Line Qty. Catalog
1 1 1.00 PROTEUS XRF ST BASE SYS
The radiographic system PROTEUS XR/f is an integrated floor mount radiographic system designed for all conventional radiographic applications in combination with a High Frequency Generator, Radiographic Table, and Wall Bucky.
Fixed elevation 4 way float patient table
- table top dimensions 220cm x 82.5cm
- 77cm table height from floor
- 300kg max weight
The Wall Stand is a heavy-duty vertical bucky stand, providing full flexibility for radiographic examinations. This stand is suitable for all varieties of examinations including horizontal, vertical and oblique angles. The bucky tray is designed for your choice of either a
14x17 or 17x17 wireless flat panel Konica Minolta AeroDR detector
Line Qty. Catalog 2 1 1.00 Prot Aero FPD 1417LT Kit
AeroDR 14x17 LT This AeroDR LT configuration combines the lightest FPD (5,35 lbs.)with a ruggedized and liquid resistant enclosure and our CS-7 S/W to deliver the only solution versatile and tough enough for the general radiology in and outside of the radiology department. Software features include Modality Worklist and Hybrid Premium Processing Algorithms.
Includes:
AeroDR LT 14x17 Cassette Sized Wireless Digital Flat Panel Detector
• 13 minute Charge
• 150 images/4.1 hours
• 4-6 Second Panel Refresh Time (Cycle Time)
• Meets International Specifications for true 14x17 cassette size
• AeroSync Auto Exposure Detection
• IPX6 Waterproof compliant
CS-7 Universal Control Station Hardware
CS-7 Universal Control Station Software
Image Quality Optimization
SHIP TO: WAREHOUSE EO B82011
V.A. Medical Center
L OGISTICS EAST ORANGE
385 TREMONT AVE
EAST ORANGE, NJ 07018-1095
PO # 561-B82011
There is a limitation of 10' on the room size The system will be required to fit in a small space.
AMMHINWARDM
Sticky Note None set by AMMHINWARDM
AMMHINWARDM
Sticky Note MigrationNone set by AMMHINWARDM
AMMHINWARDM
Sticky Note Unmarked set by AMMHINWARDM
AMMHINWARDM
Sticky Note None set by AMMHINWARDM
AMMHINWARDM
Sticky Note MigrationNone set by AMMHINWARDM
AMMHINWARDM
Sticky Note Unmarked set by AMMHINWARDM
DICOM Store (2 connections)
DICOM Modality Worklist
MPPS Software License
Aero DR Gen I/F SSRM kit
Procedure Code Mapping
8001726 DR-CS7 Dept Data Analysis Lic DR
This license analyzes and gives statistical data images taken. As a result, it provides a clear overview of quantity and reasons of non-exploitable images : too low dose level, patient motion, etc
8001762 AeroDR CS-7 Study Combine/Move Software License
Ideal for use in trauma and ER settings. This advanced software feature allows patient study data from one study (i.e. study 'A') to be transferred to a different patient study
(i.e. study 'B') on the CS-7 controller. The original study (i.e. study 'A') is removed from the local worklist native to the CS-7 after the information has been moved to the new study
(study 'B').
3 1 1.00 Proteus XRf Wall Stand Side Select
Proteus XRf Wall Stand Side Select
Line Qty. Catalog 4 1 1.00 14x17 Table bucky
14x17 rotating table bucky tray designed to auto charge Konica Minolta AeroDR detectors
Line Qty. Catalog
5 1 1.00 PROTEUS XRF HIGH PWR TUBE
X-ray tube Toshiba 12º 150 kVp dual focal spots 0,6 and 1,2 mm 400KHU LS 18/52 Kw (50Hz) 24/60 Kw (60Hz) HS 37/95 Kw
(150Hz) 40/102 Kw (180Hz)
Line Qty. Catalog 6 1 1.00 65kW 3 phase 480V 150KVP HS Gen
The high frequency x-ray generator technologies controlled by microprocessor improve the image quality and reduce the patient dose. The very low ripple and accuracy radiographic parameter KVp, mA and exposure time reduce the soft x-ray radiation and improve the homogeneity of the x-ray beam.
The high frequency x-ray generator controlled by microprocessor improve also the reliability of the whole system and reduce the maintenance cost, thanks to the constant monitoring of the system, with auto diagnostic and error code
Three phase 480VAC 65Kw maximum power
Auto exposure control tube overload protection monitoring of the remaining x-ray tube heat units
.1mAs - 640mAs (38 steps Reneard Scale)
7 1 1.00 D MDP UL 65KW 480V 60Hz 3 phases for Proteus XR-f
8 1 1.00 S S Prot XRf Tbl Pwr Select
Line Qty. Catalog 9 1 1.00 Proteus XRf Table top US Power
Proteus XRf Table top US Power
Line Qty. Catalog 10 1 1.00 ET Tech Pubs
Optional hard copy of Proteus XRF ET technical manual
Line Qty. Catalog
11 1 1.00 HAND SUPPORT WALL BUCKY
A removable overhead grip bar that mounts on the wallstand assembly.
Line Qty. Catalog
12 1 1.00 S1976HW HAND GRIPS WALLSTAND
A set of (2) hand grips that are mounted to the rear of the wall stand assembly.
Line Qty. Catalog 13 1 1.00 180cm Removable Grid
Standard 180 cm (70") removable grid - 103 lines/cm--10:1 focalized at 180 cm or 103 lines/cm--12:1 focalized at 180cm with
AMMHINWARDM
Sticky Note None set by AMMHINWARDM
AMMHINWARDM
Sticky Note MigrationNone set by AMMHINWARDM
AMMHINWARDM
carbon fiber cover.
14 1 1.00 100cm Removable Grid
Standard 100 cm (40") removable grid - 103 lines/cm--10:1 focalized at 100 cm or 103 lines/cm--12:1 focalized at 100cm with carbon fiber cover.
Line Qty. Catalog 15 1 1.00 Compression band for ST
This device supplies compression to the anatomical area of interest in order to avoid unnecessary movements
Line Qty. Catalog 16 1 1.00 Table Hand Grips for ST
The Hand Grips are used by patients and operators to keep their hands away from the Tabletop edges and make the patient feel safe while the table-top is being positioned. The Hand Grips are installed along the Tabletop rails and locked at any position with the thumbscrews.
Line Qty. Catalog
17 1 1.00 14X17 40-72 GRID W_HNDL
8001810-14x17 DR GridCap 8:1-103-40-72; 14x17
Grid-Holder with Handle
Wireless DR detector snap on grid
8:1 Ratio 103L/in.
40-72" Focal Range
With Handle
18 1 1.00 14X17 WEIGHT DIST. CAP
8001813 14x17 Aero DR Weight Distr Cap
The cap allows weight of the patient to be distributed away from the sensitive DR panel by being equally spread across the surface.
This is particularly useful in case of
•ver-weight/ obese patient and standing feet exams.
The maximum point load of the KM Aero DR panel is 150kg on a 40mm - diameter spot. This rises up to 300kg for the maximum surface load, •n total effective image area.
This cap is fitting to 14''x17'' Aero DR flat panels
Line Qty. Catalog 19 1 1.00 Proteus XR/f Expert2 Training Package New to GE Digital and Konica
Minolta
Expert 2 Package New to GE Digital and Konica Minolta
Training package includes 3 days Go Live training, plus 1 follow-up day for a total of 4 on-site days. Also includes online and remote training. Program concludes one year after the initial start date. Instruction is provided from 8 AM to 5 PM, Monday through
Friday and includes T&L expenses. Training cannot be scheduled as single day events.
AMMHINWARDM
Sticky Note None set by AMMHINWARDM
AMMHINWARDM
Sticky Note MigrationNone set by AMMHINWARDM
AMMHINWARDM
Sticky Note Unmarked set by AMMHINWARDM
AMMHINWARDM
Sticky Note None set by AMMHINWARDM
AMMHINWARDM
Sticky Note MigrationNone set by AMMHINWARDM
AMMHINWARDM
File details come from the government source that posted it.