Please note that this is not a request for proposal nor a request for quote as there is no solicitation document available. Therefore, this is a notice of intent issued by the Department of Veterans Affairs to negotiate a limited source contract with Radiological Imaging Technology for RIT Radiation Therapy QA Software. This procurement is intended to be used in Radiation Oncology Department requires the purchase of quality assurance (QA) software to assist in departmental operations.

The proposed action is IAW FAR 13.106-1(b) soliciting from a single source.

Purpose

The Clement J. Zablocki VA Medical Center Radiation Oncology department requires the purchase of quality assurance (QA) software to assist in departmental operations. The software will be installed on VA owned device by Biomedical VA Staff. Working in synchronization with the linear accelerators, this software will easily automate the LINAC image QA and obtain quantitative, repeatable analysis for all tests. The RIT Complete system can generate full reports of all imaging tests that can be automatically tracked with RIT trend. This software is necessary for maintaining the quality of diagnostic images, to minimize radiation exposure and to be cost effective. All equipment and software much follow the guidelines and meet all specifications listed below.

RIT Complete software one (1) license

General specifications

All-in-one package that covers Patient, Machine, MLC, and Imaging QA

ACR CT and ACR Large MRI phantom analysis

Patient QA

Isodose

Distance-to-agreement (DTA)

Gamma function

Subtraction

Centroid measurement

Plan based calibration (comparison between does map and film/EPID/CR/Ion chamber array images)

Automated QA checklist

Machine QA

Fully automated TG-142 test for linear accelerators

Planar kV imaging

Planar MV (EPID) Imaging

Radiation vs Light Field

Classic star shot analysis

Beam profiles

EPID Phantom tests for resolution, geometric distortion, uniformity, contrast, and noise

Multileaf collimator (MLC) QA

Elekta Leaf speed analysis

Isocenter QA

Hancock (Leaf and jaw positional accuracy) test for Elekta machines

Catphan 503 analysis for Elekta (Low contrast visibility, resolution, uniformity)

Imaging QA

Automated phantom analysis

One click batch analysis on any number of images queued into system

Automated LINAC image QA

Ability to set custom tolerance values for every measurement

Tolerance profile management ability to customize tolerance profiles per machine and per user

Statistical tracking and trending capabilities

Report generation

Technical Requirements

Must function on Windows 10 Operating System

License must be able to be stored locally (not requiring persistent outside internet connection)

Software must have an approved VA Enterprise Risk Assessment (ERA) prior to VA Staff connecting to the VA network. If the vendor does not have an approved ERA upon award, the vendor must work with VA designee to complete the template documents for system assessment and cooperate VA in all required revisions and meetings for the ERA approval process.

Deliverables or Delivery Schedule

Department of Veterans Affairs

Clement J Zablocki VA Medical Center

5000 W National Ave, Warehouse

Milwaukee, WI 53295

Period of Performance

The Contractor warrants that the software items will be free from defects from materials and workmanship under normal use for a period of not less than one (1) year for equipment defined herein. Contractor will replace or repair software if necessary, to the satisfaction of the Government. Contractor shall provide a copy of the written warranty upon acceptance of the equipment.

All labor, time, materials, equipment, travel and supplies necessary to keep equipment operational during the warranty period shall be included. Phone response to warranty issues within 1 hour and on-site response with 24-hours is required.

Information received in response to this special notice will be considered solely for the purpose of determining whether to conduct a competitive procurement. The determination to not compete the proposed acquisition is wholly within the discretion of the Government. The Government will not be responsible for paying on any costs incurred and/or associated with the development of any responses submitted to this notice. The North American Industry Classification System Code (NAICS) is 541519 (Other Computer Related Services).

Submit questions and/or comments concerning this notice to Carl Weber by email carl.weber3@va.gov no later than 11p.m. Local Time on Tuesday, August 24, 2021. Subject line shall reference notice of intent 36C25221Q1088 from beta.SAM.

Attachments

Files attached to this notice, newest first
File Type Posted
36C25221Q1088.docx DOCX document

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
6525--RIT Radiation Therapy QA Software for Clement J. Zablocki VA Medical Center Award Award Notice