12-EXHIBIT 1 - STERIS Higlighted Tech Data and Equipment Drawing.pdf
PDF 2 MB Posted
- Attached to
- Twin Lakes SSPD Equipment Replacement
- Solicitation number
- PROJ-25-134
- Issued by
- Volusia County, Florida
About this file
This file is a technical specification document for AMSCO® 400 Small Steam Sterilizers manufactured by STERIS Corporation. The document details steam sterilization equipment designed for healthcare facilities, featuring two chamber sizes (16 x 16 x 26 inches and 20 x 20 x 38 inches) with options for single or double doors, cabinet-enclosed or recessed mounting configurations. The sterilizers are equipped with prevacuum, gravity, leak test, and daily air removal test cycles, with an optional Steam Flush Pressure Pulse (SFPP) configuration. Key features include touch-screen controls with thermal printer, vertical sliding doors with hands-free operation, and electronic water-saving controls. The equipment meets ANSI/UL, AAMI, and ASME standards and can operate with either facility-supplied steam or integral electric steam generators.
The document includes comprehensive utility requirements, with sterilizers requiring 1/2" NPT steam at 50-80 psig, electrical connections ranging from 120V single-phase for controls to 480V three-phase for steam generators, and 1" NPT feed water at 30-50 psig. Water quality specifications are detailed for different configurations, with carbon steel steam generators requiring specific conductivity levels (38-500 μS/cm) and stainless steel generators requiring deionized water with minimum 1MΩ-cm resistivity. Operating weights range from 750-1726 pounds depending on configuration, with heat loss ratings from 4300-8750 BTU/hr. The equipment includes connectivity options for remote monitoring through STERIS's ConnectCare and ConnectAssure technologies, requiring either Ethernet or cellular connections for cycle data export and performance monitoring.
View the file
Other files
Show all 12
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 1
AMSCO® 400 Small Steam Sterilizer Steam Sterilization
APPLICATION
AMSCO 400® Series Small Steam Sterilizers are designed for sterilization of materials used in healthcare facilities.
The sterilizers are designed for fast, efficient sterilization of heat- and moisture-stable materials in addition to sterilization of items for immediate use. AMSCO 400 Series Small Steam Sterilizers are equipped with prevacuum, gravity, leak test, and daily air removal test cycles. An optional Steam Flush Pressure Pulse (SFPP) configuration sterilizer adds SFPP cycles.
Each sterilizer is equipped with either a single or double door, for open or recessed mounting. (Recess mounting is not available for 16 x 16 x 26" double door sterilizers.)
STANDARDS
Each sterilizer meets applicable requirements of the following listings and standards, and carries the appropriate symbols:
• ANSI/UL 61010-1 and CAN/CSA-C22.2 No. 61010-1 –
Standard for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements
• ANSI/UL 61010A-2-041 – Standard for Electrical Equipment for Measurement, Control and Laboratory Use, Part 2: Particular Requirements for Autoclaves using Steam for the Treatment of Medical Materials and Laboratory Processes
• ANSI/AAMI-ST8:2008 “Hospital Steam Sterilizers” American National Standard
• ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form U-1 is furnished. Shell and door are constructed to withstand working pressure of 50 psig (344.7 kPa).
• ASME Code, Section I, Part PMB for power boilers, if optional steam generator is supplied.
• CAN/CSA-C22.2 No. 61010-1
DESCRIPTION
AMSCO 400 Series Small Steam Sterilizers are the next advancement in the STERIS line of steam-jacketed sterilizers and are equipped with the latest features in both state-of-the-art technology and ease of use.
Primary Product Features The control system for the AMSCO 400 Series Small Steam Sterilizers features enhanced functionality and user-friendly interface screen.
• Touch-screen with 30–line x 40 character display area
• Thermal Printer
• Help screens for programming and troubleshooting alarm conditions
• Automatic check of control program and cycle data maintains process integrity
• Service reprogrammable flash ROM memory
• Available with optional water-saving electric vacuum pump
• Vertical sliding door with hands– free loading and unloading capability
• Foot pedal activated door opening and closing
• Non-lubricated, steam activated door seal
• Valve manifolds increase dependability and reduce servicing time
• Reduced piping components increase reliability
• Emergency manual exhuast valve
• Electronic water saving control Interior Chamber Dimensions
• 16 x 16 x 26" (406 x 406 x 660 mm)
• 20 x 20 x 38" (508 x 508 x 965 mm) dacciarri Highlight dacciarri Highlight dacciarri Sticky Note Sub Sterile Dimensions dacciarri Sticky Note Sterile Processing Dimensions
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 2
FEATURES
Rack and shelf design accepts wider loads: 16 x 16" sterilizers – chamber clearance is 12" (304 mm) for top shelf and 14" (357 mm) for bottom shelf. 20 x 20" sterilizers – chamber clearance is 18" (457 mm) for intermediate shelves, and 15" (381 mm) for bottom shelf.
User-programmable cycle names allow for load specific naming of cycles. These cycle names are displayed and printed in addition to the factory-default cycle type and aid in identifying the proper cycle to be used with a specific load.
Hinged front cabinet panel fully opens for convenient access to sterilizer piping and control.
Software calibration is performed in the Service Mode, accessible through the touch-screen displays, and accomplished using external or internal temperature and pressure sources. Control system provides printed record of all calibration data for verification of current readings.
Lighted DIN connectors are installed on all steam, water and exhaust valves for reliability and ease of maintenance.
Steam generator units are equipped with an automatic flush and drain system. This system helps the generator to operate at peak performance and extends the life of the heaters.
UTILITIES CONSERVATION FEATURES
Resistance Temperature Detectors (RTD) are installed for sterilizer temperature control. The chamber drain line RTD senses and controls temperature variations within the sterilizer chamber. A jacket RTD provides temperature control within the jacket space. These RTD signals, converted into electrical impulses, provide accurate control inputs and readouts throughout entire cycle and minimizes utilities usage.
Electronic water saving control includes a condenser RTD to control the amount of water used in condensing the exhausted chamber steam. Control software minimizes amount of water used to cool condensate.
Automatic utilities startup/shutdown may be programmed to activate at a specific time of day. When activated, control system automatically shuts off all utility valves, conserving steam and water usage. Sterilizer utilities can be restarted either by programmed time or manual operation. A different shutdown and restart time can be programmed for each day.
Insulation, one-inch thick, asbestos-free spin-glass (rated at 500°F [260°C] continuous) encompasses the exterior of the sterilizer vessel and is sealed in an oil and water resistant outer jacket.
Vacuum SystemWater ejector reduces chamber pressure during prevacuum and post-drying phases. Air is drawn from chamber through the vacuum system. Following dry phase, chamber vacuum is relieved to atmospheric pressure by admitting air through a bacteria-retentive filter.
An optional vacuum pump can be ordered in place of the standard water ejector. The vacuum pump provides equivalent performance, but reduces cooling water consumption by up to 60%, helping facilities to conform with Leadership in Energy and Environmental Design (LEED) requirements.
SAFETY FEATURES
Control senses when the door is closed and sealed, preventing cycle start until a limit switch signal is received. If control loses appropriate signal during cycle, alarm activates, cycle aborts and chamber safely vents with a controlled exhaust.
Chamber Float Switch activates alarm, aborts cycle and safely vents chamber with a controlled exhaust if excessive water is detected in the vessel chamber.
Pressure Relief Valve limits amount of pressure buildup so that rated pressure of vessel is not exceeded.
PROCESSING CYCLES
All processing cycles factory programmed into the sterilizer control have been validated to AAMI ST8: 2008 with 25 pound trays Prevacuum Sterilizer Models feature the following cycles:
Prevac Pre-set Cycles
Cycle Type and Use Temperature Sterilize Time Dry Time
Prevacuum Minutes
Wrapped Instrument Trays 270°F (132°C) 4 30
275°F (135°C) 3 30
Full Load Fabric 270°F (132°C) 4 20
Single Fabric Pack 270°F (132°C) 4 5
Immediate Use Prevac Minutes
Unwrapped Instrument trays (single or multiple unwrapped instruments) 270°F (132°C) 4 1
Gravity
Fabric Packs 250°F (121°C) 30 15
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 3
Wrapped Instrument Trays 270°F (132°C) 15 30
Wrapped Instrument Trays 250°F (121°C) 30 30
Fabric Packs 270°F (132°C) 25 15
Immediate Use Gravity
Unwrapped Instrument trays (single or multiple unwrapped instruments)
270°F (132°C) 3 1
270°F (132°C) 10 1
Liquid
Liquids 250°F (121°C) 45 N/A
Test
Warm-up 270°F (132°C) 3 1
Bowie-Dick/Dart 270°F (132°C) 3.5 1
Leak Test N/A
SFPP Pre-set Cycles
Cycle Type and Use Temperature Sterilize Time Dry Time
SFPP Minutes
Wrapped Instrument Trays 270°F (132°C) 4 30
275°F (135°C) 3 30
Full Load Fabric 270°F (132°C) 4 20
Single Fabric Pack 270°F (132°C) 4 5
Prevac Minutes
Wrapped Instrument Trays 270°F (132°C) 4 30
275°F (135°C) 3 30
Full Load Fabric 270°F (132°C) 4 20
Single Fabric Pack 270°F (132°C) 4 5
Immediate Use Prevac
Unwrapped Instrument trays (single or multiple unwrapped instruments) 270°F (132°C) 4 1
Gravity
Fabric Packs 250°F (121°C) 30 15
Wrapped Instrument Trays 270°F (132°C) 15 30
Wrapped Instrument Trays 250°F (121°C) 30 30
Fabric Packs 270°F (132°C) 25 15
Immediate Use Gravity
Unwrapped Instrument trays (single or multiple unwrapped instruments)
270°F (132°C) 3 1
270°F (132°C) 10 1
Liquid
Liquids 250°F (121°C) 45 N/A
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 4
Test Warm Up 270°F (132°C) 3 1
Bowie-Dick/Dart 270°F (132°C) 3.5 1
Leak Test N/A
PREVACUUM TESTING CYCLES
• Vacuum Leak Test: This cycle is used for testing the vacuum integrity of sterilizer piping. The sterilizer chamber must be empty while running this test cycle. Temperature:
270°F (132°C); all timing is preprogrammed and cannot be adjusted. This cycle is validated to AAMI standard
ST8:2008.
• A preprogrammed Bowie-Dick Test Cycle is used to test for adequate air removal from the sterilizer chamber.
Recommended load is a Dart® testing apparatus from STERIS, or a properly prepared Bowie-Dick test pack.
Preprogrammed cycle parameters cannot be adjusted by user. Sterilize exposure temperature: 270°F (132°C);
sterilize exposure time: 3-1/2 minutes; dry time: 1 minute.
This cycle is validated to AAMI standard ST8:2008.
CONTROL SYSTEM
Design Features The control system for the AMSCO 400 Series Small Steam Sterilizer monitors and controls all sterilizer operations and functions. The control system is factory-programmed with standard sterilizing cycles. Each cycle is adjustable, and cycle names are user-programmable, to meet specific processing requirements. Cycle parameters cannot be adjusted to less than preset, validated settings. All control configuring is performed through touch-screen display.
NOTE: Important: Always refer to instrument manufacturer’s instructions for use to determine processing requirements.
Cycle values and operating features may be adjusted and verified prior to cycle operation. Once cycle is started, cycles and cycle values cannot be changed until cycle is complete.
On completion of cycle, timers reset to the previously selected values, eliminating the need to reset values between repeated cycles. If chamber temperature drops below setpoint during the exposure phase the timer stops. It automatically resets once normal operating temperature is reached.
Critical control system components are housed within a compartment to protect the components from moisture and heat generated during sterilization. A cooling fan with filter maintains air flow within the compartment, keeping components cool.
Operator interface control panel, consisting of a touch-screen and a quiet and reliable thermal printer located at sterilizer’s operating end (OE – loading end or nonsterile end) of the sterilizer. If sterilizer is equipped with double doors, an additional touch-screen (but no printer) is provided on the sterilizer non-operating end (unloading or sterile).
• Touch-Sensitive Screen features a multi-color, touchscreen graphics display. Sterilizer is operated by pressing touch-sensitive areas on display (touch pads).
Appropriate touch pads, operator prompts and status messages necessary for sterilizer operation appear on display as needed. Display also indicates any alarm conditions either in or out of cycle.
• Thermal Printer High resolution (8 dots/mm) thermal printer is fast and quiet, printing 25 lines/minute on industrial grade thermal paper. Printout has a guaranteed shelf-life of 25 years and is resistant to exposure from steam, alcohol, UV and visible light, oil, heat and water.
Non-operating end (NOE) control panel, equipped on double-door sterilizers only, includes a touch-screen similar to the operating end screen, but no printer. Preprogrammed cycles can be started from the NOE control panel. Display concurrently shows the same information as the operating end screen display.
Cycle configuration is performed by accessing the Change Values menu through the operating end touch-screen. In addition to adjustment of cycle values, the following operating parameters can also be changed through the Change Values menu:
• Time Display and Printout Units – Standard AM/PM or 24-hour.
• Access Code – accessing Change Values menu causes display to request the entry of an access code. If access code is not properly entered, display returns to menu screen, denying user access to the sterilizer programming.
Supervisors can allow operators to change chosen cycle and parameters; or lock them out from making any changes.
• Audible Signals – are adjustable. Sounds made when touching the screen and for end-of-cycle signals can be adjusted to one of four sound levels (off, low, medium or high) as required for the operating environment. The alarm signal can be adjusted to low, medium or high; but cannot be turned off.
• Print Format – allows selection of either a full or condensed printout of the cycle information during processing.
• Temperature Display and Printout Units – Fahrenheit (°F) or Celsius (°C). Temperature is set, displayed, controlled and printed to the nearest 0.1°. Recalibration is not required when changing temperature units from °F to °C and vice versa.
• Pressure/Vacuum Display and Printout Units – psig/In Hg, millibar or psia. Recalibration is not required when changing pressure units.
• Utilities Control – This parameter permits the operator to program the sterilizer to automatically shut off its steam and water at the end of the work day, to conserve utilities.
It also allows control for shut down and power-up of an integral steam generator.
• Languages – This parameter can be used to select English, French or Spanish as the default for displays and printouts.
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 5
The sterilizer can also be set to allow quick changes between available languages.
• Machine Number – This parameter assigns a six-character, alphanumeric code to the sterilizer. This code appears in the heading of all printouts.
• Automatic Duplicate Print – Sterilizer can be set to automatically furnish a duplicate printout of each cycle at the end of the cycle. First line reads “DUPLICATE PRINT.”
Technical Data Control system consists of a microcomputer control board and peripheral function circuit boards, located within the control board housing behind the front cabinet service panel above the chamber.
A memory backup system maintains cycle settings indefinitely and current cycle information for approximately five days. If a power failure occurs during a cycle, the battery backup system ensures that cycle memory is retained and proper cycle completion occurs once power is restored. When power is lost, the cycle is held in phase until power is restored, exceeding the minimum government specification of one minute. Once power returns, the event is recorded on the printout and the cycle automatically resumes or restarts, depending on what phase the cycle was in at the time of power loss. If necessary, the operator can manually abort the cycle.
CONSTRUCTION
Shell Assembly Two fabricated Type 316L stainless-steel shells, welded one within the other, form the sterilizer vessel. Type 316L stainless-steel end frame(s) is welded to door end. On single door units, back of chamber is fitted with welded, 316L stainless-steel formed head.
Sterilizer vessel is ASME rated at 50 psig and insulated.
Vessel (20 x 20" units only) includes one 1.0"-NPT welded chamber bushing for Customer use.
Steam-supply opening inside the chamber is shielded by a Type 316L stainless-steel baffle.
Chamber Door(s) Door is constructed of a single formed piece of Type 316L stainless steel.
During cycle operation, door is sealed by a steam-activated door seal. Door seal is constructed of a special long-life rubber compound. When sterilizer cycle is complete, the seal retracts under vacuum into a machined groove in the sterilizer end frame. Door seal can be manually retracted to open door and remove critical load in emergency situation if loss of vacuum or loss of power occurs.
Door is suspended by cables attached to a counterweight.
Chamber door is opened (lowered) and closed (raised) by pressing a foot pedal located on the same end as the door being operated. In case of a power or mechanical failure, door can be operated manually.
A long-life proximity switch is used by the control to determine if door is closed. An additional seal pressure switch prevents inadvertent cycle initiation if door is not sealed.
The door assembly is equipped with a mechanical locking mechanism that ensures the door cannot be opened as long as the seal is intact and energized and more than 2.0 psi pressure is in the chamber.
The sterilizer door opening is fitted with a textured thermoplastic bezel. This bezel insulates the operator from the chamber end ring, lessening the chance of accidental contact with a hot metal surface.
Chamber Drain System Drain system is designed to prevent pollutants from entering into the water-supply system and sterilizer. The automatic condensing system converts chamber steam to condensate and disposes condensate to waste. Cooling water flow is regulated by the waste line RTD to minimize water usage.
Water supply shutoff valve is located behind the front cabinet service panel under the chamber.
Steam Source Sterilizers are piped, valved and trapped to receive building-supplied steam delivered at 50 to 80 psig dynamic. If building steam source is not available, an electric carbon-steel steam generator may be provided to supply steam to the sterilizer.
Steam piping is constructed of brass and includes a shutoff valve, steam strainer, flush system and a brass pressure regulator.
Piping All piping connections terminate within the confines of the sterilizer and are accessible from front and side of sterilizer.
• Manual Shutoff Valves are pressure rated at 125 psig for saturated steam. Valve handles are low-heat conducting.
• Solenoid Valves in the manifold with DIN connectors simplify sterilizer piping and can be serviced individually.
MOUNTING ARRANGEMENT
Sterilizers are arranged for either freestanding or recessed installation, as specified. Each sterilizer is equipped with a height-adjustable steel floor stand. Sterilizer subframe is equipped with a synthetic rubber gasket to ensure tight fit between the cabinet panels on freestanding units or between the front cabinet panel and wall partition on recessed units.
On freestanding units, stainless-steel side panels and a louvered top panel enclose the sterilizer body and piping.
ACCESSORY
Seismic Tie-Down Kit – conforms to California Code of Regulations.
CONNECTIVITY
ConnectCare™ Remote Monitoring, ConnectAssure™ Cycle Data Export and SPM™ Instrument Tracking The AMSCO 400 is equipped to remotely monitor connected equipment to view equipment status, current cycle and alarm information, and performance-related statistics in real time.
STERIS Service Technicians will have access to equipment performance data to facilitate a proper response and reduce downtime. Remote monitoring requires an Ethernet connection (CAT-5), or cellular module, which can be provided by STERIS following a signal strength test.
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 6
Cycle data export (connectivity) is available for the AMSCO 400 through STERIS’s ConnectAssure Technology, which can send machine cycle data directly to an instrument tracking system. ConnectAssure can be used with multiple tracking software systems, but might require an additional interface.
ConnectAssure requires an Ethernet connection (CAT-5).
SPM® Surgical Asset Tracking Software For instrument tracking and workflow management needs. SPM® Surgical Asset Tracking Software supports key aspects of the sterile processing function, guiding technicians with tools to deliver conforming work and equipping department managers with the reporting and decision-support tools necessary for managing the demands of the OR schedule.
PREVENTIVE MAINTENANCE
A global network of skilled service specialists can provide periodic inspections and adjustments to help ensure low-cost peak performance. STERIS representatives can provide information regarding annual maintenance programs.
NOTES
1. The sterilizer is not supplied with a vacuum breaker or backflow preventer and, where required by local codes, installation of such a device in the water line is not provided by STERIS.
2. Pipe sizes shown indicate terminal outlets only.
Building service lines (not provided by STERIS), must supply the specified pressures and flow rates.
3. Disconnect switches (with OFF position lockout only;
not provided by STERIS) should be installed in electric supply lines near the equipment.
4. Access to the recessing area from the control end of the sterilizer is recommended.
5. Clearances shown are minimal for installing and servicing the equipment.
6. If loading car and carriage are to be used with a 20 x 20 x 38" sterilizer, front clearance should be at least 76" (1930 mm). This permits complete withdrawal of the loading car from the chamber and allows convenient maneuverability of the transfer assembly to and from the sterilizer.
7. Floor drain should be provided within confines of sterilizer framework.
UTILITY REQUIREMENTS
Sterilizer Using Facility Steam
• Steam- 1/2” NPT, 50 to 80 psig, dynamic, 97 to 100% vapor quality.
• Drain- 1–1/2” ODT drain terminal. (Floor drain capacity must handle peak water consumption; refer to Engineering data.)
• Electrical - Controls
– 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp
• Electrical – Vacuum Pump Option
– 208 Volt, 50/60 Hz, 3-phase, 4.5 Amps
– 240 Volt, 50/60 Hz, 3-phase, 4.5 Amps or
– 480 Volt, 50/60 Hz, 3-phase, 2.2 Amps
• Sterilizer Feed Water1- 1" NPT
– 30 to 50 psig for water ejector
– 20 to 50 psig for vacuum pump or
– 40 to 50 psig for SFPP sterilizers.
NOTE: Backflow prevention (not supplied on unit) is not provided by STERIS.
Sterilizer Equipped with Integral Carbon Steel Steam Generator Every AMSCO 400 Series Small Steam Sterilizer equipped with an electric steam generator includes an automatic flush and drain package.
• Drain- Sterilizer- 1–1/2” ODT drain terminal. (Floor drain capacity must handle peak water consumption; refer to Engineering Data.)
• Drain- Steam Generator- 1/2” ODT
• Electrical - Controls
– 120 Volt, 50/60 Hz, 1-phase, 9.5 Amps2 or
– 220 Volt, 50/60 Hz, 1-phase, 5.0 Amps3
• Electrical - Steam Generator
– 208 Volt, 50/60 Hz, 3-phase, 83 Amps
– 240 Volt, 50/60 Hz, 3-phase, 72 Amps
– 380/415 Volt, 50/60 Hz, 3-phase, 38/42 Amps or
– 480 Volt, 50/60 Hz, 3-phase, 36 Amps
– 600 Volt, 50/60 Hz, 30.1 Amps
1. Water is used for ejector (creating chamber vacuum), and exhaust cooling. Refer to Table 1 for recommended water quality. Use of feed water within the nominal conditions optimizes equipment performance and helps reduce maintenance.
2. The 120V, 50/60 Hz control is always used with the 208V, 240V, and 480V integral steam generator or vacuum pump option.
3. The 220V, 50/60 Hz control is always used with the 380/415V integral steam generator.
dacciarri Highlight dacciarri Highlight
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 7
• Electrical – Vacuum Pump Option
– 208 Volt, 50/60 Hz, 3-phase, 4.5 Amps
– 240 Volt, 50/60 Hz, 3-phase, 4.5 Amps or
– 480 Volt, 50/60 Hz, 3-phase, 2.2 Amps
• Feed Water – Sterilizer4- 1" NPT
– 30 to 50 psig for water ejector
– 20 to 50 psig for vacuum pump
– 40 to 50 psig for SFPP sterilizers
• Feed Water – Steam Generator5 - 1/2" NPT
– 20 to 50 psig
4. Water is used for ejector (creating chamber vacuum), or vacuum pump option, exhaust cooling and cooling the generator drain. Refer to Table 1 for recommended water quality. Use of feed water within the nominal conditions optimizes equipment performance and helps reduce maintenance.
5. Water is used for integral steam generator.Refer to Table 2 for recommended water quality. Use of feed water within the nominal conditions optimizes equipment performance and helps reduce maintenance.
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 8
ENGINEERING DATA
Size in (mm) Heating
Maximum Operating Weight* lbs (kg)
HEAT LOSS†
BTU/hr at 70°F (21°C)
Single Door Double Door
Single Door Double Door
Cabinet Enclosed Recessed Recessed One
Wall Recessed Two
Walls
To Room Front of
Wall Back of Wall
Front of
Wall
Back of
Wall
At Each End
Between Walls
16 x 16 x 26 (406 x 406 x 660)
Steam‡ Electric**
750 (340) 890 (404)
989 (449) N/A
N/A
N/A
N/A N/A
N/A N/A
20 x 20 x 38 (508 x 508 x 965)
Steam‡ Electric**
1230 (558) 1371 (622)
1606 (728) 1726 (782)
* Based on chamber fully loaded with water flasks.
† At 70°F (21°C).
‡ In the Heating column, “Steam” refers to External Supplied Steam (Facility Steam/Stand-Alone Steam Generator).
** In the Heating column, “Electric” refers to Integral Steam Generator.
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 9
Size in (mm) Heating
Water Ejector (WE) or Optional Vacuum
Pump (VP)
UTILITIES CONSUMPTION*
Water† Steam
Cold Hot‡
Peak gpm (lpm)
Average Usage gal/cycle (l/Cycle)
Idle gph (lpm)
Peak gpm (lpm)
Per Cycle gal/cycle (l/cycle)
Idle gph (lph)
Peak** lb/hr (kg/hr)
Per Cycle lb/cycle (kg/cycle)
Idle lb/hr (kg/h)
16 x 16 x 26 (406 x 406 x
660)
Steam†† WE
VP
(57) (38)
135 (511) 50 (186)
(45) N/A N/A N/A 158 (72) 30 (14) 7 (3)
Electric‡‡ WE
VP
(57) (38)
135 (511) 50 (186)
(45) 1 (4) 3 (11) 1 (4) N/A N/A N/A
20 x 20 x 38 (508 x 508 x
965)
Steam†† WE
VP
(57) (38)
175 (662) 70 (261)
(45) N/A N/A N/A 158 (72) 42 (19) 9 (4)
Electric‡‡ WE
VP
(57) (38)
175 (662) 70 (261)
(45) 1 (4) 5 (19) 1 (4) N/A N/A N/A
* Data is based on 270°F (132°C), 4 minute sterilize, 30 minute dry cycle, processing 25 lb (11kg) instrument trays, maximum load in chamber.
† Backflow preventer device in water line, when required by local codes, is installed by others.
‡ Hot water recommended for units equipped with electric steam heat.
** Peak steam demand (lb/hr) may vary depending on operating conditions.
†† In the Heating column, “Steam” refers to External Supplied Steam (Facility Steam/Stand-Alone Steam Generator).
‡‡ In the Heating column, “Electric” refers to Integral Steam Generator.
Refer to the Following Equipment Drawing for Installation Details
Equipment Drawing Number Equipment Drawing Title
129394-044 16 x 16 x 26", single door, cabinet enclosed with steam heat
129394-045 16 x 16 x 26", single door, recessed one wall with steam heat
129394-046 16 x 16 x 26", single door, recessed one wall with electric heat
129394-047 16 x 16 x 26", single door, cabinet enclosed with electric heat
129394-048 16 x 16 x 26", double door, recessed one wall with cabinet and steam heat
129394-049 20 x 20 x 38", single door, cabinet enclosed with steam heat
129394-050 20 x 20 x 38", single door, recessed one wall with steam heat
129394-051 20 x 20 x 38", double door, recessed one wall with cabinet and steam heat
129394-052 20 x 20 x 38", double door, recessed two walls, with steam heat
129394-053 20 x 20 x 38", single door, cabinet enclosed with electric heat
129394-054 20 x 20 x 38", single door, recessed one wall with electric heat
129394-055 20 x 20 x 38", double door, recessed one wall with electric heat and cabinet
129394-056 20 x 20 x 38", double door, recessed two walls with electric heat
10066840 16 x 16 x 26", single door, recessed one wall with steam heat and vacuum pump
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 10
Refer to the Following Equipment Drawing for Installation Details
Equipment Drawing Number Equipment Drawing Title
10066841 16 x 16 x 26", single door, cabinet with steam heat and vacuum pump
10066842 16 x 16 x 26", single door, recessed one wall with electric steam heat and vacuum pump
10066843 16 x 16 x 26", single door, cabinet with electric steam heat and vacuum pump
10066844 20 x 20 x 38", single door, recessed one wall with steam heat and vacuum pump
10066845 20 x 20 x 38", single door, cabinet with steam heat and vacuum pump
10066846 20 x 20 x 38", single door, recessed one wall with electric heat and vacuum pump
10066847 20 x 20 x 38", single door, cabinet with electric heat and vacuum pump
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 11
Selections Checked Below Apply To This Equipment
Size/Type 16 x 16 x 26” Prevacuum with Liquid Cycle 20 x 20 x 38” Prevacuum with Liquid Cycle 16 x 16 x 26” SFPP and Prevacuum with Liquid Cycle 20 x 20 x 38” SFPP and Prevacuum with Liquid Cycle
Steam Source
Facility Steam6
120 Volt Control Integral Steam Generator7
Control Voltage 120 Volt Control8
220 Volt Control9
Electrical (for generator power):
208 Volts 240 Volts 480 Volts 380/415 Volts 600 Volts (not standard, requires
SSQ)
Vacuum System
Water Ejector Vacuum Pump
208–240V 480V
Doors
Single Double
Single Door Mounting
Cabinet Enclosed/Freestanding Recessed
Double Door Mounting
Recessed through one wall Recessed through two walls10
Accessories
Loading Rack and Two Shelves standard on 16 x 16 26” sterilizers (optional on 20 x 20 x 38” sterilizer)
Single Door (FV021011) Double Door (FV022011)
One Spare Shelf (20 x 20 x 38” sterilizer only) (FV020012)- Intermediate Shelf Loading Car (20 x 20” uits only)
(FV020001)
Transfer Carriage (20 x 20” units only) (FV020002) Chamber Track Assembly (20 x 20” units only)
Single Door (FV021003) Double Door (FV022003)
Loading Car, Transfer Carriage, and Track Assembly (20 x 20” units only) ◻ Single Door (FV021004) ◻ Double Door (FV022004) Seismic Tie-Down Kit11
Item:
Locations:
6. External Supplied Steam (Facility Steam/Stand-Alone Steam Generator)
7. 16 x 16 x 26” double door sterilizers are not available with integral steam generator
8. 120V control is used for 208V, 240V, and 480V powered integral steam generators
9. 220V control is always used with the 380/415V powered integral steam generator
10. Available for 20 x 20 x 38” double door sterilizers only
11. Based on CA requirements dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Sticky Note Sub Sterile dacciarri Sticky Note Sterile Processing dacciarri Highlight
Comment on Text
AMSCO® 400 Small SteamSterilizer — SD937 Rev. F 12
For Further Information, contact:
STERIS Corporation 5960 Heisley Rd.
Mentor, OH 44060–1834 • USA 440–354–2600 • 800–548–4873 www.steris.com
The base language of this document is ENGLISH. Any translations must be made from the base language document.
CUSTOMER IS RESPONSIBLE FOR
COMPLIANCE WITH APPLICABLE LOCAL
AND NATIONAL CODES AND
REGULATIONS.
©2023, STERIS Corporation.
All rights reserved.
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
www.steris.com
4 TO 8
[102-203]
[1886]
[714]
[357]
[508]
[813]
[1321]
[686]
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
SIDE VIEW
PLAN VIEW
SERVICE CLEARANCE
IN. (mm)
20 X 20 X 38
(508 X 508 X 965)
(1159 X 762 X 1891)
45 5/8 X 30 X 74 1/2
CHAMBER SIZE
MAX. OUTSIDE DIM. OF STERILIZER
NOTES:
AutoCAD SHX Text ±1/4 [±6]
AutoCAD SHX Text
WALL OPENING
AutoCAD SHX Text
OPERATING
AutoCAD SHX Text
END
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
WALL OPENING
AutoCAD SHX Text ±1/4 [±6]
AutoCAD SHX Text
STERILIZER
AutoCAD SHX Text C
AutoCAD SHX Text L
AutoCAD SHX Text
MIN.SERVICE
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
MIN.SERVICE
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
MIN.SERVICE
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
MIN.WALL CLEARANCE
AutoCAD SHX Text
TYP. BOTH SIDES
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-054
AutoCAD SHX Text
1. WALL THICKNESS: 4" TO 8" (102-203). 2. ALL DIMENSIONS IN INCHES AND (MM). 3. THESE SERVICE CLEARANCES MUST BE MAINTAINED TO ALLOW ACCESS TO STERILIZER FOR SERVICEABILITY. 4. IF LOADING CAR AND CARRIAGE ARE TO BE USED, FRONT CLEARANCE SHOULD EQUAL TWICE THE LENGTH OF THE STERILIZER.
dacciarri Highlight
[1086]
[1891]
[762]
[748]
[92]
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
SIDE VIEW
PLAN VIEW
STERILIZER INSTALLATION
FRONT VIEW
GENERAL NOTES:
AutoCAD SHX Text
OPERATING
AutoCAD SHX Text
END
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
PANEL SWING CL.
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-054
AutoCAD SHX Text
1. ALL DIMENSIONS IN INCHES AND (MM). 2. CURB WEIGHT: 20 x 20 x 38 SD VESSEL = 1240 LBS. (562 KG.) 3. MAXIMUM OPERATING WEIGHT BASED ON CHAMBER FULLY LOADED WITH WATER FLASKS: 20 x 20 x 38 SD VESSEL = 1371 LBS. (622 KG.) 4. FACILITY MUST PROVIDE REGULATED STEAM PRESSURE IN THE DYNAMIC RANGE SPECIFIED. FAILURE TO DO SO WILL RESULT IN IMPROPER EQUIPMENT OPERATION. 5. AIRFLOW GAP FROM CLEAN SIDE TO SERVICE AREA = 66.9 IN 26. HEAT LOSS AT 70°F (21°C): 20 x 20 x 38 - TO ROOM: 8750 BTU/HR (9,232 KILOJOULE) 7. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION. 8. THE TIE-DOWN OF THIS STERILIZER HAS BEEN PRE-APPROVED IN CALIFORNIA (REF. OPA-0531). SEE STERIS DWG. NO. 83280-194 FOR SEISMIC LOADING AND TIE-DOWN SPECIFICATIONS. 9. FOR SEISMIC INSTALLATIONS: A SEISMIC ADD ON KIT (REF. 146660-184) MUST BE INSTALLED ON THE STERILIZER. 10.STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE NECESSARY THROUGH FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT DRAWING AND NOTE PAGES. SPECIFICATIONS AND DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
[255]
[289] A
B
[70]
[462]
D
C
[244]
[313]
A
B
C
[301]
[29]
D
[706]
[560]
[239]
[508]
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
AC
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
SIDE VIEW
PLAN VIEW
FRONT VIEW
VIEW A-A
A STERILIZER COLD WATER
PLUMBING CONNECTIONS
B STERILIZER DRAIN
NOTE: SEE SHEET 4 FOR PLUMBING INSTALLATION SPECS
AND PLUMBING REQUIREMENTS.
C STERILIZER HOT WATER
D ELECTRIC STEAM GENERATOR
FLUSH DRAIN PORT
A A
AutoCAD SHX Text
OPERATING
AutoCAD SHX Text
END
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text 129394-054
TO BOTTOM OF FUNNEL
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
CHECK ALL NATIONAL CODES AND STANDARDS
PLUMBING REQUIREMENTS
7. FOR GENERAL INSTALLATION INFORMATION SEE STERIS DRAWING
SERVICE ENGINEERING AT 1-800-333-8848 TO OBTAIN A COPY.
EQUIPMENT DWGS.) IF DWG. IS NOT ATTACHED, CONTACT STERIS
NO. 62941-091. (THIS DWG. SHOULD ALWAYS ACCOMPANY THE
INSTALLATION SPECIFICATIONS:
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
PLACEMENT OF PIPING SHUTOFFS:
HARMS WAY
PLUMBING REQUIREMENTS
1" NPT
SFPP CYCLES
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-054
AutoCAD SHX Text
1. PIPE SIZES LISTED UNDER PLUMBING REQUIREMENTS INDICATE THE INDICATE THE EQUIPMENT TERMINATION SIZES ONLY. SIZE PIPING TO EQUIPMENT DEPENDING ON LENGTH OF PIPE RUN FROM PRESSURE REGULATING STATION FOR STEAM LINE, AND MAIN WATER HEADERS. TO SUPPLY THE SPECIFIED SERVICE PRESSURE AND FLOW RATE AT EQUIPMENT TERMINALS, INCLUDE EFFECT OF COINCIDENT DRAW OF MULTIPLE UNIT INSTALLATIONS. 2. PROVIDE PIPING, SHUT-OFF VALVE, PIPE PLUGGED TEE, AND UNION IN STEAM AND WATER SUPPLY CONNECTIONS BETWEEN EQUIPMENT AND STUB OUTS. PLUGGED TEE CAN LATER BE USED FOR TEST PRESSURE GAUGE CONNECTION. ARRANGE CONNECTION PIPING TO ALLOW ACCESS TO MACHINE COMPONENTS AND ELECTRICAL CONTROL PANEL. 3. RECOMMEND PROVISION OF BLOW DOWN VALVE AT EACH STEAM AND WATER STRAINER TO ENABLE STRAINER CLEAN OUT. 4. FOR RECOMMENDED FEED WATER QUALITY FOR STERILIZERS AND CARBON STEEL STEAM GENERATORS, SEE STERIS DWG. NO. 62941-091 5. BLOW DOWN BUILDING STEAM AND WATER SUPPLY LINES BEFORE FINAL CONNECTION TO EQUIPMENT. 6. THE STERILIZER IS NOT SUPPLIED WITH A VACUUM BREAKER OR BACKFLOW PREVENTER AND WHERE REQUIRED BY LOCAL CODES, INSTALLATION OF SUCH A DEVICE IN WATER LINE IS BY OTHERS.
AutoCAD SHX Text THE INSTALLATION OF THE CHIMERON STERILIZER MUST MEET ALL APPLICABLE REGULATIONS.
AutoCAD SHX Text INSTALLATION SPECIFICATION IS LISTED AS ENGINEERING AND INSTALLATION GUIDES. REFERENCED COMPONENTS AND SERVICE CONNECTIONS ARE NOT FURNISHED AS PART OF EQUIPMENT UNLESS UNDER WRITTEN AGREEMENT WITH STERIS.
AutoCAD SHX Text
8. PLACEMENT OF PIPING SHUTOFFS: WHEN INSTALLING;SHUTOFFS MUST WHEN INSTALLING;SHUTOFFS MUST BE LOCATED IN A SUITABLE LOCATION WITHIN LINE OF SIGHT AND CLEAR OF ANY OBSTRUCTIONS THAT WOULD PUT THE SERVICE PERSON IN HARMS WAY IN ORDER TO SHUT IT OFF. IN ORDER TO SHUT IT OFF.
AutoCAD SHX Text
SUPPLY TEMPERATURE REQUIREMENTS ARE 50-70%%DF (10-21%%DC).
AutoCAD SHX Text
VACUUM EFFICIENCY IS REDUCED AT WATER TEMPERATURES
AutoCAD SHX Text
(FLOOR DRAIN CAPACITY MUST HANDLE PEAK WATER
AutoCAD SHX Text AVERAGE-175 gal/cycle (662 l/cycle)
AutoCAD SHX Text AVERAGE-12 gal/hr (.76 lpm)
AutoCAD SHX Text
B STERILIZER DRAIN:
AutoCAD SHX Text 30-50 PSIG DYNAMIC (2.1 TO 3.5 bar)
AutoCAD SHX Text PEAK-15 gpm (57 lpm)
AutoCAD SHX Text
CONSUMPTION IN CYCLE:
AutoCAD SHX Text
ABOVE 70%%DF (21%%DC).
AutoCAD SHX Text
CONSUMPTION OUT OF CYCLE:
AutoCAD SHX Text
1 1/2" ODT
AutoCAD SHX Text
CONSUMPTION).
AutoCAD SHX Text
A STERILIZER COLD WATER: (SEE NOTE #4)
AutoCAD SHX Text
CONSUMPTION:PEAK 1 GPM, PER CYCLE 5 GAL, IDLE 1 GPH.
AutoCAD SHX Text
WATER RESISTIVITY NOT TO EXCEED 26000 OHMS/CM.
AutoCAD SHX Text
SUPPLY TEMPERATURE REQUIREMENTS ARE =< 140%%DF.
AutoCAD SHX Text
D DRAIN FROM ELECTRIC STEAM GENERATOR: 1/2" ODT.
AutoCAD SHX Text
20-50 DYNAMIC PSIG (137.9-344.7 KPA)
AutoCAD SHX Text
1/2" NPT:
AutoCAD SHX Text
C STERILIZER HOT WATER: (SEE NOTE #4)
REQUIRE A MINIMUM OF 40 psig COLD WATER PRESSURE.
[1838]
E
E
[140]
[543]
F
F
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
SIDE VIEW
PLAN VIEW
FRONT VIEW
ELECTRICAL CONNECTIONS
TO CONTROL
BOX HOLE
E
E STERILIZER POWER
NOTE: SEE SHEET 6 FOR ELECTRICAL INSTALLATION SPECS
AND ELECTRICAL REQUIREMENTS.
F ELECTRIC STEAM GENERATOR POWER
AutoCAD SHX Text
OPERATING
AutoCAD SHX Text
END
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-054
AutoCAD SHX Text
* = TO CONTROL
AutoCAD SHX Text
AutoCAD SHX Text
CHECK ALL NATIONAL CODES AND STANDARDS
20 x 20 x 38 AMSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
RECESSED ONE WALL
ELECTRIC STEAM HEAT
ELECTRICAL REQUIREMENTS
F ELECTRIC STEAM GENERATOR POWER:
BREAKER RECOMMENDED. MINIMUM RECOMMENDED LINE CONDUCTOR
208VAC, 50/60HZ, 83A, (3) PHASE. MINIMUM 90A CIRCUIT
SIZE AWG #3 COPPER (26.7 MM ) 75°C (167°F).
BREAKER RECOMMENDED. MINIMUM RECOMMENDED LINE CONDUCTOR
SIZE AWG #4 COPPER (21.5 MM ) 75°C (167°F).
240VAC, 50/60HZ, 72A, (3) PHASE. MINIMUM 80A CIRCUIT
480VAC, 50/60HZ, 36A, (3) PHASE. MINIMUM 45A CIRCUIT
BREAKER RECOMMENDED. MINIMUM RECOMMENDED LINE CONDUCTOR
SIZE AWG #8 COPPER (8.6 MM ) 75°C (167°F).
E STERILIZER POWER:
REQUIRES A FOUR (4) WIRE "DELTA" CONNECTION
(L1,L2,L3,GND.).
REQUIRES A FOUR (4) WIRE "DELTA" CONNECTION
(L1,L2,L3,GND.).
REQUIRES A FOUR (4) WIRE "DELTA" CONNECTION
(L1,L2,L3,GND.).
6. FOR GENERAL INSTALLATION INFORMATION SEE STERIS DRAWING
SERVICE ENGINEERING AT 1-800-333-8848 TO OBTAIN A COPY.
EQUIPMENT DWGS.) IF DWG. IS NOT ATTACHED, CONTACT STERIS
NO. 62941-091. (THIS DWG. SHOULD ALWAYS ACCOMPANY THE
INSTALLATION SPECIFICATIONS:
PLACEMENT OF ELECT. DISCONNECTS:
HARMS WAY
CAUTION:
ATTENTION: NEC
STERIS.
ALSO REFER TO GENERAL NOTES
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
DWG. NO. 62941-091
(MILLIMETERS)
ITEM _____________________
EQUIPMENT DRAWING NO.
LOCATION(S) ______________
SHT. OF
A U
T O
C A
D
D
W G
REV. ______
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
30 KW HEATERS
AutoCAD SHX Text
THE EQUIPMENT WITHIN EASY REACH OF THE OPERATOR AND MARKED
AutoCAD SHX Text
AS THE DISCONNECTING DEVICE FOR THE EQUIPMENT.
AutoCAD SHX Text
MINIMUM 15A CIRCUIT BREAKER IS RECOMMENDED. INSTALLED NEAR
AutoCAD SHX Text CONTROL BOX FOR: 120V, 50/60 HZ, 9.5 AMP SINGLE PHASE SERVICE. MINIMUM RECOMMENDED LINE AND GROUND CONDUCTOR SIZE AWG #12 COPPER (2.05MM). 120VAC REQUIRE A THREE (3) WIRE CONNECTION (L1, NEUT, GND.).
AutoCAD SHX Text
1. PROVIDE GROUPED OR GANGED CIRCUIT PROTECTION AND DISCONNECT FOR STERILIZER POWER AS REQUIRED BY CODES AND STANDARDS. INDIVIDUAL POWER SHUTOFFS REQUIRED NEAR EACH MACHINE FOR SERVICING. 2. PROVIDE GROUNDED METAL CONDUIT AND WIRING BETWEEN EQUIPMENT TERMINALS AND STUB OUTS OR DISCONNECTS. CHECK LOCAL CODES FOR MINIMUM AWG.WIRE SIZE, #16 AWG. MINIMUM RECOMMENDED.
AutoCAD SHX Text THE INSTALLATION OF THE CHIMERON STERILIZER MUST MEET ALL APPLICABLE REGULATIONS.
AutoCAD SHX Text INSTALLATION SPECIFICATION IS LISTED AS ENGINEERING AND INSTALLATION GUIDES. REFERENCED COMPONENTS AND SERVICE CONNECTIONS ARE NOT FURNISHED AS PART OF EQUIPMENT UNLESS UNDER WRITTEN AGREEMENT WITH STERIS.
AutoCAD SHX Text
3. PLACEMENT OF ELECT. DISCONNECTS: WHEN INSTALLING;DISCONNECTS WHEN INSTALLING;DISCONNECTS MUST BE LOCATED IN A SUITABLE LOCATION WITHIN LINE OF SIGHT AND CLEAR OF ANY OBSTRUCTIONS THAT WOULD PUT THE SERVICE PERSON IN HARMS WAY IN ORDER TO SHUT IT OFF. ALSO, THE IN ORDER TO SHUT IT OFF. ALSO, THE LOCATION OF THE DISCONNECTS SHOULD ALLOW THE SERVICE PERSON TO SHUTOFF POWER FROM THE SIDE TO PREVENT POSSIBLE ARC FLASH.
AutoCAD SHX Text
4. CAUTION: DO NOT USE GROUND FAULT CURRENT INTERRUPTERS (GFCI). DO NOT USE GROUND FAULT CURRENT INTERRUPTERS (GFCI).
AutoCAD SHX Text
5. ATTENTION: THE ELECTRICAL CLEARANCES REQUIRED BY THE NEC ARE THE ELECTRICAL CLEARANCES REQUIRED BY THE NEC ARE ARE THE RESPONSIBILITY OF THE INSTALLER. ALSO, ADHERENCE TO LOCAL CODES AND PROCUREMENT OF PERMITS ARE THE RESPONSIBILITY OF THE CUSTOMER UNLESS AGREED TO IN WRITING WITH STERIS.
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-054 dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight
16 x 16 x 26 ANSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
CABINET
ELECTRIC STEAM HEAT
SIDE VIEW
PLAN VIEW
SERVICE CLEARANCE
NOTES:
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
MIN. SERVICE
AutoCAD SHX Text
MIN. SERVICE
AutoCAD SHX Text
STERILIZER
AutoCAD SHX Text L
AutoCAD SHX Text C
AutoCAD SHX Text
FINISHED WALL
AutoCAD SHX Text
FINISHED WALL
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
SERVICE
AutoCAD SHX Text
CLEARANCE
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
ALSO REFER TO GENERAL NOTES
AutoCAD SHX Text
APPLICABLE TO EQUIPMENT DRAWINGS
AutoCAD SHX Text
ALL DIMENSIONS ARE IN INCHES
AutoCAD SHX Text
DWG. NO. 62941-091
AutoCAD SHX Text
(MILLIMETERS)
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
ITEM _____________________
AutoCAD SHX Text
EQUIPMENT DRAWING NO.
AutoCAD SHX Text
LOCATION(S) ______________
AutoCAD SHX Text
AutoCAD SHX Text
SHT. OF
AutoCAD SHX Text
AUTOCAD. DWG.
AutoCAD SHX Text
REV. ______
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-047
AutoCAD SHX Text
IN.
AutoCAD SHX Text (mm)
AutoCAD SHX Text 16 X 16 X 26
AutoCAD SHX Text (406 X 406 X 660)
AutoCAD SHX Text (937 X 660 X 1891)
AutoCAD SHX Text 36 7/8 X 26 X 74 1/2
AutoCAD SHX Text
CHAMBER SIZE
AutoCAD SHX Text
MAX. OUTSIDE DIM. OF STERILIZER
AutoCAD SHX Text
1. ALL DIMENSIONS IN INCHES AND (MM).
AutoCAD SHX Text
2. THESE SERVICE CLEARANCES MUST BE MAINTAINED TO ALLOW ACCESS TO STERILIZER FOR SERVICEABILITY.
dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight
SIDE VIEW
PLAN VIEW
STERILIZER INSTALLATION
16 x 16 x 26 SD VESSEL = 870 LBS. (394 KG.)
GENERAL NOTES:
16 x 16 x 26 ANSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
CABINET
ELECTRIC STEAM HEAT
FRONT VIEW
16 x 16 x 26 SD VESSEL = 890 LBS. (404 KG.)
16 x 16 x 26 - TO ROOM: 6050 BTU/HR (6,383 KILOJOULE)
AutoCAD SHX Text
PANEL SWING CL.
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
FINISHED WALL
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
ALSO REFER TO GENERAL NOTES
AutoCAD SHX Text
APPLICABLE TO EQUIPMENT DRAWINGS
AutoCAD SHX Text
ALL DIMENSIONS ARE IN INCHES
AutoCAD SHX Text
DWG. NO. 62941-091
AutoCAD SHX Text
(MILLIMETERS)
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
ITEM _____________________
AutoCAD SHX Text
EQUIPMENT DRAWING NO.
AutoCAD SHX Text
LOCATION(S) ______________
AutoCAD SHX Text
AutoCAD SHX Text
SHT. OF
AutoCAD SHX Text
AUTOCAD. DWG.
AutoCAD SHX Text
REV. ______
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-047
AutoCAD SHX Text
3. MAXIMUM OPERATING WEIGHT BASED ON CHAMBER FULLY LOADED WITH WATER FLASKS:
AutoCAD SHX Text
2. CURB WEIGHT:
AutoCAD SHX Text
4. FACILITY MUST PROVIDE REGULATED STEAM PRESSURE IN THE
AutoCAD SHX Text
1. ALL DIMENSIONS IN INCHES AND (MM).
AutoCAD SHX Text
6. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
AutoCAD SHX Text
9. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE NECESSARY THROUGH FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT DRAWING AND NOTE PAGES. SPECIFICATIONS AND DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
AutoCAD SHX Text
DYNAMIC RANGE SPECIFIED. FAILURE TO DO SO WILL RESULT
AutoCAD SHX Text
IN IMPROPER EQUIPMENT OPERATION.
AutoCAD SHX Text
5. HEAT LOSS AT 70°F (21°C):
AutoCAD SHX Text
8. FOR SEISMIC INSTALLATIONS: A SEISMIC ADD ON KIT
AutoCAD SHX Text
CALIFORNIA (REF. OPA-0531). SEE STERIS DWG. NO. 83280-193
AutoCAD SHX Text
FOR SEISMIC LOADING AND TIE-DOWN SPECIFICATIONS.
AutoCAD SHX Text
7. THE TIE-DOWN OF THIS STERILIZER HAS BEEN PRE-APPROVED IN
(REF. 146660-182) MUST BE INSTALLED ON THE STERILIZER.
A A
PLAN VIEW
PLUMBING CONNECTIONS
FRONT VIEW
NOTE: SEE SHEET 4 FOR PLUMBING INSTALLATION SPECS
AND PLUMBING REQUIREMENTS.
16 x 16 x 26 ANSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
CABINET
ELECTRIC STEAM HEAT
VIEW A-A
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
TO BOTTOM OF FUNNEL
AutoCAD SHX Text C
AutoCAD SHX Text D
AutoCAD SHX Text A
AutoCAD SHX Text B
AutoCAD SHX Text A
AutoCAD SHX Text D
AutoCAD SHX Text C
AutoCAD SHX Text B
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
ALSO REFER TO GENERAL NOTES
AutoCAD SHX Text
APPLICABLE TO EQUIPMENT DRAWINGS
AutoCAD SHX Text
ALL DIMENSIONS ARE IN INCHES
AutoCAD SHX Text
DWG. NO. 62941-091
AutoCAD SHX Text
(MILLIMETERS)
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
ITEM _____________________
AutoCAD SHX Text
EQUIPMENT DRAWING NO.
AutoCAD SHX Text
LOCATION(S) ______________
AutoCAD SHX Text
AutoCAD SHX Text
SHT. OF
AutoCAD SHX Text
AUTOCAD. DWG.
AutoCAD SHX Text
REV. ______
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-047
AutoCAD SHX Text
A STERILIZER COLD WATER
AutoCAD SHX Text
B STERILIZER DRAIN
AutoCAD SHX Text
C STERILIZER HOT WATER
D ELECTRIC STEAM GENERATOR DRAIN
PLUMBING REQUIREMENTS
INSTALLATION SPECIFICATIONS:
16 x 16 x 26 ANSCO 400 SERIES
PREVACUUM STERILIZER
SINGLE SLIDING DOOR
CABINET
ELECTRIC STEAM HEATT
PLACEMENT OF PIPING SHUTOFFS:
HARMS WAY
PLUMBING REQUIREMENTS
SFPP CYCLES
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
ALSO REFER TO GENERAL NOTES
AutoCAD SHX Text
APPLICABLE TO EQUIPMENT DRAWINGS
AutoCAD SHX Text
ALL DIMENSIONS ARE IN INCHES
AutoCAD SHX Text
DWG. NO. 62941-091
AutoCAD SHX Text
(MILLIMETERS)
AutoCAD SHX Text STERIS Corporation
AutoCAD SHX Text Mentor, OH
AutoCAD SHX Text
ITEM _____________________
AutoCAD SHX Text
EQUIPMENT DRAWING NO.
AutoCAD SHX Text
LOCATION(S) ______________
AutoCAD SHX Text
AutoCAD SHX Text
SHT. OF
AutoCAD SHX Text
AUTOCAD. DWG.
AutoCAD SHX Text
REV. ______
AutoCAD SHX Text
AC
AutoCAD SHX Text 129394-047
AutoCAD SHX Text
1. PIPE SIZES LISTED UNDER PLUMBING REQUIREMENTS INDICATE THE INDICATE THE EQUIPMENT TERMINATION SIZES ONLY. SIZE PIPING TO EQUIPMENT DEPENDING ON LENGTH OF PIPE RUN FROM PRESSURE REGULATING STATION FOR STEAM LINE, AND MAIN WATER HEADERS. TO SUPPLY THE SPECIFIED SERVICE PRESSURE AND FLOW RATE AT EQUIPMENT TERMINALS, INCLUDE EFFECT OF COINCIDENT DRAW OF MULTIPLE UNIT INSTALLATIONS. 2. PROVIDE PIPING, SHUT-OFF VALVE, PIPE PLUGGED TEE, AND UNION IN STEAM AND WATER SUPPLY CONNECTIONS BETWEEN EQUIPMENT AND STUB OUTS. PLUGGED TEE CAN LATER BE USED FOR TEST PRESSURE GAUGE CONNECTION. ARRANGE CONNECTION PIPING TO ALLOW ACCESS TO MACHINE COMPONENTS AND ELECTRICAL CONTROL PANEL. 3. RECOMMEND PROVISION OF BLOW DOWN VALVE AT EACH STEAM AND WATER STRAINER TO ENABLE STRAINER CLEAN OUT. 4. FOR RECOMMENDED FEED WATER QUALITY FOR STERILIZERS AND CARBON STEEL STEAM GENERATORS, SEE STERIS DWG. NO. 62941-091 5. BLOW DOWN BUILDING STEAM AND WATER SUPPLY LINES BEFORE FINAL CONNECTION TO EQUIPMENT. 6. THE STERILIZER IS NOT SUPPLIED WITH A VACUUM BREAKER OR BACKFLOW PREVENTER AND WHERE REQUIRED BY LOCAL CODES, INSTALLATION OF SUCH A DEVICE IN WATER LINE IS BY OTHERS.
AutoCAD SHX Text
7. FOR GENERAL INSTALLATION INFORMATION SEE STERIS DRAWING
AutoCAD SHX Text
SERVICE ENGINEERING AT 1-800-333-8848 TO OBTAIN A COPY.
AutoCAD SHX Text
EQUIPMENT DWGS.) IF DWG. IS NOT ATTACHED, CONTACT STERIS
AutoCAD SHX Text
NO. 62941-091. (THIS DWG. SHOULD ALWAYS ACCOMPANY THE
AutoCAD SHX Text THE INSTALLATION OF THE CHIMERON STERILIZER MUST MEET ALL APPLICABLE REGULATIONS.
AutoCAD SHX Text INSTALLATION SPECIFICATION IS LISTED AS ENGINEERING AND INSTALLATION GUIDES. REFERENCED COMPONENTS AND SERVICE CONNECTIONS ARE NOT FURNISHED AS PART OF EQUIPMENT UNLESS UNDER WRITTEN AGREEMENT WITH STERIS.
AutoCAD…
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it. Updated .