6515--RFI - Portable Dialysis Reverse Osmosis Systems

Closed Pre-Solicitation Posted

Solicitation number
36C24426Q0861
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Wilmington VA Medical Center Dialysis Department Wilmington, Delaware 19805, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs, Veterans Integrated Service Network 4, is requesting information on portable reverse osmosis water purification systems for the Wilmington VA Medical Center, with responses due August 11, 2026. This is a Request for Information only and does not constitute a solicitation or obligation to award. The RFI seeks two brand-name or equal Aqua C 15 Amp systems to replace current units experiencing frequent bacterial contamination that cannot be effectively sanitized despite maintenance efforts.

The systems must support hemodialysis operations and meet AAMI-compliant water supply standards. Each complete package includes an Aqua C 15 Amp RO unit with softener, AquaCaddy V2 backflow prevention and RPZ device, booster pump, dual carbon block AAMI water profile kit, and installation cart. Systems must achieve greater than 99 percent bacteria and endotoxin rejection rates and remain compatible with Fresenius 2008T BlueStar dialysis machines per ANSI/AAMI RD52 standards.

Contractor responsibilities include full installation, system commissioning, and on-site training for dialysis staff and biomedical technicians, with one follow-up training session within 90 days of installation. Delivery is required within 90 days of contract receipt. Place of performance is Wilmington VA Medical Center, Wilmington, Delaware. The RFI requests company information, business size status, manufacturer or distributor confirmation, pricing estimates, capabilities documentation for equivalent items, and responses to 11 specific questions addressing federal supply schedule availability, Buy American compliance, and subcontracting plans. Responses submit to Contract Specialist Theresa Murray at Theresa.Murray@va.gov by 4:00 PM Eastern on August 11, 2026.

Notice text

This is Not a Solicitation Announcement. This is a Request for Information Only. This Request for Information (RFI) is intended for information and planning purposes only at this time; and shall not be construed as a solicitation or as an obligation on the part of the Department of Veterans Affairs. Since this is Request for Information announcement, no evaluation letters and/or results will be issued to respondents. No solicitation exists currently.

The Department of Veterans Affairs VISN 4, Network Contracting Office 4 (NCO 4), is seeking information and potential qualified/verified Businesses capable of meeting requirements and salient characteristics below to provide the Wilmington VA Medical Center with Portable Reverse Osmosis Systems.

WILMGTON VA MEDICAL CENTER

SCOPE OF WORK AND GENERAL SPECIFICATIONS FOR PORTABLE REVERSE OSMOSIS SYSTEMS

Background

Wilmington VAMC requires two brand name or equal to Aqua C 15 Amp, portable reverse osmosis (RO) water purification systems.

The systems are required to support hemodialysis operations and to ensure a safe, AAMI-compliant water supply for patient care. The current portables require immediate replacement due to frequent, high-level bacterial contamination. Despite rigorous, in-depth disinfection and ongoing maintenance, the units advanced age and deterioration prevent effective sanitization.

Project Scope

The Department of Veterans Affairs, Wilmington VA Medical Center requests a brand name or equal procurement for the purchase of new Aqua C UNO PORTABLE RO 15-amp water purification system which includes equipment installation and training.

This is a brand name or equal for the following: Aqua C 15 Amp portable reverse osmosis (RO) water purification systems with Softener Portable incl. salt tablet kit, AquaCaddy V2 Backflow Prevention/RPZ, AquaCaddy V2 Bolt on HD Connection kit, AquaCaddy V2 Booster Pump, Aqua Caddy - Dual Carbon Block

AAMI Water Profile Kit and Aqua C Uno RO Install with Isopure Cart.

SPECIFIC TASK:

The contractor shall furnish and deliver the following equipment:

o Two (2) AquaC 15 Amp Portable Reverse Osmosis Systems (Product No. 24-10VA-1)

o Two Softener Portable includes salt tablet kit (Product No. 24-0479-0)

o Two AquaCaddy V2 Backflow Prevention/RPZ (Product No.

o 24-ACBF-0)

o Two AquaCaddy V2 Bolt on HD Connection. Kit (Option) (Product No. 24-ACCK-0)

o Two AquaCaddy V2 Booster Pump (Product No.24-ACBP-0)

o Two Aqua Caddy - Dual Carbon Block (Product No. 24-AC02-0)

o Two AAMI Water Profile Kit (Product No.24-AAMI-0)

o Two Aqua C Uno RO Install with Isopure Cart (Product No.111147)

The contractor shall provide complete installation of all two (2) AquaC RO units, including accessories, plumbing connections, and system commissioning.

The contractor shall provide on-site training for Medical Center clinical dialysis staff and Bio-Med techs in accordance with the following parameters:

o Service: AquaC System Training (Product No. 111122)

o One (1) follow-up training session is included within the contract price.

o Training must be initiated within thirty (30) calendar days after system installation; the follow-up training must be initiated within ninety (90) calendar days after installation. Failure to initiate within these windows shall result in expiration of included training, and additional fees will apply.

o Portable RO must be compatible with Fresenius 2008T BlueStar dialysis machines.

Required salient characteristics the unit must meet the following:

System Type

Portable, single-patient reverse osmosis (RO) water purification system

Height

50cm unit only, excluding cart

Width

52cm

Depth

20cm

Weight (Empty)

Approximately 35 kg (77.2 lb)

Weight (Full)

Approximately 37.5 kg (82.7 lb)

Portability

Must be cart-mounted and portable within the clinical environment; no permanent plumbing required

Water Quality Standard

Must meet or exceed ANSI/AAMI RD52 hemodialysis water quality standards for bacteria, endotoxins, and chemical contaminants

Permeate Output

Minimum 85 L/h (±10%) at 15°C (59°F) water inlet temperature

Bacterial / Endotoxin Rejection

Greater than 99% rejection rate for bacteria and endotoxins

Dissolved Salt Rejection

Greater than 96% average rejection rate for dissolved salts

Membrane Type

Spirally-wound RO membrane

Yield (Low Hardness)

Up to 70% at feed water hardness 17.8 ppm CaCO3)

Concentrate Pressure (Max)

Maximum 290.1 psi

Noise Level (SUPPLY mode)

44 46 dB(A) measured at 1 meter distance

Permeate Output (Disinfection)

12 L/h (±10%) during ring heat disinfection cycle

Bacterial / Endotoxin Rejection

Greater than 99% rejection rate for bacteria and endotoxins

Dissolved Salt Rejection

Greater than 96% average rejection rate for dissolved salts

Membrane Type

Spirally-wound RO membrane

Yield (Low Hardness)

Up to 70% at feed water hardness 17.8 ppm CaCO3)

Concentrate Pressure (Max)

Maximum 290.1 psi

Noise Level (SUPPLY mode)

44 46 dB(A) measured at 1 meter distance

Permeate Output (Disinfection)

12 L/h (±10%) during ring heat disinfection cycle

Bacterial / Endotoxin Rejection

Greater than 99% rejection rate for bacteria and endotoxins

Dissolved Salt Rejection

Greater than 96% average rejection rate for dissolved salts

Membrane Type

Spirally-wound RO membrane

Yield (Low Hardness)

Up to 70% at feed water hardness 17.8 ppm CaCO3)

Concentrate Pressure (Max)

Maximum 290.1 psi

Noise Level (SUPPLY mode)

44 46 dB(A) measured at 1 meter distance

Permeate Output (Disinfection)

12 L/h (±10%) during ring heat disinfection cycle

Electrical Supply

120 V AC, 60 Hz, 15.5 Amp

Nominal Power Variants

120 V 1000 VA or 120 V 1700 VA (per voltage variant)

Operating Mode

Continuous operation (Standby); SUPPLY mode used periodically during treatment

Electrical Safety Classification

Classified per EN 60601-1:2006 + A1:2013 / IEC 60601-1:2005 + A1:2012

Grounding Requirement

Must be connected to supply network with protective conductor (grounded outlet); no ungrounded connections permitted

Fuse Protection

Internal fuses for voltage component protection; replaceable only by trained/certified personnel

EMC Compliance

Compliant with applicable electromagnetic compatibility (EMC) standards; no stacked installation with other electrical devices

Interface / Connectivity

Fresenius hemodialysis device interface port; compatible with Fresenius 2008T BlueStar dialysis machines

Thermal Disinfection Module

Module heat disinfection up to 82°C (179.6°F)

Thermal Disinfection Ring

Ring heat disinfection up to 90°C (194°F)

Membrane Service Life

Minimum expected life of 120 heat disinfection cycles

Chemical Disinfection

Compatible with peracetic acid solution (~0.2% effective concentration); system volume 2.5 L (0.66 gal)

Decalcification

Compatible with approximately 20% citric acid solution; automated decalcification program required

Rinse Interval

Automated rinse cycle: minimum every 4 hours; rinse volume 10 L (3 gal) or 10-minute rinse time

AutoSTART / AutoSTOP

System shall support automatic start in SUPPLY mode via communication cable connection to dialysis machine

Disinfection Validation

System must provide auditable disinfection reporting and daily operational data acquisition logs

Biocompatibility Standard

All components in contact with permeate must be biocompatible per DIN EN ISO 10993-1

Tubing / Piping Material

FEP, silicone, or stainless steel; food-grade with chemical resistance

Sensors

Titanium (Ti), Aluminum Oxide (Al O ), PBT

Module Tube

V4A stainless steel

Membrane Material

PSU/PA (polysulfone/polyamide)

Hydraulic Blocks / Valves

iPP, PVDF, Noryl GTX

Seals

EPDM (ethylene propylene diene monomer)

Sampling Ports

Integrated permeate sampling ports for microbiological and chemical analysis

Permeate Connection

Permeate nipple; optional permeate coupling; compatible with Fresenius hemodialysis machine water inlet

Feed / Concentrate Flow

Feed flow: 0.0 20 L/min (±10%); Concentrate flow: 0.0 20 L/min (±10%)

Permeate Consumption

0 6 L/min (±10%) during normal supply operation

Drain Connection

System drain must connect to available outlet; no recontamination of permeate circuit permitted

Water Supply Hose

Must connect to potable water supply meeting pre-treatment feed water requirements

Installation Type

Portable; no permanent wall or floor mounting required;

Stacking / Adjacency

System must NOT be stacked with other electrical devices; minimum separation distances from radiation sources required per IFU

Operational Qualification

Vendor must perform Installation and Operational Qualification (IQ/OQ) at time of installation

Display

Integrated digital display with operating mode status bar; multilingual support

Permeate Conductivity Monitoring

Continuous real-time permeate conductivity measurement and display; alarm on out-of-range condition

Feed Water Conductivity

Continuous real-time feed water conductivity measurement and display

Temperature Monitoring

Permeate and concentrate temperature display; heater temperature 1.0 135°C (±1°C); concentrate temperature 1.0 135°C (±1°C)

Rejection Rate Display

Real-time rejection rate display: 0.0 99% (±1%)

Effective Yield Display

Effective yield display: 0.0 70% (±5%)

Alarm System

Audible and visual alarms for: no water supply, fault drain, conductivity fault, sensor error, and other abnormal conditions; alarm must halt dialysis supply automatically

Operational Data Logging

System must support manual and automated daily operational data acquisition and reporting per AAMI requirements

Visual Status Indicator

Green visual indicator must illuminate to confirm system is ready for treatment

TRAINING:

The contractor shall provide on-site training for Medical Center clinical dialysis staff and Bio-Med techs.

DELIVERY/PLACE OF PERFORMANCE:

Delivery shall be within 90 days after receipt of contract. Deliver equipment to:

Wilmington VA Medical Center (Attn: Shayonna Smith)

Dialysis Department

1601 Kirkwood Hwy

Wilmington, DE 19805-2917

Prior to delivery, Contractor shall verify shipment with the Point of Contact (POC). Once confirmed, tracking information, quantity and estimated arrival shall be provided.

The Contractor must comply and provide all necessary verification and information to the VAMC. Otherwise, the delivery may be refused and require redelivery at no additional cost to the government.

Responses to this RFI should include company name, address, point of contact, phone number, and point of contact e-mail, UEI Number, Cage Code, size of business pursuant to North American Industry Classification code (NAICS) 339113, The Small Business Size Standard for this NAICS code is 800 employees.

Please answer the following questions:

(1) Please indicate the size status and representations of your business in accordance with the NAICS code 339113, such as but not limited to: Service-Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), Hubzone, Woman Owned Small Business (WOSB), Large Business, etc.)

(2) Is the required equipment available on a Federal Supply Schedule or Government Wide Acquisition Contract? If so, please include the schedule number.Â

(3) Are you the manufacturer, distributor, or an equivalent solution to the items being referenced above?Â

If not, provide additional information shown below. This is to confirm compliance with the non-manufacturer rule IAW 13 CFR 121.406(b) Nonmanufacturers.Â

Does your company exceed 500 employeesÂ

Primarily engaged in the retail or wholesale trade and normally sells the type of item being supplied;Â Â Â

Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice;Â Â

Provide the items of a small business

(4) If your company is a large business, do you have any authorized small business distributors? If so, please provide their company name, telephone, point of Contact and size status (if available).Â

(5) If your company is a small business and you are an authorized distributor/reseller for the items identified above or an equivalent solution, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified?Â

(6) Is the item you are supplying made or manufactured in America? This is to ensure compliance with FAR 52.225-1 Buy American-Supplies.

(7) If you intend to subcontract any work on this contract, what portion of the total cost will be self-performed/will be performed by your organization? Please provide estimated detailed percentage breakdowns related to subcontracted work and completion of job.Â

(8) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.Â

(9) Please submit your capabilities in regard to the salient characteristics detailed above and any information pertaining to equal to items to establish capabilities for planning purposes?Â

(10) Please review salient characteristics/statement of work (if applicable) and provide feedback or suggestions. If none, please reply as N/A.Â

(11) Please provide your UEI number.Â

*** Submissions addressing number (10) should show clear, compelling and convincing*** evidence that all equal to items" meet all the salient characteristics.Â

This RFI will be conducted in accordance with Federal Acquisition Regulation (FAR) Part 12. Telephone responses will not be accepted.

Responses must be received via email to Contract Specialist Theresa Murray at Theresa.Murray@va.gov no later than 4:00 PM Eastern Standard Time on Tuesday, August 11, 2026.

This notice will help the VA in determining available potential sources only.  Do not contact VA Medical Center staff regarding this requirement, as they are not authorized to discuss this matter related to this procurement action.

All firms responding to this Request for Information are advised that their response is not a request for proposal, therefore will not be considered for a contract award. Â

Â

If a solicitation is issued, information will be posted on SAM.gov web site for all qualified interested parties at a later date and interested parties must respond to the solicitation to be considered for award.  This notice does not commit the government to contract for any supplies or services.  The government will not pay for any information or administrative cost incurred in response to this Request for Information.  Information will only be accepted in writing via email.

DISCLAIMER:

This RFI is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly.  Responses to this notice are not offers and cannot be accepted by the Government to form a binding contract.  Responders are solely responsible for all expenses associated with responding to this RFI.Â

Files

File Type Posted
36C24426Q0861.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
J046--Reverse Osmosis PM NAICS 811210 Veterans Integrated Service Network 21 Total Small Business
J065--Reverse Osmosis Systems Maintenance Services Ann Arbor VA Medical Center 2215 Fuller Road, United States · NAICS 541380 Veterans Integrated Service Network 10 No set-aside
B533--VANIHCS SPS Water Testing (VA-27-00000760) Marion, Indiana 46953, United States · NAICS 541380 Veterans Integrated Service Network 10 Service-Disabled Veteran-Owned Small Business
Z1DA--CON-NRM-668-22-106 Correct Domestic Water System Spokane, Washington 99205, United States · NAICS 238220 Veterans Integrated Service Network 20 Service-Disabled Veteran-Owned Small Business
Z1DA--Local | 460-26- 4-5074-0108 | 460-26-999 | Condenser Water Piping Modification | (VA-26-00077929) Department of Veterans Affairs Wilmington Vamc Wilmington DE 19805 · NAICS 236220 Veterans Integrated Service Network 4 Service-Disabled Veteran-Owned Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity