Granite Table inspection, calibration, and lapping/repair
Closed Pre-Solicitation Posted
- Solicitation number
- W519TC-26-SS-035003
- Agency
-
ACC Rock Island
Army Contracting Command, Department of Defense
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 541380 Testing Laboratories and Services
- Place of performance
- Rock Island, Illinois 61299, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of the Army Materiel Command Contracting Command Rock Island Arsenal is conducting a sources sought for inspection, calibration, and lapping/repair services for granite surface tables at the Rock Island Arsenal's Joint Manufacturing and Technology Center, with responses due September 1, 2026. This is a Sources Sought notice only; interested vendors must email a capability statement, company name, CAGE Code or EUI number, and point of contact to heather.j.petersen.civ@army.mil.
The contractor must hold ISO 17025 accreditation. Work includes inspection and flatness determination of granite surface plates per ASME B89.3.7 standard, with lapping of working surfaces to specification where needed. The granite inventory spans Laboratory Grade AA and Grade A tables ranging from 36 inches by 72 inches to 96 inches by 144 inches, with approximately 14 individual tabletops specified.
Performance begins immediately upon contract award for a base year plus four option years totaling five years. Invoices must provide a detailed cost breakout and require JMTC Representative review and concurrence against contract prices and JTR-approved rates. Place of performance is Rock Island Arsenal, Illinois.
NAICS 541380 (Testing Laboratories and Services); PSC H266 (Equipment and Materials Testing – Instruments and Laboratory Equipment).
Notice text
This is a Sources Sought only.
You MUST send e-mail to the POC (heather.j.petersen.civ@army.mil) that includes the following:
1) Company Name
2) Cage Code or EUI number
3) company POC, e-mail address, and phone number
4) capability statement that outlines you have and can complete the work outlined below.
___________________
Scope of Work (SOW)
1. Purpose: The purpose of this Scope of Work (SOW) is to define a performance-based specification to calibrate granite surface tables for the Rock Island Arsenal's Joint Manufacturing and Technology Center (JMTC). The work shall be performed by an ISO 17025 accredited vendor shall include the inspection, calibration, and lapping/repair of the tabletops of any that are not within specification.
2. Invoices for services must include a breakout of all costs and must be submitted to the JMTC Representative for review and concurance after services are completed. This is to verify that the invoices are in line with the contract prices and JTR approved prices.
3. Performance Period: Contractor’s effort will begin immediately after Contract/Task Order award. This will be for a Base year and Four Option Years for a total of five years.
4. Detailed Specifications: Rock Island Arsenal's JMTC includes the list of granite tables covered by this SOW and is represented below.
- Laboratory Grade AA
- 36A R-68-1439 48 X 96
- 36A 19567-M 60 X 120
- 36F R-82-791 60 X 96
- 36F 22387-M 48 X 72
- 36F RIA30627 48 X 96
- 36S 19583-M 48 X 96
- Laboratory Grade A
- 40A 19587-M 48 X 108
- 40A 19589-M 48 X 108
- 40B R-86-1412 72 X 120
- 40C 22487-M 48 X 96
- 40C R-85-3474 72 X 120
- 40C RIA31418 72 X 120
- 40C R-86-948 96 X 144
- 40M R-59-304 48 X 96
- 40R R-85-3475 72 X 120
- 40W 23415-M 72 X 120
- 299 R-85-3565 48 X 72
- 299 P-9299 48 X 72
4.1 Per ASME B89.3.7 the ISO 17025 accredited contractor shall inspect the granite surface plates and determine both its overall and local flatness. If necessary, the contractor will lap the working surface back to within specification standards.
4.2 Calibration and any necessary lapping shall be performed on an annual basis.
4.3 After the granite surface plate has been calibrated, the contractor shall affix a calibration sticker to the granite surface plate and provide a Service Acceptance Report (SAR) to the JMTC Representative within three (3) business days of completing the calibration.
5. Working Hours: 0600-1430 hours, Monday-Friday, Government Holidays excluded, unless otherwise authorized, and agreed upon, prior to deviation of the work schedule.
6. Security: The RIA-JMTC is an Army installation subject to Department of Defense safeguards, various precautions, and plant protection measures. At all times during the execution of this SOW, the contractor will maintain adequate plant protection devices to minimize espionage, sabotage, and other malicious destruction and damage. The contractor shall comply with all security requirements of the Rock Island Arsenal.
7. Final acceptance inspection shall be performed by the JMTC Representative to verify that all requirements of this SOW (and any modifications thereto) are satisfied.
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Granite Surface Plate Calibration and lapping - updated Docs | ACC Rock Island | No set-aside | |
| Wafer broach holding bar | ACC Rock Island | No set-aside | |
| Gear shaper repair | ACC Rock Island | Total Small Business |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity