6515--Palmscan Pachymeter/s with Tips
Closed Pre-Solicitation Posted
This opportunity was awarded. See the award notice from , or the latest solicitation from .
- Solicitation number
- 36C24421Q0591
- Agency
- Veterans Integrated Service Network 4 Veterans Health Administration, Department of Veterans Affairs
- Responses due
- Set-aside
- Total Small Business
Opportunity facts
- NAICS code
- 339112 Surgical and Medical Instrument Manufacturing
- Place of performance
- Lebanon VA Medical Center 1700 South Lincoln Avenue 17042 USA
Notice details come from SAM.gov. Updated .
Notice text
This is Pre-Solicitation Notice
The solicitation will be a 100% set-aside for Small Business; an offeror shall be VIP-Verified and registered at sam.gov to be eligible for award. The NAICS Code is 339112, All other personal services or All other support services.
No further information is available now; no response is expected. The solicitation will be posted at on 4/16/2021 0800 Eastern Standard Time.
Thank you for your interest.
PalmScan p2000 Fastpach Pachymeter
Statement of Need
Background:
The Lebanon VA Medical Center (LVAMC) has a requirement for ten (10) PalmScan p2000 fastpach pachymeters, and 1 box of tips per unit, to replace old and failing pachymeters. These devices are portable, battery operated units used to measure Central Corneal Thickness (CCT) of human eyes. These are used on every patient provided care in Optometry.
Scope of Work:
The device is a portable, battery operated, unit used to measure corneal thickness (CT) of human eyes. These are used on every patient provided care in Optometry.
Salient Characteristics of the p2000 Pachymeter Unit:
Battery Operated and Portable
50 Mhz Probe
Accuracy +/- 4 um
Precision +/- 1 um
Resolution +/- 1 um
Range: Corneal/Epithelium Thickness 250-1000 um
Sampling Frequency 200 MHz
Utilizes disposable pachytips
Not autoclavable or immersable
Acoustic Output Measurements:
Maximum Acoustical Output - 0.189
Maximum Acoustical Output 1.99
Mechanical Index 0.155
Palmscan Pachymeter Tip Covers:
Must fit and integrate with the p2000 Pachymeter unit.
Must be disposable
Additional Information:
The PalmScan was chosen for its accuracy, portability, functionality and durability.
Delivery:
Delivery shall be within thirty (30) days from the time of award. The Pachymeters and accessories shall be delivered to the Lebanon VAMC located at 1700 S. Lincoln Ave., Lebanon, PA 17042; Attn; Warehouse. BioMedical Engineering will need to perform a safety check and enter the equipment in to the system for annual Preventive Maintenance (PM) checks. Placement in the LVAMC clinic shall be coordinated through the Chief, Optometry, BioMed, and/or Contracting Officer Representative.
Attachments
| File | Type | Posted |
|---|---|---|
| 36C24421Q0591_1.docx | DOCX document |
Notice history
| Notice | Type | Posted |
|---|---|---|
| 6515--Palmscan Pachymeter/s with Tips | Award Notice | |