This is an exhibit of technical specifications for a Combustion Analyzer System procurement by the Oklahoma Department of Agriculture, Food & Forestry (ODAFF). The system must utilize the Dumas Combustion method with thermal conductivity detection to analyze total nitrogen in solid and liquid fertilizers and total protein in solid and liquid livestock feed, pet food, and meat samples in accordance with AOAC methods 990.03, 993.13, and 992.15. The specifications document establishes mandatory requirements across six categories: automated sample delivery system, combustion analyzer, instrument controller, system compliance, safety and environmental standards, and delivery/installation/training/support services. Bidders must initial each specification line to confirm their product meets the requirements or provide detailed explanations of any variations or exceptions. The document requires submission of product brochures and published literature detailing system specifications.
Delivery must be FOB destination to Oklahoma City, Oklahoma with shipping costs included in the bid. Installation and validation must be performed by factory-trained service personnel, accompanied by a minimum of one day of onsite training covering operation, method development, and preventive maintenance. The manufacturer must supply all start-up consumables and provide telephone support between 9:00 a.m. and 4:00 p.m. Central Time with callback times not exceeding one business day. A minimum one-year on-site warranty covering parts, labor, travel, and software updates is required. Bidders must provide detailed cost breakdowns for yearly consumable expenses based on an estimated annual usage of 2,000 animal feed samples and 1,000 fertilizer samples, with approximate gas usage information. The instrument must be new and currently in production; refurbished or demonstrator units are not acceptable. Manufacturer certification under ISO 9001 is mandatory, along with declarations of conformity and system validation documentation.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 09._OMESformCP076.pdf | ||
| 07._Exhibit_3-ExecutiveSummary.xlsx | XLSX spreadsheet | |
| 08._OMES-form-CP004_.docx | DOCX document | |
| 02._Attachment-A-agency.purpose-11.02.CZ.pdf | ||
| 01._Agency-Bid-Instructions.pdf | ||
| 04._AttachmentC-AgencyTerms.pdf | ||
| 03.Attachment-B-general-terms.docx | DOCX document | |
| 06._Exhibit_2-Price_Sheet.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
EXHIBIT 1
SPECIFICATIONS
Bid Specifications: Combustion Analyzer System
Combustion Analyzer System must be capable of using the Dumas Combustion method of analysis with thermal conductivity detector to determine total nitrogen in solid and liquid fertilizers and total protein in solid and liquid livestock feed, pet food and meat samples (in accordance with AOAC methods 990.03, 993.13, and 992.15).
For the bid to qualify for consideration the following mandatory specifications must be met OR the supplier must explain any variation or exception to these specifications in detail.
The supplier must provide product brochures and/or published literature detailing the system specifications.
Bidders must initial each line of specifications indicating product bid meets that specification.
1. Automated Sample Delivery System
____1.1. The automated sample delivery system for use with the combustion analyzer must:
____1.1.1. Automate the sample delivery as well as provide for manual sample delivery by direct operator control ____1.1.2. Deliver samples to the combustion analyzer with carry over less than 0.1% from sample to sample ____1.1.3. Not divert or lose any sample to waste during the delivery process ____1.1.4. Have a minimum capacity of 75 samples ____1.1.5. Accommodate a range of sample weights up to at least 500 mg ____1.1.6. Be constructed of finished materials such that parts in direct contact with the sample will not cause contamination
2. Combustion Analyzer
____2.1. The combustion analyzer must:
____2.1.1. Use the Dumas method of combustion ____2.1.2. Use thermal conductivity detection ____2.1.3. Provide a combustion furnace capable of independent temperature control up to 950˚C ____2.1.4. Provide a 100% Oxygen combustion environment with no interaction with other carrier gases ____2.1.5. Perform moisture removal of product gases ____2.1.6. Provide for optimization of oxygen usage for detailed combustion control ____2.1.7. Either collect all evolved combustion gases in a collection apparatus such as a ballast for aliquot analysis, OR analyze the entire volume of combustion gases directly (whole gas analysis)
____2.1.8. Be usable with at least one carrier gas other than Helium.
____2.1.9. Not exceed 6 minutes total analysis time per sample including the automated sample delivery sequence ____2.1.10. Provide and support the communication interfaces utilized by the System Controller in these bid specifications ____2.1.11. The instrument must be new and a model currently in production. Refurbished or demonstrator instruments are not acceptable.
3. Instrument Controller
____3.1. The instrument controller must:
____3.1.1. Include Operating System software and OS Restore Disk on CD(s) or flash drive.
____3.1.2. Include any peripheral devices necessary to operate the instrument ____3.1.3. The instrument controller software must meet these instrument control specifications:
____3.1.3.1. Be compatible with and provide a Windows-type graphical user interface ____3.1.3.2. Provide full functional control of the combustion analyzer and automated sample delivery system in this bid specification
____3.1.3.3. Support connectivity to an external balance, printer and Laboratory Information Management System (LIMS) using RS-232/Serial, USB or LAN industry standard hardware and protocols
____3.1.3.4. Accommodate direct electronic entry of analytical balance readings via a Serial port, USB, or LAN when a balance is connected to the system
____3.1.3.5. Perform signal acquisition, data reduction, data storage, automated reporting, and on-screen data review and modification
____3.1.3.6. Store each user-defined analytical method including combustion analyzer set points, calibration, signal acquisition conditions, and reporting parameters as a single method
____3.1.3.7. Allow priority analysis where the operator can interrupt a batch run to analyze a small number of individual samples before resuming the batch run
____3.1.3.8. Monitor the combustion analyzer during acquisition and record both combustion analyzer performance parameters (i.e. temperature, pressure) and unusual or unexpected events that would affect the integrity or quality of the results
____3.1.3.9. Reprocess results with updated calibration points selected by the operator (i.e. drift correction)
____3.1.3.10. Allow retention, recall and reprocessing of at least 1000 prior analyses
____3.1.3.11. Include graphical zoom
____3.1.3.12. Generate quantitative results in parts per million, % total nitrogen and % total protein
____3.1.3.13. Include standard reporting formats that allow operators to report analytical results with and without graphics
____3.1.3.14. Send reports to the screen, to a file and to a printer.
____3.1.3.15. Allow the system to run unattended ____3.1.3.16. Include a software-based help manual with visual aids ____3.1.3.17. Support compliance to federal electronic records and signature requirements (21 CFR Part 11 regulations) ____3.1.3.18. Export data in industry standard formats suitable for use by word processors, spreadsheets and other common applications. Standard formats include Comma Separated Values (CSV), ASCII text, Windows Metafile Format (WMF), Joint Photographic Expert Group (JPG/JPEG), Data Interchange Format (DIF), and Windows Bitmap
(BMP).
____3.1.3.19. The software must support relational database imports and exports through industry standards protocols such as Open Database Connectivity (ODBC), Extensible Mark-up Language (XML) & Dynamic Data Exchange (DDE)
____3.1.3.20. Software must be able to run without being connected to a network or internet.
4. System
The following specifications apply to all components included in this bid specification.
____4.1. GLP & Regulatory Compliance
____4.1.1. Manufacturer must be certified under ISO 9001 for both the combustion analyzer and system control software
____4.1.2. Manufacturer must provide a Declaration of Conformity describing tests performed and serial numbers for all modular components
____4.1.3. Manufacturer must provide a Declaration of System Validation for the system control software and must be able to supply the software validation processes and procedures to the end user
____4.1.4. The manufacturer must provide instructions on how to verify factory hardware performance criteria
____4.1.5. System must generate and store electronic logs of maintenance events and system errors
____4.2. Safety/Environmental:
____4.2.1. The combustion analyzer, excluding the instrument controller and input peripherals, must not exceed a countertop footprint of 28 inches deep x 32 inches wide with combustion analyzer height of no more than 34 inches
____4.2.2. Gas leak detection and safe leak handling design, including leak sensors, must be incorporated in the combustion analyzer housings
____4.2.3. Operator maintenance areas must be physically isolated from areas of high voltage
____4.2.4. Service accessibility must be from the front or side of the combustion analyzer
____4.2.5. The combustion analyzer must provide an error message to notify the operator in the event of component failure(s)
____4.2.6. The combustion furnace and casing must be constructed of insulated heat-resistant materials appropriate for the intended use and must provide automatic alert and shutdown of the furnace if temperatures exceed safe levels
____4.2.7. The combustion analyzer must operate on US standard current of 230 VAC at 60 Hz with 20A maximum current
____4.2.8. All equipment must be able to operate in an environment with a temperature range of 18-300C and with a relative humidity range of 20-80% (non-condensing)
5. Delivery, Installation, Training, Support and Service:
_____5.1 Shipping costs must be included with FOB destination to the purchasing agency in Oklahoma City, Oklahoma
_____ 5.2 Installation must be included and performed by factory trained service personnel
_____ 5.3 During installation the service personnel must demonstrate operational performance by validating the system to factory specifications
_____ 5.4 In addition to the system setup and performance demonstration, the installation must include initial onsite training on operation, method development and preventative maintenance and must be a minimum of one day. Installation and training must include all travel, labor and any other associated costs.
_____ 5.5 The manufacturer must supply all start-up supplies for initial installation and training (example items include combustion analyzer reagents, sample vials, calibration standard, furnace tube, and porous crucible)
_____ 5.6 The manufacturer must provide telephone support between the hours of 9:00 a.m. and 4:00 p.m. Central Time. Callback times must not exceed one business day.
_____ 5.7 The manufacturer must provide a minimum of a one (1) year on-site warranty including parts, travel, labor and system control software updates
_____ 5.8 The manufacturer must provide factory trained service personnel for onsite repairs
6. Estimated Annual Operating Cost Analysis
_____ 6.1. For this estimation, assume analyzing 2000 animal feed samples (100 liquid samples) for protein and 1000 fertilizer samples (50 liquid) for total nitrogen per year and running the instrument at least three times per week.
_____ 6.1.1 Yearly Instrument Consumable Cost:
Vendor must provide cost breakdown for yearly consumable cost.
Includes sample holders (either crucibles or assume tin foil), reagent tubes, reagents, and other consumables that must be changed/used at least once a year.
_____ 6.1.2 Yearly Gas Usage:
Vendor must provide approximate amount of gas usage during the year, but it does not need to provide a cost estimate.