Personal Radiation Detector

Closed Solicitation Posted

Solicitation number
W91243-22-Q-P020
Agency
Responses due
Set-aside
Total Small Business

Opportunity facts

NAICS code
334519 Other Measuring and Controlling Device Manufacturing
PSC
5860 Stimulated Coherent Radiation Devices, Components, And Accessories
Place of performance
Lincoln, Nebraska 68524, United States

Notice details come from SAM.gov. Updated .

Notice text

Personal radiation detector that distinguishes radiation sources which are industrial or medical in nature and has a highly sensitive alarm for artificial gamma. Specifications include:
Radiation Type: Gamma and X-rays plus neutrons via prompt gamma
Material: Low dose rate / Search / ID detector: CsI(Tl)
                High dose rate detector: PVT
Identification of radiation type
Extremely high dose rate measurement capabilities:  ≤1000 R/h (<10 Sv/h) for search, find, radiation identification and response missions. 
Dose Rate Range: Low dose rate / Search / ID detector: 10 nSv/h - 250 µSv/h (1 µR/h - 25 mR/h)
                             High dose rate detector: 250 µSv/h - 10 Sv/h (25 mR/h - 1000 R/h)
Energy Range Spectroscopic energy range: Low dose rate / Search / ID detector: 20 keV - 3 MeV
                                                                      High dose rate detector: 58 keV to 6 MeV
Highly sensitive alarm for artificial gamma. Supplying extra sensitivity in the range of 90 - 450 keV for effective detection of Special Nuclear Materials
Immediately separate man-made (non-NORM) gamma radiation from natural background fluctuations in real-time during operation.
Bluetooth integration with iOS and Android devices

Files

File Type Posted
LC Request_RadEye_KO signed.pdf PDF
W91243-21-22-P020_CombinedSynopsis.pdf PDF

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors