Copy_of_GP_Xray_Survey_Ser_04126-1004_Camp_Grayling_27APR2017_(002).xlsm

XLSM spreadsheet 286 KB Posted

Attached to
X Ray Machine
Solicitation number
W912JB-18-Q-0087
Issued by

About this file

Detailed description of equipment needing removal at time of installation.

View the file

Other files

File Type Posted
CDRL_.pdf PDF
IMG_0097.jpg JPG image
IMG_0096.jpg JPG image
CDRL_.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Site-Equip

SITE AND X-RAY MACHINE DATASURVEY EQUIPMENT USEDSelection
Date(s) of Survey27-Apr-2017Exposure MeterChoice
Principal Surveyor
HPOUser: The surveys described in this chapter will be performed by or under the supervision of a qualified expert. MER personnel and/or health physics technicians commonly perform many of the required tests; however, a qualified expert should review the results of these tests and perform any additional tests that are required by this chapter or that are necessary in his or her professional judgment.CPT Rodriguez, Jose H.Manufacturer:26585HeadingSurveyor2nd SurveyorInstallationECNExp. MeterModelSNCal DueECNkVp MeterModelSNCal DueECNTimerModelSNCal DueECNIon ChamberModelSNCal DueECNLight MeterSNCal DueECN
MORTONAR: Sort all remaining data to the right by MMCNManufModelSNDateMax kVpMax mAPhaseManufModelSNDateLg FSSm FSManufModelSNManufModelSNImageDisplayManufModelSNManufModelManufModel
Second SurveyorModel:UNFORSVariables
MORTONAR: Recommend listing all variables to be used in the drop down lists and then sort the variables to the order you want them to appear in the drop down list1CPT RodriguezCPT RodriguezFort Knox, KYB0944Keithley35050A600865-Dec-09B0944Keithley35050A600865-Dec-09B0944Keithley35050A600865-Dec-096073Fluke451P-RYR24106-Aug-1621039ExtechQ6193231-Nov-14869QuantumOdyssey HF GPQG4000-04A-0105Jan-041255003QuantumE72423M026Dec-031.50.6QuantumCM1100CM11395
InstallationCamp Grayling, MIECN:X22NoneCamp Atterbury, INB0962Victoreen4000M71911-Jun-10B0962Victoreen4000M71911-Jun-10B0962Victoreen4000M71911-Jun-106074Fluke451P-RYR235929-Jun-10B9277Sper Scientific2699122-Jul-10888QuantumOdyssey HF GP QX-406QG4000-03M1205Sep-031253ToshibaE72423J170Sep-03
BuildingTMCSerial Number:2255653Fort McCoy, WIG7191RADCAL2025332611-Jun-10G7191RADCAL2025332611-Jun-10G7191RADCAL2025332611-Jun-106081Fluke99054020-Aug-1026585UNFORS21271618-Jun-161158GEMPH GPAA78305MAONov-031508003VarianA19563808-65Jul-07
RoomCalibration Due Date:3-Sep-174Rock Island Arsenal, ILG9381RTIPMX-II22517-Oct-09G9381RTIPMX-II22517-Oct-09G9381RTIPMX-II22517-Oct-096082Fluke99054110-Jun-101725PlanmecaIntraIXRF0523957083PlanmecaD-0711SB
X-ray Machine Control PanelkVp Meter5Camp Grayling, MIB8923RADCAL901090-177616-Jul-10B8923RADCAL901090-177616-Jul-10B8923RADCAL901090-177616-Jul-106083Fluke9905489-Mar-101726PlanmecaIntraIXRF051968Jun-047083PlanmecaD-0711SB79082Jun-04
Calibration Due Date:Apr-17ECN:2658562145UNFORSMult-O-Meter1295951-Apr-102145UNFORSMult-O-Meter1295951-Apr-102145UNFORSMult-O-Meter1295951-Apr-106084Fluke99054219-Jun-101727PlanmecaIntraIXRF051718Jul-047083PlanmecaD-0711SB77858Jul-04
ECNCamp Grayling, MIManufacturer:UNFORS76886UNFORSXi14284710-Feb-156886UNFORSXi14284710-Feb-106886UNFORSXi14284710-Feb-106138Fluke451B-RYR103926-Jan-151728PlanmecaIntraIXRF051973Jun-047083PlanmecaD-0711SB79792Jun-04
ManufacturerDel Medical Imaging CorpModel:X2826585UNFORSX221271618-Jun-1626585UNFORSX221271618-Jun-1626585UNFORSX221271618-Jun-166139Fluke451B-RYR10385-Sep-091729PlanmecaIntraIXRF0519747083PlanmecaD-0711SB
Model NumberInfinity 400Serial Number:22556590B2701BicronMicroremB420R1-May-101730PlanmecaIntraIXRF0516757083PlanmecaD-0711SB
Serial Number04126-1004Calibration Due Date:3-Sep-17100B2703BicronRSO-5B293M24-Aug-101731PlanmecaIntraIXRF051656Jul-047083PlanmecaD-0711SB78888May-04
Date ManufacturedOct-04Timer110B2706Bicron2000B271R26-Aug-101732PlanmecaIntraIXRF0516557083PlanmecaD-0711SB
kVp (max)125ECN:26585120B2707Bicron2000B272R17-Jun-101733PlanmecaIntraIXRF051660Jul-047083PlanmecaD-0711SB79569Jun-04
mA or mAs (max)3.3Manufacturer:UNFORS130B3068BicronMicro AnalystB464R27-Aug-101734PlanmecaIntraIXRF051658Jul-047083PlanmecaD-0711SB76781May-04
Phase1Model:X2140B3070BicronRSO-50EB249Q1-May-101735PlanmecaIntraIXRF0516597083PlanmecaD-0711SB
Tube InsertSerial Number:225565150B3398BicronMicro AnalystB603R30-Apr-101736PlanmecaIntraIXRF052382Aug-047083PlanmecaD-0711SB81837Sep-04
ManufacturerEurekaCalibration Due Date:3-Sep-17160B6895Bicron2000B673Z26-Aug-101737PlanmecaIntraIXRF051971May-047083PlanmecaD-0711SB76759May-04
Model NumberUX-3Ion Chamber171738PlanmecaIntraIXRF050009Apr-047083PlanmecaD-0711SB1584Sep-04
Serial Number4J119ECN:26585181739PlanmecaPromaxRPX234734Oct-0484163PlanmecaD-054SB2Oct-04
Date ManufacturedSep-04Manufacturer:UNFORS191740PlanmecaPromaxRPX234567Oct-0484163PlanmecaD-054SB23J07069Oct-04
Large FS (mm)2Model:X2201741PlanmecaPromaxRPX234566Oct-0484163PlanmecaD-054SB23J07705Oct-04
Small FS (mm)1Serial Number:225565211743PlanmecaPromaxRPX234570Oct-0484163PlanmecaD-054SB23L08009Oct-04
Tube HousingCalibration Due Date:3-Sep-17221744PlanmecaPromaxRPX234569Oct-0484163PlanmecaD-054SB23J07769Oct-04
ManufacturerEurekaLight Meter231745PlanmecaPromaxRPX234572Oct-0484163PlanmecaD-054SB23D07088Oct-04
Model NumberUX-52H-39ECN:21039241746PlanmecaPromaxRPX234727Oct-0484163PlanmecaD-054SB2Oct-04
Serial Number4J350Manufacturer:EXTECH251886Liebel-FlarsheimFluoroSP10516-0540
Image ReceptorSerial Number:Q619323261923QuantumQG-4000 GP5C7221253QuantumE72425C0414Mar-05
ManufacturerCalibration Due Date:12-Sep-17271977PlanmecaIntraIXRF053564Nov-047083PlanmecaD-0711SBSep-04
Model NumberFocal Spot Test Tool282998GEPrecision 500D GPBRDWNVDec-031504003GE233605830238BI0May-05
Serial NumberManufacturer:292998GEPrecision 500D FluoroBRDNBDec-03120GE228914891072293
Imaging Area (in)Model:303873PlanmecaIntraIXRF061143Mar-067083PlanmecaD-0741B98628Mar-06
Display DeviceSerial Number:313874PlanmecaPromaxRPX235953Mar-0684163PlanmecaD-0545B2Jan-06
ProcessorCalibration Due Date:324149GEProteus XR/A41986461Mar-061505003GE46-155318G33131672TX8Aug-12
ManufacturerOther Phantoms or Meters334429PlanmecaPromaxRPX 236052Apr-0684163PlanmecaD-0545B24L09066Apr-06
Model NumberManufacturer:344430PlanmecaIntraIXRF062506Jun-067083PlanmecaD-0711SB1392Jun-06
Serial NumberModel:354431PlanmecaIntraIXRF062507May-067083PlanmecaD-0711SBO1721May-06
FilmSerial Number:364433PlanmecaPromaxRPX 235203May-0584163PlanmecaD-0545B24H08457May-05
ManufacturerCalibration Due Date:377594PlanmecaPromaxRPX 236914Nov-0684163PlanmecaD-0545B26C03717Nov-06
Model Number387595PlanmecaIntraIXRF 066621Dec-067083PlanmecaD-0711SB11151Dec-06
Cassettes397689Hologic FluoroscanInsight 2 Mini C-Arm09-1207-28Dec-07750.13FlouroscanASY-00871210509Dec-07
Manufacturer407711PlanmecaPromaxRPX 237248Mar-0384163PlanmecaD-0545B26DD3834Mar-03
Model Number418744GEAMX-4 Portable GP195490021-004Mar-081253203GE2208208112106TX0Mar-08
428745GEAMX-4 Portable GP195490020-001Mar-081253203GE220820812341TX3Mar-08
439733MinXrayHF100+ Mobile GP Vet33197Jul-08120703MinXrayRHF-1A33197Jul-08
449734MinXrayHF100+ Mobile GP Vet33203Jul-08120703MinXrayRHF-1A33203Jul-08
Format for the spreadsheet4510017InnovetSelect Vet GP8G0382Jul-081253InnovetE72398G0382Jul-08
Title BarVersions:MAJ Arthur Morton

MORTONAR: For more information please contact:

Major Arthur Ray Morton Chief, Health Physics Section arthur.morton@amedd.army.mil

Tripler Army Medical Center ATTN: MCHK-PVR (MAJ Morton) 1 Jarrett White Road Honolulu, HI 96859-5000

Phone (808) 433-2334 DSN 433-2334

FAX (808) 433-2440 DSN 433-2440

MORTONAR: Sort all remaining data to the right by MMCN

MORTONAR: Recommend listing all variables to be used in the drop down lists and then sort the variables to the order you want them to appear in the drop down list4610046PlanmecaIntraIXRF075280May-087083PlanmecaD-0711SB31395May-08PlanmecaIntraITHa68280
Background (Text)12-Nov-02New TB MED 5214710047PlanmecaIntraIXRF075320May-087083PlanmecaD-0711SB31158May-08PlanmecaIntraITHa68364
Data Entry12-Dec-02CF for light meters4810048PlanmecaIntraIXRF075274May-087083PlanmecaD-0711SB33400May-08PlanmecaIntraITHa68236
Tech. & Cond. Titles, Background9-Jun-03Added Macro Buttons4910049PlanmecaIntraIXRF075328May-087083PlanmecaD-0711SB33158May-08PlanmecaIntraITHa68428
Tech. & Conditions Data, Notes14-Jul-03Many small updates5010050PlanmecaIntraIXRF075281May-087083PlanmecaD-0711SB30530May-08PlanmecaIntraITHa68267
Calculations30-Jul-03Added RPP Checklist5110051PlanmecaIntraIXRF075307May-087083PlanmecaD-0711SB32879May-08PlanmecaIntraITHa68351
PASS, FAILED Box (Empty)6-Aug-03Fixed Linearity Reference5210052PlanmecaIntraIXRF075277Jun-087083PlanmecaD-0711SB33830Jun-08PlanmecaIntraITHa68975
PASS, FAILED Box (PASSED)PASSED2-Sep-03AEC Reference 4-3c5312079PlanmecaIntraIXRF 078333Jan-097083PlanmecaD-0711SB36528Dec-08PlanmecaIntraITHa 71570
PASS, FAILED Box (SHOULD)SHOULD22-Sep-03Added data cells for AEC maximum allowable techniques.5412081PlanmecaIntraIXRF 078097Jan-097083PlanmecaD-0711SB36479Dec-08PlanmecaIntraITHa 71547
PASS, FAILED Box (FAILED)FAILED6-Oct-03Modified kVp timer page5512240PlanmecaPromaxRPX 238272Sep-0884163PlanmecaD-0545B27k08703Aug-08PlanmecaD-0545132RPP0706038
References5612441PlanmecaProMaxRPX 238209Jul-0884163PlanmecaD-0545B27G08357Aug-08PlanmecaPromaxRPP0705931
57A0168SiemensMultiPro MultixS09110501LJan-011508003Siemens3345233 V2146373500 S20Jan-0110.6Siemens3345209 X19533785 S20L
58A0633GE9800 C-Arm89-1265Dec-01120GEOEC 980089-1265Dec-01
59B2234GEAdvantix GP421873WK1Feb-941506403GE46-155318633303060TV8Nov-931.250.6GE46-1550064876438EC1
60B2234GEAdvantix Fluoro421873WK1Feb-94GE211262151417WK7Apr-97GE46-275382646421549WK712
61B7465PickerMTX Chest1451927Jun-951256403PickerDU-4045111WFeb-971.20.6PickerPX-1436PQ516374BB
62B9870SiemensMultix 65 GPS08429S01LJun-001505003Siemens3345233255640520Jun-0010.6Siemens3345209X19533264520L
63CA2022PlanmecaIntraIXRF047552Dec-037083PlanmecaPlanmeca
64CA2023PlanmecaIntraIXRF047555Dec-037083PlanmecaPlanmeca
65D9637GEMVP 60 GP21232ES4Feb-911506403VarianMX100-18H4151=25Aug-091.250.6GEMaxi 100 18NR29846 A2
66E4417GEAdvantix GP65353VP5Feb-0115010003GE46-15531863329112TX0Sep-971.250.6GE46-15540064830720EX7
67E4417GEAdvantix Fluoro521427WK5Feb-01GE46-275382652521427WK512
68INGPlanmecaIntraIXRF0475517083PlanmecaPlanmeca
69INGPlanmecaProlineRCC01411784163PlanmecaPlanmeca
70INGPlanmecaProlineRCC01408084163PlanmecaPlanmeca
71NewPlanmecaProMaxRPX 238165Jun-0884163PlanmecaD-0545B27G08280Jun-08PlanmecaPromaxRPP0705862
72new installPlanmecaIntraIXRF 077335Jan-097083PlanmecaD-0711SB73270Mar-05PlanmecaIntraITHa 46805
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139

Clear Worksheet

Equip Checks

X-RAY SYSTEM EVALUATION CHECKLISTSelectionYESYESYESYESYESYESYESN / AYESYESN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / AN / A
Radiographic SystemsReference11111113113333333333333333333333333333333333
Label Conforms to 21 CFR
MORTONAR: Affix a permanent label or tag on all new x-ray systems and components indicating that the system or component conforms to the requirements of 21 CFR, Chapter 1, Subchapter J.TB MED 521, 4-2aHeadingAnswer
Warning Statement Visible

MORTONAR: Ensure that the following warning statement is displayed on the control panel containing the main power switch and is legible and accessible to view. "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."

TB MED 521, 4-2bVariables1YES
Anode, Cathode and Focal Spot marked
MORTONAR: Permanently mark the cathode and anode receptacles of the x-ray tube and the location of the apparent focal spot.TB MED 521, 4-2c2NO
Technique factors visible before exposure
MORTONAR: The technique factors to be used during an exposure shall be indicated before the exposure begins. If an automatic exposure control (AEC) is used, the technique factors that are set prior to the exposure will be indicated. For equipment with fixed technique factors, this requirement may be met by permanent markings. The indicated technique factors shall be visible from the operator's location, except in the case of spot films made by a fluoroscopist.TB MED 521, 4-3a3N / A
Means to terminate if > 0.5s
MORTONAR: Provide means to terminate the exposure after a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. Except during serial radiography, the operator will be able to terminate the exposure at any time during an exposure of greater than 1/2 second. Termination of exposure will cause the timer to automatically reset to its initial setting or to zero. It will not be possible to make an exposure when the timer is set to zero or an off position, if either position is provided. During serial radiography, the operator will be able to terminate the exposures at any time, but the x-ray system may have a means to permit completion of any single exposure in the series.TB MED 521, 4-3b4Comment
AEC mode indication on control panel

MORTONAR: For x-ray systems with AEC, indication will be made on the control panel when this mode of operation is selected. The following requirements will also be met.

(1) When the peak tube potential is 51 kVp or greater, the minimum exposure time for--

(a) Field emission equipment rated for pulsed operation will be less than or equal to the time interval equivalent to two pulses.

(b) All other equipment will be less than or equal to 1/60 second or the time interval required to deliver 5 mAs, whichever is greater.

(2) Either limit the product of peak tube potential, tube current, and exposure time to not more than 60 kilowatt-seconds per exposure or limit the product of tube current and exposure time to not more than 600 mAs per exposure. If the x-ray tube potential is less than 51 kVp, limit the product of x-ray tube current and exposure time to not more than 2,000 mAs per exposure.

(3) A visible signal will indicate when an exposure has been terminated at the limits described in (2) above. The operator will have to manually reset the AEC before further automatically timed exposures can be made.TB MED 521, 4-3c5Not Visible
Visual and audible indication of x-rays
MORTONAR: The x-ray control will provide visual indication whenever x-rays are produced. In addition, a signal audible to the operator will indicate that the exposure has terminated.TB MED 521, 4-3s6Unknown
Multiple tubes with 1 exposure switch
MORTONAR: Where two or more radiographic tubes are controlled by one exposure switch, the selected tube(s) will be clearly indicated before initiation of the exposure. This indication will be on both the x-ray control and at or near the selected tube housing assembly.TB MED 521, 4-3t7
Exposure switch > 2 m from x-ray tube

MORTONAR: The exposure switch will be--

(1) A continuous pressure "dead man" type switch.

(2) Located in such a manner as to prevent convenient use by the operator outside the protective enclosure (control booth).TB MED 521, 4-3w8
Steady mechanical support of tube head
MORTONAR: Mechanical support of the tube head and beamlimiting device will maintain its proper exposure position without drift, sag, or vibration.TB MED 521, 4-3x
Telemetric exposure switch

MORTONAR: Telemetric exposure switches controlling the tube circuit by wireless means will be used with caution. When using two or more telemetric timers within the same facility--

(1) Operate all timers on different frequencies.

(2) Label or control each telemetric switch to assure that they do not become interchanged.TB MED 521, 4-3y
Mobile and Portable Radiographic X-ray Systems
Stand at least 2 m from tube
MORTONAR: Stand as far as practical (at least 2 meters (m)) from the patient, the x-ray tube, and the primary beam.TB MED 521, 3-5a(3)
Safeguard enabling key
MORTONAR: Make provisions to safeguard the enabling key for mobile x-ray systems. Leave the key in the system only if necessary during recharging or for a quick response system (see para 3-3).TB MED 521, 3-5b(2)
Exposure switch > 2 m from x-ray tube
MORTONAR: Arrange the exposure switch so that the operator can stand at least 2 m from the x-ray tube, the primary beam, and the patient.TB MED 521, 4-4a
Battery charge indication
MORTONAR: Visual means will be provided on the x-ray control panel of the battery powered systems to indicate if the battery is charged and is adequate for proper operation. It will not be possible to produce x-rays unless there is an adequate charge for proper operation.TB MED 521, 4-4b
Enabling key
MORTONAR: An enabling key will be required to activate the system. To prevent unauthorized use, remove the key during system storage, except during battery recharging.TB MED 521, 4-4c
Minimum SSD > 30 cm
MORTONAR: Provide positive means to ensure that the minimum SSD is no less than 30 cm.TB MED 521, 4-4d
Tomographic X-ray Systems

MORTONAR: The technique factors to be used during an exposure shall be indicated before the exposure begins. If an automatic exposure control (AEC) is used, the technique factors that are set prior to the exposure will be indicated. For equipment with fixed technique factors, this requirement may be met by permanent markings. The indicated technique factors shall be visible from the operator's location, except in the case of spot films made by a fluoroscopist.

Tomographic cut level
MORTONAR: The error in the cut level indication will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 mm.TB MED 521, 4-5a
Tomographic angle

MORTONAR: The difference between the indicated tomographic angle and the measured angle will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 degrees. The symmetry of the exposure angle about the midline will not exceed the manufacturer’s specifications or, in the absence of such specifications will not exceed

2 degrees, unless the system is designed to operate in an asymmetrical manner. Since the cut thickness is dependent on exposure angle, the thickness of the cut plane should decrease as the exposure angle increases.TB MED 521, 4-5b
Tomographic plane

MORTONAR: The tomographic plane should be flat to within manufacturer’s specifications or, in the absence of such specifications, 3 mm for linear systems and 2 mm for other systems.

TB MED 521, 4-5c
Tomographic uniformity
MORTONAR: Tomographic x-ray systems will produce a uniform density along the beam path when an image of a tomographic uniformity test phantom is produced. Linear tomographic images should consist of an evenly darkened straight line without evidence of mechanical or radiation caused irregularities. Tomographic x-ray systems with circular, elliptical, hypocycloidal, or trispiral scan capability will produce an evenly darkened closed path. The path overlap will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 20 degrees.TB MED 521, 4-5d
Tomographic resolution
MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern.TB MED 521, 4-5e
Fluoroscopic X-ray Systems
Monitor kVp and mA
MORTONAR: In x-ray systems with automatic brightness control, the peak tube potential (kVp) and tube current (mA) may rise to high values without the knowledge of the operator, particularly if the gain of the intensifier diminishes. The fluoroscopist should be able to monitor the kVp and mA from his or her operating position. Systems may have meters located at the fluoroscopists operating position to assist in this monitoring.TB MED 521, 3-6e
Electronic Image Hold
MORTONAR: Fluoroscopic x-ray systems used for medical procedures will have image intensifiers. If nonradiology specialists use the system, it will have image intensifiers with electronic image-holding capabilities.TB MED 521, 3-6g
Primary beam intercepted by barrier
MORTONAR: The fluoroscopic imaging assembly will have a primary protective barrier which intercepts the entire cross section of the primary beam at any SID. The x-ray tube used for fluoroscopy will not produce x-rays unless the barrier is in position to intercept the entire primary beam. The exposure rate due to transmission through the barrier with the attenuation block in the primary beam combined with radiation from the image intensifier if provided, will not exceed 3.34 x l0-3 percent of the EER, at a distance of 10 cm from any accessible surface of the fluoroscopic imaging assembly beyond the plane of the image receptor.TB MED 521, 4-7a
Terminating exposures
MORTONAR: X-ray production in the fluoroscopic mode will be controlled by a device that requires continuous pressure by the operator for the entire time of any exposure. When recording serial fluoroscopic images, the operator will be able to terminate the x-ray exposures at any time, but means may be provided to permit completion of any single exposure of the series in process.TB MED 521, 4-7d
Indication of kVp and mA
MORTONAR: During fluoroscopy and cinefluorography, x-ray tube potential and current will be continuously indicated. Deviation of x-ray tube potential and current from the indicated values will not exceed the maximum deviation stated by the manufacturer.TB MED 521, 4-7f
Minimum SSD
MORTONAR: The equipment will have the means to limit the SSD to not less than 38 cm on stationary fluoroscopic systems and to not less than 30 cm on mobile and portable fluoroscopic systems. In addition, for image-intensified systems intended for specific surgical application that would be prohibited at the SSDs specified in this paragraph, provisions may be made for operation at shorter SSDs, but in no case less than 20 cm. When provided, the manufacturer must set forth precautions with respect to the optional means of spacing.TB MED 521, 4-7g
5 minute cumulative timing device

MORTONAR: The equipment will have the means to preset the cumulative on-time of the fluoroscopic tube. The timing device’s maximum cumulative time will not exceed 5 minutes without resetting. A signal, audible to the fluoroscopist, will indicate the completion of any preset cumulative on-time. While x-rays are being produced, this signal will continue to sound until the timing device is reset. Radiation therapy simulation systems may alternatively have a timer that--

(1) Indicates the total cumulative exposure time during which x-rays were produced.

(2) Is capable of being reset between x-ray examinations.

TB MED 521, 4-7h
Image Intensifier
MORTONAR: In addition to the above requirements, mobile and portable fluoroscopic systems will have image intensifiers.TB MED 521, 4-7i
X-ray Source Assembly
MORTONAR: Fluoroscopic x-ray systems will have a diagnostic-type protective source assembly. The apparent focal spot will be clearly indicated on the tube housing. The cathode and anode cable receptacles of the x-ray tube will be permanently identified.TB MED 521, 4-7j
Under Table shielding for bucky slot
MORTONAR: Fluoroscopic systems with the x-ray tube permanently located under a table will have a shielding device of at least one tenth value layer of material at the maximum tube potential for covering the bucky slot during fluoroscopy. If a shielded bucky slot cover is required to meet the exposure rate requirements of paragraph 4-2e, the operator should use positive means to ensure that the cover is in place before producing x-rays.TB MED 521, 4-7k
Under Table shielding drapes and panels

MORTONAR: Under table fluoroscopic equipment will have a shielding device(s) of at least 0.25 mm lead equivalent at 100 kVp. Such devices include--

(1) Overlapping protective drapes hung from the image receptor, and

(2) Hinged or sliding panels at the front of the table. When provided, a hinged or sliding panel should project at least 13 cm above the tabletop and will not interfere with the use of the table.

TB MED 521, 4-7l
Shockproof cables
MORTONAR: Each high-voltage generator assembly will have a pair of shockproof cable receptacles, per Federal Standard No. 72, for each x-ray tube assembly that the high-voltage generator is to energize. The high-voltage generator will be permanently marked "Equipped with Federal Standard Shockproof Cable Receptacles." The receptacles will be covered when not in use.TB MED 521, 4-7m
Anode Heat Calculator
MORTONAR: Cardiac x-ray systems with a tube current of 600 mA or greater that are placed in service after 1985 will have an anode heat calculator device or system. For x-ray systems installed prior to 1985, the use of a heat calculator may be beneficial, but it may not be available from the manufacturer. For such systems, the heat calculator is not required.TB MED 521, 4-8a
Tube Protective Circuit
MORTONAR: Each cardiac x-ray system will have a tube protective circuit.TB MED 521, 4-8b
Line voltage compensating device
MORTONAR: A line voltage compensating device (operable from the x-ray control panel without the aid of tools), together with a meter for determining the proper adjustment of the device, will be mounted in all cardiac x-ray control panels not having automatic voltage compensation or a prereading kVp meter that responds directly to line voltage variations.TB MED 521, 4-8c
Dental X-ray Systems
Dental Film "D" or faster
MORTONAR: Perform intraoral image receptor radiography using dental film that meets the requirements of American National Standards Institute (ANSI) speed group rating "D" or faster.TB MED 521, 3-8a(1)
Veterinary X-ray Systems
Veterinary Use Logbook
MORTONAR: Maintain a log or equivalent record on the use of the x-ray systems. Include the date of exposure, kVp, mA, exposure time, operator’s name, identification of the animal, and the name of any person used to hold the animal during the radiographic procedure.TB MED 521, 3-10i
Tomographic X-ray Systems
Tomographic cut level
MORTONAR: The error in the cut level indication will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 mm.TB MED 521, 4-5a
Tomographic angle

MORTONAR: The difference between the indicated tomographic angle and the measured angle will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 5 degrees. The symmetry of the exposure angle about the midline will not exceed the manufacturer’s specifications or, in the absence of such specifications will not exceed

2 degrees, unless the system is designed to operate in an asymmetrical manner. Since the cut thickness is dependent on exposure angle, the thickness of the cut plane should decrease as the exposure angle increases.TB MED 521, 4-5b
Tomographic plane

MORTONAR: The tomographic plane should be flat to within manufacturer’s specifications or, in the absence of such specifications, 3 mm for linear systems and 2 mm for other systems.

TB MED 521, 4-5c
Tomographic uniformity
MORTONAR: Tomographic x-ray systems will produce a uniform density along the beam path when an image of a tomographic uniformity test phantom is produced. Linear tomographic images should consist of an evenly darkened straight line without evidence of mechanical or radiation caused irregularities. Tomographic x-ray systems with circular, elliptical, hypocycloidal, or trispiral scan capability will produce an evenly darkened closed path. The path overlap will not exceed the manufacturer’s specifications or, in the absence of such specifications, will not exceed 20 degrees.TB MED 521, 4-5d
Tomographic resolution
MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern.TB MED 521, 4-5e
Comments:

Leakage-EER

LEAKAGE RADIATION FROM THE DIAGNOSTIC SOURCE ASSEMBLYCourse Manual for X-Ray Measurements, FDA, BRH July 1977
Techniques :100mA1000ms90kVpVictoreen 440 Time Response Correction Factors
Conditions :100mR/hr @1m5%duty cycle0.02Bkg (mR/hr)Seconds3 mR/hr10 mR/hr30 mR/hr100 mR/hr300 mR/hr
Meter :UnforsModelX2Manuf.225565SN0.30.2150.2450.255
uRDistancemR in 1 hrTime Resp.0.40.315
Start5mR/hrinat 1 meterC.F.
MORTONAR: Using the 450P and 1000 ms exposure time, the time response correction factor is about 0.85mR 1hr@1m0.50.1400.3200.3400.3900.390
Leakage Front4.940.00.240.850.290.80.2350.4650.5750.6200.575
Leakage Back1.334.00.050.850.051.00.3200.5950.6550.7450.700
Leakage Anode1.540.00.070.850.091.20.645
Leakage Cathode0.440.00.020.850.021.30.4150.795
Scatter Front1.50.4300.7600.8451.0000.890
Scatter Back1.80.8200.955
Scatter Anode2.00.5550.8350.9901.0500.990
Scatter Cathode2.50.6650.9151.0300.975
Max Leakage (not tube)3.00.6880.9301.0301.0001.040
3.50.7450.9651.0000.9900.990
4.00.8200.9751.0001.0001.000
4.51.0001.000
5.00.8701.0251.0001.0001.000
Maximum scatter radiation (mR in 1 hour at 1 meter)other measurements takenin mR/hr
Maximum leakage radiation (mR in 1 hour at 1 meter)0.3PASSEDat door to control booth
Maximum leakage radiation other than tube (2 mR in 1 hour at 5 cm)at door to room
at computer inside cb, door open
References:inside control booth,door closed
Leakage Radiation

MORTONAR: Construct diagnostic source assemblies so that leakage radiation exposure at a distance of 1 m in any direction from the source will not exceed 100 mR (2.58 x 10-5 coulombs/kilogram (C/kg)) in 1 hour when the

x-ray tube is operated at the leakage technique factors. If the tube housing assembly’s maximum rated peak tube potential is greater than the diagnostic source assembly’s maximum rated peak tube potential, provide positive means to limit the maximum x-ray tube potential to that of the diagnostic source assembly.TB MED 5214-2d
Leakage-other than tube

MORTONAR: Construct components other than the diagnostic source assembly so that the radiation exposure will not exceed 2 mR (5.16 x 10-7 C/kg) in 1 hour at 5 cm from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. TB MED 521 4-2e

FLOURO ENTRANCE EXPOSURE RATES AND SYSTEM RESOLUTION
Techniques :kVpSID (in)SSD (in)
Conditions :10R/min EER2mR/hr PB30Mesh
ModesEERAudible
Max. Cont.PhantomkVpmAR/minSignal

MORTONAR: Yes or No MORTONAR: Using the 450P and 1000 ms exposure time, the time response correction factor is about 0.85 MORTONAR: Construct diagnostic source assemblies so that leakage radiation exposure at a distance of 1 m in any direction from the source will not exceed 100 mR (2.58 x 10-5 coulombs/kilogram (C/kg)) in 1 hour when the x-ray tube is operated at the leakage technique factors. If the tube housing assembly’s maximum rated peak tube potential is greater than the diagnostic source assembly’s maximum rated peak tube potential, provide positive means to limit the maximum x-ray tube potential to that of the diagnostic source assembly.

AEC/ no HLCLead
AEC/HLCLead
ManualLead
Max. Pulsed Fluoro
Pulsed 3.75 fpsLead
Pulsed 7.5 fpsLead
Pulsed 15 fpsLead
Pulsed 30 fpsLead
Patient Equivalent Phantom
ManualPatient Eq
AECPatient Eq
Pulsed 3.75 fpsPatient Eq
Pulsed 7.5 fpsPatient Eq
Pulsed 15 fpsPatient Eq
Pulsed 30 fpsPatient Eq
System ResolutionPantomMeshPhantomResolution
IR 6"
IR 9"
IR 15"
IR 16"
Tomographic
Maximum Entrance Exposure Rate (R/min)
References:
Fluoroscopic EER

MORTONA: The following EERs apply to fluoroscopic x-ray systems.

(1) Fluoroscopic equipment that is provided with automatic exposure rate control (AERC) will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 roentgens per minute (R/min) (2.58xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional high-level control (HLC) is provided.

(2) Fluoroscopic equipment that is not provided with AERC will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R/min (1.29x10-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional HLC is provided.

(3) Equipment with both an AERC mode and a manual mode will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R/min (2.58xl0-3 C/kg×min) in either mode at the point where the center of the primary beam enters the patient, except during recording of fluoroscopic images or when an optional HLC is provided.

(4) If HLC is provided, the system will not be operable, in either AERC or manual mode, at any combination of tube potential and current that will result in an exposure rate in excess of 5 R/min (1.29xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient, unless the HLC is activated. The system will have a special means to activate the HLC. The HLC will be operable only when the operator provides continuous manual activation. A continuous signal audible to the fluoroscopist will indicate that the HLC is activated.

(5) Fluoroscopic equipment manufactured on and after 19 May 1995 will be equipped with AERC if operable at any combination of tube potential and current that results in an exposure rate greater than 5 R/min (1.29xl0-3 C/kg×min) at the point where the center of the primary beam enters the patient. Provision for manual selection of technique factors may be provided.

(a) This equipment will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R/min (2.58x10-3 C/kg×min) at the point where the center of the primary beam enters the patient, except during the recording of images from an x-ray image-intensifier tube using photographic film or a video camera when the x-ray source is operated in a pulsed mode or when an optional HLC is activated. Special means will be required to activate the HLC.

(b) When the HLC is activated, the equipment will not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 20 R/min (5.16x10-3 C/kg×min) at the point where the center of the primary beam enters the patient.

(c) The HLC will be operable only when the operator provides continuous manual activation. A continuous signal audible to the fluoroscopist will indicate that the HLC is activated.

(6) Fluoroscopic radiation therapy simulation systems are exempt from the requirements set forth in (1) through (5) above concerning EERs.

MORTONAR: Construct components other than the diagnostic source assembly so that the radiation exposure will not exceed 2 mR (5.16 x 10-7 C/kg) in 1 hour at 5 cm from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. TB MED 521 4-7e EER Notes MORTONA: If the x-ray tube is under the table, position the ion chamber 1 cm above the table top, otherwise measure the exposure rate 30 cm from the input surface of the image receptor. In manual mode, set the contols to 90 kVp and 2 mA. Center the attenuation block over the ion chamber, and make an exposure and collimate the beam the dimensions of the attenuation block and verify the ion chamber is centered in the primary beam. Repeat measurement with 1/8 inch of lead over the attenuation block. Test high level control and pulsed modes (note the presence of an audible signal).

Tomo resolution MORTONAR: All tomographic systems will be capable of resolving a mesh pattern with 30 hole per inch (30 mesh) over the entire length in the plane of the cut. Specialized equipment should be capable of resolving a 50 mesh pattern.

MORTONA: Leakage Notes: To check diagnostic source assembly, close the shutters and add 1/8" of lead to block the primary beam. Make exposure measurements of approximately 1s duration at the maximum rated tube current for the maximum rated kVp. Measure the leakage radiation using a suitable ionization chamber at approximately 1 meter from the tube focal spot in 4 cardinal directions. At these settings, the tube can be operated only a limited number of seconds each minute or minutes per hour. This is the % on time per hour figure above or otherwise known as the duty cycle. Note that for many ion chambers the response time is greater than 1 second and either the exposure time will need to be increased or a time response correction factor applied or the meter is operated in integrated exposure mode. The radiation emitted from a component other than the diagnostic source assembly shall not exceed 2 mR in 1 hour at 5 cm from any accessible surface of the component. An example is the standby radiation from capacitor energy storage equipment without activating the exposure switch.

MORTONA: If the x-ray tube is under the table, position the ion chamber 1 cm above the table top, otherwise measure the exposure rate 30 cm from the input surface of the image receptor. In manual mode, set the contols to 90 kVp and 2 mA. Center the attenuation block over the ion chamber, and make an exposure and collimate the beam the dimensions of the attenuation block and verify the ion chamber is centered in the primary beam. Repeat measurement with 1/8 inch of lead over the attenuation block. Test high level control and pulsed modes (note the presence of an audible signal). TB MED 521 4-5e

RepLin

REPRODUCIBILITY / LINEARITYReference: NCRP 102
Techniques:1Phase90kVp24SSD (in)24"1 Phase3 Phase
Conditions:0.05CV30%Output0.1CLkVpmR/mAsmR/mAs
Focal SpotLargeLarge403.05.1
ms10001000505.69.6
mA100200608.715.0
mAs100.0200.07011.519.6
Exposure (mR)8015.025.3
X1809.001724.009018.432.2
X2803.501720.0010021.936.8
X3799.901722.0011025.342.6
X4793.101718.0012029.950.6
X513035.761.0
X614039.165.6
X715043.773.6
X8
X9Reference: AAPM Report 25
X1024"1 Phase3 Phase
ReproducibilitykVpmR/mAsmR/mAs
mean exposure (mR)801.381721.0403.25.5
standard deviation6.66602.5820505.28.5
coefficient of variation (CV)0.00830.0015607.612.0
ReproducibilityPASSEDPASSED7010.516.0
Output8013.920.6
mean (mR/mAs)8.018.619017.825.8
mean (mR/mAs) at 24"8.018.6110022.231.5
NCRP % output difference-73.5%-71.5%11027.137.8
AAPM % output difference-68.8%-66.5%12032.544.6
NCRP % outputShould <Should <13038.551.9
AAPM % outputShould <Should <14045.059.8
Linearity15052.068.2
coefficient of linearity (CL)0.0356
LinearityPASSED
References:Coefficient of VariationLinearity
Reproducibility

MORTONA: When the equipment is operated on an adequate power supply per the manufacturer’s specifications, the estimated coefficient of variation of radiation exposures will be no greater than 0.05 for any combination of selected technique factors. That is:

(See Equation 1) where: Xi = the ith exposure measurement _ = the average of the exposure measurements n = the number of measurements taken

TB MED 5214-3e
Linearity

MORTONA: When the equipment is operated on an adequate power supply per the manufacturer’s specifications and at any fixed tube potential within the range of 40 percent to 100 percent of the maximum, the following requirements will be met.

(1) If the equipment has independent selection of x-ray tube current (mA) and exposure time (s), the average radiation output (mR/mAs) at any two consecutive tube current settings will not differ by more than 0.10 times their sum. That is:

(See equation 2) where: X1 and X2 = the average radiation output values obtained at each of two consecutive tube current settings or at two settings differing by no more than a factor of 2 if the tube current selection is continuous.

(2) If the equipment has selection of tube current-exposure time product (mAs) only, the average radiation output (mR/mAs) at any two consecutive mAs selector settings will not differ by more than 0.10 times their sum. That is:

(See equation 2) where: X1 and X2 = the average radiation output values obtained at each of two consecutive mAs selector settings or at two settings differing by no more than a factor of 2 where the mAs selector provides continuous selection.

TB MED 521 4-3f

Output NCRP

1 PhasemR/mAs
4050607080901001101201301401502.9919447640966635.63866513233601918.745684695051783911.50747986191024114.95972382048331418.41196777905638621.86421173762946125.31645569620253229.91944764096662735.67318757192175139.12543153049482243.7284234752589213 PhasemR/mAs
4050607080901001101201301401505.06329113924050679.551208285385500914.95972382048331419.56271576524741125.31645569620253232.22094361334867336.82393555811277342.57767548906789350.63291139240506360.98964326812428365.59263521288838273.647871116225545kVp

mR/mAs

Output AAPM

1 PhasemR/mAs
4050607080901001101201301401503.23932123108591125.17565172819909597.590067907259213910.49141145662621213.8872741284330917.78432169182985922.18850469445560627.10520364309244832.53933344739519438.49542117093238944.97766553504480751.989983511144753 PhasemR/mAs
4050607080901001101201301401505.53202313744991348.453041722578008811.95246223202225816.01977014197168420.64624038874289625.82443165155908831.54786722400969337.81082088498462944.60816637029773851.93526760293332259.78789623718331168.162168184576302kVp

mR/mAs kVpTimer

TUBE POTENTIALESE Reference Table
Techniques:1000ms24SSD (in)100mA 1st Condition
MORTONAR: Measurements 1-824SSD (in)
Condition +/-:5.0%ShallShould3.0%200mA 2nd Condition

MORTONAR: Measurements 9-16

SetMeasuredMult CorrAbsolutePercentSetMeasuredSetSetOutput
kVp

MORTONAR: Table is set up to analyze kVp for two conditions such as focal spot or mA

kVpFactor
MORTONAR: Insert the multiplicative correction factor from the calibration certificate or 1.00 for no correction.Diff (kVp)DifferencekVpkVpmAmsmR
X16667.01.001.01.5%PASSEDX166.0067.001001000386.40
X27072.41.002.43.4%SHOULDX270.0072.401001000463.40
X39093.81.003.84.2%SHOULDX390.0093.801001000812.80
X4110114.11.004.13.7%SHOULDX4110.00114.1010010001235.00
X58082.61.002.63.2%SHOULDX580.0082.601001000628.00
X65759.41.002.44.2%SHOULDX657.0059.401001000266.70
X77577.81.002.83.7%SHOULDX775.0077.801001000544.70
X88587.11.002.12.5%PASSEDX885.0087.101001000705.70
X96667.71.001.72.6%PASSEDX966.0067.702001000829.30
X107072.91.002.94.1%SHOULDX1070.0072.9020010001009.00
X119092.91.002.93.2%SHOULDX1190.0092.9020010001717.00
X12110115.61.005.65.1%FAILEDX12110.00115.6020010002628.00
X138082.51.002.53.1%SHOULDX1380.0082.5020010001318.00
X145759.91.002.95.1%FAILEDX1457.0059.902001000552.20
X157577.01.002.02.7%PASSEDX1575.0077.0020010001171.00
X168587.81.002.83.3%SHOULDX1685.0087.8020010001526.00

kilovolt peak (kVp) passed all conditions? FAILED

SHIMABUKULN: Paste Here Paste Keithley Triad Data from Radiographic window here

TIMERmsmR
100mA90kVp24SSD (in)
Condition +/-:5.0%ShallShould3.0%AND0ms
SetMeasuredSubt CorrAbsolutePercent
msmsFactor (ms)