FDA Berthold Centro or equal microplate luminometer

Closed Solicitation Posted

Solicitation number
75F40126Q134302
Agency
Responses due
Set-aside
Total Small Business

Opportunity facts

Notice details come from SAM.gov. Updated .

About this opportunity

The FDA's Center for Biologics Evaluation and Research seeks one Berthold Centro microplate luminometer or equal product, with quotes due September 4, 2026 at 2:00 pm CDT. The procurement is set aside for small business concerns under NAICS 334516 (Analytical Laboratory Instrument Manufacturing), with a size standard of 1,000 employees.

Place of performance is the FDA facility in Silver Spring, Maryland, with FOB Destination delivery expected within 30 days after award. The luminometer must read less than 5 amol ATP per well and detect less than 2 Zeptomoles of firefly luciferase. Required salient characteristics include a laboratory-appropriate benchtop footprint, variable-volume injection system with a minimum of two injectors, spectral range of 340–630 nm, greater than 6 orders of magnitude dynamic range, PC-based software for operation and data analysis, and a minimum one-year full manufacturer's warranty with installation and staff training. For products offered as "equal" to the Berthold Centro, offerors must submit a narrative demonstrating how each salient characteristic is met. An Accessibility Conformance Report based on the Voluntary Product Accessibility Template is required for items not in compliance with Section 508 accessibility standards.

The contract type is Firm Fixed-Price. Offerors must be registered and active in SAM.gov with a valid UEI and submit technical acceptability documentation sufficient to demonstrate compliance with minimum requirements. Technical acceptability is evaluated on a pass/fail basis; award goes to the lowest-priced technically acceptable and responsible offeror. Offer prices remain firm through September 30, 2026. Payment terms are Net 30 days after government acceptance via the Department of Treasury's Invoice Processing Platform. Quotes must be submitted by email to timothy.walbert@fda.hhs.gov.

Notice text

This FDQ requirement is for a Berthold Centro or equal microplate luminometer.  Please see attached PDF file titled 75F40126Q134302.

Files

File Type Posted
75F40126Q134302.pdf PDF

Similar open opportunities

Opportunity Agency Set-aside Responses due
Hidex 600SLe Liquid Scintillation Counter United States · NAICS 334516 National Institute on Drug Abuse No set-aside
Low Background Smear Counter New York 11973, United States · NAICS 334511 Office of Science No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity