Copy_of_GP_Xray_Survey_Ser_04126-1004_Camp_Grayling_27APR2017_(002).xlsm
XLSM spreadsheet 286 KB Posted
- Attached to
- X Ray Machine
- Solicitation number
- W912JB-18-Q-0087
About this file
Detailed description of equipment needing removal at time of installation.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| CDRL_.pdf | ||
| IMG_0097.jpg | JPG image | |
| IMG_0096.jpg | JPG image | |
| CDRL_.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Site-Equip
| SITE AND X-RAY MACHINE DATA | SURVEY EQUIPMENT USED | Selection | |||||||||||||||||||||||||||||||||||||||||||||||
| Date(s) of Survey | 27-Apr-2017 | Exposure Meter | Choice | ||||||||||||||||||||||||||||||||||||||||||||||
| Principal Surveyor | |||||||||||||||||||||||||||||||||||||||||||||||||
| HPOUser: The surveys described in this chapter will be performed by or under the supervision of a qualified expert. MER personnel and/or health physics technicians commonly perform many of the required tests; however, a qualified expert should review the results of these tests and perform any additional tests that are required by this chapter or that are necessary in his or her professional judgment. | CPT Rodriguez, Jose H. | Manufacturer: | 26585 | Heading | Surveyor | 2nd Surveyor | Installation | ECN | Exp. Meter | Model | SN | Cal Due | ECN | kVp Meter | Model | SN | Cal Due | ECN | Timer | Model | SN | Cal Due | ECN | Ion Chamber | Model | SN | Cal Due | ECN | Light Meter | SN | Cal Due | ECN | |||||||||||||||||
| MORTONAR: Sort all remaining data to the right by MMCN | Manuf | Model | SN | Date | Max kVp | Max mA | Phase | Manuf | Model | SN | Date | Lg FS | Sm FS | Manuf | Model | SN | Manuf | Model | SN | Image | Display | Manuf | Model | SN | Manuf | Model | Manuf | Model | |||||||||||||||||||||
| Second Surveyor | Model: | UNFORS | Variables | ||||||||||||||||||||||||||||||||||||||||||||||
| MORTONAR: Recommend listing all variables to be used in the drop down lists and then sort the variables to the order you want them to appear in the drop down list | 1 | CPT Rodriguez | CPT Rodriguez | Fort Knox, KY | B0944 | Keithley | 35050A | 60086 | 5-Dec-09 | B0944 | Keithley | 35050A | 60086 | 5-Dec-09 | B0944 | Keithley | 35050A | 60086 | 5-Dec-09 | 6073 | Fluke | 451P-RYR | 2410 | 6-Aug-16 | 21039 | Extech | Q619323 | 1-Nov-14 | 869 | Quantum | Odyssey HF GP | QG4000-04A-0105 | Jan-04 | 125 | 500 | 3 | Quantum | E7242 | 3M026 | Dec-03 | 1.5 | 0.6 | Quantum | CM1100 | CM11395 | ||||
| Installation | Camp Grayling, MI | ECN: | X2 | 2 | None | Camp Atterbury, IN | B0962 | Victoreen | 4000M | 719 | 11-Jun-10 | B0962 | Victoreen | 4000M | 719 | 11-Jun-10 | B0962 | Victoreen | 4000M | 719 | 11-Jun-10 | 6074 | Fluke | 451P-RYR | 2359 | 29-Jun-10 | B9277 | Sper Scientific | 26991 | 22-Jul-10 | 888 | Quantum | Odyssey HF GP QX-406 | QG4000-03M1205 | Sep-03 | 125 | 3 | Toshiba | E7242 | 3J170 | Sep-03 | ||||||||
| Building | TMC | Serial Number: | 225565 | 3 | Fort McCoy, WI | G7191 | RADCAL | 2025 | 3326 | 11-Jun-10 | G7191 | RADCAL | 2025 | 3326 | 11-Jun-10 | G7191 | RADCAL | 2025 | 3326 | 11-Jun-10 | 6081 | Fluke | 990 | 540 | 20-Aug-10 | 26585 | UNFORS | 212716 | 18-Jun-16 | 1158 | GE | MPH GP | AA78305MAO | Nov-03 | 150 | 800 | 3 | Varian | A195 | 63808-65 | Jul-07 | ||||||||
| Room | Calibration Due Date: | 3-Sep-17 | 4 | Rock Island Arsenal, IL | G9381 | RTI | PMX-II | 225 | 17-Oct-09 | G9381 | RTI | PMX-II | 225 | 17-Oct-09 | G9381 | RTI | PMX-II | 225 | 17-Oct-09 | 6082 | Fluke | 990 | 541 | 10-Jun-10 | 1725 | Planmeca | Intra | IXRF052395 | 70 | 8 | 3 | Planmeca | D-0711SB | ||||||||||||||||
| X-ray Machine Control Panel | kVp Meter | 5 | Camp Grayling, MI | B8923 | RADCAL | 9010 | 90-1776 | 16-Jul-10 | B8923 | RADCAL | 9010 | 90-1776 | 16-Jul-10 | B8923 | RADCAL | 9010 | 90-1776 | 16-Jul-10 | 6083 | Fluke | 990 | 548 | 9-Mar-10 | 1726 | Planmeca | Intra | IXRF051968 | Jun-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 79082 | Jun-04 | ||||||||||||||
| Calibration Due Date: | Apr-17 | ECN: | 26585 | 6 | 2145 | UNFORS | Mult-O-Meter | 129595 | 1-Apr-10 | 2145 | UNFORS | Mult-O-Meter | 129595 | 1-Apr-10 | 2145 | UNFORS | Mult-O-Meter | 129595 | 1-Apr-10 | 6084 | Fluke | 990 | 542 | 19-Jun-10 | 1727 | Planmeca | Intra | IXRF051718 | Jul-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 77858 | Jul-04 | |||||||||||||
| ECN | Camp Grayling, MI | Manufacturer: | UNFORS | 7 | 6886 | UNFORS | Xi | 142847 | 10-Feb-15 | 6886 | UNFORS | Xi | 142847 | 10-Feb-10 | 6886 | UNFORS | Xi | 142847 | 10-Feb-10 | 6138 | Fluke | 451B-RYR | 1039 | 26-Jan-15 | 1728 | Planmeca | Intra | IXRF051973 | Jun-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 79792 | Jun-04 | |||||||||||||
| Manufacturer | Del Medical Imaging Corp | Model: | X2 | 8 | 26585 | UNFORS | X2 | 212716 | 18-Jun-16 | 26585 | UNFORS | X2 | 212716 | 18-Jun-16 | 26585 | UNFORS | X2 | 212716 | 18-Jun-16 | 6139 | Fluke | 451B-RYR | 1038 | 5-Sep-09 | 1729 | Planmeca | Intra | IXRF051974 | 70 | 8 | 3 | Planmeca | D-0711SB | ||||||||||||||||
| Model Number | Infinity 400 | Serial Number: | 225565 | 9 | 0B2701 | Bicron | Microrem | B420R | 1-May-10 | 1730 | Planmeca | Intra | IXRF051675 | 70 | 8 | 3 | Planmeca | D-0711SB | |||||||||||||||||||||||||||||||
| Serial Number | 04126-1004 | Calibration Due Date: | 3-Sep-17 | 10 | 0B2703 | Bicron | RSO-5 | B293M | 24-Aug-10 | 1731 | Planmeca | Intra | IXRF051656 | Jul-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 78888 | May-04 | ||||||||||||||||||||||||||||
| Date Manufactured | Oct-04 | Timer | 11 | 0B2706 | Bicron | 2000 | B271R | 26-Aug-10 | 1732 | Planmeca | Intra | IXRF051655 | 70 | 8 | 3 | Planmeca | D-0711SB | ||||||||||||||||||||||||||||||||
| kVp (max) | 125 | ECN: | 26585 | 12 | 0B2707 | Bicron | 2000 | B272R | 17-Jun-10 | 1733 | Planmeca | Intra | IXRF051660 | Jul-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 79569 | Jun-04 | ||||||||||||||||||||||||||||
| mA or mAs (max) | 3.3 | Manufacturer: | UNFORS | 13 | 0B3068 | Bicron | Micro Analyst | B464R | 27-Aug-10 | 1734 | Planmeca | Intra | IXRF051658 | Jul-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 76781 | May-04 | ||||||||||||||||||||||||||||
| Phase | 1 | Model: | X2 | 14 | 0B3070 | Bicron | RSO-50E | B249Q | 1-May-10 | 1735 | Planmeca | Intra | IXRF051659 | 70 | 8 | 3 | Planmeca | D-0711SB | |||||||||||||||||||||||||||||||
| Tube Insert | Serial Number: | 225565 | 15 | 0B3398 | Bicron | Micro Analyst | B603R | 30-Apr-10 | 1736 | Planmeca | Intra | IXRF052382 | Aug-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 81837 | Sep-04 | |||||||||||||||||||||||||||||
| Manufacturer | Eureka | Calibration Due Date: | 3-Sep-17 | 16 | 0B6895 | Bicron | 2000 | B673Z | 26-Aug-10 | 1737 | Planmeca | Intra | IXRF051971 | May-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 76759 | May-04 | ||||||||||||||||||||||||||||
| Model Number | UX-3 | Ion Chamber | 17 | 1738 | Planmeca | Intra | IXRF050009 | Apr-04 | 70 | 8 | 3 | Planmeca | D-0711SB | 1584 | Sep-04 | ||||||||||||||||||||||||||||||||||
| Serial Number | 4J119 | ECN: | 26585 | 18 | 1739 | Planmeca | Promax | RPX234734 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | Oct-04 | ||||||||||||||||||||||||||||||||||
| Date Manufactured | Sep-04 | Manufacturer: | UNFORS | 19 | 1740 | Planmeca | Promax | RPX234567 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | 3J07069 | Oct-04 | |||||||||||||||||||||||||||||||||
| Large FS (mm) | 2 | Model: | X2 | 20 | 1741 | Planmeca | Promax | RPX234566 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | 3J07705 | Oct-04 | |||||||||||||||||||||||||||||||||
| Small FS (mm) | 1 | Serial Number: | 225565 | 21 | 1743 | Planmeca | Promax | RPX234570 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | 3L08009 | Oct-04 | |||||||||||||||||||||||||||||||||
| Tube Housing | Calibration Due Date: | 3-Sep-17 | 22 | 1744 | Planmeca | Promax | RPX234569 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | 3J07769 | Oct-04 | ||||||||||||||||||||||||||||||||||
| Manufacturer | Eureka | Light Meter | 23 | 1745 | Planmeca | Promax | RPX234572 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | 3D07088 | Oct-04 | ||||||||||||||||||||||||||||||||||
| Model Number | UX-52H-39 | ECN: | 21039 | 24 | 1746 | Planmeca | Promax | RPX234727 | Oct-04 | 84 | 16 | 3 | Planmeca | D-054SB2 | Oct-04 | ||||||||||||||||||||||||||||||||||
| Serial Number | 4J350 | Manufacturer: | EXTECH | 25 | 1886 | Liebel-Flarsheim | Fluoro | SP10516-0540 | |||||||||||||||||||||||||||||||||||||||||
| Image Receptor | Serial Number: | Q619323 | 26 | 1923 | Quantum | QG-4000 GP | 5C722 | 125 | 3 | Quantum | E7242 | 5C0414 | Mar-05 | ||||||||||||||||||||||||||||||||||||
| Manufacturer | Calibration Due Date: | 12-Sep-17 | 27 | 1977 | Planmeca | Intra | IXRF053564 | Nov-04 | 70 | 8 | 3 | Planmeca | D-0711SB | Sep-04 | |||||||||||||||||||||||||||||||||||
| Model Number | Focal Spot Test Tool | 28 | 2998 | GE | Precision 500D GP | BRDWNV | Dec-03 | 150 | 400 | 3 | GE | 2336058 | 30238BI0 | May-05 | |||||||||||||||||||||||||||||||||||
| Serial Number | Manufacturer: | 29 | 2998 | GE | Precision 500D Fluoro | BRDNB | Dec-03 | 120 | GE | 2289148 | 91072293 | ||||||||||||||||||||||||||||||||||||||
| Imaging Area (in) | Model: | 30 | 3873 | Planmeca | Intra | IXRF061143 | Mar-06 | 70 | 8 | 3 | Planmeca | D-0741B | 98628 | Mar-06 | |||||||||||||||||||||||||||||||||||
| Display Device | Serial Number: | 31 | 3874 | Planmeca | Promax | RPX235953 | Mar-06 | 84 | 16 | 3 | Planmeca | D-0545B2 | Jan-06 | ||||||||||||||||||||||||||||||||||||
| Processor | Calibration Due Date: | 32 | 4149 | GE | Proteus XR/A | 41986461 | Mar-06 | 150 | 500 | 3 | GE | 46-155318G33 | 131672TX8 | Aug-12 | |||||||||||||||||||||||||||||||||||
| Manufacturer | Other Phantoms or Meters | 33 | 4429 | Planmeca | Promax | RPX 236052 | Apr-06 | 84 | 16 | 3 | Planmeca | D-0545B2 | 4L09066 | Apr-06 | |||||||||||||||||||||||||||||||||||
| Model Number | Manufacturer: | 34 | 4430 | Planmeca | Intra | IXRF062506 | Jun-06 | 70 | 8 | 3 | Planmeca | D-0711SB | 1392 | Jun-06 | |||||||||||||||||||||||||||||||||||
| Serial Number | Model: | 35 | 4431 | Planmeca | Intra | IXRF062507 | May-06 | 70 | 8 | 3 | Planmeca | D-0711SB | O1721 | May-06 | |||||||||||||||||||||||||||||||||||
| Film | Serial Number: | 36 | 4433 | Planmeca | Promax | RPX 235203 | May-05 | 84 | 16 | 3 | Planmeca | D-0545B2 | 4H08457 | May-05 | |||||||||||||||||||||||||||||||||||
| Manufacturer | Calibration Due Date: | 37 | 7594 | Planmeca | Promax | RPX 236914 | Nov-06 | 84 | 16 | 3 | Planmeca | D-0545B2 | 6C03717 | Nov-06 | |||||||||||||||||||||||||||||||||||
| Model Number | 38 | 7595 | Planmeca | Intra | IXRF 066621 | Dec-06 | 70 | 8 | 3 | Planmeca | D-0711SB | 11151 | Dec-06 | ||||||||||||||||||||||||||||||||||||
| Cassettes | 39 | 7689 | Hologic Fluoroscan | Insight 2 Mini C-Arm | 09-1207-28 | Dec-07 | 75 | 0.1 | 3 | Flouroscan | ASY-00871 | 210509 | Dec-07 | ||||||||||||||||||||||||||||||||||||
| Manufacturer | 40 | 7711 | Planmeca | Promax | RPX 237248 | Mar-03 | 84 | 16 | 3 | Planmeca | D-0545B2 | 6DD3834 | Mar-03 | ||||||||||||||||||||||||||||||||||||
| Model Number | 41 | 8744 | GE | AMX-4 Portable GP | 195490021-004 | Mar-08 | 125 | 320 | 3 | GE | 2208208 | 112106TX0 | Mar-08 | ||||||||||||||||||||||||||||||||||||
| 42 | 8745 | GE | AMX-4 Portable GP | 195490020-001 | Mar-08 | 125 | 320 | 3 | GE | 2208208 | 12341TX3 | Mar-08 | |||||||||||||||||||||||||||||||||||||
| 43 | 9733 | MinXray | HF100+ Mobile GP Vet | 33197 | Jul-08 | 120 | 70 | 3 | MinXray | RHF-1A | 33197 | Jul-08 | |||||||||||||||||||||||||||||||||||||
| 44 | 9734 | MinXray | HF100+ Mobile GP Vet | 33203 | Jul-08 | 120 | 70 | 3 | MinXray | RHF-1A | 33203 | Jul-08 | |||||||||||||||||||||||||||||||||||||
| Format for the spreadsheet | 45 | 10017 | Innovet | Select Vet GP | 8G0382 | Jul-08 | 125 | 3 | Innovet | E7239 | 8G0382 | Jul-08 | |||||||||||||||||||||||||||||||||||||
| Title Bar | Versions: | MAJ Arthur Morton |
MORTONAR: For more information please contact:
Major Arthur Ray Morton Chief, Health Physics Section arthur.morton@amedd.army.mil
Tripler Army Medical Center ATTN: MCHK-PVR (MAJ Morton) 1 Jarrett White Road Honolulu, HI 96859-5000
Phone (808) 433-2334 DSN 433-2334
FAX (808) 433-2440 DSN 433-2440
MORTONAR: Sort all remaining data to the right by MMCN
| MORTONAR: Recommend listing all variables to be used in the drop down lists and then sort the variables to the order you want them to appear in the drop down list | 46 | 10046 | Planmeca | Intra | IXRF075280 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 31395 | May-08 | Planmeca | Intra | ITHa68280 | ||||||||||||||||||||
| Background (Text) | 12-Nov-02 | New TB MED 521 | 47 | 10047 | Planmeca | Intra | IXRF075320 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 31158 | May-08 | Planmeca | Intra | ITHa68364 | ||||||||||||||||||
| Data Entry | 12-Dec-02 | CF for light meters | 48 | 10048 | Planmeca | Intra | IXRF075274 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 33400 | May-08 | Planmeca | Intra | ITHa68236 | ||||||||||||||||||
| Tech. & Cond. Titles, Background | 9-Jun-03 | Added Macro Buttons | 49 | 10049 | Planmeca | Intra | IXRF075328 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 33158 | May-08 | Planmeca | Intra | ITHa68428 | ||||||||||||||||||
| Tech. & Conditions Data, Notes | 14-Jul-03 | Many small updates | 50 | 10050 | Planmeca | Intra | IXRF075281 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 30530 | May-08 | Planmeca | Intra | ITHa68267 | ||||||||||||||||||
| Calculations | 30-Jul-03 | Added RPP Checklist | 51 | 10051 | Planmeca | Intra | IXRF075307 | May-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 32879 | May-08 | Planmeca | Intra | ITHa68351 | ||||||||||||||||||
| PASS, FAILED Box (Empty) | 6-Aug-03 | Fixed Linearity Reference | 52 | 10052 | Planmeca | Intra | IXRF075277 | Jun-08 | 70 | 8 | 3 | Planmeca | D-0711SB | 33830 | Jun-08 | Planmeca | Intra | ITHa68975 | ||||||||||||||||||
| PASS, FAILED Box (PASSED) | PASSED | 2-Sep-03 | AEC Reference 4-3c | 53 | 12079 | Planmeca | Intra | IXRF 078333 | Jan-09 | 70 | 8 | 3 | Planmeca | D-0711SB | 36528 | Dec-08 | Planmeca | Intra | ITHa 71570 | |||||||||||||||||
| PASS, FAILED Box (SHOULD) | SHOULD | 22-Sep-03 | Added data cells for AEC maximum allowable techniques. | 54 | 12081 | Planmeca | Intra | IXRF 078097 | Jan-09 | 70 | 8 | 3 | Planmeca | D-0711SB | 36479 | Dec-08 | Planmeca | Intra | ITHa 71547 | |||||||||||||||||
| PASS, FAILED Box (FAILED) | FAILED | 6-Oct-03 | Modified kVp timer page | 55 | 12240 | Planmeca | Promax | RPX 238272 | Sep-08 | 84 | 16 | 3 | Planmeca | D-0545B2 | 7k08703 | Aug-08 | Planmeca | D-0545132 | RPP0706038 | |||||||||||||||||
| References | 56 | 12441 | Planmeca | ProMax | RPX 238209 | Jul-08 | 84 | 16 | 3 | Planmeca | D-0545B2 | 7G08357 | Aug-08 | Planmeca | Promax | RPP0705931 | ||||||||||||||||||||
| 57 | A0168 | Siemens | MultiPro Multix | S09110501L | Jan-01 | 150 | 800 | 3 | Siemens | 3345233 V2146 | 373500 S20 | Jan-01 | 1 | 0.6 | Siemens | 3345209 X1953 | 3785 S20L | |||||||||||||||||||
| 58 | A0633 | GE | 9800 C-Arm | 89-1265 | Dec-01 | 120 | GE | OEC 9800 | 89-1265 | Dec-01 | ||||||||||||||||||||||||||
| 59 | B2234 | GE | Advantix GP | 421873WK1 | Feb-94 | 150 | 640 | 3 | GE | 46-155318633 | 303060TV8 | Nov-93 | 1.25 | 0.6 | GE | 46-15500648 | 76438EC1 | |||||||||||||||||||
| 60 | B2234 | GE | Advantix Fluoro | 421873WK1 | Feb-94 | GE | 2112621 | 51417WK7 | Apr-97 | GE | 46-275382646 | 421549WK7 | 12 | |||||||||||||||||||||||
| 61 | B7465 | Picker | MTX Chest | 1451927 | Jun-95 | 125 | 640 | 3 | Picker | DU-404 | 5111W | Feb-97 | 1.2 | 0.6 | Picker | PX-1436PQ | 516374BB | |||||||||||||||||||
| 62 | B9870 | Siemens | Multix 65 GP | S08429S01L | Jun-00 | 150 | 500 | 3 | Siemens | 3345233 | 255640520 | Jun-00 | 1 | 0.6 | Siemens | 3345209X1953 | 3264520L | |||||||||||||||||||
| 63 | CA2022 | Planmeca | Intra | IXRF047552 | Dec-03 | 70 | 8 | 3 | Planmeca | Planmeca | ||||||||||||||||||||||||||
| 64 | CA2023 | Planmeca | Intra | IXRF047555 | Dec-03 | 70 | 8 | 3 | Planmeca | Planmeca | ||||||||||||||||||||||||||
| 65 | D9637 | GE | MVP 60 GP | 21232ES4 | Feb-91 | 150 | 640 | 3 | Varian | MX100-18 | H4151=25 | Aug-09 | 1.25 | 0.6 | GE | Maxi 100 18NR | 29846 A2 | |||||||||||||||||||
| 66 | E4417 | GE | Advantix GP | 65353VP5 | Feb-01 | 150 | 1000 | 3 | GE | 46-155318633 | 29112TX0 | Sep-97 | 1.25 | 0.6 | GE | 46-155400648 | 30720EX7 | |||||||||||||||||||
| 67 | E4417 | GE | Advantix Fluoro | 521427WK5 | Feb-01 | GE | 46-275382652 | 521427WK5 | 12 | |||||||||||||||||||||||||||
| 68 | ING | Planmeca | Intra | IXRF047551 | 70 | 8 | 3 | Planmeca | Planmeca | |||||||||||||||||||||||||||
| 69 | ING | Planmeca | Proline | RCC014117 | 84 | 16 | 3 | Planmeca | Planmeca | |||||||||||||||||||||||||||
| 70 | ING | Planmeca | Proline | RCC014080 | 84 | 16 | 3 | Planmeca | Planmeca | |||||||||||||||||||||||||||
| 71 | New | Planmeca | ProMax | RPX 238165 | Jun-08 | 84 | 16 | 3 | Planmeca | D-0545B2 | 7G08280 | Jun-08 | Planmeca | Promax | RPP0705862 | |||||||||||||||||||||
| 72 | new install | Planmeca | Intra | IXRF 077335 | Jan-09 | 70 | 8 | 3 | Planmeca | D-0711SB | 73270 | Mar-05 | Planmeca | Intra | ITHa 46805 | |||||||||||||||||||||
| 73 | ||||||||||||||||||||||||||||||||||||
| 74 | ||||||||||||||||||||||||||||||||||||
| 75 | ||||||||||||||||||||||||||||||||||||
| 76 | ||||||||||||||||||||||||||||||||||||
| 77 | ||||||||||||||||||||||||||||||||||||
| 78 | ||||||||||||||||||||||||||||||||||||
| 79 | ||||||||||||||||||||||||||||||||||||
| 80 | ||||||||||||||||||||||||||||||||||||
| 81 | ||||||||||||||||||||||||||||||||||||
| 82 | ||||||||||||||||||||||||||||||||||||
| 83 | ||||||||||||||||||||||||||||||||||||
| 84 | ||||||||||||||||||||||||||||||||||||
| 85 | ||||||||||||||||||||||||||||||||||||
| 86 | ||||||||||||||||||||||||||||||||||||
| 87 | ||||||||||||||||||||||||||||||||||||
| 88 | ||||||||||||||||||||||||||||||||||||
| 89 | ||||||||||||||||||||||||||||||||||||
| 90 | ||||||||||||||||||||||||||||||||||||
| 91 | ||||||||||||||||||||||||||||||||||||
| 92 | ||||||||||||||||||||||||||||||||||||
| 93 | ||||||||||||||||||||||||||||||||||||
| 94 | ||||||||||||||||||||||||||||||||||||
| 95 | ||||||||||||||||||||||||||||||||||||
| 96 | ||||||||||||||||||||||||||||||||||||
| 97 | ||||||||||||||||||||||||||||||||||||
| 98 | ||||||||||||||||||||||||||||||||||||
| 99 | ||||||||||||||||||||||||||||||||||||
| 100 | ||||||||||||||||||||||||||||||||||||
| 101 | ||||||||||||||||||||||||||||||||||||
| 102 | ||||||||||||||||||||||||||||||||||||
| 103 | ||||||||||||||||||||||||||||||||||||
| 104 | ||||||||||||||||||||||||||||||||||||
| 105 | ||||||||||||||||||||||||||||||||||||
| 106 | ||||||||||||||||||||||||||||||||||||
| 107 | ||||||||||||||||||||||||||||||||||||
| 108 | ||||||||||||||||||||||||||||||||||||
| 109 | ||||||||||||||||||||||||||||||||||||
| 110 | ||||||||||||||||||||||||||||||||||||
| 111 | ||||||||||||||||||||||||||||||||||||
| 112 | ||||||||||||||||||||||||||||||||||||
| 113 | ||||||||||||||||||||||||||||||||||||
| 114 | ||||||||||||||||||||||||||||||||||||
| 115 | ||||||||||||||||||||||||||||||||||||
| 116 | ||||||||||||||||||||||||||||||||||||
| 117 | ||||||||||||||||||||||||||||||||||||
| 118 | ||||||||||||||||||||||||||||||||||||
| 119 | ||||||||||||||||||||||||||||||||||||
| 120 | ||||||||||||||||||||||||||||||||||||
| 121 | ||||||||||||||||||||||||||||||||||||
| 122 | ||||||||||||||||||||||||||||||||||||
| 123 | ||||||||||||||||||||||||||||||||||||
| 124 | ||||||||||||||||||||||||||||||||||||
| 125 | ||||||||||||||||||||||||||||||||||||
| 126 | ||||||||||||||||||||||||||||||||||||
| 127 | ||||||||||||||||||||||||||||||||||||
| 128 | ||||||||||||||||||||||||||||||||||||
| 129 | ||||||||||||||||||||||||||||||||||||
| 130 | ||||||||||||||||||||||||||||||||||||
| 131 | ||||||||||||||||||||||||||||||||||||
| 132 | ||||||||||||||||||||||||||||||||||||
| 133 | ||||||||||||||||||||||||||||||||||||
| 134 | ||||||||||||||||||||||||||||||||||||
| 135 | ||||||||||||||||||||||||||||||||||||
| 136 | ||||||||||||||||||||||||||||||||||||
| 137 | ||||||||||||||||||||||||||||||||||||
| 138 | ||||||||||||||||||||||||||||||||||||
| 139 |
Clear Worksheet
Equip Checks
| X-RAY SYSTEM EVALUATION CHECKLIST | Selection | YES | YES | YES | YES | YES | YES | YES | N / A | YES | YES | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | N / A | ||||
| Radiographic Systems | Reference | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 3 | 1 | 1 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | ||||
| Label Conforms to 21 CFR | |||||||||||||||||||||||||||||||||||||||||||||||||
| MORTONAR: Affix a permanent label or tag on all new x-ray systems and components indicating that the system or component conforms to the requirements of 21 CFR, Chapter 1, Subchapter J. | TB MED 521, 4-2a | Heading | Answer | ||||||||||||||||||||||||||||||||||||||||||||||
| Warning Statement Visible |
MORTONAR: Ensure that the following warning statement is displayed on the control panel containing the main power switch and is legible and accessible to view. "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
| TB MED 521, 4-2b | Variables | 1 | YES | |
| Anode, Cathode and Focal Spot marked | ||||
| MORTONAR: Permanently mark the cathode and anode receptacles of the x-ray tube and the location of the apparent focal spot. | TB MED 521, 4-2c | 2 | NO | |
| Technique factors visible before exposure | ||||
| MORTONAR: The technique factors to be used during an exposure shall be indicated before the exposure begins. If an automatic exposure control (AEC) is used, the technique factors that are set prior to the exposure will be indicated. For equipment with fixed technique factors, this requirement may be met by permanent markings. The indicated technique factors shall be visible from the operator's location, except in the case of spot films made by a fluoroscopist. | TB MED 521, 4-3a | 3 | N / A | |
| Means to terminate if > 0.5s | ||||
| MORTONAR: Provide means to terminate the exposure after a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. Except during serial radiography, the operator will be able to terminate the exposure at any time during an exposure of greater than 1/2 second. Termination of exposure will cause the timer to automatically reset to its initial setting or to zero. It will not be possible to make an exposure when the timer is set to zero or an off position, if either position is provided. During serial radiography, the operator will be able to terminate the exposures at any time, but the x-ray system may have a means to permit completion of any single exposure in the series. | TB MED 521, 4-3b | 4 | Comment | |
| AEC mode indication on control panel |
MORTONAR: For x-ray systems with AEC, indication will be made on the control panel when this mode of operation is selected. The following requirements will also be met.
(1) When the peak tube potential is 51 kVp or greater, the minimum exposure time for--
(a) Field emission equipment rated for pulsed operation will be less than or equal to the time interval equivalent to two pulses.
(b) All other equipment will be less than or equal to 1/60 second or the time interval required to deliver 5 mAs, whichever is greater.
(2) Either limit the product of peak tube potential, tube current, and exposure time to not more than 60 kilowatt-seconds per exposure or limit the product of tube current and exposure time to not more than 600 mAs per exposure. If the x-ray tube potential is less than 51 kVp, limit the product of x-ray tube current and exposure time to not more than 2,000 mAs per exposure.
| (3) A visible signal will indicate when an exposure has been terminated at the limits described in (2) above. The operator will have to manually reset the AEC before further automatically timed exposures can be made. | TB MED 521, 4-3c | 5 | Not Visible |
| Visual and audible indication of x-rays | |||
| MORTONAR: The x-ray control will provide visual indication whenever x-rays are produced. In addition, a signal audible to the operator will indicate that the exposure has terminated. | TB MED 521, 4-3s | 6 | Unknown |
| Multiple tubes with 1 exposure switch | |||
| MORTONAR: Where two or more radiographic tubes are controlled by one exposure switch, the selected tube(s) will be clearly indicated before initiation of the exposure. This indication will be on both the x-ray control and at or near the selected tube housing assembly. | TB MED 521, 4-3t | 7 | |
| Exposure switch > 2 m from x-ray tube |
MORTONAR: The exposure switch will be--
(1) A continuous pressure "dead man" type switch.
| (2) Located in such a manner as to prevent convenient use by the operator outside the protective enclosure (control booth). | TB MED 521, 4-3w | 8 |
| Steady mechanical support of tube head | ||
| MORTONAR: Mechanical support of the tube head and beamlimiting device will maintain its proper exposure position without drift, sag, or vibration. | TB MED 521, 4-3x | |
| Telemetric exposure switch |
MORTONAR: Telemetric exposure switches controlling the tube circuit by wireless means will be used with caution. When using two or more telemetric timers within the same facility--
(1) Operate all timers on different frequencies.
| (2) Label or control each telemetric switch to assure that they do not become interchanged. | TB MED 521, 4-3y |
| Mobile and Portable Radiographic X-ray Systems | |
| Stand at least 2 m from tube | |
| MORTONAR: Stand as far as practical (at least 2 meters (m)) from the patient, the x-ray tube, and the primary beam. | TB MED 521, 3-5a(3) |
| Safeguard enabling key | |
| MORTONAR: Make provisions to safeguard the enabling key for mobile x-ray systems. Leave the key in the system only if necessary during recharging or for a quick response system (see para 3-3). | TB MED 521, 3-5b(2) |
| Exposure switch > 2 m from x-ray tube | |
| MORTONAR: Arrange the exposure switch so that the operator can stand at least 2 m from the x-ray tube, the primary beam, and the patient. | TB MED 521, 4-4a |
| Battery charge indication | |
| MORTONAR: Visual means will be provided on the x-ray control panel of the battery powered systems to indicate if the battery is charged and is adequate for proper operation. It will not be possible to produce x-rays unless there is an adequate charge for proper operation. | TB MED 521, 4-4b |
| Enabling key | |
| MORTONAR: An enabling key will be required to activate the system. To prevent unauthorized use, remove the key during system storage, except during battery recharging. | TB MED 521, 4-4c |
| Minimum SSD > 30 cm | |
| MORTONAR: Provide positive means to ensure that the minimum SSD is no less than 30 cm. | TB MED 521, 4-4d |
| Tomographic X-ray Systems |
MORTONAR: The technique factors to be used during an exposure shall be indicated before the exposure begins. If an automatic exposure control (AEC) is used, the technique factors that are set prior to the exposure will be indicated. For equipment with fixed technique factors, this requirement may be met by permanent markings. The indicated technique factors shall be visible from the operator's location, except in the case of spot films made by a fluoroscopist.
| Tomographic cut level | |
| MORTONAR: The error in the cut level indication will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 mm. | TB MED 521, 4-5a |
| Tomographic angle |
MORTONAR: The difference between the indicated tomographic angle and the measured angle will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 degrees. The symmetry of the exposure angle about the midline will not exceed the manufacturer’s specifications or, in the absence of such specifications will not exceed
| 2 degrees, unless the system is designed to operate in an asymmetrical manner. Since the cut thickness is dependent on exposure angle, the thickness of the cut plane should decrease as the exposure angle increases. | TB MED 521, 4-5b |
| Tomographic plane |
MORTONAR: The tomographic plane should be flat to within manufacturer’s specifications or, in the absence of such specifications, 3 mm for linear systems and 2 mm for other systems.
| TB MED 521, 4-5c | |
| Tomographic uniformity | |
| MORTONAR: Tomographic x-ray systems will produce a uniform density along the beam path when an image of a tomographic uniformity test phantom is produced. Linear tomographic images should consist of an evenly darkened straight line without evidence of mechanical or radiation caused irregularities. Tomographic x-ray systems with circular, elliptical, hypocycloidal, or trispiral scan capability will produce an evenly darkened closed path. The path overlap will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 20 degrees. | TB MED 521, 4-5d |
| Tomographic resolution | |
| MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern. | TB MED 521, 4-5e |
| Fluoroscopic X-ray Systems | |
| Monitor kVp and mA | |
| MORTONAR: In x-ray systems with automatic brightness control, the peak tube potential (kVp) and tube current (mA) may rise to high values without the knowledge of the operator, particularly if the gain of the intensifier diminishes. The fluoroscopist should be able to monitor the kVp and mA from his or her operating position. Systems may have meters located at the fluoroscopists operating position to assist in this monitoring. | TB MED 521, 3-6e |
| Electronic Image Hold | |
| MORTONAR: Fluoroscopic x-ray systems used for medical procedures will have image intensifiers. If nonradiology specialists use the system, it will have image intensifiers with electronic image-holding capabilities. | TB MED 521, 3-6g |
| Primary beam intercepted by barrier | |
| MORTONAR: The fluoroscopic imaging assembly will have a primary protective barrier which intercepts the entire cross section of the primary beam at any SID. The x-ray tube used for fluoroscopy will not produce x-rays unless the barrier is in position to intercept the entire primary beam. The exposure rate due to transmission through the barrier with the attenuation block in the primary beam combined with radiation from the image intensifier if provided, will not exceed 3.34 x l0-3 percent of the EER, at a distance of 10 cm from any accessible surface of the fluoroscopic imaging assembly beyond the plane of the image receptor. | TB MED 521, 4-7a |
| Terminating exposures | |
| MORTONAR: X-ray production in the fluoroscopic mode will be controlled by a device that requires continuous pressure by the operator for the entire time of any exposure. When recording serial fluoroscopic images, the operator will be able to terminate the x-ray exposures at any time, but means may be provided to permit completion of any single exposure of the series in process. | TB MED 521, 4-7d |
| Indication of kVp and mA | |
| MORTONAR: During fluoroscopy and cinefluorography, x-ray tube potential and current will be continuously indicated. Deviation of x-ray tube potential and current from the indicated values will not exceed the maximum deviation stated by the manufacturer. | TB MED 521, 4-7f |
| Minimum SSD | |
| MORTONAR: The equipment will have the means to limit the SSD to not less than 38 cm on stationary fluoroscopic systems and to not less than 30 cm on mobile and portable fluoroscopic systems. In addition, for image-intensified systems intended for specific surgical application that would be prohibited at the SSDs specified in this paragraph, provisions may be made for operation at shorter SSDs, but in no case less than 20 cm. When provided, the manufacturer must set forth precautions with respect to the optional means of spacing. | TB MED 521, 4-7g |
| 5 minute cumulative timing device |
MORTONAR: The equipment will have the means to preset the cumulative on-time of the fluoroscopic tube. The timing device’s maximum cumulative time will not exceed 5 minutes without resetting. A signal, audible to the fluoroscopist, will indicate the completion of any preset cumulative on-time. While x-rays are being produced, this signal will continue to sound until the timing device is reset. Radiation therapy simulation systems may alternatively have a timer that--
(1) Indicates the total cumulative exposure time during which x-rays were produced.
(2) Is capable of being reset between x-ray examinations.
| TB MED 521, 4-7h | |
| Image Intensifier | |
| MORTONAR: In addition to the above requirements, mobile and portable fluoroscopic systems will have image intensifiers. | TB MED 521, 4-7i |
| X-ray Source Assembly | |
| MORTONAR: Fluoroscopic x-ray systems will have a diagnostic-type protective source assembly. The apparent focal spot will be clearly indicated on the tube housing. The cathode and anode cable receptacles of the x-ray tube will be permanently identified. | TB MED 521, 4-7j |
| Under Table shielding for bucky slot | |
| MORTONAR: Fluoroscopic systems with the x-ray tube permanently located under a table will have a shielding device of at least one tenth value layer of material at the maximum tube potential for covering the bucky slot during fluoroscopy. If a shielded bucky slot cover is required to meet the exposure rate requirements of paragraph 4-2e, the operator should use positive means to ensure that the cover is in place before producing x-rays. | TB MED 521, 4-7k |
| Under Table shielding drapes and panels |
MORTONAR: Under table fluoroscopic equipment will have a shielding device(s) of at least 0.25 mm lead equivalent at 100 kVp. Such devices include--
(1) Overlapping protective drapes hung from the image receptor, and
(2) Hinged or sliding panels at the front of the table. When provided, a hinged or sliding panel should project at least 13 cm above the tabletop and will not interfere with the use of the table.
| TB MED 521, 4-7l | |
| Shockproof cables | |
| MORTONAR: Each high-voltage generator assembly will have a pair of shockproof cable receptacles, per Federal Standard No. 72, for each x-ray tube assembly that the high-voltage generator is to energize. The high-voltage generator will be permanently marked "Equipped with Federal Standard Shockproof Cable Receptacles." The receptacles will be covered when not in use. | TB MED 521, 4-7m |
| Anode Heat Calculator | |
| MORTONAR: Cardiac x-ray systems with a tube current of 600 mA or greater that are placed in service after 1985 will have an anode heat calculator device or system. For x-ray systems installed prior to 1985, the use of a heat calculator may be beneficial, but it may not be available from the manufacturer. For such systems, the heat calculator is not required. | TB MED 521, 4-8a |
| Tube Protective Circuit | |
| MORTONAR: Each cardiac x-ray system will have a tube protective circuit. | TB MED 521, 4-8b |
| Line voltage compensating device | |
| MORTONAR: A line voltage compensating device (operable from the x-ray control panel without the aid of tools), together with a meter for determining the proper adjustment of the device, will be mounted in all cardiac x-ray control panels not having automatic voltage compensation or a prereading kVp meter that responds directly to line voltage variations. | TB MED 521, 4-8c |
| Dental X-ray Systems | |
| Dental Film "D" or faster | |
| MORTONAR: Perform intraoral image receptor radiography using dental film that meets the requirements of American National Standards Institute (ANSI) speed group rating "D" or faster. | TB MED 521, 3-8a(1) |
| Veterinary X-ray Systems | |
| Veterinary Use Logbook | |
| MORTONAR: Maintain a log or equivalent record on the use of the x-ray systems. Include the date of exposure, kVp, mA, exposure time, operator’s name, identification of the animal, and the name of any person used to hold the animal during the radiographic procedure. | TB MED 521, 3-10i |
| Tomographic X-ray Systems | |
| Tomographic cut level | |
| MORTONAR: The error in the cut level indication will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 mm. | TB MED 521, 4-5a |
| Tomographic angle |
MORTONAR: The difference between the indicated tomographic angle and the measured angle will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 degrees. The symmetry of the exposure angle about the midline will not exceed the manufacturer’s specifications or, in the absence of such specifications will not exceed
| 2 degrees, unless the system is designed to operate in an asymmetrical manner. Since the cut thickness is dependent on exposure angle, the thickness of the cut plane should decrease as the exposure angle increases. | TB MED 521, 4-5b |
| Tomographic plane |
MORTONAR: The tomographic plane should be flat to within manufacturer’s specifications or, in the absence of such specifications, 3 mm for linear systems and 2 mm for other systems.
| TB MED 521, 4-5c | |
| Tomographic uniformity | |
| MORTONAR: Tomographic x-ray systems will produce a uniform density along the beam path when an image of a tomographic uniformity test phantom is produced. Linear tomographic images should consist of an evenly darkened straight line without evidence of mechanical or radiation caused irregularities. Tomographic x-ray systems with circular, elliptical, hypocycloidal, or trispiral scan capability will produce an evenly darkened closed path. The path overlap will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 20 degrees. | TB MED 521, 4-5d |
| Tomographic resolution | |
| MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern. | TB MED 521, 4-5e |
| Comments: |
Leakage-EER
| LEAKAGE RADIATION FROM THE DIAGNOSTIC SOURCE ASSEMBLY | Course Manual for X-Ray Measurements, FDA, BRH July 1977 | |||||||||||||
| Techniques : | 100 | mA | 1000 | ms | 90 | kVp | Victoreen 440 Time Response Correction Factors | |||||||
| Conditions : | 100 | mR/hr @1m | 5% | duty cycle | 0.02 | Bkg (mR/hr) | Seconds | 3 mR/hr | 10 mR/hr | 30 mR/hr | 100 mR/hr | 300 mR/hr | ||
| Meter : | Unfors | Model | X2 | Manuf. | 225565 | SN | 0.3 | 0.215 | 0.245 | 0.255 | ||||
| uR | Distance | mR in 1 hr | Time Resp. | 0.4 | 0.315 | |||||||||
| Start | 5 | mR/hr | in | at 1 meter | C.F. | |||||||||
| MORTONAR: Using the 450P and 1000 ms exposure time, the time response correction factor is about 0.85 | mR 1hr@1m | 0.5 | 0.140 | 0.320 | 0.340 | 0.390 | 0.390 | |||||||
| Leakage Front | 4.9 | 40.0 | 0.24 | 0.85 | 0.29 | 0.8 | 0.235 | 0.465 | 0.575 | 0.620 | 0.575 | |||
| Leakage Back | 1.3 | 34.0 | 0.05 | 0.85 | 0.05 | 1.0 | 0.320 | 0.595 | 0.655 | 0.745 | 0.700 | |||
| Leakage Anode | 1.5 | 40.0 | 0.07 | 0.85 | 0.09 | 1.2 | 0.645 | |||||||
| Leakage Cathode | 0.4 | 40.0 | 0.02 | 0.85 | 0.02 | 1.3 | 0.415 | 0.795 | ||||||
| Scatter Front | 1.5 | 0.430 | 0.760 | 0.845 | 1.000 | 0.890 | ||||||||
| Scatter Back | 1.8 | 0.820 | 0.955 | |||||||||||
| Scatter Anode | 2.0 | 0.555 | 0.835 | 0.990 | 1.050 | 0.990 | ||||||||
| Scatter Cathode | 2.5 | 0.665 | 0.915 | 1.030 | 0.975 | |||||||||
| Max Leakage (not tube) | 3.0 | 0.688 | 0.930 | 1.030 | 1.000 | 1.040 | ||||||||
| 3.5 | 0.745 | 0.965 | 1.000 | 0.990 | 0.990 | |||||||||
| 4.0 | 0.820 | 0.975 | 1.000 | 1.000 | 1.000 | |||||||||
| 4.5 | 1.000 | 1.000 | ||||||||||||
| 5.0 | 0.870 | 1.025 | 1.000 | 1.000 | 1.000 |
| Maximum scatter radiation (mR in 1 hour at 1 meter) | other measurements taken | in mR/hr | |||
| Maximum leakage radiation (mR in 1 hour at 1 meter) | 0.3 | PASSED | at door to control booth | ||
| Maximum leakage radiation other than tube (2 mR in 1 hour at 5 cm) | at door to room | ||||
| at computer inside cb, door open | |||||
| References: | inside control booth,door closed | ||||
| Leakage Radiation |
MORTONAR: Construct diagnostic source assemblies so that leakage radiation exposure at a distance of 1 m in any direction from the source will not exceed 100 mR (2.58 x 10-5 coulombs/kilogram (C/kg)) in 1 hour when the
| x-ray tube is operated at the leakage technique factors. If the tube housing assembly’s maximum rated peak tube potential is greater than the diagnostic source assembly’s maximum rated peak tube potential, provide positive means to limit the maximum x-ray tube potential to that of the diagnostic source assembly. | TB MED 521 | 4-2d |
| Leakage-other than tube |
MORTONAR: Construct components other than the diagnostic source assembly so that the radiation exposure will not exceed 2 mR (5.16 x 10-7 C/kg) in 1 hour at 5 cm from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. TB MED 521 4-2e
| FLOURO ENTRANCE EXPOSURE RATES AND SYSTEM RESOLUTION | ||||||
| Techniques : | kVp | SID (in) | SSD (in) | |||
| Conditions : | 10 | R/min EER | 2 | mR/hr PB | 30 | Mesh |
| Modes | EER | Audible | ||||
| Max. Cont. | Phantom | kVp | mA | R/min | Signal |
MORTONAR: Yes or No MORTONAR: Using the 450P and 1000 ms exposure time, the time response correction factor is about 0.85 MORTONAR: Construct diagnostic source assemblies so that leakage radiation exposure at a distance of 1 m in any direction from the source will not exceed 100 mR (2.58 x 10-5 coulombs/kilogram (C/kg)) in 1 hour when the x-ray tube is operated at the leakage technique factors. If the tube housing assembly’s maximum rated peak tube potential is greater than the diagnostic source assembly’s maximum rated peak tube potential, provide positive means to limit the maximum x-ray tube potential to that of the diagnostic source assembly.
| AEC/ no HLC | Lead | |||
| AEC/HLC | Lead | |||
| Manual | Lead | |||
| Max. Pulsed Fluoro | ||||
| Pulsed 3.75 fps | Lead | |||
| Pulsed 7.5 fps | Lead | |||
| Pulsed 15 fps | Lead | |||
| Pulsed 30 fps | Lead | |||
| Patient Equivalent Phantom | ||||
| Manual | Patient Eq | |||
| AEC | Patient Eq | |||
| Pulsed 3.75 fps | Patient Eq | |||
| Pulsed 7.5 fps | Patient Eq | |||
| Pulsed 15 fps | Patient Eq | |||
| Pulsed 30 fps | Patient Eq | |||
| System Resolution | Pantom | Mesh | Phantom | Resolution |
| IR 6" | ||||
| IR 9" | ||||
| IR 15" | ||||
| IR 16" | ||||
| Tomographic |
| Maximum Entrance Exposure Rate (R/min) | |
| References: | |
| Fluoroscopic EER |
MORTONA: The following EERs apply to fluoroscopic x-ray systems.
(1) Fluoroscopic equipment that is provided with automatic exposure rate control (AERC) will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 roentgens per minute (R/min) (2.58xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional high-level control (HLC) is provided.
(2) Fluoroscopic equipment that is not provided with AERC will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R/min (1.29x10-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional HLC is provided.
(3) Equipment with both an AERC mode and a manual mode will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R/min (2.58xl0-3 C/kg×min) in either mode at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional HLC is provided.
(4) If HLC is provided, the system will not be operable, in either AERC or manual mode, at any combination of tube potential and current that will result in an exposure rate in excess of 5 R/min (1.29xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient, unless the HLC is activated. The system will have a special means to activate the HLC. The HLC will be operable only when the operator provides continuous manual activation. A continuous signal audible to the fluoroscopist will indicate that the HLC is activated.
(5) Fluoroscopic equipment manufactured on and after 19 May 1995 will be equipped with AERC if operable at any combination of tube potential and current that results in an exposure rate greater than 5 R/min (1.29xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient. Provision for manual selection of technique factors may be provided.
(a) This equipment will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R/min (2.58x10-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during the recording of images from an x-ray image-intensifier tube using photographic film or a video camera when the x-ray source is operated in a pulsed mode or when an optional HLC is activated. Special means will be required to activate the HLC.
(b) When the HLC is activated, the equipment will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 20 R/min (5.16x10-3 C/kg×min) at the point where the center of the primary beam enters the patient.
(c) The HLC will be operable only when the operator provides continuous manual activation. A continuous signal audible to the fluoroscopist will indicate that the HLC is activated.
(6) Fluoroscopic radiation therapy simulation systems are exempt from the requirements set forth in (1) through (5) above concerning EERs.
MORTONAR: Construct components other than the diagnostic source assembly so that the radiation exposure will not exceed 2 mR (5.16 x 10-7 C/kg) in 1 hour at 5 cm from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. TB MED 521 4-7e EER Notes MORTONA: If the x-ray tube is under the table, position the ion chamber 1 cm above the table top, otherwise measure the exposure rate 30 cm from the input surface of the image receptor. In manual mode, set the contols to 90 kVp and 2 mA. Center the attenuation block over the ion chamber, and make an exposure and collimate the beam the dimensions of the attenuation block and verify the ion chamber is centered in the primary beam. Repeat measurement with 1/8 inch of lead over the attenuation block. Test high level control and pulsed modes (note the presence of an audible signal).
Tomo resolution MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern.
MORTONA: Leakage Notes: To check diagnostic source assembly, close the shutters and add 1/8" of lead to block the primary beam. Make exposure measurements of approximately 1s duration at the maximum rated tube current for the maximum rated kVp. Measure the leakage radiation using a suitable ionization chamber at approximately 1 meter from the tube focal spot in 4 cardinal directions. At these settings, the tube can be operated only a limited number of seconds each minute or minutes per hour. This is the % on time per hour figure above or otherwise known as the duty cycle. Note that for many ion chambers the response time is greater than 1 second and either the exposure time will need to be increased or a time response correction factor applied or the meter is operated in integrated exposure mode. The radiation emitted from a component other than the diagnostic source assembly shall not exceed 2 mR in 1 hour at 5 cm from any accessible surface of the component. An example is the standby radiation from capacitor energy storage equipment without activating the exposure switch.
MORTONA: If the x-ray tube is under the table, position the ion chamber 1 cm above the table top, otherwise measure the exposure rate 30 cm from the input surface of the image receptor. In manual mode, set the contols to 90 kVp and 2 mA. Center the attenuation block over the ion chamber, and make an exposure and collimate the beam the dimensions of the attenuation block and verify the ion chamber is centered in the primary beam. Repeat measurement with 1/8 inch of lead over the attenuation block. Test high level control and pulsed modes (note the presence of an audible signal). TB MED 521 4-5e
RepLin
| REPRODUCIBILITY / LINEARITY | Reference: NCRP 102 | ||||||||
| Techniques: | 1 | Phase | 90 | kVp | 24 | SSD (in) | 24" | 1 Phase | 3 Phase |
| Conditions: | 0.05 | CV | 30% | Output | 0.1 | CL | kVp | mR/mAs | mR/mAs |
| Focal Spot | Large | Large | 40 | 3.0 | 5.1 | ||||
| ms | 1000 | 1000 | 50 | 5.6 | 9.6 | ||||
| mA | 100 | 200 | 60 | 8.7 | 15.0 | ||||
| mAs | 100.0 | 200.0 | 70 | 11.5 | 19.6 | ||||
| Exposure (mR) | 80 | 15.0 | 25.3 | ||||||
| X1 | 809.00 | 1724.00 | 90 | 18.4 | 32.2 | ||||
| X2 | 803.50 | 1720.00 | 100 | 21.9 | 36.8 | ||||
| X3 | 799.90 | 1722.00 | 110 | 25.3 | 42.6 | ||||
| X4 | 793.10 | 1718.00 | 120 | 29.9 | 50.6 | ||||
| X5 | 130 | 35.7 | 61.0 | ||||||
| X6 | 140 | 39.1 | 65.6 | ||||||
| X7 | 150 | 43.7 | 73.6 | ||||||
| X8 | |||||||||
| X9 | Reference: AAPM Report 25 | ||||||||
| X10 | 24" | 1 Phase | 3 Phase | ||||||
| Reproducibility | kVp | mR/mAs | mR/mAs | ||||||
| mean exposure (mR) | 801.38 | 1721.0 | 40 | 3.2 | 5.5 | ||||
| standard deviation | 6.6660 | 2.5820 | 50 | 5.2 | 8.5 | ||||
| coefficient of variation (CV) | 0.0083 | 0.0015 | 60 | 7.6 | 12.0 | ||||
| Reproducibility | PASSED | PASSED | 70 | 10.5 | 16.0 | ||||
| Output | 80 | 13.9 | 20.6 | ||||||
| mean (mR/mAs) | 8.01 | 8.61 | 90 | 17.8 | 25.8 | ||||
| mean (mR/mAs) at 24" | 8.01 | 8.61 | 100 | 22.2 | 31.5 | ||||
| NCRP % output difference | -73.5% | -71.5% | 110 | 27.1 | 37.8 | ||||
| AAPM % output difference | -68.8% | -66.5% | 120 | 32.5 | 44.6 | ||||
| NCRP % output | Should < | Should < | 130 | 38.5 | 51.9 | ||||
| AAPM % output | Should < | Should < | 140 | 45.0 | 59.8 | ||||
| Linearity | 150 | 52.0 | 68.2 | ||||||
| coefficient of linearity (CL) | 0.0356 | ||||||||
| Linearity | PASSED |
| References: | Coefficient of Variation | Linearity |
| Reproducibility |
MORTONA: When the equipment is operated on an adequate power supply per the manufacturer’s specifications, the estimated coefficient of variation of radiation exposures will be no greater than 0.05 for any combination of selected technique factors. That is:
(See Equation 1) where: Xi = the ith exposure measurement _ = the average of the exposure measurements n = the number of measurements taken
| TB MED 521 | 4-3e |
| Linearity |
MORTONA: When the equipment is operated on an adequate power supply per the manufacturer’s specifications and at any fixed tube potential within the range of 40 percent to 100 percent of the maximum, the following requirements will be met.
(1) If the equipment has independent selection of x-ray tube current (mA) and exposure time (s), the average radiation output (mR/mAs) at any two consecutive tube current settings will not differ by more than 0.10 times their sum. That is:
(See equation 2) where: X1 and X2 = the average radiation output values obtained at each of two consecutive tube current settings or at two settings differing by no more than a factor of 2 if the tube current selection is continuous.
(2) If the equipment has selection of tube current-exposure time product (mAs) only, the average radiation output (mR/mAs) at any two consecutive mAs selector settings will not differ by more than 0.10 times their sum. That is:
(See equation 2) where: X1 and X2 = the average radiation output values obtained at each of two consecutive mAs selector settings or at two settings differing by no more than a factor of 2 where the mAs selector provides continuous selection.
TB MED 521 4-3f
Output NCRP
| 1 Phase | mR/mAs | ||||||||||||||||||||||||
| 40 | 50 | 60 | 70 | 80 | 90 | 100 | 110 | 120 | 130 | 140 | 150 | 2.991944764096663 | 5.6386651323360191 | 8.7456846950517839 | 11.507479861910241 | 14.959723820483314 | 18.411967779056386 | 21.864211737629461 | 25.316455696202532 | 29.919447640966627 | 35.673187571921751 | 39.125431530494822 | 43.728423475258921 | 3 Phase | mR/mAs |
| 40 | 50 | 60 | 70 | 80 | 90 | 100 | 110 | 120 | 130 | 140 | 150 | 5.0632911392405067 | 9.5512082853855009 | 14.959723820483314 | 19.562715765247411 | 25.316455696202532 | 32.220943613348673 | 36.823935558112773 | 42.577675489067893 | 50.632911392405063 | 60.989643268124283 | 65.592635212888382 | 73.647871116225545 | kVp |
mR/mAs
Output AAPM
| 1 Phase | mR/mAs | ||||||||||||||||||||||||
| 40 | 50 | 60 | 70 | 80 | 90 | 100 | 110 | 120 | 130 | 140 | 150 | 3.2393212310859112 | 5.1756517281990959 | 7.5900679072592139 | 10.491411456626212 | 13.88727412843309 | 17.784321691829859 | 22.188504694455606 | 27.105203643092448 | 32.539333447395194 | 38.495421170932389 | 44.977665535044807 | 51.98998351114475 | 3 Phase | mR/mAs |
| 40 | 50 | 60 | 70 | 80 | 90 | 100 | 110 | 120 | 130 | 140 | 150 | 5.5320231374499134 | 8.4530417225780088 | 11.952462232022258 | 16.019770141971684 | 20.646240388742896 | 25.824431651559088 | 31.547867224009693 | 37.810820884984629 | 44.608166370297738 | 51.935267602933322 | 59.787896237183311 | 68.162168184576302 | kVp |
mR/mAs kVpTimer
| TUBE POTENTIAL | ESE Reference Table | ||||||||
| Techniques: | 1000 | ms | 24 | SSD (in) | 100 | mA 1st Condition | |||
| MORTONAR: Measurements 1-8 | 24 | SSD (in) | |||||||
| Condition +/-: | 5.0% | Shall | Should | 3.0% | 200 | mA 2nd Condition |
MORTONAR: Measurements 9-16
| Set | Measured | Mult Corr | Absolute | Percent | Set | Measured | Set | Set | Output |
| kVp |
MORTONAR: Table is set up to analyze kVp for two conditions such as focal spot or mA
| kVp | Factor | ||||||||||||||
| MORTONAR: Insert the multiplicative correction factor from the calibration certificate or 1.00 for no correction. | Diff (kVp) | Difference | kVp | kVp | mA | ms | mR | ||||||||
| X1 | 66 | 67.0 | 1.00 | 1.0 | 1.5% | PASSED | X1 | 66.00 | 67.00 | 100 | 1000 | 386.40 | |||
| X2 | 70 | 72.4 | 1.00 | 2.4 | 3.4% | SHOULD | X2 | 70.00 | 72.40 | 100 | 1000 | 463.40 | |||
| X3 | 90 | 93.8 | 1.00 | 3.8 | 4.2% | SHOULD | X3 | 90.00 | 93.80 | 100 | 1000 | 812.80 | |||
| X4 | 110 | 114.1 | 1.00 | 4.1 | 3.7% | SHOULD | X4 | 110.00 | 114.10 | 100 | 1000 | 1235.00 | |||
| X5 | 80 | 82.6 | 1.00 | 2.6 | 3.2% | SHOULD | X5 | 80.00 | 82.60 | 100 | 1000 | 628.00 | |||
| X6 | 57 | 59.4 | 1.00 | 2.4 | 4.2% | SHOULD | X6 | 57.00 | 59.40 | 100 | 1000 | 266.70 | |||
| X7 | 75 | 77.8 | 1.00 | 2.8 | 3.7% | SHOULD | X7 | 75.00 | 77.80 | 100 | 1000 | 544.70 | |||
| X8 | 85 | 87.1 | 1.00 | 2.1 | 2.5% | PASSED | X8 | 85.00 | 87.10 | 100 | 1000 | 705.70 | |||
| X9 | 66 | 67.7 | 1.00 | 1.7 | 2.6% | PASSED | X9 | 66.00 | 67.70 | 200 | 1000 | 829.30 | |||
| X10 | 70 | 72.9 | 1.00 | 2.9 | 4.1% | SHOULD | X10 | 70.00 | 72.90 | 200 | 1000 | 1009.00 | |||
| X11 | 90 | 92.9 | 1.00 | 2.9 | 3.2% | SHOULD | X11 | 90.00 | 92.90 | 200 | 1000 | 1717.00 | |||
| X12 | 110 | 115.6 | 1.00 | 5.6 | 5.1% | FAILED | X12 | 110.00 | 115.60 | 200 | 1000 | 2628.00 | |||
| X13 | 80 | 82.5 | 1.00 | 2.5 | 3.1% | SHOULD | X13 | 80.00 | 82.50 | 200 | 1000 | 1318.00 | |||
| X14 | 57 | 59.9 | 1.00 | 2.9 | 5.1% | FAILED | X14 | 57.00 | 59.90 | 200 | 1000 | 552.20 | |||
| X15 | 75 | 77.0 | 1.00 | 2.0 | 2.7% | PASSED | X15 | 75.00 | 77.00 | 200 | 1000 | 1171.00 | |||
| X16 | 85 | 87.8 | 1.00 | 2.8 | 3.3% | SHOULD | X16 | 85.00 | 87.80 | 200 | 1000 | 1526.00 |
kilovolt peak (kVp) passed all conditions? FAILED
SHIMABUKULN: Paste Here Paste Keithley Triad Data from Radiographic window here
| TIMER | ms | mR | |||||||||
| 100 | mA | 90 | kVp | 24 | SSD (in) | ||||||
| Condition +/-: | 5.0% | Shall | Should | 3.0% | AND | 0 | ms | ||||
| Set | Measured | Subt Corr | Absolute | Percent | |||||||
| ms | ms | Factor (ms) |