3 Soliciation Barometer Specifications Document Revised.pdf
PDF 15 KB Posted
- Attached to
- Barometric Pressure Sensors Federal contract opportunity
- Solicitation number
- 1305M224Q0101
About this file
This document is a Barometer Specifications Revised document that provides details on the technical specifications for barometric pressure sensors required for a federal contract opportunity.
The document outlines the specifications for two different models of barometric pressure sensors, the Druck Paroscientific TERPS RPS/DPS 8100 MET 1-2 and an unspecified model. Key details include the output format, operating range, resolution, accuracy, operating temperature range, long-term stability, power requirements, and physical dimensions. The document indicates that the government has a justification for specifying these particular manufacturer models.
The related federal contract opportunity is a Request for Quote (RFQ) for barometric pressure sensors for the National Data Buoy Center at Stennis Space Center, MS. The NAICS code is 334513, and the government intends to use Simplified Acquisition Procedures under FAR Part 13. Contractors must be registered in the System for Award Management (SAM) to be eligible for award.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 3 Soliciation Barometer Specifications Document Revision 3.pdf | ||
| 3 Soliciation Barometer Specifications Document Revision 2.pdf | ||
| 3 Solicitation Amended.pdf | ||
| Solicitation .pdf | ||
| Sensor Brand and Model Justification.pdf | ||
| Barometer Specifications Document.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Barometer Specifications Revised 4/24/2024
Manufacturer Druck Paroscientific
Model TERPS RPS/DPS 8100 MET 1-2
Output Format Digital - RS-232 Digital - RS-232
Range 750-1150 hPa 800-1100 hPa
Resolution 0.01 hPa 0.001 hPa
Accuracy (-40°C to +60°C) Improved Accuracy - 0.04 hPa 0.1 hPa
Operating Temperature Range -10°C to +50°C -40°C to +50°C
Long Term Stability 0.04 hPa /year 0.1 hPa /year
Supply Voltage 7 to 28 VDC 6 to 16 VDC
Maximum Current Draw 32 mA 43 16 mA
Maximum Sensor Length (to fit in enclosure) 4.815 inches N/A - Enclosure provided
Maximum Sensor Diameter (to fit in enclosure) 0.98 inch N/A - Enclosure provided
Electrical Connector Polyurethane Cable N/A Cable internal to enclosure
Cable Length 3 feet N/A Cable internal to enclosure
Pressure Connector 1/4 NPT Female N/A Pressure Connector internal to enclosure
File details come from the government source that posted it. Updated .