This solicitation seeks proposals for two microwave measurement receivers with frequency ranges up to 26.5 GHz and 50 GHz respectively. Offerors must meet technical performance requirements for analog modulation analysis including amplitude, frequency, and phase modulation measurements. The receivers will be used by the Air Force Materiel Command Lifecycle Management Center and its Precision Measurement Equipment Laboratories to calibrate signal generators, attenuators, and other test equipment. Proposals are due by the date specified in the solicitation documents. The incumbent equipment being replaced is a Keysight microwave measuring receiver purchased in 2005.
View the file
Other files
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
21E‐102A‐RF & 21E‐102B‐RF, Rev3, August 31st 2020
PURCHASE DESCRIPTION
FOR
Microwave Measurement Receiver
20E-102A-RF & 20E-102B-RF
1.1. SCOPE: AFMETCAL is seeking to purchase Microwave Measurement Receivers (MMR) for its Precision Measurement Equipment Laboratories (PMEL’s) and Air Force Primary Standards Laboratory (AFPSL). The MMR shall be used in calibration of Signal Generators and Fixed & Variable Attenuators. Nearly all the parameters of a Signal Generator shall be calibrated with the Microwave Measurement Receiver; RF Output Accuracy, Flatness, Spectral Purity (Harmonics, Non-Harmonics, Sub-Harmonics), Analog Modulation (AM, FM & PM).
AFMETCAL is also seeking to purchase two Alignment Systems and/or equipment or accessories required for periodic recertification and adjustment of the MMR.
1.2. HISTORICAL BACKGROUND: This procurement seeks to replace and improve upon the Keysight N5531S Microwave Measuring Receiver which was procured in 2005. This was the first Microwave Measuring Receiver to reside on a Spectrum Analyzer platform, unlike its previous predecessor the Agilent 8902A (procured in the 1980’s).
2. PERFORMANCE REQUIREMENTS:
2.1. 50 GHz Microwave Measurement Receiver (20E-102B-RF)
2.1.1. Frequency Range (Carrier): 100 kHz to 50 GHz
2.1.2. Connector Type: Precision 2.4 mm male or precision 1.85mm male, as preferred by the vendor.
2.1.3. Analog Modulation Analysis
2.1.3.1. Amplitude Modulation
2.1.3.1.1. AM Carrier Frequency Range: 100 kHz to 50 GHz
2.1.3.1.2. AM Depth: 0% to 100%
2.1.3.1.3. AM Rate Range: up to 1 MHz
2.1.3.1.4. AM Depth Uncertainty: ≤ ±(0.25% + 0.0075 X measured values for AF
≤ 100 kHz and CF ≤ 50 GHz)
2.1.3.1.5. Residual AM: 30 Hz-23 kHz
2.1.3.1.5.1. ≤ 0.01% rms for RF carrier frequencies ≤ 8 GHz
2.1.3.1.5.2. ≤ 0.05% rms for RF carrier frequencies > 8 GHz
2.1.3.1.6. FM Rejection:
2.1.3.1.7. AM Distortion Measurement Range & Distortion Accuracy:
2.1.3.1.7.1. AM Distortion Measurement Range: -80 dB to 0 dB
2.1.3.1.7.2. AM Distortion Accuracy: ≤ ±1 dB
2.1.3.2. Frequency Modulation
2.1.3.2.1. FM Carrier Frequency Range: 100 kHz to 50 GHz
2.1.3.2.2. FM Max Deviation:
2.1.3.2.2.1. 100 kHz ≤ RF < 10 MHz, max deviation 50 kHz peak
2.1.3.2.2.2. 10 MHz ≤ RF < 1 GHz max deviation 5 MHz peak
2.1.3.2.2.3. 1 GHz ≤ RF ≤ 50 GHz max deviation 16 MHz peak
2.1.3.2.3. FM Rate Range: 10 Hz to 1 MHz.
2.1.3.2.4. FM Deviation Accuracy: (Utilizing an rms detector)
RF Freq. BW Rate Peak Dev. FM Rejection 10 MHz to 8 GHz
3 kHz 400 Hz to 1 kHz
< 50 kHz < 0.025%
FM Dev Accy RF Freq Rate Pk Dev
2.1.3.2.5. FM Distortion Measurement Range & Distortion Accuracy:
2.1.3.2.5.1. FM Distortion Measurement Range: -80 db to 0 dB
2.1.3.2.5.2. FM Distortion Accuracy: ≤ ±1 dB
2.1.3.3. Phase Modulation
2.1.3.3.1. PM Carrier Frequency Range: 100 kHz to 50 GHz
2.1.3.3.2. PM Max Deviation: < 10k rad, or <16MHz/AF whichever is smaller.
2.1.3.3.3. PM Rate Range: up to 1 MHz
2.1.3.3.4. PM Deviation Accuracy: (Utilizing an RMS detector)
2.1.3.3.5. PM Distortion Measurement Range & Distortion Accuracy:
2.1.3.3.5.1. PM Distortion Measurement Range: -80 dB to 0 dB
2.1.3.3.5.2. PM Distortion Accuracy: ≤ ±1 dB
2.2. Tuned RF
2.2.1. Relative Measurement Uncertainty
2.2.1.1. Residual Noise Threshold (Linearity Uncertainty) to Max Power: ±(0.015 +
0.005 dB/10 dB step) The Residual Noise Threshold is defined as the input power level at which the uncertainty switches from linearity dominated to noise dominated.
2.2.1.2. Minimum Power to Residual Noise Threshold: ±(cumulative error + 0.0012 x (Input Power – Residual Noise Threshold Power)2)
2.2.1.3. Range to Range Uncertainty (approximately 60 dB): ≤ ±0.031 dB
2.2.1.4. Range to Range Uncertainty (approximately 80 dB): ≤ ±0.031 dB
2.2.2. Minimum Power Level
2.2.2.1. At 10 Hz IFBW: (no sensor, direct to RF Input)
Frequency Range Min Pwr (dBm) 100 kHz to 2 MHz ≤ -137 2 to 10 MHz ≤ -140 10 MHz to 3.05 GHz ≤ -148
3.05 to 6.6 GHz ≤ -140
6.6 to 13.2 GHz ≤ -140
13.2 to 19.2 GHz ≤ -140
19.2 to 26.5 GHz ≤ -123
26.5 to 31.15 GHz ≤ -136
31.15 to 41 GHz ≤ -126 41 to 45 GHz ≤ -118 45 to 50 GHz ≤ -106
2.2.3. Selectable IFBW’s: 10 Hz, 75 Hz, 30 kHz, 200 kHz (degradation of Min Pwr is expected).
It is acceptable to present other IFBW’s if they result in lower S/N for the power measurement.
2.2.3.1. Absolute Measurement Accuracy:
Residual Noise Threshold Power to +20 dBm
≤ ±(Power Meter Range Uncertainty +
0.005 dB/10 dB step)
Minimum Power to Residual Noise Threshold Power
≤ ±(cumulative error + 0.0012 x (Input Power – Residual Noise Threshold Power)2)
≤ 0.5% of (AF + dev.) + 5 Hz
100 kHz to 50 GHz ≤ 1 MHz ≤ 16 MHz
PM Dev Accy RF Freq Rate Pk Dev ≤ 0.5% of rdg +
0.002 rad
100 kHz to 50 GHz ≤ 1 MHz ≤ 600 rads and 16 MHz/AF whichever is smaller
Residual Noise Threshold Power (dBm) Residual Noise Threshold Power ≤ Minimum Power +30 dB
Range 2 Uncertainty ≤ ±0.031 dB Range 3 Uncertainty ≤ ±0.031 dB
2.3. 26.5 GHz Microwave Measurement Receiver (20E-102A-RF)
2.3.1. Frequency Range (Carrier): 100 kHz to 26.5 GHz
2.3.2. Analog Modulation Analysis:
2.3.2.1. Amplitude Modulation
2.3.2.1.1. AM Carrier Frequency Range: 100 kHz to 26.5 GHz
2.3.2.1.2. AM Depth: 0% to 100%
2.3.2.1.3. AM Rate Range: up to 1 MHz
2.3.2.1.4. AM Depth Accuracy: ≤ (±(0.25% + 0.0075 X measured values for
AF ≤ 100 kHz and CF ≤ 26.5 GHz)
2.3.2.1.5. Residual AM: 30 Hz-23 kHz
2.3.2.1.5.1. ≤ 0.01% rms for RF carrier frequencies ≤ 8 GHz
2.3.2.1.5.2. ≤ 0.05% rms for RF carrier frequencies > 8 GHz
2.3.2.1.6. FM Rejection:
2.3.2.1.7. AM Distortion Measurement Range & Distortion Accuracy:
2.3.2.1.7.1. AM Distortion Measurement Range: -80 dB to 0 dB
2.3.2.1.7.2. AM Distortion Accuracy: ≤ ±1 dB
2.3.2.2. Frequency Modulation
2.3.2.2.1. FM Carrier Frequency Range: 250 KHz to 26.5 GHz
2.3.2.2.2. FM Max Deviation: 10 MHz to 26.5 GHz,
2.3.2.2.2.1. 100 kHz ≤ RF < 10 MHz, max deviation 50 kHz peak
2.3.2.2.2.2. 10 MHz ≤ RF < 1 GHz max deviation 5 MHz peak
2.3.2.2.2.3. 1 GHz ≤ RF ≤ 26.5 GHz max deviation 16 MHz peak
2.3.2.2.3. FM Rate Range: 10 MHz ≤ RF ≤ 50 GHz, supported modulation rates 10 Hz to 1 MHz.
2.3.2.2.4. FM Deviation Accuracy: (Utilizing an rms detector) FM Dev Accy RF Freq Rate Pk Dev ≤ 0.5% of (AF + dev.) + 5 Hz
100 kHz to 26.5 GHz ≤ 1 MHz ≤ 16 MHz
2.3.2.2.5. FM Distortion Measurement Range & Distortion Accuracy:
2.3.2.2.5.1. FM Distortion Measurement Range: -80 dB to 0 dB
2.3.2.2.5.1.1. FM Distortion Accuracy: ≤ ±1 dB
2.3.2.3. Phase Modulation
2.3.2.3.1. PM Carrier Frequency Range: 100 kHz to 26.5 GHz
2.3.2.3.2. PM Max Deviation: < 10k rad, or <16MHz/AF whichever is smaller.
2.3.2.3.3. PM Rate Range: up to 1 MHz
2.3.2.3.4. PM Deviation Accuracy: (utilizing RMS detector)
PM Dev Accy RF Freq Rate Pk Dev
RF Freq. BW Rate Peak Dev. FM Rejection 10 MHz to 8 GHz
3 kHz 400 Hz to 1 kHz
< 50 kHz < 0.025%
≤ 0.5% of rdg +
0.002 rad
100 kHz to 26.5 GHz ≤ 1 MHz ≤ 600 rads and 16 MHz/AF whichever is smaller
2.3.2.3.5. PM Distortion Measurement Range & Distortion Accuracy:
2.3.2.3.5.1. PM Distortion Measurement Range: -80 dB to 0 dB
2.3.2.3.5.1.1. PM Distortion Accuracy: ≤ ± 1 dB
2.3.3. Tuned RF
2.3.4. Relative Measurement Uncertainty
2.3.4.1. Residual Noise Threshold to Max Power: ±(0.015 + 0.005 dB/10 dB step)
2.3.4.2. Minimum Power to Residual Noise Threshold: ±(cumulative error + 0.0012 x
(Input Power – Residual Noise Threshold Power)2)
2.3.4.3. Range to Range Uncertainty (approximately 60 dB): ≤ ±0.031 dB
2.3.4.4. Range to Range Uncertainty (approximately 80 dB): ≤±0.031 dB
2.3.5. Minimum Power Level
2.3.5.1. At 10 Hz IFBW: (no sensor, direct to RF Input)
Frequency Range Min Pwr (dBm) 100 kHz to 2 MHz ≤ -137 2 to 10 MHz ≤ -140 10 MHz to 3.05 GHz ≤ -148
3.05 to 6.6 GHz ≤ -140
6.6 to 13.2 GHz ≤ -140
13.2 to 19.2 GHz ≤ -140
19.2 to 26.5 GHz ≤ -121.5
2.3.5.2. Selectable IFBW’s: 10 Hz, 75 Hz, 30 kHz, 200 kHz (degradation of Min Pwr expected)
2.4. Modulation Filters
2.4.1.
Description Specification 50 Hz High-Pass Filter <±1% at rates > 50 Hz 300 Hz High-Pass Filter <±1% at rates > 300 Hz 400 Hz High-Pass Filter <±1% at rates > 400 Hz 3 kHz Low-Pass Filter <±1% at rates < 3,030 Hz 15 kHz Low-Pass Filter <±1% at rates < 15,030 Hz 30 kHz Low-Pass Filter <±1% at rates < 30,000 Hz 80 kHz Low-Pass Filter <±1% at rates < 80,000 Hz 300 kHz Low-Pass Filter <±1% at rates < 300,000 Hz
2.5. Absolute RF Power: The MMR shall measure RF Power over the frequency range of 100 kHz to 26.5 GHz (20E-102A-RF) & 100 kHz to 50 GHz (20E-102B-RF), and an RF Power Range of +30 to -20 dBm. The RF Power Measurement function shall be a selectable function from the Spectrum Analyzer (or MMR personality).
2.5.1. Power Sensor Connectors
2.5.1.1. The 50 GHz Microwave Measurement Receiver (20E-102B-RF) shall have the following power sensors: Precision N-Type, Precision 3.5mm (SMA is not acceptable) & Precision 2.4mm or 1.85mm sensors. Multiple sensors in one connector type can be provided to meet the specifications.
2.5.1.2. The 26.5 GHz Microwave Measurement Receiver (20E-102A-RF) shall have the following power sensors: Precision N-Type & Precision 3.5mm (SMA is not acceptable). Multiple sensors in one connector type can be provided to meet the specifications.
2.5.2. Calibration Factor Flatness: Contiguous calibration factor points shall not change more than the OEM uncertainty of the highest contiguous calibration factor. In other words the frequency response will be linear in nature.
2.5.3. Sensors VSWR: (at 50 ohms)
Frequency VSWR 100 kHz to 2 GHz <1.10:1 2 GHz to 18 GHz <1.28:1 18 to 26.5 GHz <1.40:1
26.5 GHz to 33GHz <1.55:1
33 to 40 GHz <1.7:1 40 to 50 GHz <1.75:1
2.6. Spurious Signals: The MMR shall be capable of making spurious (harmonics, non-harmonics, sub-harmonic) measurements. This is a functional requirement. It is left to the submitter to identify the technical specifications needed to meet the functional requirement.
2.6.1. The maximum measurable 2nd harmonic for the 26.5 GHz version (20E-102A-RF) will be
26.5 GHz (13.25 GHz Carrier Frequency).
2.6.2. The maximum measurable 2nd harmonic for the 50 GHz version (20E-102B-RF) will be 50 GHz (25 GHz Carrier Frequency).
2.6.3. Range: < -80 dBc from the carrier value across the frequency range.
2.7. Accessory Kit: Contains a set of metrology grade adapters used to connect the Microwave
Measurement Receiver. It should contain the following adapters:
2.7.1. 26.5 GHz Accessory Kit (20E-102A-RF): A storage container to be included.
2.7.1.1. 3.5mm (f) to APC-7, Qty 2
2.7.1.2. 3.5mm (f) to Type-N (f), Qty 2
2.7.1.3. 3.5mm (m) to 3.5mm (f), Qty 2
2.7.1.4. 26.5 GHz Cable, 3.5mm (m) to 3.5mm (m)
2.7.2. 50 GHz Accessory Kit (20E-102B-RF): A storage container to be included. 1.85mm can be substituted in place of a 2.4mm connector.
2.7.2.1. 2.4mm (m) – APC 7mm, Qty 2
2.7.2.2. 2.4mm (f) – Type-N (f), Qty 2
2.7.2.3. 2.4mm (f) – APC 3.5mm (m), Qty 2
2.7.2.4. 2.4mm (f) – APC 3.5mm (f), Qty 2
2.7.2.5. 2.4mm (f) – 2.4mm (f), Qty 2
2.7.2.6. 50 GHz Cable, 2.4mm (m) to 2.4mm (f)
2.8. 10 MHz Timebase Reference Output/Input
2.8.1. Reference Oscillator Range: 10 MHz
2.8.2. Reference Oscillator Output Level:
2.8.3. Aging/Year: ≤ ± 5 x 10-8
2.9. Y-Axis Video Output: This is a spectrum analyzer video output. (Not an HDMI or VGA output). This is a functional requirement. It is left to the submitter to identify the technical specifications needed to meet the functional requirement.
3. FUNCTIONAL REQUIREMENTS:
3.1. Architecture: The MMR shall reside on a Spectrum Analyzer or Signal Analyzer platform.
The benefits to this are the ability use a spectrum analyzer to make measurements for Spectral Purity parameters of Signal Generators. It may include Power Sensors, Power Meters and other instruments to complete its needed measurements.
3.2. Automated and Manual Control: The MMR shall be able to be operated from a front panel and remotely. This includes both hard keys and/or touch screen display.
3.2.1. Communications Interface: The MMR shall be able to be controlled through a GPIB IEEE-488 interface, as well as any other peripherals as needed (ex: power meter).
3.2.2. Automated Command Set: The contractor shall provide full command set and operating control codes.
3.3. Power: An assembled power cord of at least 6 feet in length for 110 to 120 volt operation shall be provided with each AC powered instrument. Each AC powered instrument shall have an AC power switch and a front panel indication when AC power is applied.
3.4. Warm-Up: The MMR shall warm-up in 4 hours or less. Once complete all warranted measurement functions should be available without further warm-up needed (ex: any pre-amplifiers installed).
3.5. Timebase: The timebase of the MMR shall have the capability to be adjusted at the field level. Adjustment of the timebase must be able to be accomplished in one of the following methods: a manual tweaker, through front panel control or through GPIB controller.
3.6. Rack Mounting: The MMR shall be capable of being rack mountable into a standard 19-inch (483 mm) rack mount frame. Hardware shall be provided to rack mount the system.
4. Commercial Warranty:
4.1. AFMETCAL is seeking a 10 year commercial warranty on the MMR that will warrant the unit to be free from defects in workmanship, manufacturing, and design and will perform to or exceed specifications.
4.2. Operational Usage: The intended usage of this instrument is that it will operate in a laboratory environment. Its movement will be minimal, in that it will rarely leave the laboratory, most likely installed in a console. It will typically be powered on (possibly in standby mode) the majority of the time. The only time it won’t be powered up is during scheduled power outages.
5. Alignment System:
5.1. The Alignment System: a suite (qty: 2) of any necessary Test Measurement & Diagnostic
Equipment (TMDE), hardware, adapters & cables needed to perform alignment and maintenance shall be provided along with the MMR.
5.1.1. Standard commercial warranties for the alignment system components shall be provided.
5.2. Automated Performance Testing and Adjustment Software and License Keys.
5.2.1. If alignment system for the MMR uses any form of software for automated performance testing and/or adjustment, software and updates with all necessary license keys shall be included with the MMR.
5.2.1.1. Software will be perpetual in use, without a need to renew a subscription.
5.2.2. If software is required to nominalize the timebase of the reference oscillator it shall be provided.
6. PHYSICAL/MECHANICAL REQUIREMENTS
6.1. Physical Size: W x H x D, ≤ 18.75 in x 24 in x 30 in
6.2. Weight: ≤ 70 lbs
6.3. Operating Temperature: 73° ± 6° f
6.4. Operating Humidity: 20 to 50 %
7. DISPLAY REQUIREMENTS:
7.1. Front Panel Display: MMR shall have a front panel display for operation.
8. SOFTWARE:
8.1. If software is used to make measurements for parameters in this purchase description, it shall be provided.
8.2. Software Identification: All software used in conjunction with instrument shall be identified by Name, Version, and Date; to be displayed upon startup or located in an “About” menu in the software.
8.3. All software must run under Microsoft Windows 10 (Professional operating system, (64bit, latest version release). All software must run Windows 10 as a non-elevated user.
8.3.1. AFMETCAL approval is needed if during the period of performance the operating system should change from Windows 10.
8.3.2. All software provided under this contract shall be at the same version, revision and revision date at the completion of the contract
9. UNCERTAINTY ANALYSIS (20E-102A-RF & 20E-102B-RF): The total Absolute Uncertainty for all functions shall be the worst case Root-Square-Sum of Reference Standard absolute uncertainty, standard-deviation, stability, temperature/power coefficients (where applicable), all other system Type A and all other system Type B uncertainties are displayed in terms of Coverage Factor of two (k=2). All sources of uncertainty/error shall be included in the analysis. Any contractor caveat conditions or requirements to meet/perform an Absolute Uncertainty Analysis shall be clearly identified and documented by the contractor.
10. Interchangeability (20E-102A-RF & 20E-102B-RF): The first system shall be the same as the last system in terms of firmware, software, functionality, and configuration control. All replaceable assemblies, subassemblies, components, and parts of designated model provided under a single contractor or ordered from any one manufacturer shall be fully interchangeable without interference or degradation of performance.
11. FIRMWARE (20E-102A-RF & 20E-102B-RF): When firmware or software upgrade is anticipated by the manufacturer during the life of this contract, the contractor shall contact AFMETCAL for approval. Contractor shall provide AFMETCAL documentation detailing the changes made to the revision/upgrade. If AFMETCAL concurs with the proposed upgrade all MMRs delivered under this contract shall be updated to this upgrade. All firmware or software provided under this contract shall be at the same version, revision and revision date at the completion of the contract.
12. LICENSING (20E-102A-RF & 20E-102B-RF): Where licensing issues apply to any part of the MMR, such licensing and associated documents shall be obtained and provided by the contractor. Licenses shall not contravene federal law, federal procurement regulations, or the requirements of this acquisition.
13. TESTING (20E-102A-RF & 20E-102B-RF): This item shall be inspected and tested at destination, as specified elsewhere in this contract. This item shall be checked with test equipment with parameters equal to or better than those required to assure that specified performance requirements are met. Test equipment used shall have first echelon certification directly traceable to a National Metrology Institute or other AFMETCAL approved equivalent laboratory when required. Tests shall be performed to assure the item meets the requirements of this purchase description. Such testing at destination does not relieve the contractor of performing all inspections and quality checks at the point of fabrication, as specified elsewhere in this contract, or as necessary, to ensure performance, as specified herein. In addition to all elements of this purchase description, all commercially published components provided with this system must, as a minimum, satisfy their commercially published performance specifications (and must be provided by the contractor).
Attachment:
Data Requirements
Attachment to Purchase Description 20E-102A/B-RF
Microwave Measurement Receiver (26.5 GHz and 50 GHz)
DATA REQUIREMENTS
MANUALS: A set of operational and maintenance manuals shall be provided. As a minimum, the manuals shall include an explanation of features, performance specifications, operational instructions, theory of operation, calibration procedures, and an illustrated parts breakdown. Wherever possible, use of pictures, diagrams, schematics and mathematical formulas shall be included such that any text material may be better understood. Specifically, the calibration procedure shall include a list of required calibration equipment, by part number, connection diagrams and step-by-step instructions to follow for completing this task. In addition, a system- level manual explaining the system operation and software shall also be provided. Manuals shall be on CD/DVD-ROM in Indexed Portable Document Format (iPDF). A validation and verification of the manuals will be required. The manual shall be in accordance with the Air Force Technical Manual Contract Requirements (TMCR) TM86-01N and Contract Data Requirements List (DD Form 1423).
CALIBRATION CERTIFICATE: A calibration certificate, traceable to the National Institute of Standards and Technology (NIST) or other national metrology institutes (NMIs) shall be provided with each instrument. This calibration certificate must comply with Data Item Description (DID) DI- QCIC-80798B and the Contract Data Requirements List (DD Form 1423).
FAILURE MODES, EFFECTS, AND CRITICALITY ANALYSIS: A failure modes, effects and criticality analysis shall be provided. The analysis data shall show and document the relationship of the FMECA to the systems architecture and engineering artifacts and analysis. The analysis shall identify independent single item failures and the resulting potential impact on mission success, performance, safety, and maintainability. The report shall be in accordance with Data Item Description (DID) DI- SESS-81495B and Contract Data Requirements List (DD Form 1423).
LOGISTICS PRODUCT DATA SUMMARIES: A summary of sufficient information to assist in logistics planning and analysis, program decision making, design status assessment and verification of contractor performance shall be provided. The report shall be in accordance with Data Item Description (DID) DI-SESS-81759A and Contract Data Requirements List (DD Form 1423).