FY26 USAFA DFBI Hyperbaric Chamber
Closed Pre-Solicitation Posted
- Solicitation number
- FY26_USAFA_DFBI_Hyperbaric_Chamber
- Agency
-
Air Force Academy
Headquarters Air Force, Department of Defense
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 339112 Surgical and Medical Instrument Manufacturing
- Place of performance
- Usaf Academy, Colorado 80840, United States
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of the Air Force Academy is conducting market research to identify vendors capable of providing, delivering, and installing a Hyperbaric Chamber System, with a sources-sought response deadline of June 17, 2026. No set-aside is anticipated. The government intends to solicit a firm-fixed-price contract.
The system will support the Dean of Faculty Department of Biology at the Air Force Academy in Colorado Springs, Colorado, with FOB destination delivery to USAF Academy, CO 80840. The requirement is for a commercially available or modified-commercial hyperbaric chamber featuring a dynamic oxygen and carbon dioxide controller. Core specifications include oxygen control range of 0.1–99.9% (±1% accuracy), carbon dioxide control range of 0.1–20% (±1.5% accuracy), gas control resolution of 0.1%, and a digital controller displaying process values with interface for gas infusion control. The system must support dynamic programming and profile creation with multiple setpoints, ranges, and cycle configurations, along with pod-based sensors that monitor chamber conditions and communicate to the gas controller.
This is a sources-sought notice and request for information only; it does not constitute a formal solicitation or commitment to award. The government is soliciting input from qualified vendors, including small businesses designated as 8(a), Service-Disabled Veteran-Owned, HUBZone, or Women-Owned. All respondents must be registered in SAM.gov prior to responding.
NAICS 339112; PSC 6515.
Notice text
SOURCES SOUGHT SYNOPSIS: Hyperbaric Chamber System
- Notice Type: Sources Sought / Request for Information (RFI)
- Title: FY26 USAFA DFBI Hyperbaric Chamber
- NAICS Code: 339112 - Surgical and Medical Instrument Manufacturing
- Product Service Code (PSC): 6515 - Medical and Surgical Instruments, Equipment, and Supplies
1.0 Introduction
This is a Sources Sought Notice only. This notice is issued solely for information and planning purposes - it does not constitute a Request for Quote (RFQ), Request for Proposal (RFP), or a promise to issue a formal solicitation in the future.
The Department of War (DoW) is conducting market research to identify qualified and capable vendors, including small businesses (e.g., 8(a), Service-Disabled Veteran-Owned, HUBZone, Women-Owned), that can provide, deliver, and install a Hyperbaric Chamber System.
The government intends to solicit a firm-fixed-price contract. Delivery is FOB destination to USAF Academy, CO 80840. All contractors doing business with the Federal Government must be registered in the System for Award Management (SAM) database. The website for registration is www.sam.gov.
2.0 Background and Requirement Description
The US Air Force Academy has a requirement for a commercially available or modified-commercial Hyperbaric Chamber System to support the Dean of Faculty Department of Biology.
The anticipated core requirements for the system include, but are not limited to:
SALIENT CHARACTERISTICS:
1. Dynamic O2 and CO2 Controller with the following requirements:
a. O2 Control Range: 0.1% - 99.9% with accuracy of ± 1%
b. CO2 Control Range: 0.1% - 20% with accuracy of ± 1.5%
c. Gas control resolution of 0.1%
d. Digital Controller that displays process values and provides interface to control gas infusion
e. Dynamic programming and profile creation enabling multiple setpoints, ranges and number of cycles
f. Pod-based sensor that monitors/measures conditions within chamber (discussed below) and communicates to the gas controller to maintain desired gas concentrations
g. Connections for O2, N2 and CO2. Gas source delivered through compressed gas cylinders
h. Bi-directional control action with O2 gas infusion for raising O2 and nitrogen gas infusion for lowering O2.
i. Bi-directional control action with CO2 gas infusion for raising CO2 and nitrogen gas infusion for lowering CO2.
j. Graphical trending and data logging capabilities
2. System Chamber with the following requirements:
a. Enclosure with a polycarbonate and stainless steel construction for durability and corrosion resistance
b. Enclosure is compatible with the gas controller listed above
c. Enclosure with 2 shelves, each shelf has a minimum dimension of 12.75" W × 10.25" D
d. Enclosure dimensions are no larger than 14"W x 13"D x 6"H
e. Enclosure protects samples of interest from room air exposure/environmental instability and permits the precise control of oxygen, nitrogen and carbon dioxide gas concentrations
f. Enclosure fits within most third-party incubators
3.0 Requested Information (Capability Statement)
Interested parties are requested to submit a Capability Statement demonstrating their ability to fulfill this requirement. Please include the following information structured in your response:
1. Corporate Profile - Company Name, Address, CAGE Code, UEI Number, and Point of Contact (Name, Phone, Email).
2. Business Size - Business size standard and socioeconomic status under NAICS 339112.
3. Technical Capability - A brief description of the hyperbaric chamber(s) your company manufactures or distributes that meet or exceed the requirements in Section 2.0. Include spec sheets if available.
4. Past Performance - Examples of prior government or commercial contracts for similar hyperbaric systems (include contract numbers, scope, and value).
5. Supply Chain / Lead Time - Estimated manufacturing and delivery lead time from date of contract award.
6. GSA / GWAC Vehicles - Identify if your product is available on a GSA Multiple Award Schedule (MAS) or other existing government-wide acquisition contracts.
4.0 Response Instructions
Responses must be submitted via email to the Contracting Office Point of Contact (POC) listed below.
- Due Date/Time: 17 June 2026 by 10:00 am Mountain Time
- Page Limit: Submissions should not exceed 5 pages, excluding standard product specification sheets.
- Primary POC: Danielle Hyde, Contracting Officer, danielle.hyde.2@us.af.mil
- Secondary POC: Missy Morris, Contract Specialist, marissa.morris.4@us.af.mil
Disclaimer: The Government will not reimburse any costs associated with the preparation or submission of responses to this Sources Sought notice.