J065--Intent to Sole Source Medrad Injector Service

Closed Special Notice Posted

A newer pre-solicitation was posted. See the latest pre-solicitation from .

Solicitation number
36C24126Q0001
Agency
Veterans Integrated Service Network 1 Veterans Health Administration, Department of Veterans Affairs
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
811210 Electronic and Precision Equipment Repair and Maintenance
PSC
J065 Maint/Repair/Rebuild Of Equipment- Medical, Dental, And Veterinary Equipment And Supplies
Points of contact

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 1 has issued a Special Notice announcing its intent to award a sole-source contract for Medrad Injector Service at the Manchester VA Medical Center in New Hampshire. This notice is published in accordance with Federal Acquisition Regulation requirements for disseminating information on proposed contract actions. The VA intends to award a firm-fixed-price contract under the authority of 10 U.S.C 2304 (C)(1) and FAR 6.302-1(a)(2), citing that only one responsible source exists to satisfy the agency's requirements. The procurement is classified under NAICS code 811210 for Electronic and Precision Equipment Repair and Maintenance and product service code J065 for maintenance and repair of medical equipment. The contracting officer for this requirement is Manases Cabrera, who can be contacted at manases.cabrera@va.gov.

The purpose of this contract is to provide comprehensive maintenance, service, and support for three contrast injectors at the Manchester VA Medical Center: one Medrad MRXPERION and two Medrad Stellant Flex units. The contract will cover a three-year performance period from October 1, 2025, to September 30, 2028, with requirements including on-site repairs within 24 hours, telephone technical support, preventive maintenance visits, software updates and patches, and comprehensive documentation. Interested parties have until August 15, 2025, at 2:00 PM Eastern Standard Time to respond to this notice, though this is not a request for competitive proposals but rather a notification of sole-source intent. Any response must provide clear and convincing evidence that competition would be advantageous to the government. The determination whether to compete this contract remains solely within the government's discretion, and all required services must be provided at no additional cost to the government.

Notice text

This notice is being published in accordance with Federal Acquisition Regulation (FAR) Part 5.101(a)(2) requiring the dissemination of information on proposed contract actions. This is a notice of intent to award a sole-source contract pursuant to 10 U.S.C 2304 (C)(1), as implemented by FAR 6.302-1(a)(2) and 6.301-1(c), only one responsible source and no other supplies or services will satisfy agency requirements, and brand-name descriptions.

The Department of Veterans Affairs Medical Center located at 718 Smyth Road, Manchester, NH, 03104-7007 has a requirement Medrad Injector Service.

The Government intends to award a firm-fixed-price sole-source contract.

A determination by the Government not to compete with this proposed contract based upon responses to this notice is solely within the discretion of the Government. This notice of intent is not a request for competitive proposals. This is a written notice to inform the public of the government's intent to award on a sole-source basis.

Interested parties may identify their interest and capability to respond to the requirement. Any response to this notice must show clear and convincing evidence that competition will be advantageous to the Government.

Please contact Manases Cabrera, Contracting Officer, with the responses or questions regarding this requirement at manases.cabrera@va.gov.

This notice will close on August 15, 2025, at 2:00 PM Eastern Standard Time (EST)

STATEMENT OF WORK FOR SERVICE AND SUPPORT OF

Bayer Equipment Service Contract

BACKGROUND

Manchester VA Medical Center has one Medrad MRXPERION contrast injector and two Medrad Stellant Flex contrast injectors. The devices will need to be maintained, by a qualified and reliable company, for ongoing quality assurance, software package modifications including upgrades & patches, telephone technical support, and future onsite repairs, calibrations, and preventive maintenance.

OBJECTIVES

The contract effort goals are to maintain, upgrade, service and support all components of the equipment listed in Section F throughout the total contract period at the Manchester VA Medical Center. Objectives to meet the service and support goal include, but are not limited to, replacement parts coverage, service support, telephone support, software and hardware support, continuing education for VA staff and support service, full maintenance coverage by providing all replacement parts, labor, travel expenses, tools, test equipment and other incidentals necessary to maintain or restore the listed equipment to operating specifications according to the manufacturer s recommendations throughout the term of the contract.

PERIOD OF PERFORMANCE

After the contract has been awarded, the period of performance will be defined within the following time parameters.

3 Years: 10/1/2025-9/30/2028

PLACE OF PERFORMANCE

Delivery of services in this case will be conducted virtually and/or at the following address:

Manchester VA Medical Center

718 Smyth Road, Manchester, NH 03104

SCOPE OF WORK (TASKS FOR SERVICES)

The work requirements for this contract performance period are introduced as stated in the preceding sections. However, it is here within the scope of work that they will be captured explicitly. The contractor will meet the following requirements as bulleted below: THE CONTRACTOR WILL:

REPAIRS

All Field Engineer on site visits shall be performed during the normal VA business hours

8:00am-5:00pm, Monday through Friday, excluding Federal holidays

On-site response time shall be within 24 hours of request

All Field Engineers or company representative must check in with Healthcare Technology Management (HTM) (aka: Biomedical Engineering or Clinical Engineering) prior to starting their work at the facility.

All telephone technical support shall be provided at no additional cost

Repair parts shall be provided and shipped, next day 10:30 am at no extra cost to the Government

SCHEDULE MAINTENANCE (PREVENTIVE MAINTENANCE)

One Planned Maintenance (PM) visit with complete inspection of listed instrumentation. Service (work) includes labor, travel, a PM kit, complete inspection of the instrument with cleaning. Includes one additional PM or cleaning if necessary.

Preventive maintenance shall be scheduled during normal VA business hours as identified above.

Preventive maintenance shall be coordinated with the Service Line and HTM so that there is minimal negative impact to patient care.

All parts and equipment necessary to complete the preventive maintenance shall be provided and shipped, next day 10:30 am at no extra cost to the Government

SOFTWARE

All software updates, inclusive of installation of any minor patches or major revisions, shall be included at no extra cost to the Government

The Contractor shall assist in any/all software and/or network vulnerabilities identified by the Government that need remediation as expressed by the Government.

All software will be approved and/or scanned by HTM (aka: Biomedical Engineering or Clinical Engineering) prior to installation.

SERVICE DOCUMENTATION

At the conclusion of each repair and scheduled maintenance call, the Contractor shall provide a written service report to HTM (aka: Biomedical Engineering or Clinical Engineering) within 3 business days indicating:

Date(s) of service

Make, model, serial number, and barcode/ID number of the equipment serviced

Name of FSE who performed the work

Hours worked

Description of the reported problem, the services performed, and parts replaced

Make, model, serial number, barcode/ID number, and proof of up-to-date calibration of the test equipment used in the repair or preventive maintenance service.

Quantifiable measurement results, inclusive of range and tolerance, for all tests conducted

EQUIPMENT LIST

Item #

Manufacturer

Serial Number

Product

43534

Medrad

301346U1

Medrad Stellant Flex

39928

Medrad

300820U1

Medrad Stellant Flex

43375

Medrad

84284254 112579

Medrad MRXPERION

Attachments

Files attached to this notice, newest first
File Type Posted
36C24126Q0001.docx DOCX document

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
J065--Medrad Injector Service Latest pre-solicitation Pre-Solicitation
J065--Intent to Sole Source Medrad Injector Service This notice · Original Special Notice