Attachment 1- Technical Specification Sheet.pdf

PDF 221 KB Posted

Attached to
Hydrazine Detection System
Solicitation number
W50S6S24Q0022
Issued by

About this file

This document is a technical specification sheet for a Hydrazine Detection System solicited under U.S. Army contract W50S6S24Q0022. The solicitation is a 100% small business set-aside with a NAICS code of 334519 and a 600-employee size standard. The required equipment includes the Drager X-am 5100 (or equal) gas detection monitor and the Drager X-zone 5800 (or equal) portable gas detection system, with detailed technical specifications provided. The contract requires compliance with various environmental, electrical, and safety standards. Quotes are due by the date listed on SAM.gov and should be emailed to the provided contracting points of contact. The government will respond to any questions by June 2, 2024. Vendors must submit completed vendor information and comply with all stated terms and conditions.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

W50S6S24Q0022

Hydrazine Detection System

MFG: Drager or equal

Model: Drager X-am 5100 & Drager X-zone 5800 or equal

Part number- 8326839 (Drager X-am 5100) , 8329040 (Drager X-zone 5800) or equal Conforming to the below specifications:

CLIN 0001(Drager X-am 5100) or equal

• Dimensions (W x H x D)- 48 x 130 x 61 mm; 1.89 x 5.12 x 2.4 inches

• Weight- 220 – 250 g; 7.8 – 8.8 oz.

• Ambient conditions Temperature - 20 to + 50 °C; - 4 to + 122 °F

• Pressure 700 to 1300 mbar; 20.7 to 38.4 inch Hg

• Relative humidity 10 to 90 % RH

• Alarms Visual 360°

• Audible Multi-tone > 90 dB at 30 cm; 1 ft.

• Vibration

• Ingress Protection IP 54

• Operating times

• ~150 h (alkaline battery), ~150 h (NiMH rechargeable battery), or

• ~180 h (NiMH high capacity rechargeable battery)

• Charging times < 4 h

• Data logger Retrievable using an infrared interface

• > 1000 h at a recording interval of 1 value per minute

• CE mark Electromagnetic compatibility

• Guideline 2004/108/EC

• Explosion protection – Guideline 94/9/EC

• The Drager X-am 5100 or Equal must include compatible power pack and charging set.

CLIN 0002 (Drager X-zone 5800) or equal

• Dimensions (W x H x D)- 490 x 300 x 300 mm (20 x 12 x 12 in)

• Weight- 10 kg (22 lbs.) (24 Ah battery)

• Ambient condition

• Temperature max. +50 °C (+122 °F), depending on sensors used

• Pressure 700 to 1,300 hPa

• Humidity 10 to 95 % r.h.

• Ambient conditions

• Temperature as low as -50 °C (-58 °F), depending on sensors used

• Ingress Protection- IP67

Alarm

• Visual 360° LED (illuminated ring)

• Audible 360°; >108 dB in a distance of 1 m (3 ft), 120 dB in a distance of 30 cm (1 ft)

• Configurable alarm patterns, frequencies and volumes

Life signal

• Green status display (360°),

• illuminated ring (LED)

Operating time

• Approx. 120 h (24 Ah battery)

• Depending on sensor equipment / configuration of the life signal, Extended run time is possible with optional Power Supply Ex

Charging time

• Approx. 14 h

• Flexible power supply: External 100 – 240 V charger (worldwide) or inductive wireless charger

Pump operation.

• Maximum hose length: 45 m (150 ft); 30 m (98 ft) for Zone 0

Alarm output.

• Potential-free alarm contact for intrinsically safe circuits (6-pole); <20 V to 0.25 A (0.15

A constant current); Resisting load.

Radio transmission

• Worldwide license-free ISM frequencies

• Digital radio, robust and interference-free transmission up to 100 m (328 ft)

RF approvals

• 915 MHz (USA, Canada)

Approvals

• ATEX I M1 Ex ia I Ma, II 1G Ex ia IIC T4 Ga

• IECEx Ex ia I Ma, Ex ia IIC T4 Ga

• CSA (Canada and US) [Ex ia Ga] IIC, [AEx ia Ga] IIC

• The Drager X-Zone 5800 or Equal must include the following accessories

1. Power Supply 2.400A 15V

2. Power Cord USA 3 m

3. Power Supply Ex 110V (w/o power cord)

4. X-zone 5x00 alarm-attenuation ring

5. Grease 3g

6. Spare set pin block (X-zone 5500)

7. Spare part gas inlet filter 5100 holder

8. USB DIRA with USB cable (interface with USB-Cable fore infrared data transfer for all portable X-am series instruments).

All items are brand name or equal. Refer to FAR 52.211-6 for additional or equal requirements.