TEAM A
| Consolidation Report TEAM A | |||||
| HTME VISN | Station | Requisition Number in Action | Consolidation Period | Consolidation Type | Equipment Type |
| VISN 4: VA Stars & Stripes Healthcare Network | 503 ALTOONA, PA | 503-B80019 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 8: VA Sunshine Healthcare Network | 516 ST.PETERSBURG,FL | 516-B90014 | MAY 2019 | CONSOLIDATION | XR CT |
| VISN 5: VA Capitol Health Care Network | 540 CLARKSBURG, WV | 540-B81004 | MAY 2019 | CONSOLIDATION | XR CATH LAB |
| VISN 5: VA Capitol Health Care Network | 540 CLARKSBURG, WV | 540-B81007 | MAY 2019 | CONSOLIDATION | NM SPECT CT |
| VISN 5: VA Capitol Health Care Network | 540 CLARKSBURG, WV | 540-B81008 | MAY 2019 | CONSOLIDATION | NM SPECT CT |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541-B83032 | MAY 2019 | CONSOLIDATION | XR CT |
| VISN 7: VA Southeast Health Care Network | 544 COLUMBIA, SC | 544-B92030 | MAY 2019 | CONSOLIDATION | NM SPECT |
| VISN 11: VA Veterans In Partnership | 550 DANVILLE, IL | 550-B80018 | MAY 2019 | CONSOLIDATION | NM BONE DEN |
| VISN 11: VA Veterans In Partnership | 550 DANVILLE, IL | 550-B80023 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 10: VA Healthcare System of Ohio | 553 DETROIT, MI | 553-B90039 | MAY 2019 | CONSOLIDATION | XR MRI |
| VISN 10: VA Healthcare System of Ohio | 553 DETROIT, MI | 553-B90040 | MAY 2019 | CONSOLIDATION | XR CATH LAB |
| VISN 10: VA Healthcare System of Ohio | 553 DETROIT, MI | 553-B90041 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 10: VA Healthcare System of Ohio | 553 DETROIT, MI | 553-B90043 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 10: VA Healthcare System of Ohio | 553 DETROIT, MI | 553-B90044 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 12: VA Great Lakes Health Care System | 556 NORTH CHICAGO IL | 556-B93167 | MAY 2019 | CONSOLIDATION | XR-CT |
| VISN 2: VA Healthcare Network Upstate New York | 561 EAST ORANGE, NJ | 561-B82011 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 16: South Central VA Health Care Network | 580 HOUSTON, TX | 580-B80022 | MAY 2019 | CONSOLIDATION | NM SPECT |
| VISN 16: South Central VA Health Care Network | 598 N.LITTLE ROCK, A | 598-B80014 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 16: South Central VA Health Care Network | 598 N.LITTLE ROCK, A | 598-B80021 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 22: VA Desert Pacific Healthcare Network | 600 LONG BEACH, CA | 600-B90005 | MAY 2019 | CONSOLIDATION | XR-CT |
| VISN 9: VA Mid South Healthcare Network | 614 MEMPHIS, TN | 614-B83004 | MAY 2019 | CONSOLIDATION | NM SPECT |
| VISN 9: VA Mid South Healthcare Network | 626 CHATTANOOGA, TN | 626-B90046 | MAY 2019 | CONSOLIDATION | XR CT |
| VISN 9: VA Mid South Healthcare Network | 626 CHATTANOOGA, TN | 626-B90049 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 9: VA Mid South Healthcare Network | 626 CHATTANOOGA, TN | 626-B90050 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 9: VA Mid South Healthcare Network | 626 CHATTANOOGA, TN | 626-B90051 | MAY 2019 | CONSOLIDATION | XR MAMMO |
| VISN 23: VA Midwest Health Care Network | 636 Des Moines, IA | 636-B94033 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 6: VA Mid-Atlantic Health Care Network | 637 ASHEVILLE, NC | 637-B80012 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 4: VA Stars & Stripes Healthcare Network | 642 PHILADELPHIA, PA | 642-B80023 | MAY 2019 | CONSOLIDATION | XR CT |
| VISN 4: VA Stars & Stripes Healthcare Network | 642 Philadelphia, PA | 642-B90009 | MAY 2019 | CONSOLIDATION | NM SPECT |
| VISN 4: VA Stars & Stripes Healthcare Network | 642 PHILADELPHIA, PA | 642-B90011 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646-B92001 | MAY 2019 | CONSOLIDATION | XR HYBRID |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652-B80059 | MAY 2019 | CONSOLIDATION | NM SPECT CT |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652-B80065 | MAY 2019 | CONSOLIDATION | XR CT |
| VISN 21: VA Sierra Pacific Network | 654 RENO, NV | 654-B99012 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 15: VA Heartland Network | 657 POPLAR BLUFF, MO | 657-B93008 | MAY 2019 | CONSOLIDATION | NM SPECT |
| VISN 20: VA Northwest Network | 663 SEATTLE, WA | 663-B80032 | MAY 2019 | CONSOLIDATION | XR MOBILE MRI |
| VISN 5: VA Capitol Health Care Network | 688 WASHINGTON, DC | 688-B80805 | MAY 2019 | CONSOLIDATION | XR RF |
| VISN 5: VA Capitol Health Care Network | 688 WASHINGTON, DC | 688-B80808 | MAY 2019 | CONSOLIDATION | XR RAD |
| VISN 5: VA Capitol Health Care Network | 688 Washington DC | 688-B80818 | MAY 2019 | CONSOLIDATION | XR CT |
Q&A
| Service Responsible for providing answers | PO# | QUESTIONS | ANSWERS | |
| HTM | General | If the CAD drawings of all the location spaces are posted, does that change the requirments of the submissions to include site-specific CAD drawings? | No. No site specific CAD drawings are required for any line item | |
| HTM | General | In reviewing the FIPS requirements, we see there are 4 security levels. What level do you require medical devices to conform to? Also, if we find our equipment lacking in a particular secuirty requirement, do you accept the addition of a 'compensating control' such as a firewall, to augment the security of the medical device? | The wireless capability meets the FIPS 140-2 requirement if it has a “validated” or “certified” status. NIST provides a searchable database of FIPS 140-2 compliance certificates at https://csrc.nist.gov/Projects/Cryptographic-Module-Validation-Program/Validated-Modules/Search. A device with 802.11 wireless communication must be FIPS 140-2 certified or validated to be procured by the VA. | |
| HTM | General | Can we include the typical drawings in each quote rather than pulling and putting in info folder? | Typical drawings should be included in the information folder by equipment type per the instructions to offers | |
| NAC Team A | 553-B90044 | VA Detroit coded as XR RAD, the actual benchmark is for a Siemens Agile RF system. It's the same benchmark quote for the XR RF Room Obligation 553-B90041. NAC needs to verify if the Agile Benchmark is correct, then they need to adjust the category to XR RF. If it's supposed to be a RAD spec, then they need to post a corrected benchmark. | The benchmark posted is for an R/F, the spreadsheet with the list of line items has been updated. | |
| NAC Team A | General | Can we do a contractor walkthrough at any of the facilities at this stage? | No walk throughs will be conducted at this time. | |
| VISN 02 | 561-B82011 | 1. Can we please have the exact specifications of the room? 10' is very small room for an R/F unit | ||
| 2. The specs show that only 1 detector is required. Please confirm. Does the benchmarked system include 1 detector inherent in the configuration and then another detector is quoted or is only ne detector being specified? | 1. Room is 10’x10’ with an ~3’x6’ control room space. Please note that this request is for a RAD system not an RF system. |
2. One inherent and one additional detector is required.
| VISN 04 | 503-B80019 | 1. What level/type of shielding is rquired, if any? | |
| 2. What does minimum control room computer with equivalent to 25k images mean? | 1. | Shielding is not required. | |
| 2. | The hard drive of the computer should be appropriately sized to hold at least 25,000 images. | ||
| VISN 04 | 642-B90009 | Can we get a CAD drawing for VA Philadelphia, 642-B90009, and location of the current treadmill system located in same room? |
Technical Requirement: Additional Specifications Section i.
- Upright and supine imaging cabaility. Request - The "upright" imaging capability limits competition. Please remove this word from technical specification. A vendor had contacted us on 5/23/19 and requested a drawing of the Nuc Med room so we provided one to the NAC. Please see revised Technical Specifications. The treadmill system is not currently located in this room, but will be moving into the room with the new cardiac camera. The stress test system cart is approximately 43” wide x 31” deep and the associated treadmill is approximately 30” wide x 80” long. The treadmill should have approximately 3 feet of clearance on each side and 6 feet from the end.
The “upright” imaging capability requirement was included in the specifications because we need to scan many patients either prone or upright. Many of our patients are overweight and scanning prone is difficult due to their size and breathing difficulties.
| VISN 04 | 642-B90011 | How many detectors is the minimum specifications calling for - 2 or 3 detectors? If 3 - If there's an option to have a removable detector (that can be fixed if required) at both wall stand and table, it will eliminate the need for a 3rd detector. Is that of value? | The minimum number of detectors is 2 if one of them is a removable detector. |
| VISN 04 | 642-B80023 | The same system was awarded in the September Consolidation. Please confirm that this is a valid opportunity for the May Consolidation. | The need has changed and requirement was deferred to the May 2019 consolidation. |
| VISN 04 | 646-B92001 | The technical elements in this specification are very specific and may be from one vendor. To allow for full competition please consider the following changes. |
Technical Specifications
1. 1.d. Single Plane Floor Mounted System Request - Please consider changing this requirement to a Single Plane, Non-ceiling mounted system. This would allow vendors to offer floor mounted system or a system that is robotic and not mounted to the floor.
2. 1.e Heat dissipation rate must be 1,500,000 HU per min.
Request - To allow for full competition, please remove this as a minimum requirement or change to 544KHU/min (6.72kW)
3. 1.h. Ability for ultra-low dosage during 3D image capture with acquisition rate of at least 90 frames per second Request - Please remove this requirement, it is specific to one vendor General Software Specifications.
4. 2.a. Ability to acquire 3D images in less than 5 seconds Request - Please remove this requirement, it is specific to one vendor
5. 2.b. Ability to automatically acquire 3D images using EKG data Request - Please remove this requirement, it is specific to one vendor
6. 2.d. Ability to generate 3D images at a rate of at least 85/second Request - Please remove this requirement, it is specific to one vendor
7. 2.l. Bi-directional interface with Viista/CPRS Request - Please remove this item Vista/CPRS interaces are afunction of the hemodynamic system, not the imaging system, so this is not relevant
8. 2.q. Ability to integrate with any ultrasound system Request - If the intent is to obtain a video feed, please change this requirement to can integrate with industry standard video feed
9. 2. Ability to provide continuous physiological monitoring during procedures to include but not limited to: NIBP, heart rate, pulse oximetry, respiration rate, end tiday CO2 (ddEtCO2) Request - Please confirm that you are requesting vendors quote hemodynamic monitoring with the Imaging system.
General Hardware Requirements
10. 3.b. Foot-switch activated, overhead intercom system enabling bi-directional communication between control and exam rooms Request - Please change this to must have the ability to intercom to/from control room
11. 3.c. Ability to accept at least 8 different video signals in control room and display on multiple screens Request - Please remove this requirement, it is more of a construction project instead of imaging.
12. 3.e. Radiation shield to protect staff from radiation exposure Request - The Radiation shield is normally quoted by the lighting vendor. Please confirm with the customer and remove if possible.
13. 3.g. Include mobile or table-mounted MEDRAD Stellant injector for contrast delivery: Must have capability of automatic reloading of contrast agent, must have minimum dose delivery of 180cc, must have capability to store 250 cc, must haver interface with table Request - Please change this to a generic Power Injector (The Stellant injector is a CT injector)
14. 3.j. Professional-grade image printer Request - Please remove this requirement
C-arm specifications
15. 4.d. Floor-mounted on retractable, multi-axis robotic arm Request - Please change this to Non-Ceiling mount retractable C-Arm
16. 4.j. Usable space greater than 95 centimeters Request - Please remove this requirement is specific to one vendor
17. 4.k. Arm will be fixed and hermetically sealed at floor-unit juncture Request - Please change this requirement to read, if floor mounted Arm will be fixed and hermetically sealed at floor-unit juncture
Power Requirements
18. 5.a. Include a 480-volt output uninterruptable power supply (UPS) capable of lasting at lease 60 minutes to power the imaging modality to its fullest capacity Request - This is a very expensive UPS requirement, please validate this type of UPS and ensure the facility has the power to support or update to a more generic Full System UPS.
19. 5.b. Bi-directional intercom system and all ancillary components between control room and procedure room to include at least 4 ceiling integrated speakers that terminate in control room Request - Please change the requirement to at least 2 ceiling integrated speakers Catherization/Electrophysiology Capabilities
20. 6.a.Requires 510(k) approved interface with: GE CardioLabs, GE micropace, St. Jude EnSite cardiac mapping, Carto cardiac mapping, Karl Storz AIDA HD Request - Please remove this requirement, 510k is not required.
21. 6.c. Ability to interface with any ultrasound modality Request - If the intent is to obtain a video feed, please change this requirement to can integrate with industry standard video feed.
Video Sources
22. 9.a-l - Video Source Request - This is the function of a video integration unit. All those inputs are not necessary to display on the large display monitor for the procedure. They are used for other in the case on monitors around the room. They are the function of a video integration unit that should be evaluated separately as an independent purchase.
Video Destinations
23. 10b Wall Mounted surgical monitor display 10c Video Monitor included into system Request - the two-video monitor requirements are part of a video integration that should be purchased outside of this purchase please remove.
Security Connectivity Requirements
| 24. | 13.e. Karl AiDA HD and Stream Connect Server | ||
| Request - Please clarify this request. This appears to be an item specific to one vendor, please remove. | 1. | Will retain requirement for floor-mounted | |
| 2. | This will be updated in specifications | ||
| 3. | Updated specification | ||
| 4. | Suggestion noted; specification removed. | ||
| 5. | Updated specification | ||
| 6. | Suggestion noted; specification removed. | ||
| 7. | Suggestion noted; specification removed. | ||
| 8. | Updated specification | ||
| 9. | Updated specification | ||
| 10. | Updated specification | ||
| 11. | Suggestion noted; specification removed. | ||
| 12. | Suggestion noted and moved to value added section | ||
| 13. | Updated specification | ||
| 14. | Suggestion noted and moved to value added section | ||
| 15. | Will retain requirement for floor-mounted | ||
| 16. | Suggestion noted; specification removed | ||
| 17. | Suggestion noted; specification removed | ||
| 18. | Noted and validated this is required. | ||
| 19. | Suggestion noted; specification removed | ||
| 20. | Updated specification | ||
| 21. | Suggestion noted; specification removed | ||
| 22. | Suggestion noted; specification removed | ||
| 23. | Suggestion noted; specification removed | ||
| 24. | Updated specification | ||
| VISN 05 | 540-B81004 | 1. XR Cath Lab for VA Clarksburg, must fit in floor plan provided in the benchmark PDF. Please provide Auto-CAD drawing to facilitate the planners being able to generate site specific drawings in a timely manner to suppor the submmittal deadline. |
2. Can you please provide the following information for the trade-in at VA Clarksburg Product: GE Serial Number:
Manufacturer: 1. CAD drawings will not be provided at this time. Site specific drawings are not required from vendors at this time. Typical drawings of vendor equipment is suggested to be provided with vendor proposals.
2. GE Innova 4100-IQ digital vascular and interventional system S/N 429442, acquired on 9/18/08.
VISN 05 540-B81007 1. What is the Vintage of BOTH GE Infinia nd what CT is on it?
2. Can we get an actual CAD drawing for room 2170 and room 2171?
3. Is this turnkey? If so, will the room renovations be authorized via a modification? 1. The current equipment is a GE INFINIA HAWKEYE installed in 2010.
2. CAD drawings will not be provided at this time.
3. The method of site prep will be discussed with the awarded vendor. No site prep proposal should be included at this time.
VISN 05 540-B81008 1. What is the Vintage of BOTH GE Infinia nd what CT is on it?
2. Can we get an actual CAD drawing for room 2170 and room 2171?
3. Is this turnkey? If so, will the room renovations be authorized via a modification? 1. The current equipment is a GE INFINIA HAWKEYE installed in 2010.
2. CAD drawings will not be provided at this time.
3. The method of site prep will be discussed with the awarded vendor. No site prep proposal should be included at this time.
VISN 05 688-B80805 1. Does the site require long bone/long length imaging capabilities?
2. What level/type of shielding is required, if any?
The attached benchmark is on the master spreadsheet as a RAD system. Can we please have the spreadsheet updated so that it gets submitted into the correct folder. 1. Yes.
2. Shielding requirements of selected unit will be reviewed following award.
688-B80808 is for a RAD room.
688-B80805 is for an RF system VISN 05 688-B80808 This opportunity pushed from September and was listed and benchmarked as RAD opportunity. It now on the May consolidation listed as a RAD spreadsheet opportunity, however, the benchmark calls for RF. Want to confirm that this is indeed calling for an RF. 688-B80808 is for a RAD room.
688-B80805 is for an RF system
| VISN 05 | 688-B80818 | Technical Requirement: Unit Physical Specifications Section d. Minimum detector extended field of view 78cm. | |
| Request - Diagnostic Field of View is 50 cm and that is identified in technical requirement d. The Expanded Field of view requirement (requirement d.) is a specification from one vendor and is unduly restrictive. Please remove requirement the extended FOV from section d. | 1. Specification has been updated |
| VISN 06 | 652-B80059 | 1. | Can we get a CAD drawing of the room for the SPECT/CT? | |
| 2. | Is this a turnkey project? | |||
| 3. | Regarding trade-in of GE Hawkeye 4: |
- Manufacturing date of trade-in?
| - Can we get a list of options? | 1. | No. CAD drawings will not be provided at this time. | ||
| 2. | The method of site prep will be discussed with the awarded vendor | |||
| 3. | The only options are 4 collimators hi, medium, low energy, and pin hole; Purchased June 1 2007 | |||
| VISN 06 | 652-B80065 | 1. | Technical Requirement: Unit Physical Specifications Section e. Minimum detector extended field of view 78cm. |
Request - Diagnostic Field of View is 50cm and that is identified in technical requirement d.
| 2. | The Expanded Field of view requirement (requirement e.) is a specification from one vendor and is unduly restrictive. Please remove requirement e. | |||
| 3. | Technical Requirement: Unit Physical Specifications Section g. Minimum 360 rotations time is .25s. Request - to allow for full competition please change this requirement to .28s. | 1. | Unit Physical Specification requirement “1e” removed. | |
| 2. | Unit Physical Specification requirement “1e” removed | |||
| 3. | 360 rotation time adjusted to 0.28s | |||
| VISN 08 | 516-B90014 | 1. | Specifications 2 a&b (CT Fluoroscopy) - Is this an essential specification? | |
| 2. | Can you please clarify if we need to quote new LAP lasers or will the end user keep the existing LAP lasers? |
Additional Specifications
3. 2.a. The CT scanner shall provide the ability for the user to independently operate the table, gantry, and CT fluoroscopy mode from the control room.
Request - Please confirm if the site is planning to do interventional CT procedures ("Fluoro" - Biopsies, drainages, ablations) on this asset along with the plans to do radiation therapy treatment planning procedures"? If yes, will the interventional procedures be performed from our side of the scan room?
4. 2.b. The CT scanner shall provide the ability for the user to independently operate the table, gantry, and CT fluoroscopy mode while performing biopsy or patient-related exams.
Request - Please confirm if the site is planning to do interventional CT procedures ("Flouro" - Biopsies, drainages, ablations) on this asset along with the plans to do radiation therapy treatment planning drainages, ablations) on this asset along with the plans to do radiation therapy treatment planning procedures"? If yes, will the interventional procedures be performed from our side of the scan room?
| 5. | 2.e. Full UPS |
| Request - Please confirm that the facility would like a Full System UPS instead of a partial UPS. If yes, please confirm that the facility has room and power requirements for a full UPS. | 1. No |
2. Will use existing LAP lasers
3. No interventional CT procedures will be performed.
4. No interventional CT procedures will be performed.
5. Currently have a partial UPS with the current system and would like another partial UPS with the new system. Not enough incoming power for a full UPS.
| VISN 09 | 614-B83004 | 1. | Is this a turnkey that the vendor must include the construction in their proposal? | |
| 2. | What space is available? Can we get a CAD drawing of space? | |||
| 3. | Are they willing to expand the space available if it is not large enough for the equipment? | |||
| 4. | What other cameras do they have today? | |||
| 5. | Please confirm the training we need to include (and option), both for biomedical and technologists | |||
| 6. | Regarding trade-in of GE Hawkeye 4: |
- Age of system
- What collimators, computers included
- Does it include all hard drives, and software with all patient data removed? 1.The method of site prep will be discussed with the awarded vendor. No site prep proposal should be included at this time.
2. N/A - space conversations will take place with the awarded vendors. Space is not a driving factor for this line item. Technology is the driving factor. CAD drawings will not be provided at this time.
3. N/A - space conversations will take place with the awarded vendors. Space is not a driving factor for this line item. Technology is the driving factor.
4. The facility currently has three existing gamma cameras which is inclusive of the specified trade-in. This will be a replacement for one of these units.
5. Please reference the “Instructions to Offers” section 2.8.g for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” section 7.3.3. for response format. Technical training information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type. Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order.
6. GE Infinia Hawkeye 4 Acquisition Date: 04/01/2009 2 collimators included.
Equipment will be positioned as required for deinstallation. After positioning, the hard drives will be removed. The vendor/contractor will then be allowed to remove the equipment from the facility.
| VISN 09 | 626-B90046 | 1. | There is a full UPS for the Gantry, does the facility have the space and power requirements for a full UPS? | |
| 2. | Workstation UPS Requirements, please confirm that this is partial UPS | |||
| 3. | Please confirm that the customer plans to use the Coronary Gating Section 2, f., g. and h. | 1. Yes, site is to be built and space was planned for a full UPS |
2. Partial operation; computer running.
3.Yes, we plan to use Coronary Gating on this unit VISN 09 626-B90046 1. 2. Additional Specifications Section:
Line J/K: UPS, Power Conditioner. Please clarify on type of UPS required, e.g. ability to keep scanning or partial operation of computer / monitor only?
| 2. | Line O: Ceiling-mounted contrast injector. Preference on type of injector? Single or dual head injector? |
| 3. | 3. Workstation Requirements Section: |
| Specific software applications to run on workstation, e.g. calcium scoring, vascular, etc.? | 1. Partial operation; computer running. |
2. MedRad injector preferred; dual head injector needed
3. Software applications: Vascular, calcium scoring, metal artifact reduction (MAR)
| VISN 09 | 626-B90049 | 1. | RFO states "System Configuration" is "Fixed wireless detector", but the Physical Specs sections states "2nd wireless detector". I need to know if they want a single fixed detector in the wall and a single wireless detector for the table. Or do they want (2) wireless detectors and a single fixed detector? | |
| 2. | RFO states "automated system movement (option)". Do we option this or add it to the body of the proposal? | 1. Need (2) wireless detectors and a single fixed detector |
2. Please leave this as an option.
| VISN 10 | 553-B90039 | 1. | Benchmark calls for Multinuclear Support (aka MNS) (Multinuclear Support #Vi) - multinuclear spectroscopy is generally considered a research tool versus single voxel spectroscopy. Please clarify your intent around spectroscopy (both multi and single voxel) so we can provide a thorough response? | |
| 2. | Benchmark calls for a Peripheral Vascular Coil (Peripheral Angio 36 #Sk)- We rarely see this coil used in clinical practice (time consuming for technologist, hard to store, uncomfortable for patient wtih minimal added clinical benefit) and as it tends to add considerable cost. Please explain how this will be used for patients, there may be a better solution? | |||
| 3. | Benchmark calls for a FUll UPS (Eaton 93PM-200 kW UPS) - Please confirm that a full UPS (Easton 93PM-200 kW UPS) – Please confirm that a full UPS is required and there is room and power to install the full UPS or would a partial UPS be preferred? | |||
| 4. | Benchmark calls for some cardiac related acquisition packages, please describe the type and volume of cardiac MRI scanning you are planning to perform and what, if any post processing you will require | 1. | VA research is planned, with clinical applications in Neurology/Psychiatry, Oncology, Cardiology, and MSK. In general, 1H MR spectroscopy translation to clinical practice has been limited, as in many instances the proton nucleus alone is not enough. For example, liver imaging using a combination spectroscopy of H1, P31, and C13 together has been shown to enhance the biochemical analysis of liver imaging. Multinuclear spectroscopy enables imaging of Flourine19, Sodium23, Carbon13, Phosphorous31, Xenon129, Helium3, Lithium7. | |
| 2. | Peripheral runoffs are preferable to CTA in our veteran patient population, as a renal image-softly solution to renal insufficiency. Plan is to convert routine runoff CTAs to runoff MRAs. | |||
| 3. | A Full UPS is required | |||
| 4. | We have had a robust Cardiology-Radiology collegial cardiac CT program here since spring 2016, with cardiac CTs occurring several times a week and all five cardiac imagers reviewing each study collaboratively. Together we decided to hold off with our cardiac MR program until such time as this magnet becomes available. There are two credentialed cardiac cardiologists and three credentialed cardiac radiologists. Plan is to offer the full array of structural and physiologic cardiac clinical MR, as well as planned research production. | |||
| VISN 12 | 550-B80023 | The technical elements in this specification are very specific and may be from one vendor. To allow for full competition please consider the following changes. |
Technical Requirements 1.a. Minimum Wall detector size 17x17 Request - To allow for full competition fo all vendors, please chane the minimum requiremetn to 14" x 17".
1.c. Maximum pixel size 50 Request - This may be an error, please change this requirment to 148 1.e. Height range for wall Bucky 15.75-75.39 Request - Please change this range to 15.75 Bucky Center - 75.5 1.f. Maximum Patient Weight 600 lbs.
Request - Please clarify if this weight is static or dynamic. If dynamic, please change to 586 lbs.
1.p. Heat Unit X-ray Tubes 783 Request - Please change this requirement to 600 1.s. Minimum control room computer 200 Request - Please explain this requirement and measurement Additional Specifications 2.o. Table and upright auto tracking package Request - This is a very specific vendor option, please remove or make optional 2.p. In-room protocol adjustment via tube head or other in-room mounted system Request - This is a very specific vendor option, please remove or make optional 2.s. - Full articulation
| Request - Please explain this requirement | 1. | Updated | ||
| 2. | Was a typo, has been updated | |||
| 3. | Updated | |||
| 4. | Updated to indicate it’s a static weight. | |||
| 5. | Updated | |||
| 6. | Figured this would be where images are initially stored and needs enough storage for OS and to hold reasonable amount of images prior to transfer. | |||
| 7. | Made Optional – Moved to Added Value Section | |||
| 8. | Made Optional – Newer R/F configurations utilize an Overhead Fluoro Tube at the Face, Eye, and Neck level with a digital detector underneath. The under table tubehead configuration has the tube at much lower level, beneath the patient and away from the end user. This reduces possible radiation dose to the lens of the eyes which is advantageous for VA employed Radiologists. Minimal Radiation doses to this area has been found to cause cataracts. Local physicians have also found it much easier to get the desired images with in-room controls for patient safety. Physician is at the patients side for any possible adverse patient reactions. | |||
| 9. | We interpreted it to mean that the table could be adjusted for every possible patient position needed for fluoro imaging. | |||
| VISN 15 | 657-B93008 | 1. | Is this a turnkey that the vendor must include the construction in their proposal? | |
| 2. | What space is available? | |||
| 3. | Can we get a CAD drawing of the space? | |||
| 4. | Are they willing to expand the space available if it is not large enough for the equipment? | |||
| 5. | What other cameras do they have today? | |||
| 6. | Please confirm the training we need to include (and option), both for biomedical and technologists | |||
| 7. | Regarding trade-in of Siemens Symbia S: |
- Age of system?
- What collimators, computers included?
- Does it include all hard drives, and software with all patient data removed?
In general, the specifications for this generic benchmark are very specific to one vendor, as it seems to match a vendor datasheet and may not have much clinical value. Most of the elements can be met by other vendors, but a few are limited to the one vendor; please consider changing the elements below to allow for open competition.
8. Section A., last sentence: Will be equipped with Auto QC and housed table collimator exchanger. Request - Auto QC and Housed table collimator exchanger is a description used by one vendor. Please change this to a generic requirement to allow for full competition and remove this sentence.
9. B. Technical Requirements 1.c. - Intrinsic spatial resolution 3.0mm Request - This is a vendor specific Requirement, please change the specification to ≤ 3.7mm
10. 1.j. Minimum Workstation hard drive space [GB] 1TB Request - Please change this to 750GB
11. 1.g. - Maximum whole-body scan rate [cm/min] - 40 cm/min Request - This is a vendor specific requirement, please change to 30cm/min to allow for full competition
12. 1.l. - Minimum bed weight capacity (lbs.) - 500 lb.
Request - Please change this to a Maximum bed weight capacity (lbs.) - 500 lb.
13. 2. Additional Specifications 2.d. - UPS that provides 10 minutes of backup power to the gantry and workstations to allow for proper shut down.
| Request - Please remove the 10 minutes from the description. Or change the description to a UPS that will allow enough time to remove the patient from the scanner and shut the system down properly. The 10 minutes can't be measured properly. | 1. | The method of site prep will be discussed with the awarded vendor. No site prep proposal should be included at this time. |
| 2. | N/A - space conversations will take place with the awarded vendors. Space is not a driving factor for this line item. Technology is the driving factor. | |
| 3. | CAD drawings will not be provided at this time. | |
| 4. | N/A - space conversations will take place with the awarded vendors. Space is not a driving factor for this line item. Technology is the driving factor. | |
| 5. | The facility currently has two gamma cameras. This will be a replacement for one of these units. | |
| 6. | Please reference the “Instructions to Offers” section 2.8.g for further information about the type of information to provide by equipment type not by specific request. Please also reference the “Instructions to Offers” section 7.3.3. for response format. Technical training information to include detailed information about the curriculum and length of the biomedical technical training required for each equipment type. Although the NAC will not award this training along with the equipment, it is imperative that the customer is informed that this training is available. Vendors must demonstrate that they can provide any required off-site training, therefore off-site training should be quoted as an optional item. Off-site training will be purchased at the time of need via a modification (if the original order remains open) or via a separate order. No travel expenses for any VA employees will be included in any HTME equipment or training order. |
7. The system age is 11 years.
2 Low Energy High Resolution Collimators 1 High Energy Pinhole Collimator 2 Medium Energy Collimators 1 system workstation/computer
- This does not include hard drives. Hard drives will be retained by the VA facility.
| 8. | Specification changed to state: Will be equipped with the ability to conduct QC checks. | |||
| 9. | Suggested change made to updated specifications. | |||
| 10. | Suggested change made to updated specifications. | |||
| 11. | Suggested change made to updated specifications. | |||
| 12. | Suggested change made to updated specifications. | |||
| 13. | Suggested change made to updated specifications. | |||
| VISN 16 | 580-B80022 | 1. | Is this a turnkey that the vendor must include the construction in their proposal? | |
| 2. | If so, can we do a contractor walkthrough? | 1. | Method of site prep will be discussed with the awarded vendor. No site prep proposals are required at this time. | |
| 2. | No walk throughs will be conducted at this time | |||
| VISN 20 | 663-B80032 | 1.Technical Specifications |
1.f. Maximum Patient Weight withstood 550lb
| Request - To all for full competition from multiple vendors please change the minimum requirement to 500lb. | 1. | The minimum requirement has been changed to 500lb | ||
| VISN 22 | 600-B90005 | 1. For the Toshiba Trade In, our trade in desk will want to know if the tube has been replaced, and when (or not), plus current scan seconds on the tube. | 1. Yes, the tube was replaced on 10/11/2017. Tube Scan Seconds: 26,078 |