Notice text

A Gas Chromatograph (GC) instrument is needed for air monitoring in replacement of MINICAMS instruments

    • Flame Ionization Detector (FID)/Dual Flame Photometric Detector (DFPD)
    • Air sampling pre-concentrator trap capability with flow rate through the trap of at least 500mL/min
    • Capability to add mass flow controller and collect flow data
    • Oven programmable from ambient to 400℃
    • Cycle time from injection to injection of 15 minutes or less
    • Electronic pressure control (EPC) of gases
    • Displayed control zones, valve status, temperature, pressure, and voltage.
    • Built in data acquisition system
    • GC chassis 12” x 14” x 15” (width x height x depth)
    • Instrument weight approximately 40 pounds or less.
    • Alarm capability based on analyte response

Shipping: Shall be shipped within 30 days of contract award.