Surgical Stool

Closed Solicitation Posted

Solicitation number
36C26022Q0355
Agency
Responses due
Set-aside
Service-Disabled Veteran-Owned Small Business

Opportunity facts

NAICS code
339112 Surgical and Medical Instrument Manufacturing
PSC
6515 Medical And Surgical Instruments, Equipment, And Supplies
Place of performance
Portland, Oregon 97207, United States

Notice details come from SAM.gov. Updated .

Notice text

Salient Characteristics

Procedural Stool with Hydraulic Lift and Floor Lock
 

  1. Procedural Stool must have height range of 21” to 28-1/2“
  2. Must have five legs on the stool base for stability
  3. Must have built-in floor lock on the chair that can be engaged easily by the clinician without the clinician having to leave the stool seat.  This floor lock must prevent chair movement during procedures.
  4. Must have at least 2-1/2” Rubber casters for smooth mobility when floor lock is not engaged
  5. Seat must swivel 360 degrees without affecting its height
  6. Must have foot-activated hydraulic for hands-free adjustment of stool height
  7. Must have upholstered backrest that is adjustable both vertically and forward tilt, and locks in place 
  8. Must have upholstered armrests that are adjustable vertically and lock in place
  9. Angle of upholstered armrests must be adjustable and lock in place