About this opportunity
The U.S. Geological Survey (USGS) Oklahoma-Texas Water Science Center has issued a Sources Sought Notice to identify potential contractors capable of providing submersible pCO2 sensors and pumps for measuring dissolved carbon dioxide in seawater. The procurement requires two pCO2 sensors and two pumps with antifouling caps, with specifications mandating infrared detection capabilities, factory calibration using WMO traceable standards, automatic zero compensation, 0.5% accuracy, and 0.01 ppm resolution. The sensors must operate at depths up to 600 meters, temperatures from 0°C to 40°C, and integrate with existing USGS equipment including Campbell CR1000x Dataloggers and Seabird SeapHOx systems. Currently, only the Pro-Oceanus CO2-Pro CV is known to be compatible with existing equipment, though the agency seeks "Pro-Oceanus Systems brand name or equal" products. Responses must be submitted by July 21, 2025, and should detail the extent to which specifications can be met, business classification and small business status, and relevant project experience.
This opportunity operates under NAICS code 334513 with a 750-employee size standard and includes no set-aside designation, though small businesses including 8(a), HUBZone, SDVOSB, WOSB, and EDWOSB firms are strongly encouraged to participate. The procurement specifically seeks two pCO2 sensors that must not exceed 38 cm in length and 10 cm in diameter, along with two corresponding pumps weighing no more than 0.5kg in air and 0.1kg in water. Pro-Oceanus Systems appears to be the incumbent or preferred supplier, as their CO2-Pro CV model is referenced as the baseline for compatibility requirements. The sensors must sample every second and provide data via RS-232 connectivity while operating on no more than 18 VDC input voltage. No award values or budget ranges are specified in this market research notice, which explicitly states it does not constitute a commitment to purchase and respondents will not be notified of evaluation results.
Notice text
This is a Sources Sought Notice only. This is not a request for proposal, but a survey to locate potential sources. This Sources Sought does not constitute an Invitation for Bids, Request for Proposals or a Request for Quotations, and is not to be construed as a commitment by the Government to issue an order or otherwise pay for the information solicited, nor is it a guarantee of a forthcoming solicitation or contract. It is for market research purposes only. Respondents will not be notified of the results of the evaluation.
The purpose of this Sources Sought Notice is to gain knowledge of interest, capabilities, and qualifications of various members of industry, to include the Small Business Community: Small Business, Section 8(a), Historically Underutilized Business Zones (HUB-Zone), Service-Disabled Veteran-Owned Small Business (SDVOSB), Women-Owned Small Business (WOSB), and Economically Disadvantaged Women-Owned Small Business (EDWOSB). The Government must ensure there is adequate competition among the potential pool of responsible contractors. Small business, Section 8(a), HUBZone, SDVOSB, WOSB, & EDWOSB businesses are highly encouraged to participate.
NAICS Code: 334513 Instruments and Related Products Manufacturing for Measuring, Displaying, and Controlling Industrial Process Variables
NAICS Size Standard: 750 Employees
Subject: Sources Sought for Submersible pCO2 sensor and pump Pro-Oceanus Systems brand name or an equal to product
Sources Sought Closing Response Date: July 21, 2025
Contact Point: Kimberly Schneider, krschneider@usgs.gov
Objective: To find sources that are qualified to meet the supplies/services as listed below. Note that the specific requirements are subject to change prior to the release of any solicitation.
The U.S. Geological Survey (USGS) Oklahoma-Texas Water Science Center (OTWSC) is seeking potential sources Provide a submersible pCO2 sensor and pump and pump with accessories. Currently the only known source that is compatible with existing equipment is the Pro-Oceanus CO2-Pro CV.
Specifications/Salient Characteristics: We are seeking sources for Pro-Oceanus Systems brand name or an equal to product.
The following salient characteristics are required:
1. Specifications: Item 1: 2x PCO2 sensors
1. Must measure the partial pressure of CO2 gas dissolved in seawater using infrared detection.
2. The sensor shall be factory calibrated using WMO traceable standards.
3. Must have an automatic zero compensation feature.
4. Must have a range of 0-2,000 parts per million.
5. Must have an accuracy of ±0.5%.
6. Must have a resolution of 0.01 ppm.
7. Must not exceed 38 cm in length and 10 cm in diameter.
8. Must have a depth rating of at least 600 meters.
9. Must be able to operate at temperatures ranging from 0ºC to 40ºC.
10. Must be able to connect to water pump.
11. Must sample every second.
12. The sensor shall provide data via RS-232 and have the capability to integrate with a user provided datalogger.
13. Must not require input voltage exceeding 18 VDC.
14. Must include a water-pumped interface head.
15.
2. Specifications: Item 2: 2x Pump with antifouling cap
1. Must be a centrifugal pump.
2. Pump must have a depth rating of 600 meters.
3. Must provide constant flow to sensors at 2000 RPMs.
4. Must include a 1-meter sensor to pump cable.
5. Must not require input voltage exceeding 18 VDC.
6. Must have a plastic housing and weigh no more than 0.5kg in air/0.1kg in water.
7. Must include an antifouling cap that can be attached to pump to prevent debris from being pumped into sensors.
8. Must have a wet-pluggable MCBH Connector.
9.
3. Accuracy and Precision:
1. methods used for carbonate system data collection must provide data at high accuracy and precision to detect small changes over decadal timescales.
2. pCO2 at an accuracy of ±0.5% and a resolution of 0.01 ppm while requiring minimal regular maintenance and no use of liquid chemical reagents, resulting in data of appropriate quality to meet project objectives.
3. a tubular interface and enhanced biofouling resistance to ensure stability for long-term field deployments and is factory calibrated using WMO traceable standard gases to improve accuracy.
4. Data Comparability:
1. ProOceanus CO2-Pro CV data comparability and consistency is essential that the measurement instruments, methods, and accuracy for pCO2 data remain the same.
5. Compatibility:
1. Integrates effectively with the existing equipment being installed at the station and with established USGS telemetry procedures. Its dimensions and deployment specifications are suitable for the enclosure where the sensors will be deployed.
2. Sensor must be compatible with a Campbell CR1000x Datalogger and allow physical integration with a Seabird SeapHOx to allow external pump use.
3. The sensor must be able to deploy horizontally and fit in the enclosure with other sensors, thus diameter should be no greater than 5 inches.
Responses: All interested parties should notify this office via e-mail within the posted date. Responses shall include:
(I) To what extent each of the specifications can be met.
(II) Include your type of business (i.e. commercial, academia), whether your organization is classified as a large or small business. If a small business, you must also list any small disadvantaged status you hold [HUBZone, 8(a), Service Disabled Veteran Owned Small Business (SDVOSB), Women Owned Small Business (WOSB), Economically Disadvantaged Women-Owned Small Business (EDWOSB), etc.].
(III) Experience/ performance through the description of completed projects.
Notice history
| Notice | Type | Posted |
|---|---|---|
| OTWSC Pro-Oceanus Systems pCO2 Sensor and Pump | Award Notice | |
| OTWSC Pro-Oceanus Systems pCO2 Sensor and Pump | Solicitation | |
| OTWSC pCO2 Sensor and Pump | Pre-Solicitation |
On GovTribe
Work this opportunity on GovTribe
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity