6515--Sources Sought Ceiling lift for Milwaukee VA.

Closed Pre-Solicitation Posted

Solicitation number
36C25226Q0596
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Milwaukee VA 53295

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs, NCO 12 Great Lakes Acquisition Center, is conducting market research to identify qualified vendors for ceiling lift equipment at the Milwaukee VA Medical Center (VAMC), with responses due July 22, 2026. This sources sought notice is not a request for quote but seeks to determine if sufficient capacity exists among Service Disabled Veteran Owned Small Businesses, Veteran Owned Small Businesses, small or emerging small businesses, or other than small business firms to supply two ceiling lift units meeting specified requirements.

The lift units must feature steel frame construction with a gear-reduced, high-torque horizontal axis motor, minimum 770 pounds weight capacity per lift using a single motor, and an integrated Class III scale for weight reading via hand control. Required safety and operational features include electronic soft-start and soft-stop motor control, emergency manual lowering device with red cord allowing continuous-pull descent, emergency stopping device, current limiter for circuit protection, low-battery motor shutdown, emergency brake for mechanical failure, and hanger bar reaching floor level. The hand controller must store securely on the hanger bar. Rechargeable batteries shall provide up to 120 transfers at 200-pound load and comply with IEC 6100-1-2 life-cycle requirements. Respondents must submit company information, SAM Unique ID, tax identification, business classification, capability statement, manufacturer authorization documentation, and manufacturer location to Contracting Officer Matthew Hohl at Matthew.hohl@va.gov by July 21, 2026 at 11:59 PM Central Time. NAICS code 339112 applies with a 1,000-employee size standard.

Notice text

Page 2 of 3

THIS IS A SOURCES SOUGHT NOTICE ONLY. THIS IS NOT A REQUEST FOR QUOTE.

The NCO 12, Great Lakes Acquisition Center, 115 S. 84th Street, Suite 101, Milwaukee, WI 53214, is performing market research to determine if there is a sufficient number of qualified: (1) Service Disabled Veteran Owned Small Business (SDVOSB); (2) Veteran Owned Small Business (VOSB); or (3) Small or emerging small business firms, or (4) Other Than Small business firms, to provide Ceiling Lift for the Milwaukee VAMC.

General Requirements:

Salient Characteristics:

i. The Lift Unit shall be constructed of a steel frame system driven by a gear-reduced, high-torque, horizontal axis motor.

1. The ceiling lift motor shall have an integrated Class III scale that allows users to read weights off the hand control.

2. 2 lifts with minimum 770 lbs weight capacity

a. Weight capacity shall be achieved using a single lift motor

3. Electronic soft-start and soft-stop motor control.

4. Shall have an emergency manual lowering device. This emergency manual lowering device cord shall be red. Operation of the emergency manual lowering device shall be in such a way that users can apply continuous pull to the cord to facilitate slow descent of the patient.

5. Emergency stopping device.

6. Current limiter for circuit protection in case of overload.

7. Safety device that stops the motor to lift when batteries are low.

8. Emergency brake (in case of mechanical failure).

9. Hanger bar shall be able to reach the floor.

10. Cab: VO plastic fire retardant, UL 94.

11. Hand controller shall be able to be stored securely to the hanger bar.

12. The safe working load of the ceiling lift motor shall be clearly labeled on the outside of the ceiling lift motor.

13. The life cycle (number of charging cycles) for batteries shall comply with IEC 6100-1-2.

14. Provide rechargeable batteries with up to 120 transfers with a load of 200lbs (74kg) (for repositioning)

15. The ceiling lift system (motor/hanger assembly) shall be approved for use with Guldmann Repositioning, High Back, Twin Turner and Limb Lifter loop slings.

ii. Two (2) Basic High Slings:

1. The straps shall meet ISO 10535, Appendix A guidelines. The straps shall ensure the patient s safety by preventing the patient from falling out of the sling.

2. The sling shall meet ISO 10535, Appendix A guidelines. The sling shall cradle the body of the patient. Bariatric slings shall be rated to a minimum of 770 lbs.

iii. Two (2) Hanger Bars: The hanger bar style shall be the four (4) suspension point style compatible with the lift strap.

iv. Continuous charge track systems

1. Two (2) Double Rail continuous charge track systems with bathroom access

2. The Ceiling Track shall be made from high strength extruded aluminum. Provide anchor supports at ceiling substrate

3. Installed rail shall be security tested for 1.5 times greater than the motor s weight capacity and maximum allowable deflection of a horizontal rail is no more than 1/16th inch per 7.87 inch of track length per ISO 10535 standards.

All interested firms who can meet the requirements stated above should respond, in writing,

INCLUDING ALL THE FOLLOWING:

Company name.

Address.

Point of contact (name/title/telephone number/e-mail address).

FSS / SEWP Contract number, if applicable.

SAM Unique ID number.

Tax ID number.

Indication of which business category (SDVOSB, VOSB, Small Business or Other Than Small Business) for which the organization qualifies.

A capability statement that addresses the organizations qualifications and ability to provide the requirement depicted above.

Documentation from manufacturer (i.e. correspondence from manufacturer) of your firm being an authorized distributor for the manufacturer.

Location of Manufacturer

Submit responses to the Contracting Specialist, Matthew Hohl, at Matthew.Hohl@va.gov no later than 11:59 PM (Central Standard Time), July 21st 2026.

36C25226Q0596 Ceiling lift-3215 shall be referenced on all correspondence regarding this announcement.

TELEPHONIC INQUIRIES WILL NOT BE ACCEPTED.

The NAICS Code 339112, Surgical and Medical Instrument Manufacturing, is applicable to this acquisition; the size standard is 1,000 employees. Interested firms are reminded that in accordance with FAR 4.12, prospective contractors shall complete electronic annual representations and certifications in conjunction with FAR 4.11 required registration in the Systems for Award Management (SAM) database prior to award of a contract. The Government will use responses to this notice to make an appropriate acquisition decision. This is the only notice; no other copies of this notice will be provided.

Files

File Type Posted
36C25226Q0596.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
Z1DA--537-26-101 Construction of Ogden Elevator Cab Interiors Jesse Brown VA Medical Center 820 S. Damen Ave Chicago IL 60612 · NAICS 238290 Veterans Integrated Service Network 12 Service-Disabled Veteran-Owned Small Business
Expand 111 OR Department 2AS Phase 1 - Construction - Milwaukee, WI - Project:695-501 Milwaukee, Wisconsin 53295, United States · NAICS 236220 VA Technology Acquisition Center Service-Disabled Veteran-Owned Small Business
WTX Elevator Inspections and Load Testing Services Texas 79720, United States · NAICS 238290 Veterans Integrated Service Network 17 Service-Disabled Veteran-Owned Small Business
6530--Rehabilitation Equipment - High Low Beds and Parallel Bars - for the East Indianapolis VA clinic East Indianapoli VA Clinic Lawrenceburg, Indiana 46229, United States · NAICS 339113 Veterans Integrated Service Network 10 No set-aside
J039--561A4-25-500 Elevator Upgrade & Modernization Repairs at VA NJ HCS Lyons Buildings 6 & 53 Department of Veterans Affairs New Jersey Healthcare System Lyons, New Jersey 07939, United States · NAICS 238290 Veterans Integrated Service Network 2 Service-Disabled Veteran-Owned Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity