6515--temperature, MRN 657-23-3-2775-0194 (VA-23-00077903)
Closed Pre-Solicitation Posted
- Solicitation number
- 36C25523Q0526
- Agency
-
Veterans Integrated Service Network 15
Veterans Health Administration, Department of Veterans Affairs
- Responses due
- Set-aside
- Total Small Business
Opportunity facts
- NAICS code
- 339112 Surgical and Medical Instrument Manufacturing
- Place of performance
- Marion VA Medical Center 2401 W Main St.
Notice details come from SAM.gov. Updated .
Notice text
2 versions
Update #2 · Latest ·
.
Update #1 ·
Scope of Work Laboratory Steam Sterilizer
Objective
The Marion VA Health Care System intends to procure a Steam Sterilizer for its Pathology and Laboratory Service. This document highlights the technical specifications requested by the facility. Vendors under this proposal shall provide all equipment and accessories, installation services, and project management support.
Equipment Specifications
26 x26 x49 capacity
Power- 480v
Cycle Temps-110C-140C
Energy Efficient/Water saving
Auto sliding door
Touch Screen programming
Security access codes
Programmable Cycles for pre-vacuum, gravity, liquid, leak test, waste bag, USP 660, BA3 and BA4, continuous cycle
Printable cycle records
Visible pressure gauges
Emergency stop button
Control lockout switch
One-year limited warranty in parts and labor against manufacturing defects form the date of delivery
System functionality
Unit to be used for sterilization of heat and moisture-stable materials such as media, specimens or equipment used by Pathology and Laboratory.
Delivery/Performance Information
The products will be provided within 120 days from the issue of the PO
The vendor shall provide all related manuals, specification sheets, instructions, and warranty information with the equipment/materials delivery.
Location of Work, the equipment will be in the new lab expansion of Marion VAMC located at 2401 W. Main Street, Marion, IL 62959
Period of Performance, Length of equipment life or until deemed obsolete.
Deliverables Schedule, the products will be delivered to Building 16 (Warehouse) at Marion VAMC at 2401 W. Main, Marion, IL 62959.
Acceptance Criteria, A warehouse materials handler shall determine if the products are acceptable at the time of delivery.