DEPARTMENT OF THE AIR FORCE

HEADQUARTERS, 673D AIR BASE WING

JOINT BASE ELMENDORF-RICHARDSON, ALASKA

9 August 2022

MEMORANDUM FOR Veterinary Treatment Facility/MCHB-RW-JBR

FROM: 673 OMRS/SGXB

SUBJECT: Semi-Annual Ventilation Survey Results, Portable Laboratory Hood, 0687-HOSY- 727A

1. On 9 August 2022, SrA Andrew Cantrell from Bioenvironmental Engineering (BE) performed a semi-annual ventilation survey of the portable laboratory hood. The purpose of this survey was to determine if the system is operating at the recommended air flow rates per the UFC-4-510-10 Design: Military Medical Facilities and the manufacturer’s recommendations.

Equipment used to conduct this survey was a TSI VelociCalc (SN:9565P1818041), calibrated 27 May 2022 and Electronic Micromanometer ADM-880C (SN:M20243), calibrated on 14 September 2020.

2. The purpose of this ventilation system is to prevent worker exposure to Formaldehyde during the mass preservation process and determine whether the system meets the design criteria directed by UFC-4-510-10 Military Medical Facilities. The measurement is expressed in feet per minute (FPM); results are listed below for your information.

SYSTEM

SURVEYED HAZARD

MEASURED

FACE

VELOCITY

REQUIRED RANGE

FPM WITHIN LIMITS?

Portable Lab Hood Formaldehyde 68 fpm 67.5 - 82.5

100 maximum Yes Yes

Rooms Pressure Required

(POS/NEG)

Ventilation Recommended

(AC/hr)

Minimum Pressure

Differential

Ventilation Measurement

(AC/hr)/ Pressure

Measured Pressure

Differential Pass/ Fail

Room 2, Surgery POS 15 10% 21/POS 61% Pass Room 4, Treatment POS 15 10% 17/POS 43% Pass

3. The lab hood is operating at the recommended criteria. Be sure to keep the lab hood clear of clutter and maintain hazardous chemicals in the proper storage. The result of this survey is invalidated if the system is relocated, the filters are changed, or if repairs/maintenance are made.

Please contact BE at 384-3985 if any of the listed actions occur or if you have any questions/concerns regarding this evaluation.

ALEX B. SHUTTLEWORTH, TSgt, USAF Bioenvironmental Engineering Flight

SHUTTLEWORT

H.ALEX.BRAND

ON.1078634126

Digitally signed by

SHUTTLEWORTH.ALEX.BRA

NDON.1078634126

Date: 2022.08.29 13:25:53 -08'00'Controlled by: Bassett Army Community Hospital

Controlled by: Equipment Management Branch CUI Category: General Procurement and Acquisition

LDC: FEDCONONLY