3 Phase Power Supply Test System
Closed Pre-Solicitation Posted
Not the latest version. See the latest solicitation from .
- Solicitation number
- SP4706-26-Q-00XX
- Agency
-
Defense Logistics Agency
Department of Defense
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 334516 Analytical Laboratory Instrument Manufacturing
- PSC
- 6640 Laboratory Equipment And Supplies
- Place of performance
- Columbus, Ohio 43213, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Defense Logistics Agency intends to solicit for two 3-phase power supply test systems to be delivered to Defense Supply Center Columbus, Columbus, Ohio, with a pre-solicitation response due by May 15, 2026.
The procurement covers one 18kVA and one 1kVA system, each capable of supplying DC, single-phase, and 3-phase power from DC to 5kHz in sinusoidal, square wave, and arbitrary waveform configurations. Both systems must be new (not refurbished) and include operating manuals. Performance specifications include accuracy of at least 0.15%, distortion no more than 0.3% at 50/60Hz and no more than 1% above those frequencies, a minimum power factor of 0.98, and resolution of at least 0.1V. Each system must carry NIST-traceable and ISO 17025-accredited calibration certificates with data and uncertainties.
The 18kVA system requires one 480V/3-phase delta hard-wired disconnect; the 1kVA system operates from a standard 120V/single-phase 20A outlet. Both systems must mount on wheeled frames and fit through standard commercial double doors (approximately 83 inches height by 70 inches width). Deliverables include all cables, sub-equipment, and accessories necessary for reliable operation, plus two Windows 11 PCs with permanent software licenses and GPIB, ethernet, and USB control interfaces. Contractors must offer optional extended warranty and multi-year service contracts.
No set-aside applies to this procurement. NAICS 334516 (Analytical Laboratory Instrument Manufacturing); PSC 6640 (Laboratory Equipment and Supplies).
Notice text
2 versions
Update #2 · Latest ·
The contractor shall provide all labor, materials, tools, transportation, supplies, supervision, etc. (except as noted), necessary to provide all components for a 18kVA 3Φ Power Supply Test System and 1kVA 3Φ Power Supply Test System to the Defense Supply Center Columbus (DSCC), Columbus, Ohio. The salient characteristics of the 3Φ Power Supply Test Systems are as listed below:
A. DC, 1Φ and 3Φ power system with support equipment with the following salient characteristics:
1.Shall be new, not refurbished.
2.Shall be supplied with Operating manual.
3.First system shall be capable of supplying: DC, 1Φ and 3Φ power from DC to 5kHz (sinusoidal, square wave, and arbitrary waveform) rated to 18kVA
4.Second system shall be capable of supplying: DC, 1Φ and 3Φ power from DC to 5kHz (sinusoidal, square wave, and arbitrary waveform) rated to 1kVA
5.Shall have accuracy of at least 0.15%.
6.Shall have no more than 0.3% distortion at 50/60Hz and no more than 1% distortion above these frequencies.
7.Shall have no less than 0.98 power factor.
8.Shall have resolution of no less than 0.1V.
9.Optional extended warranty and multi-year service contract must be offered.
10.Shall have: NIST Traceable Calibration Certificate and ISO 17025 accredited calibration certificate with data and uncertainties.
11.Shall include all the cables, sub-equipment and accessories for the system to reliably operate.
12.Shall be on a wheeled frame for easy movement.
13.Shall be able to fit through standard commercial double door (approximately 83”H x 70”W).
14.First system shall work as a single system to only require one 480V/3φ delta hard-wired disconnect for the entire system. Seconds system shall work as a single system to only require standard 120V/1φ 20A outlet.
B. Required Upgrades and Accessories - Brand Name or Equal:
1.GPIB, ethernet, USB Interfaces.
2.2 Windows 11 PCs with all necessary peripherals, hardware, ports, et cetera with software and permanent license to reliably control system.
3.8 hours of on-site training and setup.
C. Place of Performance
The instrument shall be delivered to:
Defense Supply Center Columbus
300 North James Road, Bldg 17-3
Mark for: DLA Product Test Center, Bldg 11-7 (TE Lab)
Columbus, Ohio 43213
Other Requirements:
1.Delivery: 30 days from date of the award.
2.Contractor shall give notification when the instrument is ready to be delivered.
3.All shipping costs are to be included in the contracted price.
Update #1 ·
The contractor shall provide all labor, materials, tools, transportation, supplies, supervision, etc. (except as noted), necessary to provide all components for a 18kVA 3Φ Power Supply Test System and 1kVA 3Φ Power Supply Test System to the Defense Supply Center Columbus (DSCC), Columbus, Ohio. The salient characteristics of the 3Φ Power Supply Test Systems are as listed below:
A. DC, 1Φ and 3Φ power system with support equipment with the following salient characteristics:
1.Shall be new, not refurbished.
2.Shall be supplied with Operating manual.
3.First system shall be capable of supplying: DC, 1Φ and 3Φ power from DC to 5kHz (sinusoidal, square wave, and arbitrary waveform) rated to 18kVA
4.Second system shall be capable of supplying: DC, 1Φ and 3Φ power from DC to 5kHz (sinusoidal, square wave, and arbitrary waveform) rated to 1kVA
5.Shall have accuracy of at least 0.15%.
6.Shall have no more than 0.3% distortion at 50/60Hz and no more than 1% distortion above these frequencies.
7.Shall have no less than 0.98 power factor.
8.Shall have resolution of no less than 0.1V.
9.Optional extended warranty and multi-year service contract must be offered.
10.Shall have: NIST Traceable Calibration Certificate and ISO 17025 accredited calibration certificate with data and uncertainties.
11.Shall include all the cables, sub-equipment and accessories for the system to reliably operate.
12.Shall be on a wheeled frame for easy movement.
13.Shall be able to fit through standard commercial double door (approximately 83”H x 70”W).
14.First system shall work as a single system to only require one 480V/3φ delta hard-wired disconnect for the entire system. Seconds system shall work as a single system to only require standard 120V/1φ 20A outlet.
B. Required Upgrades and Accessories - Brand Name or Equal:
1.GPIB, ethernet, USB Interfaces.
2.2 Windows 11 PCs with all necessary peripherals, hardware, ports, et cetera with software and permanent license to reliably control system.
3.8 hours of on-site training and setup.
C. Place of Performance
The instrument shall be delivered to:
Defense Supply Center Columbus
300 North James Road, Bldg 17-3
Mark for: DLA Product Test Center, Bldg 11-7 (TE Lab)
Columbus, Ohio 43213
Other Requirements:
1.Delivery: 30 days from date of the award.
2.Contractor shall give notification to Bradley Deslich (bradley.deslich@dla.mil/614-693-0182)
when the instrument is ready to be delivered.
3.All shipping costs are to be included in the contracted price.
Notice history
| Notice | Type | Posted |
|---|---|---|
| 3 Phase Power Supply Test System | Solicitation | |
| This notice | Pre-Solicitation |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Power supply | Land and Maritime | No set-aside | |
| Power supply | Land and Maritime | No set-aside | |
| Power supply | Land and Maritime | No set-aside | |
| Power supply | Land and Maritime | No set-aside | |
| Power supply | Land and Maritime | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity