Statement of Work Cell Analyzer.pdf

PDF 156 KB Posted

Attached to
Cell Analyzer Federal contract opportunity
Solicitation number
12505B24Q0261
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Midwest Area

About this file

This document is a Statement of Work for a federal contract opportunity to provide a Cell Analyzer (or equal) to the USDA Agricultural Research Service. The overall goal is to use the cell analyzer to phenotype genome-edited cells in poultry and livestock to identify improved traits that can enhance production, animal welfare, and farm profitability. The key technical requirements for the cell analyzer include having two lasers, four fluorescence detectors, a 200-mm quartz flow cell, a sample concentration capacity of over 5 x 10^6 cells per mL, and fluorescence sensitivity for FITC. The associated federal contract opportunity is a Request for Quote (RFQ) issued by the USDA Agricultural Research Service for a Cell Analyzer, laptop with compatible software, and installation/warranty. The NAICS code is 334516 (Laboratory Equipment and Supplies) with a small business size standard of 1,000 employees. The basis for award is Lowest Price Technically Acceptable, and the period of performance is 60 days from time of award.

View the file

Other files for this federal contract opportunity

Other files attached to Cell Analyzer, newest first.
File Type Posted
SF 30 12505B24Q0261 - Amendment 0001.pdf PDF
Provisions and Clauses Updated - Cell Analyzer.pdf PDF
Provisions and Clauses.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work (SOW) Project: Validate Causative Mutations in Agriculturally Important Vertebrates

GENERAL INFORMATION

1.0 Scope of Work:

Provide a Cell Analyzer Brand Name (or Equal) to: BD Accuri C6 Flow Cytometer System.

2.0 Background:

The overall goal is to phenotype genome-edited cells to identify improved traits in pigs and poultry leading to increased production, improved animal welfare enhancing farm profitability and sustainability. Genome editing to produce changes in DNA thought to enhance production agriculture traits will be analyzed by culturing edited cells in vitro to determine if there is a positive effect on cell phenotype. The objective of this work is to develop poultry and livestock genotypes that will provide improved disease resistance traits to viral and/or bacterial pathogens. A new Research Virologist position at USDA-ARS was recently filled. The scientist will use gene editing and other technologies to systematically alter DNA sequences predicted to improve traits to cause beneficial changes in poultry and livestock production and test these changes in vitro and in vivo where warranted. The scientist will investigate and identify host variants associated with disease susceptibility or resistance. The outcome will be improved understanding of host dependencies of pathogens for both poultry and livestock. The scientist will contribute towards identifying and establishing poultry and porcine cell lines that can be used to phenotype agriculturally important traits in vitro while identifying the effects of mutations on these cell lines. Poultry and livestock genotypes that will provide improved disease resistance traits to viral and/or bacterial pathogens will be developed. The cell analyzer with live-cell imaging offers real-time cellular analysis and evaluates cell numbers, cell viability, apoptosis induction, morphology, cell health, proliferation, and cell-mediated toxicity.

CONTRACTOR REQUIREMENTS

3.0 Technical Requirements/Tasks:

• The cell analyzer should be Brand Name (or Equal) to BD Accuri C6 Flow Cytometer System

• The cell analyzer has a minimum of 2 standard lasers (a blue and red laser) offering simultaneous parameter detection for maximum coverage of the fluorochrome emission spectrum. The system should be equipped with two scatter detectors, and four fluorescence detectors with interference filters optimized for the detection of FITC, PE, PerCP-Cy™5.5, and APC

• Should have 200-μm ID quartz flow cell with minimum detectable particle size of 0.5 μm with a minimum sample volume of 50 μL

• Each detector array has to be equipped with appropriately matched optical filters for light collection

• A sheath focused core enables event acquisition rates of up to 10,000 events per second and a sample concentration over 5 x 106 cells per mL

• Should offer a fluorescence sensitivity of FITC <75; PE <50; fluorescence linearity 2 ±0.05% for chicken erythrocyte nuclei (CEN) and fluorescence precision <3% CV for CEN

• Should include tubing peristaltic pump, sheath additive, flow cell clean solution, test beads.

• Must include laptop with software compatible with device

• Must include installation

• Must provide 1-year extended warranty.

4.0 Government Furnished: Not applicable

5.0 Deliverables / Schedule:

Must be delivered within 60 days of award. The equipment must be delivered to:

USDA ARS

University of Missouri

162 ASRC

920 East Campus Dr Columbia, MO 65211

6.0 Travel: Travel is not required

7.0 Contractor’s Key Personnel: NA

8.0 Security Requirements: There are no security requirements.

9.0 Data Rights: There are no data rights to be acquired.

10.0 Section 508 – Electronic and Information Technology Standards: NA

CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
6.0 Travel: Travel is not required

File details come from the government source that posted it. Updated .