This document is a Description of Requirement (DOR) for a Stability Chamber solicitation by the US Army Engineer Research and Development Center. The Cement and Pozzolan Laboratory within the Concrete and Materials Branch requires a specialized environmental chamber for curing and storing testing specimens, with specific technical requirements including maintaining 95% relative humidity at 23.0 ± 2.0 °C, utilizing a 208v / 60 Hz electrical supply, having a minimum of 45 cubic feet of internal storage, and featuring stainless steel wire shelves.

The solicitation (W912HZ25Q8243) is a 100% small business set-aside using NAICS code 333415, with a simplified acquisition approach. Key requirements include a maximum chamber dimension of 73in wide x 40in deep x 82in tall, no glass door, and delivery within 12 weeks of contract award to the U.S. Army Engineer Research and Development Center in Vicksburg, MS. The government will evaluate offers based on lowest price and technical capability, with the intent to award without discussions and requiring vendors to be registered in the System for Award Management (SAM).

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Description of Requirement

Stability Chamber

Specifications:

The Cement and Pozzolan Laboratory within the Concrete and Materials Branch (CMB) of the Geotechnical and Structures Laboratory (GSL) of the US Army Engineer Research and Development Center (ERDC) requires a stability chamber for curing and storage of testing specimens. Stability chambers are specialized equipment that will maintain specific environmental conditions by controlling temperature and humidity at precise levels as to follow ASTM and other standards.

The chamber selected must meet or exceed the following requirements:

1. Maintain a minimum of 95% relative humidity at 23.0 ± 2.0 °C

2. 208v / 60 Hz electrical supply; less than 20 Amps draw

3. Minimum of 45 cubic feet of internal storage

4. Must come with stainless steel wire shelves

5. No glass door

6. Maximum overall dimensions of 73 in wide x 40 in deep x 82 in tall

7. Must be delivered within 12 weeks of contract award date

8. Shipping and all fees must be included in quote

Delivery Location:

U.S. Army Engineer Research and Development Center 3909 Halls Ferry Road Bldg 6000 Vicksburg, MS 39180

Delivery Date:

The chamber must be delivered within 12 weeks of contract award date.