Seawater Appendix1a.pdf
PDF 12 MB Posted
- Attached to
- Seawater CO2 System Federal contract opportunity
- Solicitation number
- 1333ND24QNB130376
About this file
This document appears to be an appendix related to a federal contract opportunity for a Seawater CO2 System. The solicitation is being issued by the National Institute of Standards and Technology (NIST) Chemical Sciences Division (CSD) to procure a custom designed borosilicate vacuum extraction line. The key objectives are for the contractor to provide scientific glass blowing services to fabricate this custom vacuum extraction line, which is designed to minimize adsorption of carbon dioxide and consists of four modular units on the front side and a mirrored non-modular rear vacuum line. The line includes an acidification module, gas purification traps, and a sample collection module with specified components such as pressure gauges and diffusion pumps. Preliminary schematics and design specifications are provided in the appendix, though the contractor may recommend design changes as needed. The overall purpose is to support NIST's measurement services for ocean carbon metrology and greenhouse gas metrology.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Response to Questions.pdf | ||
| Seawater CO2 System SOW -6-28-2024.pdf | ||
| Appendix 2 - Seawater vacuum line (vessel mark-up).pdf | ||
| Appendix 3 - Seawater CO2 System Parts List - Glassblower -6-28-24.xlsx | XLSX spreadsheet | |
| RFQ -1333ND24QNB130376-Seawater CO2 System-Final.pdf | ||
| Appendix 1 - Seawater system schematics (Drawings).pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
none Appendix 1
Acid reservoir, 3,000 mL Round bottom flask Excel C5
Needle valve w/o-ring Excel C8
14/35, Excel C9, C10
Total height wet side approximately 106 cm
Excel C# references part numbers given in Appendix 2
Section # reference sections listed in part list given in Appendix 2
Page numbers correspond to figures and pictures in this appendix
Wong pipette (completed), page 15
Acid chamber, 40 mL
Extraction flask 500 mL fliter flask w/ side arm
Thimble trap
29/40, Excel C6, C7
14/35, Excel C9, C10
Excel part list section 1.1 page 3, 8, 9, 16, 17
8/19 ball/socket Excel C11-13
29/40, Excel C6, C7
To upper manifold
To dry side, section 2
50/30 ball/socket Excel 27-29
Excel part list section 1.2 page 4, 15, 16, 17
Wet side extraction line
50 cm
0 cm
100 cm
Piriani gauge Excel C51
Reference picture 12 removing ln2 from t2
Connection for cylinder ϭͬϰ͟�ƐƚĂŝŶůĞƐƐ�ƐƚĞĞů�
*Connection to piriani gauge suggested parts: Kovar CG- 306-06 to Swagelok NPT
Stopcock ϰ�ŵŵ͕�ϵϬ϶ Excel C45
8/19 ball/socket Excel C39-41
To upper manifold
*Design input needed, originally part is no longer available, suggested part is listed
Stopcock ϰ�ŵŵ͕�ϵϬ϶ Excel C45
Piriani gauge Excel C51
Piriani gauge Excel C51
*Connection to piriani gauge suggested parts: Kovar CG- 306-06 to Swagelok NPT
*Connection to piriani gauge suggested parts: Kovar CG- 306-06 to Swagelok NPT
To upper manifold
29/43 ground glass joint Excel C37, C38 Round bottom flask
500 mL, Excel C46
Round bottom flask 1,000 mL, Excel C47 Pages 5, 19
Stopcock ϰ�ŵŵ͕�ϵϬ϶ Excel C45
All stopcocks ϰ�ŵŵ͕�ϵϬ϶ Excel C45
8/19 ball/socket Excel C39-41
All stopcocks ϰ�ŵŵ͕�ϵϬ϶ Excel C45
� Excel C# references part numbers given in Appendix 2 � Section # reference sections designated part list given in Appendix 2 � Page number correspond to figures and picture in this appendix � Section brackets also designate suggested break points in extraction line � Total length of wet and dry side is approximately 85 cm
Dry side extraction line
Excel part list section 2.1 Pages 5, 18, 19, 21, 22, 23
Excel part list section 3.1 Pages 6, 18, 20, 21, 22
Excel part list section 4.1 Pages 7, 20, 21
To wet side
10 cm 42 cm 86 cm
Extraction line, wet side
Extraction line, dry side, Traps 1 and 2
Extraction line, dry side, U traps
Extraction line, dry side, sample ports
Extraction line, wet side, back of manifold
Extraction line, connection to manifold
Diffusion pump bypass and main trap
Oil trap
Filling rig
Vacuum gauges and controller
Wong pipette schematic
Extraction line, wet and dry sides
Extraction line, dry side trap 2 in use
Extraction line
Extraction line, in use
Extraction line, dry side, U trap
Extraction line, dry side
Diffusion pump, oil trap
Diffusion pump schematics
Back of extraction line, manifold
File details come from the government source that posted it. Updated .