1232SA25Q0474 ECS_Specifications_.pdf
PDF 167 KB Posted
- Solicitation number
- 1232SA25Q0474
About this file
This document is a detailed technical specification for an Eddy Covariance System, specifying requirements for scientific measurement equipment from Campbell Scientific. The system includes an IRGASON Integrated Open Path 3-D Sonic Anemometer & CO2/H2O Gas Analyzer, a CR6 Measurement & Control Datalogger with RF407 900MHz Radio Telemetry, a GRANITE Series VOLT116 Analog Input Module, and a 20 ft Stainless Steel Tripod with ground and guy kit. Key technical specifications include operating temperature ranges of -30° to 50°C, low power requirements for remote deployment, compatibility with Campbell Scientific software and equipment, and rugged design capable of prolonged exposure to ambient weather conditions. The system is designed for continuous research-grade measurements of wind velocity, water and carbon dioxide concentration, air temperature, and barometric pressure in agricultural and natural systems.
The document emphasizes the system's ability to be deployed with minimal training, its compatibility with existing Campbell Scientific instruments, and the research team's extensive experience with this equipment. Technical details include precise measurement accuracies, such as sonic anemometer accuracy of ±1 mm/s horizontally and 0.5 mm/s vertically, carbon dioxide accuracy of 0.2 mg/m3, and water vapor accuracy of 0.004 g/m3. The components feature multiple communication ports, low power consumption, and are designed to withstand sustained wind loads up to 40 mph, making them suitable for long-term environmental monitoring in remote locations.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 1232SA25Q0474 SF 1449 (11_2021) SOL.pdf | ||
| 1232SA25Q0474 RFQ TC.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Eddy Covariance Systems Specifications and Requirements
Technical Specifications & Requirements:
IRGASON Integrated Open Path 3-D Sonic Anemometer & CO2/H2O Gas Analyzer General Specifications/Requirements:
▪ Must be of fully integrated design to minimize flow distortion and measurement errors that adversely affect measurement accuracy.
▪ Mus be capable of collecting continuous research-grade measurements of the orthogonal wind velocity components, water and carbon dioxide concentration, air temperature, and barometric pressure.
▪ Must be of rugged design, capable of prolonged (> 1 yr) exposure to ambient weather conditions for long-term deployment in the field
▪ Must be fully compatible and interoperable with Campbell Scientific dataloggers, micrometeorological equipment, and both programming and data processing software.
▪ Must have low power requirements in order to ensure that it can be deployed in remote locations using batteries and solar panels as a part of an automated system.
Technical Specifications/Requirements
▪ Operating Temperature Range: -30° to 50°C
▪ Input Voltage Range 10 to 16 Vdc
▪ Internal Sampling Rate: 60 Hz
▪ Output Sampling Rate: 5, 10, 12.5, or 20 Hz (user-programmable)
▪ Output Options: SDM, RS-485, USB, analog (CO2 and H2O only)
▪ Sonic Anemometer Accuracy: Horizontal: ±1 mm/s Vertical: 0.5 mm/s
▪ Sonic Temperature Accuracy: 0.025 °C
▪ Gas Analyzer Pathlength: 15.37 cm
▪ Carbon Dioxide Accuracy: 0.2 mg/m3 (0.15 μmol/mol)
▪ Carbon Dioxide Calibration Range: 0 to 1,000 μmol/mol
▪ Water Vapor Accuracy: 0.004 g/m3 (0.006 mmol/mol)
▪ Water Vapor Calibrated Range: 0 to 72 mmol/mol
▪ Basic Barometer Accuracy: ±1.5 kPa (0° to 50°C) CR6 Measurement & Control Datalogger with RF407 w/900MHz Radio Telemetry General Specifications/Requirements:
▪ Must be of rugged design, capable of prolonged (> 1 yr) exposure to ambient weather conditions for long-term deployment in the field
▪ Must be fully compatible and interoperable with Campbell Scientific and other micrometeorological equipment.
▪ Must be programmable using Campbell Scientific LoggerNet software and capable of operating the Campbell Scientific data processing software, Easy Flux.
▪ Must have low power requirements in order to ensure that it can be deployed in remote locations using batteries and solar panels as a part of an automated system.
▪ Must be capable of transmitting data via radio telemetry.
Technical Specifications/Requirements
▪ Operating Temperature Range: -40° to 70°C
▪ Universal Input Channels: 12
▪ Number of Channels: 6 differential or 12 single-ended inputs
▪ Number of Excitation Channels: 4
▪ Excitation Voltage: ±5 V
▪ Number of 12 V Outputs: 2
▪ Pulse Counter Channels: 16
▪ Static Vibrating Wire Measurements Supported
▪ Input Ranges: ±5000 mV, ±1000 mV, ±200 mV (analog w/ 24 bit ADC)
▪ Maximum Scan Rate: 1 KHz (1 channel)
▪ Communications Ports: Ethernet, USB Micro B, CS I/O, CPI, RS-485, RS-422, SDI-12
▪ Data Storage Port: microSD
▪ Processor: Renesas RX63N (32-bit at 100 MHz)
▪ Onboard Memory: 128 MB of flash and 4 MB SRAM
▪ Data Storage expandable via removable microSD flash memory, up to 16 GB
▪ Active Power Consumption: 67 mA
▪ Internet Protocols Supported: Ethernet, PPP, RNDIS, ICMP/Ping, Auto-IP (APIPA), IPv4, IPv6, UDP, TCP, TLS (v1.2), DNS, DHCP, SLAAC, Telnet, HTTP(S), SFTP, FTP(S), POP3/TLS, NTP, SMTP/TLS, SNMPv3, CS I/O IP, MQTT
▪ Communications Protocols Supported: CPI, PakBus, SDM, SDI-12, Modbus, TCP, DNP3, UDP, NTCIP, NMEA 0183, I2C, SPI
GRANITE Series VOLT116 16 or 32 Channel Analog Input Module General Specifications/Requirements:
▪ Must be of rugged design, capable of prolonged (> 1 yr) exposure to ambient weather conditions for long-term deployment in the field
▪ Must be fully compatible and interoperable with Campbell Scientific and other micrometeorological equipment.
▪ Must be programmable using Campbell Scientific LoggerNet software and capable of operating the Campbell Scientific data processing software, Easy Flux.
▪ Must have low power requirements in order to ensure that it can be deployed in remote locations using batteries and solar panels as a part of an automated system.
Technical Specifications/Requirements:
▪ Operating Temperature Range: -40° to 70°C
▪ Power Requirements: 9.6 V
▪ Active Power Consumption: 20 mA
▪ Number of Channels: 16 differential or 32 single-ended inputs
▪ Number of Excitation Channels: 4
▪ Excitation Voltage: ±5 V
▪ Number of 12 V Outputs: 2
▪ Input Ranges: ±5000 mV, ±1000 mV, ±200 mV (analog w/ 24 bit ADC)
▪ Maximum Scan Rate: 1 KHz (1 channel); 167 Hz (16 channel)
Campbell Scientific 20 ft Stainless Steel Tripod w/ Ground & Guy Kit and both the 4 ft and 6 ft Sensor Crossarms
General Specifications/Requirements:
▪ Must be portable to facilitate transportation and field deployment at remote locations.
▪ Must be 20 ft in height and capable of tilting to facilitate deployment and maintenance in agricultural environments.
▪ Must be of rugged design, capable of prolonged (> 1 yr) exposure to ambient weather conditions for long-term deployment in the field
▪ Must be fully compatible with Campbell Scientific and other micrometeorological equipment. Peripherals, and hardware including mounts, crossarms, and enclosures.
Technical Specifications/Requirements
▪ Tripod Material: Stainless steel
▪ Crossarm Material: Anodized aluminum
▪ Height: 20 ft
▪ Crossarm Lengths: 4 ft and 6 ft
▪ Tripod Outer Diameter: 1.5 – 2 in
▪ Crossarm Outer Diameter: 1.1315 in
▪ Vertical Load Limit: 100 lb
▪ Base Diameter :7 ft with legs fully extended
▪ Sustained Wind Load Tolerance (guyed at feet): 40 mph
▪ Gust Wind Load Tolerance (guyed at feet): 55 mph
16 ×18 in Enclosure w/ dual Conduits, Tripod Mast Mounting, & Standard Backplate General Specifications/Requirements:
▪ Must be of rugged design, capable of prolonged (> 1 yr) exposure to ambient weather conditions for long-term deployment in the field
▪ Must be fully compatible with Campbell Scientific and other micrometeorological equipment. Peripherals, and hardware including mounts, crossarms, and enclosures.
Technical Specifications/Requirements
▪ Internal Dimensions: 18 × 16 × 9 in
▪ External Dimensions: 19.5 × 17.5 × 10.6 in
▪ Material: Fiberglass-reinforced polyester
▪ Number of Cable Conduits: 2
▪ Lockable Hasps
▪ Weather-resistant with door gasket
▪ Grounding Lug
▪ Reflective white exterior
Other General Considerations for all Eddy Covariance System Components
▪ Weather-resistant and rugged construction capable of prolonged (> 1 yr) exposure to ambient weather conditions to facilitate long-term deployment in agricultural and natural systems.
▪ Fully compatible/interoperable with Campbell Scientific instruments, dataloggers, and software currently used by SWAR and NLAE scientists.
▪ Must have low power requirements in order to ensure that it can be deployed in remote locations.
▪ Can be deployed and maintained with minimal training or supervision by personnel with limited expertise in micrometeorology and environmental measurements.
▪ The research team has extensive experience programming, deploying, and collecting data with Campbell Scientific instrumentations and software. The continued use of Campbell Scientific equipment facilitates deployment and maintenance of equipment, data collection and processing, and when necessary, troubleshooting and field repairs. It also ensures the accuracy, consistency, and integrity of the data.
File details come from the government source that posted it. Updated .