Purchase Description ATC Laser-based Elemental Analyzer..pdf

PDF 136 KB Posted

Attached to
1. Each. Item EA-300 Keyence Laser-based Elemental Analyzer
Solicitation number
W91ZLK-25-Q-0007
Issued by

About this file

This is a Product Description document detailing requirements for one Keyence Laser-based Elemental Analyzer to be purchased by the Chemical Sampling and Analysis Branch (CSAB) laboratory at Aberdeen Test Center. The analyzer will attach to an existing VHX microscope purchased in 2024 and must perform non-destructive elemental analysis with AI-based material identification capabilities.

The equipment specifications include laser-induced breakdown spectroscopy supporting elements H1 to U92, Nd:YAG laser type at 355nm wavelength, observation magnification of 300x to 1000x, and Class 1 laser safety rating. Key performance requirements include one-click elemental analysis, seamless transition between observation and analysis modes, no pre-processing or vacuum requirements, and real-time analysis capabilities. Delivery and installation are required within 30 days after contract award at the Aberdeen Test Center in Maryland. The purchase was endorsed by YPG and NATICK Army organizations following vendor demonstrations.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PRODUCT DESCRIPTION

Keyence Laser-based Elemental Analyzer

1. INTRODUCTION

The Chemical Sampling and Analysis Branch (CSAB) laboratory requires the Keyence

Laser-based Elemental Analyzer for use with VHX microscope that was purchased in

2024. This attachment will perform elemental analysis with no destruction, deposition, or vacuum required. AI-based recommendations make it possible to identify materials or foreign substances. Seamless transition from magnified observation to elemental analysis all in one unit.

2. PURPOSE AND BACKGROUND

This acquisition is for the purchase of the Keyence Laser-based Elemental Analyzer for use with VHX microscope that was purchased in 2024. The recommendations from YPG and NATICK Army organizations as well has several vendor demonstrations solidified the purchase plans for the Keyence Laser-based Elemental Analyzer. Our requirements to test the same as YPG our sister lab has also endorsed this purchase within our command.

3. PRODUCT SPECIFICATIONS:

This acquisition includes the following product specification and requirements:

QTY One (1) – Purchase and Installation of Laser-based Elemental Analyzer:

Equipment Specific Requirements:

1. Laser-induced breakdown spectroscopy

2. Supported elements H1 to U92

3. Laser type Nd: YAG laser

4. Laser class Class 1 laser product (IEC/EN 60825-1, JIS C 6802, FDA (CDRH) Part 1040.10)

5. Laser wavelength 355 nm

6. Observation magnification 300× to 1000×

7. High dynamic range 16-bit intensity range through RGB data from each pixel

8. Power supplied via dedicated cable from VHX Series controller camera port

4. DELIVERY, INSTALLATION AND PERFORMANCE:

The expected delivery and installation 30 days after contract award.

5. PLACE OF DELIVERY, INSTALLATION

W4QU USA ABERDEEN TEST CENTER

Attn: Mike Long

6850 Lanyard Road, BLDG 358, Aberdeen

Proving Ground, MD 21005-5059

4. PERFORMANCE REQUIREMENTS

4.1 Objectives

1. One-click elemental analysis is possible.

2. Users can easily go from magnified observation to elemental analysis without needing to re-align or focus the laser objective lens.

3. Non-destructive, alteration-free analysis is possible on targets of any size.

4. An integrated AI function will recommend the most likely detected material.

5. No pre-processing or vacuum required.

6. Simple attachment to microscope

7. Both magnified observation and elemental analysis can be done in real-time with a single device.

8. Analysis of foreign particles in a multi-layer film

File details come from the government source that posted it. Updated .