. Do you need a tilting table for a bi-plane EP lab configuration? Please move this requirement to optional | i.k updated 1.w not changed 2.j. Tiliting table required | |||
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20034 | 1.a. Please reduce the minimum PA gantry depth to 94CM to allow for fair competition. |
1.g. Please clarify. Lateral gantry is found on a biplane system. The requirement includes specifications for a ceiling mounted single plane system.
1.h. Please clarify. Lateral gantry is found on a biplane system. The requirement includes specifications for a ceiling mounted single plane system.
1.i. Please clarify. Lateral gantry is found on a biplane system. The requirement includes specifications for a ceiling mounted single plane system.
1.n. Optimal detector for cardiac cath lab is 20x20cm. Please consider adjusting this requirement to at least 20x20? 1a The technical specification has been updated.
1g The technical specification has been removed.
1h The technical specification has been removed.
1i The technical specification has been removed 1n The technical specification has been updated
| VISN 10: VA Healthcare System of Ohio | 583 INDIANAPOLIS, IN | 583B20034 | Trade - In. Please confirm the Trade In is correct. | |
| 28k. Please change table lateral motion range to 29 cm to allow full and fair competition. | Trade In has been confirmed to be correct |
1k The technical specification will remain as is due to clinical requirements.
VISN 10: VA Healthcare System of Ohio 583 Indianapolis 583B20036 Request 1.d. Minimum monitor rotation 360 degrees be revised to 180 degrees.
Request 1.k Minimum battery life 120 minutes be revised to 60 min.
1d The technical specification will remain as is due to clinical requirements.
1k The technical specification changed to 60 minutes- added 120 as a value add
| VISN 16: South Central VA Health Care Network | 598 N. LITTLE ROCK, AR | 598B20017 | 4b. Please remove Touchscreen requirement to allow full and fair competition | Dropped from Attached Requirements; updated tech spec |
| VISN 16: South Central VA Health Care Network | 598 N. LITTLE ROCK, AR | 598B20017 | Is there full leg full spine or stiching required? | Not Required |
| VISN 16: South Central VA Health Care Network | 598 N. LITTLE ROCK, AR | 598B20017 | Who is responsible Turnkey be Performed? By Vendor or the VA Site? If by Vendor, a date will have to be set for Walk through to calculate cost of Trunkey. | Questions are not applicable, as turnkey is not part of this requirement. |
| VISN 22: VA Desert Pacific Healthcare Network | VA LONG BEACH | 600B20017 | 1) Does the customer require a new contrast injector? |
2) How many slices do you require for the CT?
3) Regarding required Benchmark Trade-in - What is the serial number and installation date of the system being traded in? Has it been under a full service agreement? 1) No
2) 128
3) Serial #: 269717HM5, installation date: 6/2011; equipment has been under full service contract
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20026 | 1d - Please clarify - is a 10"x12" being requested in the specification? Or is the request that the current 10"x12" the site has will work with the new system? If the request is that the current 10"x12" detector be shared with the new system, this is a lockout spec and should be removed. | 1 additional wireless detector is requested. Tech specs updated to reflect the minimum and maximum dimensions. | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20026 | 3t. Please remove preview image on the OTS to allow full and fair competition. | ||
| 4f. Please make touchscreen optional to allow full and fair competition. | No change; these items are standard per the VA tech spec template | ||||
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600B20026 | No files were provided for this RNA. Please provide Technical Specifications and Compliance Matrix documents | Please clarify. Technical specifications and compliance matrix have been provided. | |
| VISN 22: VA Desert Pacific Healthcare Network | 600 VAMC LONG BEACH | 600B20026 | 2.d Please take a look 1000mA is really required at 100kVp. It means 100kW system is required for this |
2.j Minimum patient table load capacity(static) [Ibs] 820lbs --> Please change from 820lbs to 771lbs to ensure maximum competition.
2.l Minimum table height from floor [in] 20 --> Please change from 20" to 21.5" for maximum competition.
3.m Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity --> Motor-driven X-Y table top movement is not a spec for general radiography rooms, more for fluoroscopy. 2d) Tech specs updated to reflect 80 kVp 2j) Tech specs updated to reflect minimum static patient table load capacity of 771lbs 2l) No change; minimum table height from floor of 20” includes a table height of 21.5” 3m) No change; this item is standard per the VA tech spec template for digital radiography equipment
| VISN 16: South Central VA Health Care Network | 629 New Orleans | 629B20018 | Request 1.i Minimum battery life 180 min be revised to 60. | |
| Request 8.e. and 8.f. and 8.g. be removed or added to Added Value. | 1.i. Minimum battery life revised to 60 minutes. 8e, 8f, and 8g moved to Added Value | |||
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B20027 | Can you please clarify the max system dimensions; it is 700 inches? |
| Requirement removed from specifications | |||
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B20027 | B. Technical Requirements |
1. Unit physical specifications a.(line 18) minimum number of slices acquired simultaneously—640. Could you please clarify what slice count is required for reconstruction slice count and acquired slice count?
c.(line 20) minimum detector width z-axis--.5mm. Could you please clarify if you are requiring a minimum detector width? Please allow for detector width of 38.4mm to ensure fair competition.
d.(line 21) minimum scan field of view—16cm. Could you please clarify what specification is needed? Did you intend to specify the maximum scan field of view?
f.(line23) minimum table height range from floor---430mm. Please allow for floor to table range of 39.4 in minimum to ensure fair competition?
g.(line24) minimum patient table load capacity—20 lbs. Please clarify if you are requiring a minimum table load or maximum table load. Tables are not generally limited to a minimum patient load as staff will conduct QA procedure with no weight on the table.
h.(line25) maximum 360 rotation time---.27s. Could you please clarify if you are requesting to limit the rotation time to a maximum of .27s (meaning not to exceed .27s). Would you allow for a minimum rotation time of .28s? See also E. Added Value. D. Vendors are encourage to propose the fastest rotation time available.
i. (line 26) Maximum system dimensions---700. Could you please clarify what unit of measurement is 700?
2. Additional specifications
o. (line 42) Gantry tilt +/- 30 degrees. Is gantry tilt a necessary specification? Would you accept a CT system without gantry tilt?
C. Training Requirements
1. Clinical Training Per System
c. (line 74) OFFSITE clinical applications training for 2 technologist. Please remove this requirement, per NAC training memo. 1a. reduced to a minimum slice count of 512.
1c. increased minimum detector width to 80 mm
1d. Approved to be removed from requirements.
1f. increased minimum table to floor height range to 1000 mm.
1g. removed from tech spec
1.h Clarification: maximum amount of time to make a 360-degree rotation ie. (360-degree rotation cannot be 0.29 s or greater).
Increased to 0.28s.
1.i. Requirement to be removed from specifications.
2. +/- 30 degrees gantry tilt is a necessary for the proposed clinical applications.
C.1.c.moved to the added value section.
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B20027 | 1.a. This specification is unduly restrictive as it only allows for one vendor to meet the specification. Please change to minimum 512 slices acquired simultaneously to allow full competition between vendors. | |
| 1. c. This specification is unduly restrictive as it only allows for one vendor to meet the specification. Please change this specification to the industry standard of .625mm to allow full competition | 1.a. reduced to a minimum slice count of 512. |
1.c. increased minimum detector width to 80 mm VISN 19: VA Rocky Mountain Network 635 OKLAHOMA CTY, OK 635B20007 • Can you identify if the trade-in is required in the Compliance Matrix - Line 172 - Trade-in? If so, can you confirm the linac equipment specifications, such as the vendor, installed year, serial number, and included software and configuration? Will the vault be empty, and if so, what specifications will be provided?
• We have reviewed the Compliance Matrix provided and need clarity on Appendix A. We are proposing QA equipment based on our equipment recommendations, please confirm this is allowable. Please advise best way to incorporate this into the document as these cells are locked and not able to make edits to existing or insert additional lines. Can you please provide a new unprotected Compliance Matrix?
• Appendix A of Compliance Matrix requests QA…