The Department of Veterans Affairs, Veterans Integrated Service Network 22, is conducting a sources-sought market research for the procurement of a Hamilton-MR1 ventilator transport system, with responses due by August 4, 2026 at 6:00 am Pacific Time. This is an information-gathering RFI only and does not constitute a solicitation; however, the VA has indicated that a formal solicitation may be published on SAM and GSA eBuy following review of responses. VA Prescott Healthcare System requires one Hamilton-MR1 Package for providing mechanical ventilation to patients requiring respiratory support during Magnetic Resonance Imaging procedures. The requirement is brand name or equal, and the VA encourages identification of domestic alternatives manufactured in the United States.
The ventilator must be MR-conditional to 50 millitesla, compatible with both 1.5T and 3T scanners, weigh approximately 6.8 kilograms, support patient populations from neonatal through adult, operate on battery power for 8 to 9 hours, and be FDA 510(k) cleared. The package includes clinical training in Reno, oxygen hose, adult and pediatric configuration, trolley with intelligent brake system, lithium-ion batteries, and applications including trends/loops, NIV/NIV-ST, and DuoPAP/APRV. Responses must address business size classification (with emphasis on identifying VOSB or SDVOSB contractors), GSA FSS contract status, manufacturing capability, and Buy American Act compliance. If market research identifies two or more certified VOSB or SDVOSB firms able to perform the work, VA will set aside the underlying solicitation per 38 U.S.C. § 8127. Submit responses to Rochelle Malveaux at Rochelle.Malveaux@va.gov.
Notice text
Sources Sought Synopsis Only
THIS REQUEST FOR INFORMATION (RFI)/SOURCES SOUGHT/RFI IS ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES ONLY. THIS IS NOT A SOLICITATION.
THE SUBMISSION OF PRICING, CAPABILITIES FOR PLANNING PURPOSES, AND OTHER MARKET INFORMATION IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI IN ACCORDANCE WITH (IAW) FAR 15.201(e).
The purpose of this synopsis is to gain knowledge of potential qualified sources and their size classification (hubzone, 8(a), small, small disadvantage, veteran owned small business (VOSB), service-disabled veteran owned small business (SDVOSB), or large business relative to NAICS 334510 with size standards of 1250 employees. Note that if VA s pending market research establishes that two or more VOSB or SDVOSB, who are certified in U.S Small Business Administration site https://www.sba.gov can perform the project, VA will set aside the underlying solicitation per 38 U.S.C. §8127.Â
Responses to this synopsis will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Synopsis, a solicitation announcement may be published on SAM, and/or GSA eBuy. Responses to this Sources Sought Synopsis are not considered adequate responses to the solicitation announcement.
General Requirements:
The Department of Veterans Affairs (VA), VA Prescott Healthcare System, has a need for the HAMILTON-MR1 PACKAGE. The Hamilton-MR1 will be used to provide mechanical ventilation to patients who require respiratory support while undergoing diagnostic or therapeutic procedures in a Magnetic Resonance Imaging environment.
The requirement is for brand name or equal to: MR1 Ventilator Transport
Please Identify any domestic alternatives (meaning manufactured in the United States)
Nomenclature: Transport Ventilator
MANUFACTURE AND BRAND NAME PRODUCT INFORMATION Hamilton Medical Inc
CLIN
Description
Qty
Unit of Measure
01
Hamilton-MR1 Package
1
EA
- Clinical Training
1
- Reno Service Training Tuition
1
- O2 Hose
1
- Adult/Pediatric Configuration
1
- US Power Cord
1
- Trolley (with Intelli brake)
1
- Diss Connector 02-Platform C
1
- 02 Cell-Initial Order
1
- Li-Ion Batteries (2 pieces)
1
- English Language Kit
1
- USA Customization
1
02
Hamilton C3/C2/C1/T1/MR1 battery calibrator w/cord
1
EA
03
Trends/Loops Application
1
EA
04
NIV/NIV-ST Application
1
EA
05
DuoPAP/APRV Application
1
EA
REQUIREMENTS: The Radiology department requests the acquisition of a Mechanical Ventilator transport for the purpose of providing continuous respiratory support to patients during MRI imaging procedures. The Salient Characteristics must include:
MR-Conditional to 50 mT (compatible with 1.5 T & 3 T scanners) TeslaSpy continuously monitors magnetic field and alerts if too close even when off.
Portable ICU ventilator weighing ~6.8 kg ( 21 kg with trolley); compact size supports ICU MRI transport with quick-release trolley and auto-lock brakes.
Broad patient range: neonatal (tidal down to 2 mL), pediatric, adult.
Ventilation modes: full suite volume & pressure-controlled, ASV®, NIV, nCPAP, HFOT, duet modes like PSIMV+, APRV, DuoPAP.
Turbine-driven, no external compressed air needed; delivers up to 260 L/min flow.
Battery life 8 9 hr (72 Wh, 2× lithium-ion); recharge time ~3 6 hrs.
Ventilation Cockpit UI with 8.4 touchscreen showing waveforms, loops, trends, lung mechanics.
Safety & monitoring: alarms for pressures, volumes, disconnections, apnea; leak compensation; dynamic lung panels show compliance, resistance, stress/strain
FDA 510(k) cleared in 2016 (K153046) for invasive and noninvasive use in MRI, with neonatal capability and enhanced battery runtime
Participating Healthcare Facilities:
VISN 22 West
VA Prescott Healthcare System 500 HW 89N; Prescott, AZ 86313-5001
In response to this announcement, please complete the information below:
Company:
Address:
Contact Name:
Phone No:
Email:
DUNS:
Business size information
Select all that applies:
Small Business
Emerging Small Business
Small Disadvantaged Business
Certified under Section 8(a) of the Small Business Act
HubZone
Woman Owned
Certified Service-Disabled Veteran Owned Small in Vetbiz.gov
Certified Veteran Owned Small Business in Vetbiz.gov
Large Business
Other:____________
All offered items must be listed on vendor FSS contract
FSS/ GSA Contract Holder of the requested item(s)
Yes No
FSS/ GSA Contract Number
Effective Date / Expiration Date
Does your company manufacture the requested product(s)? If no, answer the questions below.
Yes No
Does your company exceed the 500-employee alternative size standard for nonmanufacturers?
Yes No
Is your company primarily engaged in the retail or wholesale trade and normally sell the type of product(s) being supplied?
Yes No
Does your company take ownership or possession of the products(s) with its personnel, equipment, or facilities in a manner consistent with industry practice?
Yes No
Does your company supply the end product(s) of a small business manufacturer or processor made in the US, or obtain a waiver of such requirement?
Yes No
Identify any other companies owned (wholly or in-part) by the owners of this business who provide goods or services under the same or a related NAICS codes.
Not applicable
List:
Buy American Act Information
Are all items manufactured in the U.S? If not, list the items not manufactured in the US and the country of origin in the table provided below.
Yes No
Item #
Description
Country of Origin
Domestic Alternative Product
Do you have domestic alternatives? (Equal product that is manufactured within the United States) If so, please state the item, description and place of manufacture (country of origin) in the spaces below
Yes No
Item #
Description
Country of Origin
This notice is neither a request for competitive proposals or solicitation of offerors. This notice is to assist the VA in determining sources only and a solicitation is not currently available. If a solicitation is issued it will be announced at a later date, and all interested parties must respond to that solicitation announcement separately from this response. Responses to this Sources Sought is not a request to be added to a prospective bidders list or to receive a copy of the solicitation. Information received as a result of this notice will normally be considered solely for the purpose of determining whether to conduct competitive procurement. This notice does not represent a commitment by the Government to pay for costs incurred in the preparation and submission of information or any other costs incurred as a response to this announcement. Inquiries will only be accepted in writing via email to Rochelle.Malveaux@va.gov on or before Tuesday, August 4th, 2026, by 6:00 am PST Local Time.
Files
| File | Type | Posted |
|---|---|---|
| 36C26226Q1109.docx | DOCX document |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| 6515--Anesthesia Machines - Cheyenne VA Medical Center | Veterans Integrated Service Network 19 | Total Small Business | |
| 6515--Anesthesia Units Extend due date/time | Veterans Integrated Service Network 1 | No set-aside | |
| 6515 - Intent to Sole Source - Siemens Skyra Coldhead RDE-412L4 (11472997 | Veterans Integrated Service Network 21 | No set-aside | |
| J065--Imaging Service & Preventative Maintenance EHCC & MTV | Veterans Integrated Service Network 15 | No set-aside | |
| Sleep studies CLE | Veterans Integrated Service Network 10 | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity