solicitation questions and answers.docx

DOCX document 17 KB Posted

Attached to
Gas Chromatograph with Flame Ionization Detector Federal contract opportunity
Solicitation number
1333ND26QNB030133
Issued by
Department of Commerce National Institute of Standards and Technology

About this file

This is a Questions and Answers document for a federal solicitation (1333ND26QNB030133) issued by the National Institute of Standards and Technology (NIST) under the Department of Commerce for a Gas Chromatograph with Flame Ionization Detector.

The Q&A addresses the technical specifications and sampling requirements for the solicitation. The government will primarily sample bulk gases including nitrogen, air, helium, and argon at ultra-high purity levels of 99.999% or better. Air samples will contain approximately 78% nitrogen, 21% oxygen, and 0.9% argon. The analytical work will focus on detecting trace impurities such as nitrogen, argon, hydrogen, oxygen, and hydrocarbons, with measurement requirements at sub-ppb (parts per billion) levels. The document indicates that specific compounds of interest may vary depending on the bulk gas being analyzed, and no additional combination of sample gases beyond those specified is anticipated. The required lower detection limits are referenced but not explicitly detailed in this excerpt of the Q&A document.

View the file

Other files for this federal contract opportunity

Other files attached to Gas Chromatograph with Flame Ionization Detector, newest first.
File Type Posted
Attachment 2 Clauses and Provisions.docx DOCX document
attachment 2 diagram.jpg JPG image
CSS 1333ND26QNB030133 Gas Chromatograph .docx DOCX document
Attachment 1 SOW.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1333ND26QNB030133 Questions and Answers Type of gas to be sampled

1. Primarily bulk gases, e.g., nitrogen, air, helium, argon, etc.

1. · General sample composition

1. Ultra-high purity, inert bulk gases, expected purity 99.999 % or better

1. Air samples would contain approx. 78 % nitrogen, 21 % oxygen and 0.9 % argon

1. · Combination of sample gases

1. Air samples would contain approx. 78 % nitrogen, 21 % oxygen and 0.9 % argon

1. Additional combination of sample gases not anticipated

1. · Specific compounds to be measured

1. Any trace impurities we can see, e.g., nitrogen, argon, hydrogen, oxygen, hydrocarbons, etc.

1. Compounds of interest may likely vary depending on bulk gas being analyzed.

1. · Desired measurement ranges

1. Sub-ppb levels

1. · Required lower detection limits (LDL)

1. <1 ppb

1. · Operating voltage

1. 120 volts

1. · Requirement for a rack mount kit

1. No requirement

1. · Requirement for an internal blender

1. No requirement

1. · Converter and VBC (VDC/mA) analog requirements

1. No requirement

File details come from the government source that posted it. Updated .