Updated Room List TB sheet.xlsx
XLSX spreadsheet 149 KB Posted
- Solicitation number
- 36C24825Q0173
About this file
This document is an extensive Air Balance Summary Sheet detailing comprehensive ventilation system testing across multiple buildings and floors at the Bruce W. Carter Veterans Affairs Medical Center. The testing, performed in April 2023, systematically evaluated air pressure, air changes, supply, return, and exhaust rates in various clinical and research spaces including operating rooms, patient rooms, laboratories, clinics, and specialized research areas.
The detailed testing revealed numerous rooms failing air balance requirements, with issues including taped-off exhaust grilles, offline fume hoods, improper air pressure differentials, and inconsistent air exchange rates. The summary covers buildings including main medical center facilities, animal research spaces, and multiple clinical sites such as Broward, Pembroke, Hollywood, and Deerfield Clinics. The testing was likely conducted in preparation for the VA Healthcare System's 5-Year Testing and Balancing of Ventilation System solicitation (36C24825Q0173), which seeks to ensure proper ventilation and air quality across their medical facilities. Specific areas of concern included research laboratories, operating rooms, patient rooms, and specialized clinical spaces where precise air handling is critical for patient and research safety.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SCHEDULE SITE VISIT.docx | DOCX document | |
| 36C24825Q0173 0002_1.docx | DOCX document | |
| Combined Synopsis 2025.docx | DOCX document | |
| 36C24825Q0173_1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1st FL & Other Areas
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/21/23
| 1ST FLOOR & OTHER AREAS | |||||||||||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | |||||||||||
| Room Volume Cu. Ft. | NEG. POS. EQUAL NR | CODE Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual | |||||
| Room # | Room Description | Area Sq. Ft. | Ceiling Height |
| B1214 | STORAGE ENT (P/TMP/RH) | 182 | 7.5 | 1,365 | 0.002 | 4.0 | N/A | 6.81 | 9300.00 | N/A | 155 | N/A | 0 | N/A | 0 | N/A | 155 | PASS | *Exhaust grille taped off |
| B1208 | STORAGE DERM (P/TMP/RH) | 59 | 8.0 | 468 | 0.002 | 4.0 | N/A | 38.46 | 18000.00 | N/A | 300 | N/A | 0 | N/A | 0 | N/A | 300 | PASS | *Exhaust grille taped off |
| C803 | 8C803 Podiatry (P/TMP/RH) | 109 | 9.0 | 981 | 0.001 | 4.0 | N/A | 3.36 | 3300.00 | N/A | 55 | N/A | 0 | N/A | 18 | N/A | 37 | FAIL | |
| 1E128 | AUDIOLOGY CONTROL RM | 55 | 7.3 | 402 | 0.001 | 6.0 | N/A | 15.84 | 6360.00 | N/A | 106 | N/A | 0 | N/A | 0 | N/A | 106 | PASS | |
| 1E128 | AUDIOLOGY BOOTH | 49 | 6.6 | 321 | -0.001 | 6.0 | N/A | 2.80 | 0.00 | N/A | 0 | N/A | 0 | N/A | 15 | N/A | -15 | FAIL | *No airflow |
| 1E017 | TRIAGE ROOM | 169 | 8.3 | 1,394 | 0.003 | 6.0 | N/A | 10.54 | 10620.00 | N/A | 177 | N/A | 245 | N/A | 0 | N/A | -68 | PASS | |
| 1B147 | SUPPLY ROOM | 90 | 9.0 | 810 | 0.011 | 4.0 | N/A | 8.30 | 6720.00 | N/A | 112 | N/A | 0 | N/A | 0 | N/A | 112 | PASS | |
| 1A-106 A-L | PHARMACY | - | - | 15,713 | 0.015 | 10.0 | N/A | 13.09 | 205620.00 | N/A | 3,427 | N/A | 0 | N/A | 871 | N/A | 2,556 | PASS | |
| CLC204 | RESEARCH | 620 | 8.5 | 5,270 | 0.002 | 6.0 | N/A | 11.13 | 58680.00 | N/A | 978 | N/A | 0 | N/A | 397 | N/A | 581 | PASS | |
| CLC205 | RESEARCH | 152 | 8.5 | 1,292 | 0.006 | 6.0 | N/A | 13.79 | 17820.00 | N/A | 297 | N/A | 0 | N/A | 0 | N/A | 297 | PASS | *No return / exhaust grilles |
| CLC305 | DENTAL OPERATORY | 179 | 8.0 | 1,432 | 0.001 | 6.0 | N/A | 5.07 | 7260.00 | N/A | 121 | N/A | 0 | N/A | 108 | N/A | 13 | FAIL | |
| D912 | VISION | 242 | 8.0 | 1,932 | 0.002 | 6.0 | N/A | 3.60 | 6960.00 | N/A | 116 | N/A | 21 | N/A | 0 | N/A | 95 | FAIL | |
| D912A | OFFICE | 113 | 8.0 | 900 | -0.001 | 6.0 | N/A | 2.60 | 1920.00 | N/A | 32 | N/A | 39 | N/A | 0 | N/A | -7 | FAIL | |
| 13 |
Total Rooms = 13
Grand Total of Rooms Tested = 287 Will include 15 rooms testing, if needed. For Total of 302 Rooms.
| Highly Critical Rooms =71 |
| Lower Critical Rooms = 216 |
2nd FL AUDIOLOGY
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/21/23
| AUDIOLOGY LEVEL 2 | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
Room Volume Cu. Ft. NEG. POS. EQUAL NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| 2A151 | RECEPTION | 281 | 8.0 | 2,244 | 0.003 | 4.0 | N/A | 3.26 | 7320.00 | N/A | 122 | N/A | 0 | N/A | 0 | N/A | 122 | FAIL | *Exhaust grille taped off |
| 2A151A | AUDIOLOGY | 124 | 9.5 | 1,176 | 0.002 | 6.0 | N/A | 5.72 | 6720.00 | N/A | 112 | N/A | 0 | N/A | 0 | N/A | 112 | FAIL | |
| 2A151B | AUDIOLOGY / SPEECH | 198 | 8.0 | 1,584 | 0.000 | 6.0 | N/A | 1.21 | 1920.00 | N/A | 32 | N/A | 52 | N/A | 0 | N/A | -20 | FAIL | |
| 2A151C | AUDIOLOGY / SPEECH | 121 | 9.0 | 1,089 | 0.001 | 6.0 | N/A | 2.26 | 2460.00 | N/A | 41 | N/A | 18 | N/A | 0 | N/A | 23 | FAIL | |
| 2A151D | AUDIOLOGY / SPEECH | 110 | 9.0 | 990 | 0.001 | 6.0 | N/A | 3.76 | 3720.00 | N/A | 62 | N/A | 22 | N/A | 0 | N/A | 40 | FAIL | |
| 2A151E | AUDIOLOGY / SPEECH | 110 | 8.0 | 880 | 0.002 | 6.0 | N/A | 2.86 | 2520.00 | N/A | 42 | N/A | 21 | N/A | 0 | N/A | 21 | FAIL | |
| 2A151F | AUDIOLOGY / SPEECH | 101 | 8.0 | 810 | 0.002 | 6.0 | N/A | 3.33 | 2700.00 | N/A | 45 | N/A | 15 | N/A | 0 | N/A | 30 | FAIL | |
| 2A151G | AUDIOLOGY / SPEECH | 98 | 8.0 | 780 | 0.001 | 6.0 | N/A | 2.54 | 1980.00 | N/A | 33 | N/A | 13 | N/A | 0 | N/A | 20 | FAIL | |
| 2A151H | AUDIOLOGY / SPEECH | 153 | 8.0 | 1,224 | 0.005 | 6.0 | N/A | 4.12 | 5040.00 | N/A | 84 | N/A | 24 | N/A | 0 | N/A | 60 | FAIL | |
| 2A151 G1 | BOOTH 1 CONTROL RM | 80 | 7.3 | 580 | 0.000 | 6.0 | N/A | 1.86 | 1080.00 | N/A | 18 | N/A | 0 | N/A | 0 | N/A | 18 | FAIL | |
| 2A151 G2 | BOOTH 1 | 77 | 6.5 | 501 | 0.000 | 6.0 | N/A | 0.72 | 360.00 | N/A | 6 | N/A | 0 | N/A | 0 | N/A | 6 | FAIL | |
| 2A151 G4 | BOOTH 2 CONTROL RM | 80 | 7.3 | 580 | 0.000 | 6.0 | N/A | 1.76 | 1020.00 | N/A | 17 | N/A | 0 | N/A | 0 | N/A | 17 | FAIL | |
| 2A151 G3 | BOOTH 2 | 78 | 6.5 | 507 | 0.000 | 6.0 | N/A | 0.47 | 240.00 | N/A | 4 | N/A | 0 | N/A | 0 | N/A | 4 | FAIL | |
| 2A151 G6 | BOOTH 3 CONTROL RM | 44 | 6.5 | 286 | 0.000 | 6.0 | N/A | 0.84 | 240.00 | N/A | 4 | N/A | 0 | N/A | 0 | N/A | 4 | FAIL | |
| 2A151 G5 | BOOTH 3 | 54 | 6.5 | 348 | 0.000 | 6.0 | N/A | 2.59 | 900.00 | N/A | 15 | N/A | 0 | N/A | 0 | N/A | 15 | FAIL | |
| 2A157C | SUPPLY ROOM | 202 | 7.8 | 1,566 | 0.002 | 4.0 | N/A | 3.64 | 5700.00 | N/A | 95 | N/A | 0 | N/A | 0 | N/A | 95 | FAIL | |
| 16 |
Total Rooms = 16
2ND FL DENTAL
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/19/23
| DENTAL LEVEL 2 | |||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | |||
| Room Volume | |||||||||
| Cu. Ft. | NEG. |
POS.
EQUAL
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| 2A150 C3 | DENTAL OPERATORY | 144 | 8.8 | 1,260 | 0.006 | 6.0 | N/A | 5.00 | 6300.00 | N/A | 105 | N/A | 0 | N/A | 0 | N/A | 105 | FAIL | |
| 2A150 C6 | DENTAL OPERATORY | 126 | 7.8 | 977 | 0.000 | 6.0 | N/A | 1.47 | 1440.00 | N/A | 24 | N/A | 0 | N/A | 17 | N/A | 7 | FAIL | |
| 2A150 C9 | DENTAL OPERATORY | 126 | 8.0 | 1,008 | 0.001 | 6.0 | N/A | 2.26 | 2280.00 | N/A | 38 | N/A | 0 | N/A | 14 | N/A | 24 | FAIL | |
| 2A150 C10 | DENTAL OPERATORY | 126 | 8.0 | 1,008 | 0.001 | 6.0 | N/A | 1.96 | 1980.00 | N/A | 33 | N/A | 0 | N/A | 17 | N/A | 16 | FAIL | |
| 2A150 C12 | STAFF PERIODONTIST | 138 | 8.8 | 1,208 | 0.014 | 6.0 | N/A | 11.28 | 13620.00 | N/A | 227 | N/A | 0 | N/A | 71 | N/A | 156 | PASS | |
| 2A150 C2 | DENTAL OPERATORY | 125 | 9.0 | 1,125 | 0.001 | 6.0 | N/A | 2.13 | 2400.00 | N/A | 40 | N/A | 0 | N/A | 20 | N/A | 20 | FAIL | |
| 2A150 C5 | DENTAL OPERATORY | 126 | 7.5 | 945 | 0.001 | 6.0 | N/A | 1.84 | 1740.00 | N/A | 29 | N/A | 0 | N/A | 16 | N/A | 13 | FAIL | |
| 2A150 14 | DENTAL OPERATORY | 98 | 8.0 | 784 | 0.055 | 6.0 | N/A | 10.64 | 8340.00 | N/A | 139 | N/A | 0 | N/A | 0 | N/A | 139 | PASS | |
| 2A150 12 | DENTAL OPERATORY | 96 | 8.0 | 768 | 0.008 | 6.0 | N/A | 6.33 | 4860.00 | N/A | 81 | N/A | 0 | N/A | 18 | N/A | 63 | PASS | |
| 2A150 10 | DENTAL OPERATORY | 96 | 8.0 | 768 | 0.002 | 6.0 | N/A | 13.20 | 10140.00 | N/A | 169 | N/A | 0 | N/A | 155 | N/A | 14 | PASS | |
| 2A150 B4 | DENTAL OPERATORY | 110 | 8.0 | 880 | 0.012 | 6.0 | N/A | 12.20 | 10740.00 | N/A | 179 | N/A | 0 | N/A | 20 | N/A | 159 | PASS | |
| 2A152 | MEDICAL SURGICAL SUPPLY RM | 250 | 8.0 | 2,000 | 0.005 | 4.0 | N/A | 2.22 | 4440.00 | N/A | 74 | N/A | 0 | N/A | 15 | N/A | 59 | FAIL | |
| 2A153 | DENTAL HYGIENIST | 131 | 8.0 | 1,048 | 0.003 | 6.0 | N/A | 7.16 | 7500.00 | N/A | 125 | N/A | 0 | N/A | 0 | N/A | 125 | PASS | |
| 2A150C11 | DENTAL OPERATORY | 127 | 8.0 | 1,016 | 0.001 | 6.0 | N/A | 1.18 | 1200.00 | N/A | 20 | N/A | 0 | N/A | 29 | N/A | -9 | FAIL | |
| 2A150 11 | DENTAL OPERATORY | 96 | 8.0 | 768 | 0.008 | 6.0 | N/A | 7.50 | 5760.00 | N/A | 96 | N/A | 0 | N/A | 21 | N/A | 75 | PASS | |
| 2A150 13 | DENTAL OPERATORY | 96 | 8.0 | 768 | 0.011 | 6.0 | N/A | 9.77 | 7500.00 | N/A | 125 | N/A | 0 | N/A | 18 | N/A | 107 | PASS | |
| 2A150 B2 | DENTAL OPERATORY | 116 | 8.0 | 928 | 0.029 | 6.0 | N/A | 10.15 | 9420.00 | N/A | 157 | N/A | 0 | N/A | 15 | N/A | 142 | PASS | |
| 2A150 B8 | EQUIPMENT ROOM | 68 | 9.0 | 612 | -0.001 | 6.0 | N/A | 2.75 | 1680.00 | N/A | 28 | N/A | 0 | N/A | 39 | N/A | -11 | FAIL | *Exhaust grille taped off |
| 2A150 B9 | ORAL SURGERY | 210 | 8.0 | 1,680 | 0.019 | 15.0 | N/A | 9.11 | 15300.00 | N/A | 255 | N/A | 0 | N/A | 35 | N/A | 220 | FAIL | |
| 19 |
Total Rooms = 19
2ND FL RESEARCH
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/18/23
| RESEARCH BRAVO SIDE LEVEL 2 | |||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | |||
| Room Volume | |||||||||
| Cu. Ft. | NEG. |
POS.
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| 2B112 | COPIER ROOM | 80 | 7.8 | 620 | -0.001 | 6.0 | N/A | 0.00 | 0.00 | N/A | 51 | N/A | 83 | N/A | 0 | N/A | -32 | FAIL | |
| 2B11 | OFFICE | 85 | 8.8 | 744 | -0.002 | 6.0 | N/A | 6.13 | 4560.00 | N/A | 0 | N/A | 0 | N/A | 76 | N/A | -76 | PASS | |
| 2B110 | MANAGEMENT | 138 | 8.3 | 1,139 | 0.003 | 6.0 | N/A | 8.85 | 10080.00 | N/A | 168 | N/A | 67 | N/A | 0 | N/A | 101 | PASS | |
| 2B109 | PHSYSIOLOGY LAB | 212 | 9.0 | 1,908 | 0.040 | 6.0 | N/A | 10.35 | 19740.00 | N/A | 329 | N/A | 0 | N/A | 0 | N/A | 329 | PASS | |
| 2B106 | NEUROCHEMISTRY | 203 | 9.0 | 1,827 | 0.000 | 6.0 | N/A | 16.58 | 30300.00 | N/A | 505 | N/A | 0 | N/A | 375 | N/A | 130 | PASS | |
| 2B104 | RESEARCH OFFICE | 214 | 8.5 | 1,819 | 0.024 | 6.0 | N/A | 10.39 | 18900.00 | N/A | 315 | N/A | 0 | N/A | 54 | N/A | 261 | PASS | |
| 2B105 | RESEARCH DATA | 91 | 7.5 | 683 | -0.011 | 6.0 | N/A | 0.00 | 0.00 | N/A | 0 | N/A | 78 | N/A | 0 | N/A | -78 | FAIL | |
| 2B103 | MOLECULAR BIO LAB | 351 | 8.3 | 2,896 | 0.030 | 6.0 | N/A | 8.41 | 24360.00 | N/A | 406 | N/A | 0 | N/A | 25 | N/A | 381 | PASS | |
| 2B102 | RESEARCH FILE | 257 | 11.0 | 2,827 | -0.023 | 6.0 | N/A | 5.69 | 16080.00 | N/A | 85 | N/A | 0 | N/A | 268 | N/A | -183 | FAIL | |
| 2B115 | IRB CHAIR | 105 | 8.0 | 840 | -0.004 | 6.0 | N/A | 9.14 | 7680.00 | N/A | 81 | N/A | 0 | N/A | 128 | N/A | -47 | PASS | |
| 2B130 | SPINAL CORD PHARM LAB | 332 | 8.8 | 2,905 | 0.007 | 6.0 | N/A | 17.10 | 49680.00 | N/A | 828 | N/A | 0 | N/A | 285 | N/A | 543 | PASS | |
| 2B132 | LAB | 244 | 8.0 | 1,952 | -0.009 | 6.0 | N/A | 17.06 | 33300.00 | N/A | 371 | N/A | 0 | N/A | 555 | N/A | -184 | PASS | |
| 2B133 | OFFICE | 333 | 9.0 | 2,997 | 0.003 | 6.0 | N/A | 11.99 | 35940.00 | N/A | 599 | N/A | 145 | N/A | 0 | N/A | 454 | PASS | |
| 2B133A | RESEARCH LAB | 116 | 9.0 | 1,044 | 0.001 | 6.0 | N/A | 6.95 | 7260.00 | N/A | 121 | N/A | 0 | N/A | 55 | N/A | 66 | PASS | |
| 2B133B | OFFICE | 107 | 8.8 | 936 | 0.005 | 6.0 | N/A | 22.49 | 21060.00 | N/A | 351 | N/A | 0 | N/A | 0 | N/A | 351 | PASS | |
| 2B131 | LAB | 86 | 8.8 | 753 | 0.000 | 6.0 | N/A | 25.59 | 19260.00 | N/A | 321 | N/A | 105 | N/A | 100 | N/A | 116 | PASS | |
| 2B129 | SPINAL CORD PHARM LAB | 450 | 8.5 | 3,825 | 0.030 | 6.0 | N/A | 14.98 | 57300.00 | N/A | 955 | N/A | 0 | N/A | 49 | N/A | 906 | PASS | |
| 2B127 | ONCOLOGY RESEARCH | 886 | 9.0 | 7,974 | 0.038 | 6.0 | N/A | 4.36 | 34740.00 | N/A | 579 | N/A | 0 | N/A | 76 | N/A | 503 | FAIL | *Fume hood is offline |
| 2B126 | RESEARCH OFFICE | 106 | 9.0 | 954 | 0.000 | 6.0 | N/A | 6.73 | 6420.00 | N/A | 107 | N/A | 0 | N/A | 0 | N/A | 107 | PASS | |
| 2B125 | RESEARCH OFFICE | 328 | 9.0 | 2,952 | 0.056 | 6.0 | N/A | 13.05 | 38520.00 | N/A | 642 | N/A | 0 | N/A | 0 | N/A | 642 | PASS | *Fume hood is offline |
| 2B117 | CLINICAL LAB | 262 | 8.8 | 2,293 | 0.015 | 6.0 | N/A | 16.10 | 36900.00 | N/A | 615 | N/A | 0 | N/A | 375 | N/A | 240 | PASS | |
| 2B118 | ONCOLOGY RESEARCH LAB | 226 | 9.0 | 2,034 | 0.015 | 6.0 | N/A | 13.33 | 27120.00 | N/A | 452 | N/A | 0 | N/A | 0 | N/A | 452 | PASS | *Fume hood is offline |
| 2B120A | GLASS WASH | 280 | 9.0 | 2,520 | 0.000 | 10.0 | N/A | 12.24 | 30840.00 | N/A | 514 | N/A | 0 | N/A | 655 | N/A | -141 | PASS | |
| 2B120 | GLASS WASH | 74 | 8.0 | 592 | -0.016 | 10.0 | N/A | 0.00 | 0.00 | N/A | 110 | N/A | 0 | N/A | 0 | N/A | 110 | FAIL | *Washer units offline |
| 24 |
Total Rooms = 24
2ND FL RESEARCH (2)
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/18/23
| RESEARCH BRAVO SIDE LEVEL 2 | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG. |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 2B108 | OFFICE | 93 | 8.5 | 791 | 0.000 | 6.0 | N/A | 19.13 | N/A | 252 | N/A | 136 | N/A | 0 | N/A | 116 | PASS | |
| 2B107 | OFFICE | 93 | 8.5 | 791 | 0.005 | 6.0 | N/A | 24.97 | N/A | 329 | N/A | 253 | N/A | 0 | N/A | 76 | PASS | |
| 2B101 | PULMONARY RESEARCH | 328 | 8.5 | 2,788 | 0.004 | 6.0 | N/A | 9.36 | N/A | 435 | N/A | 55 | N/A | 0 | N/A | 380 | PASS | *Fume hood is offline |
| 2B128 | BIOCHEMISTRY | 87 | 9.0 | 783 | 0.056 | 6.0 | N/A | 2.53 | N/A | 33 | N/A | 0 | N/A | 0 | N/A | 33 | FAIL | |
| 2B128A | BIOCHEMISTRY | 262 | 9.0 | 2,358 | 0.002 | 6.0 | N/A | 11.65 | N/A | 458 | N/A | 55 | N/A | 0 | N/A | 403 | PASS | *Fume hood is offline |
| 2B128B | BIOCHEMISTRY | 44 | 8.0 | 352 | 0.006 | 6.0 | N/A | 18.75 | N/A | 110 | N/A | 0 | N/A | 0 | N/A | 110 | PASS | |
| 2B128C | BIOCHEMISTRY | 147 | 9.0 | 1,323 | 0.003 | 6.0 | N/A | 12.70 | N/A | 280 | N/A | 85 | N/A | 0 | N/A | 195 | PASS | *Fume hood is offline |
| 2B128C1 | BIOCHEMISTRY | 75 | 8.8 | 656 | 0.011 | 6.0 | N/A | 17.55 | N/A | 192 | N/A | 100 | N/A | 0 | N/A | 92 | PASS | |
| 2B123 | RESEARCH LAB | 165 | 9.0 | 1,485 | 0.022 | 6.0 | N/A | 13.74 | N/A | 340 | N/A | 0 | N/A | 195 | N/A | 145 | PASS | |
| 2B124 | TISSUE CULTURE | 41 | 8.0 | 328 | 0.001 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 40 | N/A | 0 | N/A | -40 | FAIL | |
| 2B125A | OFFICE | 84 | 8.0 | 672 | 0.055 | 6.0 | N/A | 59.55 | N/A | 667 | N/A | 25 | N/A | 0 | N/A | 642 | PASS | *Fume hood is offline |
| 2B127B | OFFICE | 112 | 8.0 | 896 | 0.001 | 6.0 | N/A | 14.80 | N/A | 221 | N/A | 0 | N/A | 0 | N/A | 221 | PASS | |
| 2B118A | COMMON LAB | 140 | 9.0 | 1,260 | 0.000 | 6.0 | N/A | 9.52 | N/A | 200 | N/A | 404 | N/A | 0 | N/A | -204 | PASS | |
| 2B118B | COMMON LAB | 120 | 9.0 | 1,080 | 0.000 | 6.0 | N/A | 1.11 | N/A | 20 | N/A | 151 | N/A | 0 | N/A | -131 | FAIL | |
| 14 |
Total Rooms =14
2ND FL RESEARCH (3)
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/17/23
| RESEARCH ALPHA SIDE LEVEL 2 | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG. |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 2A127 | BIOCHEMICAL LAB | - | - | 3900 | 0.000 | 6.0 | N/A | 0.95 | N/A | 62 | N/A | 0 | N/A | 0 | N/A | 62 | FAIL | |
| 2A126 | BIOCHEMISTRY LAB | - | - | 3430 | 0.000 | 6.0 | N/A | 4.76 | N/A | 272 | N/A | 0 | N/A | 0 | N/A | 272 | FAIL | *Fume hood is offline |
| 2A103 | RESEARCH FINANCIAL | - | - | 3293 | 0.001 | 6.0 | N/A | 2.24 | N/A | 123 | N/A | 0 | N/A | 47 | N/A | 76 | FAIL | |
| 2A102 | RESEARCH ADMIN | - | - | 1254 | 0.002 | 6.0 | N/A | 4.59 | N/A | 96 | N/A | 0 | N/A | 0 | N/A | 96 | FAIL | |
| 2A102A | CHIEF | - | - | 1025 | 0.001 | 6.0 | N/A | 4.68 | N/A | 80 | N/A | 0 | N/A | 0 | N/A | 80 | FAIL | |
| 2A102B | SCIENCE INF. OFFICE | - | - | 1380 | 0.001 | 6.0 | N/A | 3.91 | N/A | 90 | N/A | 0 | N/A | 0 | N/A | 90 | FAIL | |
| 2A104A | MALE TOILET | - | - | 905 | 0.001 | 6.0 | N/A | 3.25 | N/A | 49 | N/A | 0 | N/A | 80 | N/A | -31 | FAIL | |
| 2A125 | WOMENS LOCKERS | - | - | 926 | 0.001 | 6.0 | N/A | 5.83 | N/A | 90 | N/A | 0 | N/A | 105 | N/A | -15 | FAIL | |
| 2A125A | WOMENS TOILET | - | - | 750 | 0.001 | 6.0 | N/A | 8.00 | N/A | 100 | N/A | 0 | N/A | 109 | N/A | -9 | PASS | |
| 2A105 | 2ND FLOOR | - | - | 640 | 0.005 | 6.0 | N/A | 6.28 | N/A | 67 | N/A | 0 | N/A | 0 | N/A | 67 | PASS | |
| 2A104 | MENS LOCKERS | - | - | 780 | 0.001 | 6.0 | N/A | 3.08 | N/A | 40 | N/A | 0 | N/A | 99 | N/A | -59 | FAIL | |
| 2A110 | CONFERENCE ROOM | - | - | 2475 | 0.002 | 6.0 | N/A | 2.04 | N/A | 84 | N/A | 0 | N/A | 25 | N/A | 59 | FAIL | |
| 2A128 | BIOCHEMISTRY LAB | - | - | 3512 | 0.000 | 6.0 | N/A | 2.08 | N/A | 122 | N/A | 0 | N/A | 0 | N/A | 122 | FAIL | *Fume hood is offline |
| 2A121 | HISTOLOGY RESEARCH LAB | - | - | 4200 | 0.001 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 0 | N/A | 0 | FAIL | *Diffusers capped off |
| 2A123 | ENDO, POLYP, CAN, INST | - | - | 1200 | 0.002 | 6.0 | N/A | 5.00 | N/A | 100 | N/A | 0 | N/A | 0 | N/A | 100 | FAIL | *Exhaust register taped off |
| 2A123A | OFFICE | - | - | 1313 | -0.001 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 0 | N/A | 0 | FAIL | *Exhaust register taped off |
| 2A129 | LABORATORY OFFICE | - | - | 4333 | 0.001 | 6.0 | N/A | 1.59 | N/A | 115 | N/A | 0 | N/A | 0 | N/A | 115 | FAIL | |
| 2A130 | RESEARCH BUDGET OFFICE | - | - | 3713 | 0.006 | 6.0 | N/A | 2.49 | N/A | 154 | N/A | 0 | N/A | 0 | N/A | 154 | FAIL | |
| 2A120 | IMMUNOLOGY RESEARCH | - | - | 1480 | 0.001 | 6.0 | N/A | 9.32 | N/A | 230 | N/A | 0 | N/A | 0 | N/A | 230 | PASS | |
| 2A120A | 2ND FLOOR | - | - | 180 | 0.000 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 90 | N/A | -90 | FAIL | |
| 2A120B | 2ND FLOOR | - | - | 1600 | 0.000 | 6.0 | N/A | 13.13 | N/A | 350 | N/A | 0 | N/A | 0 | N/A | 350 | PASS | |
| M251 | 2ND FLOOR | - | - | 2066 | 0.003 | 6.0 | N/A | 13.65 | N/A | 470 | N/A | 0 | N/A | 0 | N/A | 470 | PASS | |
| 22 |
Total Rooms = 22
2ND FL RESEARCH (4)
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/17/23
| RESEARCH ALPHA SIDE LEVEL 2 | ||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | ||
| Room Volume Cu. Ft. | NEG. |
POS.
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| 2A124 | OFFICE | - | - | 1908 | 0.002 | 6.0 | N/A | 8.74 | N/A | 278 | N/A | 0 | N/A | 0 | N/A | 278 | PASS | |
| 2A111 | HISTOPATOLOGY LAB | - | - | 4593 | -0.001 | 6.0 | N/A | 0.56 | N/A | 398 | N/A | 0 | N/A | 43 | N/A | 355 | FAIL | *Fume hood is offline |
| 2A108 | RESEARCH LAB | - | - | 4020 | 0.001 | 6.0 | N/A | 0.76 | N/A | 51 | N/A | 0 | N/A | 79 | N/A | -28 | FAIL | *Fume hood is offline |
| 2A107 | RESEARCH LAB | - | - | 4530 | 0.014 | 6.0 | N/A | 5.66 | N/A | 427 | N/A | 0 | N/A | 0 | N/A | 427 | FAIL | *Fume hood is offline |
| 2A106 | GULF WAR ILLNESS | - | - | 840 | 0.005 | 6.0 | N/A | 12.86 | N/A | 180 | N/A | 0 | N/A | 0 | N/A | 180 | PASS | |
| 2A114 | CELL CRYPTOPRESERVATION | - | - | 3945 | 0.014 | 6.0 | N/A | 6.31 | N/A | 415 | N/A | 0 | N/A | 55 | N/A | 360 | PASS | |
| 2A122 | TISSUE CULTURE RESEARCH | - | - | 2100 | 0.034 | 6.0 | N/A | 14.91 | N/A | 522 | N/A | 0 | N/A | 0 | N/A | 522 | PASS | *Fume hood is offline |
| 2A122A | TISSUE CULTURE RESEARCH | - | - | 2345 | 0.001 | 6.0 | N/A | 5.88 | N/A | 230 | N/A | 0 | N/A | 0 | N/A | 230 | FAIL | *Fume hood is offline |
| 2A115 | LAB | - | - | 4075 | 0.003 | 6.0 | N/A | 5.01 | N/A | 340 | N/A | 0 | N/A | 75 | N/A | 265 | FAIL | |
| 2A118 | CLINICAL LAB | - | - | 786 | 0.001 | 6.0 | N/A | 3.89 | N/A | 51 | N/A | 0 | N/A | 0 | N/A | 51 | FAIL | |
| 2A116 | IMMUNOLOGY LAB | - | - | 984 | 0.006 | 6.0 | N/A | 9.76 | N/A | 160 | N/A | 0 | N/A | 80 | N/A | 80 | PASS | |
| 11 |
Total Rooms = 11
ADD ISO RM
AIR BALANCE SUMMARY SHEET
PROJECT NAME: DATE:
| ADDITIONASL ISOLATION ROOMS | ||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | ||
| Room Volume Cu. Ft. | NEG. |
POS.
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| B126-1 | HEMO BED STA | - | - | 6390 | 0.041 | 20.0 | N/A | 20.61 | N/A | 2,195 | 0 | 0 | N/A | 990 | N/A | 1,205 | PASS |
| B127-1 | HEMO BED STA | - | - | 6390 | 0.041 | 20.0 | N/A | 20.61 | N/A | 2,195 | 0 | 0 | N/A | 991 | N/A | 1,204 | PASS |
| B128-1 | HEMO BED STA | - | - | 6390 | 0.041 | 20.0 | N/A | 20.61 | N/A | 2,195 | 0 | 0 | N/A | 992 | N/A | 1,203 | PASS |
| C302A4 | MICU PT BED | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,108 | N/A | 472 | FAIL |
| C302A5 | PCU PT BED | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,109 | N/A | 471 | FAIL |
| C302A6 | PCU PT BED | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,110 | N/A | 470 | FAIL |
| 1.00E+78 | ED PT BED | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,111 | N/A | 469 | FAIL |
| 1.00E+78 | ED PT BED | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,112 | N/A | 468 | FAIL |
| 8 |
Total Rooms = 8
3RD FL OR
AIR BALANCE SUMMARY SHEET
PROJECT NAME: DATE:
| 3RD FLOOR OR's | ||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | ||
| Room Volume Cu. Ft. | NEG. |
POS.
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| OR-1 (B354) | OPERATING ROOM | - | - | 6390 | 0.041 | 20.0 | N/A | 20.61 | N/A | 2,195 | 0 | 0 | N/A | 990 | N/A | 1,205 | PASS |
| OR-2 (B351) | OPERATING ROOM | - | - | 6770 | 0.039 | 20.0 | N/A | 20.43 | N/A | 2,305 | 0 | 0 | N/A | 855 | N/A | 1,450 | PASS |
| OR-3 (B340) | OPERATING ROOM | - | - | 5150 | 0.035 | 20.0 | N/A | 27.73 | N/A | 2,380 | 0 | 0 | N/A | 1,235 | N/A | 1,145 | PASS |
| OR-4 (B342) | OPERATING ROOM | - | - | 5450 | 0.057 | 20.0 | N/A | 26.43 | N/A | 2,401 | 0 | 0 | N/A | 908 | N/A | 1,493 | PASS |
| OR-5 (B344) | OPERATING ROOM | - | - | 7740 | 0.057 | 20.0 | N/A | 20.58 | N/A | 2,655 | 0 | 0 | N/A | 1,155 | N/A | 1,500 | PASS |
| OR-6 (B347) | OPERATING ROOM | - | - | 7620 | 0.069 | 20.0 | N/A | 20.51 | N/A | 2,605 | 0 | 0 | N/A | 278 | N/A | 2,327 | PASS |
| OR-7 (B348) | OPERATING ROOM | - | - | 7650 | 0.032 | 20.0 | N/A | 20.02 | N/A | 2,552 | 0 | 0 | N/A | 1,109 | N/A | 1,443 | PASS |
| SUPPORT OR | SUPPORT OR 1 to 4 | - | - | 5630 | -0.002 | 4.0 | N/A | 4.38 | N/A | 398 | 0 | 0 | N/A | 411 | N/A | -13 | PASS |
| OR-8 | OPERATING ROOM | - | - | 5761 | 0.018 | 20.0 | N/A | 16.46 | N/A | 1,580 | 0 | 0 | N/A | 1,108 | N/A | 472 | FAIL |
| OR-9 | OPERATING ROOM | - | - | 5254 | 0.048 | 20.0 | N/A | 16.62 | N/A | 1,455 | 0 | 0 | N/A | 178 | N/A | 1,277 | FAIL |
| B331 | PACU | - | - | 18292 | -0.001 | 4.0 | N/A | 3.92 | N/A | 871 | 0 | 0 | N/A | 1,195 | N/A | -324 | FAIL |
| B331B | BED 10 | - | - | 1018 | -0.015 | 20.0 | N/A | 10.61 | N/A | 67 | 0 | 0 | N/A | 180 | N/A | -113 | FAIL |
| OR-10 (B337B) | OPERATING ROOM | - | - | 3932 | 0.013 | 20.0 | N/A | 23.00 | N/A | 1,507 | 0 | 0 | N/A | 700 | N/A | 807 | PASS |
| OR-11 (B337A) | OPERATING ROOM | - | - | 2187 | 0.015 | 20.0 | N/A | 44.58 | N/A | 1,625 | 0 | 0 | N/A | 1,255 | N/A | 370 | PASS |
| B346C | SUPPORT OR 5 & 6 | - | - | 1312 | -0.003 | 4.0 | N/A | 8.19 | N/A | 108 | 0 | 0 | N/A | 179 | N/A | -71 | PASS |
| 15 |
Total Rooms = 15
3RD FL LAB
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/21/23
| 3RD FLOOR | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG. |
POS.
NR CODE Req'd Design Actual Design Actual Design Actual Design Actual Design Actual
Room # Room Description Area Sq. Ft. Ceiling Height
| A305 | LABORATORY AREA | 9,939 | 8.5 | 84,482 | 0.054 | 6.0 | N/A | 7.69 | N/A | 10,821 | N/A | N/A | 5,507 | N/A | 5,314 | PASS | |
| A305Q | COPY & FILE ROOM | 223 | 8.8 | 1,951 | 0.036 | 4.0 | N/A | 5.97 | N/A | 194 | N/A | N/A | 232 | N/A | -38 | PASS | |
| A306A | READING ROOM | 63 | 8.8 | 551 | 0.014 | 4.0 | N/A | 30.91 | N/A | 284 | N/A | N/A | 165 | N/A | 119 | PASS | |
| A305A | BLOOD DRAWING | 87 | 9.0 | 783 | -0.014 | 6.0 | N/A | 14.41 | N/A | 139 | N/A | N/A | 188 | N/A | -49 | PASS | |
| A305K | N/A | 368 | 7.8 | 2,852 | 0.015 | 4.0 | N/A | 19.61 | N/A | 932 | N/A | N/A | 0 | N/A | 932 | PASS | *No exhaust / return grilles |
| B325B | SPS PREP AREA | 1,399 | 9.0 | 12,591 | -0.006 | 4 | N/A | 0.12 | N/A | 915 | N/A | N/A | 25 | N/A | 890 | FAIL | *No fume hood or other exhaust equipment found |
| B325E | SPS STERILIZATION RM | 713 | 8.0 | 5,704 | 0 | 10 | N/A | 6.89 | N/A | 655 | N/A | N/A | 1,339 | N/A | -684 | FAIL | *Door not installed |
| B325 | SPS PASS THRU | 252 | 8.0 | 2,016 | 0.001 | 4 | N/A | 0.00 | N/A | 0 | N/A | N/A | 10 | N/A | -10 | FAIL | |
| A305P | FILES | 175 | 8.0 | 1,400 | 0.000 | 4 | N/A | 5.96 | N/A | 139 | N/A | N/A | 130 | N/A | 9 | PASS | |
| A305M | ELECTRON MICROSCOPE | 189 | 8.0 | 1,512 | 0.000 | 4 | N/A | 7.78 | N/A | 196 | N/A | N/A | 293 | N/A | -97 | PASS | |
| A305N | GROSSING | 227 | 8.0 | 1,816 | 0.004 | 4 | N/A | 21.57 | N/A | 653 | N/A | N/A | 55 | N/A | 598 | PASS | |
| B325 | SPS OR ENTRANCE | 276 | 8.0 | 2,208 | 0 | 4 | N/A | 0.00 | N/A | 0 | N/A | N/A | 15 | N/A | -15 | FAIL | |
| B325D5 | DECONTAMINATION | 850 | 8.0 | 6,800 | -0.011 | 6.0 | N/A | 9.75 | N/A | 215 | N/A | N/A | 1,105 | N/A | -890 | PASS | |
| B317A | MEDICATION PREP. | 112 | 8.0 | 896 | 0.005 | 6.0 | N/A | 19.75 | N/A | 295 | N/A | N/A | 119 | N/A | 176 | PASS | *Room under construction |
| 14 | |||||||||||||||||
| *Doors removed from several sps areas |
Total Rooms = 14
12C WING
AIR BALANCE SUMMARY SHEET
| PROJECT NAME: | BRUCE W CARTER VETERANS AFFAIR MEDICAL CENTER | DATE: | 4/17/23 | |||||||||||||
| TEST & BALANCE - BUILDING 1 | ||||||||||||||||
| PATIENT ROOMS LEVEL 12 | ||||||||||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | ||||||||||
| Room Volume Cu. Ft. | NEG. POS. EQUAL NR | CODE Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| C1201A | PATIENT ROOM | 171 | 7.5 | 1,283 | EQUAL | 4.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 0 | N/A | 0 | FAIL | *Under construction | |
| C1202 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 2.84 | N/A | 39 | N/A | 59 | N/A | 58 | N/A | -78 | FAIL | ||
| C1206 | PATIENT ROOM | 157 | 7.5 | 1,178 | EQUAL | 4.0 | N/A | 4.94 | N/A | 97 | N/A | 55 | N/A | 55 | N/A | -13 | PASS | ||
| C1207 | PATIENT ROOM | 171 | 7.5 | 1,283 | EQUAL | 4.0 | N/A | 2.20 | N/A | 25 | N/A | 42 | N/A | 47 | N/A | -64 | FAIL | ||
| C1208 | PATIENT ROOM | 172 | 7.5 | 1,290 | EQUAL | 4.0 | N/A | 5.35 | N/A | 115 | N/A | 38 | N/A | 70 | N/A | 7 | PASS | ||
| C1209 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 1.76 | N/A | 37 | N/A | 22 | N/A | 38 | N/A | -23 | FAIL | ||
| C1210 | PATIENT ROOM | 169 | 7.5 | 1,268 | EQUAL | 4.0 | N/A | 1.85 | N/A | 21 | N/A | 39 | N/A | 15 | N/A | -33 | FAIL | ||
| C1211 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 7.48 | N/A | 157 | N/A | 35 | N/A | 36 | N/A | 86 | PASS | ||
| C1212 | PATIENT ROOM | 169 | 7.5 | 1,268 | EQUAL | 4.0 | N/A | 0.71 | N/A | 0 | N/A | 0 | N/A | 15 | N/A | -15 | FAIL | ||
| C1213 | PATIENT ROOM | 170 | 7.5 | 1,275 | EQUAL | 4.0 | N/A | 2.07 | N/A | 44 | N/A | 25 | N/A | 30 | N/A | -11 | FAIL | ||
| C1214 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 8.90 | N/A | 25 | N/A | 187 | N/A | 44 | N/A | -206 | PASS | ||
| C1215 | PATIENT ROOM | 167 | 7.5 | 1,253 | EQUAL | 4.0 | N/A | 7.66 | N/A | 18 | N/A | 160 | N/A | 10 | N/A | -152 | PASS | ||
| C1216 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 20.77 | N/A | 0 | N/A | 431 | N/A | 12 | N/A | -443 | PASS | *No supply grille installed | |
| C1217 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 6.89 | N/A | 36 | N/A | 143 | N/A | 18 | N/A | -125 | PASS | ||
| C1218 | PATIENT ROOM | 165 | 7.5 | 1,238 | EQUAL | 4.0 | N/A | 3.30 | N/A | 40 | N/A | 68 | N/A | 20 | N/A | -48 | FAIL | ||
| C1219 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 5.54 | N/A | 55 | N/A | 115 | N/A | 15 | N/A | -75 | PASS | ||
| C1220 | PATIENT ROOM | 167 | 7.5 | 1,253 | EQUAL | 4.0 | N/A | 2.16 | N/A | 18 | N/A | 45 | N/A | 10 | N/A | -37 | FAIL | ||
| C1221 | PATIENT ROOM | 164 | 7.5 | 1,230 | EQUAL | 4.0 | N/A | 21.85 | N/A | 25 | N/A | 448 | N/A | 18 | N/A | -441 | PASS | ||
| C1222 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 13.64 | N/A | 33 | N/A | 283 | N/A | 35 | N/A | -285 | PASS | ||
| 19 |
Total Rooms = 19
12D WING
AIR BALANCE SUMMARY SHEET
| PROJECT NAME: | BRUCE W CARTER VETERANS AFFAIR MEDICAL CENTER | DATE: | 4/17/23 | |||||||||||||
| TEST & BALANCE - BUILDING 1 | ||||||||||||||||
| PATIENT ROOMS LEVEL 11 | ||||||||||||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS or FAIL | ||||||||||
| Room Volume Cu. Ft. | NEG. POS. EQUAL NR | CODE Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| D1202 | PATIENT ROOM | 170 | 7.5 | 1,275 | EQUAL | 4.0 | N/A | 1.32 | N/A | 28 | N/A | 35 | N/A | 48 | N/A | -55 | FAIL | ||
| D1203 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 4.86 | N/A | 102 | N/A | 22 | N/A | 96 | N/A | -16 | PASS | ||
| D1204 | PATIENT ROOM | 159 | 7.5 | 1,193 | EQUAL | 4.0 | N/A | 15.60 | N/A | 310 | N/A | 18 | N/A | 15 | N/A | 277 | PASS | ||
| D1205 | PATIENT ROOM | 171 | 7.5 | 1,283 | EQUAL | 4.0 | N/A | 10.53 | N/A | 225 | N/A | 55 | N/A | 55 | N/A | 115 | PASS | ||
| D1206 | PATIENT ROOM | 169 | 7.5 | 1,268 | EQUAL | 4.0 | N/A | 0.47 | N/A | 10 | N/A | 0 | N/A | 10 | N/A | 0 | FAIL | ||
| D1207 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 4.14 | N/A | 87 | N/A | 65 | N/A | 14 | N/A | 8 | PASS | ||
| D1208 | PATIENT ROOM | 169 | 7.5 | 1,268 | EQUAL | 4.0 | N/A | 5.11 | N/A | 108 | N/A | 72 | N/A | 20 | N/A | 16 | PASS | ||
| D1209 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 4.10 | N/A | 55 | N/A | 86 | N/A | 0 | N/A | -31 | PASS | ||
| D1210 | PATIENT ROOM | 169 | 7.5 | 1,268 | EQUAL | 4.0 | N/A | 8.85 | N/A | 187 | N/A | 112 | N/A | 0 | N/A | 75 | PASS | ||
| D1211 | PATIENT ROOM | 170 | 7.5 | 1,275 | EQUAL | 4.0 | N/A | 3.06 | N/A | 65 | N/A | 46 | N/A | 26 | N/A | -7 | FAIL | ||
| D1212 | PATIENT ROOM | 168 | 7.5 | 1,260 | EQUAL | 4.0 | N/A | 2.14 | N/A | 45 | N/A | 45 | N/A | 10 | N/A | -10 | FAIL | ||
| D1213 | PATIENT ROOM | 167 | 7.5 | 1,253 | EQUAL | 4.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 0 | N/A | 0 | FAIL | *Under construction | |
| D1214 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 3.18 | N/A | 25 | N/A | 66 | N/A | 55 | N/A | -96 | FAIL | ||
| D1215 | PATIENT ROOM | 166 | 7.5 | 1,245 | EQUAL | 4.0 | N/A | 2.31 | N/A | 48 | N/A | 40 | N/A | 60 | N/A | -52 | FAIL | ||
| D1216 | PATIENT ROOM | 165 | 7.5 | 1,238 | EQUAL | 4.0 | N/A | 5.09 | N/A | 0 | N/A | 45 | N/A | 105 | N/A | -150 | PASS | ||
| 15 |
Total Rooms = 15
ANIMAL RESEARCH 1ST FL
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/14/23
| BLDG. 7 ANIMAL RESEARCH 1ST FLOOR | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 115 | STORAGE ROOM | - | - | 1336 | 0.002 | 4.0 | N/A | 4.54 | N/A | 101 | N/A | 0 | N/A | 89 | N/A | 12 | PASS | |
| 117 | STORAGE ROOM | - | - | 1215 | -0.002 | 4.0 | N/A | 4.59 | N/A | 88 | N/A | 0 | N/A | 93 | N/A | -5 | PASS | |
| 120 | NECROPSY | - | - | 1638 | 0.001 | 15.0 | N/A | 1.83 | N/A | 50 | N/A | 0 | N/A | 45 | N/A | 5 | FAIL | |
| 123 | STORAGE ROOM | - | - | 1215 | -0.200 | 4.0 | N/A | 27.46 | N/A | 63 | N/A | 0 | N/A | 556 | N/A | -493 | PASS | *Temporary AC unit installed in exhaust |
| 124 | PROCEDURE LAB | - | - | 1275 | 0.001 | 15.0 | N/A | 4.09 | N/A | 87 | N/A | 0 | N/A | 35 | N/A | 52 | FAIL | |
| 126 | VETERINARIAN | - | - | 1275 | 0.001 | 4.0 | N/A | 3.11 | N/A | 66 | N/A | 0 | N/A | 15 | N/A | 51 | FAIL | |
| 127 | SUPERVISOR | - | - | 1336 | 0.001 | 4.0 | N/A | 2.60 | N/A | 58 | N/A | 0 | N/A | 0 | N/A | 58 | FAIL | |
| 128 | ANIMAL RECEIVING | - | - | 2457 | 0.006 | 10.0 | N/A | 2.10 | N/A | 86 | N/A | 0 | N/A | 36 | N/A | 50 | FAIL | |
| 136 | ANIMAL WARD | - | - | 4725 | 0.002 | 15.0 | N/A | 2.97 | N/A | 234 | N/A | 0 | N/A | 190 | N/A | 44 | FAIL | *Temporary AC unit installed in exhaust |
| 139 | PROCEDURE LAB | - | - | 2457 | -0.035 | 15.0 | N/A | 16.02 | N/A | 261 | N/A | 0 | N/A | 656 | N/A | -395 | PASS | |
| 116 | ANIMAL WARD | - | - | 4725 | 0.000 | 15.0 | N/A | 2.68 | N/A | 211 | N/A | 0 | N/A | 212 | N/A | -1 | FAIL | |
| 110 | OPERATING ROOM | - | - | 2475 | 0.007 | 15.0 | N/A | 3.88 | N/A | 160 | N/A | 0 | N/A | 0 | N/A | 160 | FAIL | |
| 108 | OPERATING ROOM | - | - | 2474 | 0.006 | 15.0 | N/A | 4.00 | N/A | 165 | N/A | 0 | N/A | 69 | N/A | 96 | FAIL | |
| 109 | PRE-OPERATION ROOM | - | - | 1080 | 0.000 | 4.0 | N/A | 4.50 | N/A | 81 | N/A | 0 | N/A | 61 | N/A | 20 | PASS | |
| 107 | POST-OPERATION ROOM | - | - | 1080 | 0.000 | 10.0 | N/A | 4.61 | N/A | 83 | N/A | 0 | N/A | 67 | N/A | 16 | FAIL | |
| 106 | CONFERENCE ROOM | - | - | 2079 | -0.001 | 4.0 | N/A | 2.91 | N/A | 56 | N/A | 0 | N/A | 101 | N/A | -45 | FAIL | |
| 149 | STORAGE ROOM | - | - | 2457 | -0.004 | 4.0 | N/A | 6.03 | N/A | 245 | N/A | 0 | N/A | 247 | N/A | -2 | PASS | |
| 148 | STORAGE ROOM | - | - | 1170 | 0.033 | 4.0 | N/A | 3.85 | N/A | 75 | N/A | 0 | N/A | 36 | N/A | 39 | FAIL | |
| 148A | STORAGE ROOM | - | - | 1170 | 0.002 | 4.0 | N/A | 5.03 | N/A | 98 | N/A | 0 | N/A | 76 | N/A | 22 | PASS | |
| 147 | STORAGE ROOM | - | - | 2268 | -0.006 | 4.0 | N/A | 4.68 | N/A | 109 | N/A | 0 | N/A | 177 | N/A | -68 | PASS | *Diffuser taped off |
| 144 | PROCEDURE LAB | - | - | 2457 | -0.004 | 15.0 | N/A | 12.94 | N/A | 246 | N/A | 0 | N/A | 530 | N/A | -284 | FAIL | |
| 143 | ANIMAL WARD | - | - | 2457 | 0.001 | 15.0 | N/A | 4.96 | N/A | 203 | N/A | 0 | N/A | 152 | N/A | 51 | FAIL | |
| 22 |
Total Rooms = 22
ANIMAL RESEARCH 1ST FL (2)
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/13/23
| BLDG. 7 ANIMAL RESEARCH 1ST FLOOR | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 142 | PROCEDURAL ROOM | - | - | 2457 | 0.004 | 15.0 | N/A | 4.74 | N/A | 194 | N/A | 0 | N/A | 129 | N/A | 65 | FAIL | |
| 140 | ANIMAL WARD | - | - | 2457 | -0.002 | 15.0 | N/A | 2.71 | N/A | 270 | N/A | 0 | N/A | 277 | N/A | -7 | FAIL | |
| 138 | STORAGE ROOM | - | - | 2457 | 0.000 | 4.0 | N/A | 1.86 | N/A | 76 | N/A | 0 | N/A | 111 | N/A | -35 | FAIL | *Light removed from ceiling, creating large opening |
| 137 | CAGE WASHING ROOM | - | - | 2547 | 0.055 | 20.0 | N/A | 14.96 | N/A | 635 | N/A | 0 | N/A | 0 | N/A | 635 | FAIL | *Autoclave down for repair. Large opening in ceiling |
| 125 | QUARANTINE | - | - | 889 | 0.008 | 15.0 | N/A | 13.36 | N/A | 198 | N/A | 0 | N/A | 101 | N/A | 97 | FAIL | |
| 122 | FEED ROOM | - | - | 525 | 0.001 | 4.0 | N/A | 7.43 | N/A | 65 | N/A | 0 | N/A | 0 | N/A | 65 | PASS | |
| 121 | STORAGE ROOM | - | - | 551 | 0.016 | 4.0 | N/A | 22.76 | N/A | 209 | N/A | 0 | N/A | 0 | N/A | 209 | PASS | |
| 112 | STORAGE ROOM | - | - | 2452 | 0.000 | 4.0 | N/A | 5.16 | N/A | 211 | N/A | 0 | N/A | 161 | N/A | 50 | PASS | |
| 111 | AUTOCLAVE | - | - | 1066 | 0.002 | 4.0 | N/A | 8.11 | N/A | 144 | N/A | 0 | N/A | 88 | N/A | 56 | PASS | |
| N/A | OR WASHING ROOM | - | - | 891 | -0.004 | 4.0 | N/A | 16.84 | N/A | 190 | N/A | 0 | N/A | 250 | N/A | -60 | PASS | |
| 135 | CLEAN CAGE ROOM | - | - | 4061 | 0.055 | 6.0 | N/A | 9.09 | N/A | 615 | N/A | 0 | N/A | 0 | N/A | 615 | PASS | *Cage wash under repair. Unable to take measurement |
| 134 | LOUNGE | - | - | 1179 | -0.022 | 6.0 | N/A | 1.22 | N/A | 0 | N/A | 0 | N/A | 24 | N/A | -24 | FAIL | |
| CORRIDOR | 1ST FLOOR | - | - | 15300 | 0.007 | 4.0 | N/A | 2.00 | N/A | 510 | N/A | 0 | N/A | 0 | N/A | 510 | FAIL | |
| NOTE: 1*: | 13 |
Total Rooms = 13
ANIMAL RESEARCH 3RD FL
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/12/23
| BLDG. 7 ANIMAL RESEARCH 3RD FLOOR | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 307 | TISSUE CULTURE | - | - | 1384 | 0.001 | 15.0 | N/A | 4.77 | N/A | 110 | N/A | 0 | N/A | 175 | N/A | -65 | FAIL | |
| 308 | PHYSIOLOGICAL LAB | - | - | 3369 | -0.019 | 15.0 | N/A | 10.93 | N/A | 304 | N/A | 0 | N/A | 614 | N/A | -310 | FAIL | |
| 309 | PHYSIOLOGICAL LAB | - | - | 3380 | 0.001 | 15.0 | N/A | 7.49 | N/A | 422 | N/A | 0 | N/A | 675 | N/A | -253 | FAIL | |
| 312 | BIOCHEMICAL LAB | - | - | 3377 | 0.002 | 15.0 | N/A | 5.54 | N/A | 312 | N/A | 0 | N/A | 738 | N/A | -426 | FAIL | |
| 313 | BIOCHEMICAL LAB | - | - | 3377 | 0.000 | 15.0 | N/A | 3.86 | N/A | 217 | N/A | 0 | N/A | 731 | N/A | -514 | FAIL | |
| 316 | RESEARCH LAB | - | - | 3377 | 0.008 | 15.0 | N/A | 4.12 | N/A | 232 | N/A | 0 | N/A | 712 | N/A | -480 | FAIL | |
| 318 | BIOCHEMICAL LAB | - | - | 3322 | 0.005 | 15.0 | N/A | 3.38 | N/A | 187 | N/A | 0 | N/A | 0 | N/A | 187 | FAIL | *Fume hood not working |
| 324 | NEPHROLOGY RESEARCH | - | - | 3541 | 0.022 | 15.0 | N/A | 5.98 | N/A | 353 | N/A | 0 | N/A | 877 | N/A | -524 | FAIL | |
| 321 | PHYSIOLOGICAL LAB | - | - | 1734 | -0.001 | 15.0 | N/A | 12.91 | N/A | 80 | N/A | 0 | N/A | 373 | N/A | -293 | FAIL | |
| 320 | PHYSIOLOGICAL LAB | - | - | 1661 | -0.021 | 15.0 | N/A | 16.07 | N/A | 211 | N/A | 0 | N/A | 445 | N/A | -234 | PASS | |
| 315 | BIOCHEMICAL LAB | - | - | 2599 | 0.006 | 15.0 | N/A | 6.81 | N/A | 295 | N/A | 0 | N/A | 0 | N/A | 295 | FAIL | |
| 314 | PHYSIOLOGICAL LAB | - | - | 5301 | -0.024 | 15.0 | N/A | 13.13 | N/A | 421 | N/A | 0 | N/A | 1160 | N/A | -739 | FAIL | |
| 311 | GENERAL LAB | - | - | 5301 | -0.004 | 15.0 | N/A | 13.84 | N/A | 290 | N/A | 0 | N/A | 1223 | N/A | -933 | FAIL | |
| 310 | HISTOPATOLOGY, ONC LAB | - | - | 3649 | -0.008 | 15.0 | N/A | 14.75 | N/A | 415 | N/A | 0 | N/A | 897 | N/A | -482 | FAIL | |
| 344 | OPHTALMOLOGY, ENVI RE | - | - | 3181 | 0.006 | 15.0 | N/A | 3.70 | N/A | 196 | N/A | 0 | N/A | 670 | N/A | -474 | FAIL | |
| 342 | PEPTIDE RECEPTOR LAB | - | - | 3349 | 0.004 | 15.0 | N/A | 3.28 | N/A | 183 | N/A | 0 | N/A | 610 | N/A | -427 | FAIL | |
| 341 | MOLECULAR BIO INST | - | - | 3350 | 0.005 | 15.0 | N/A | 6.59 | N/A | 368 | N/A | 0 | N/A | 441 | N/A | -73 | FAIL | |
| 340 | STEM CELL BIOLOGY | - | - | 3422 | 0.000 | 15.0 | N/A | 2.89 | N/A | 165 | N/A | 0 | N/A | 584 | N/A | -419 | FAIL | |
| 339 | BRAIN DEV, MND LAB | - | - | 3422 | 0.002 | 15.0 | N/A | 3.23 | N/A | 184 | N/A | 0 | N/A | 561 | N/A | -377 | FAIL | |
| 338 | CMJ STEM CELL, BRAIN RE | - | - | 3422 | 0.005 | 15.0 | N/A | 5.17 | N/A | 295 | N/A | 0 | N/A | 451 | N/A | -156 | FAIL | |
| 335 | SKELETAL CELL, MOL BIO RM | - | - | 3422 | 0.005 | 15.0 | N/A | 4.63 | N/A | 264 | N/A | 0 | N/A | 595 | N/A | -331 | FAIL | |
| 333 | PROSTATE CANCER | - | - | 3422 | 0.003 | 15.0 | N/A | 2.52 | N/A | 144 | N/A | 0 | N/A | 475 | N/A | -331 | FAIL | |
| 332 | GLASS WASH, AUTO CL RM | - | - | 2473 | -0.08 | 15.0 | N/A | 35.18 | N/A | 1,060 | N/A | 0 | N/A | 1450 | N/A | -390 | PASS | |
| 23 |
Total Rooms = 23
ANIMAL RESEARCH 3RD FL (2)
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/12/23
| BLDG. 7 ANIMAL RESEARCH 3RD FLOOR | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 331 | COMMON INSTRUMENTS | - | - | 2198 | -0.008 | 15.0 | N/A | 16.60 | N/A | 130 | N/A | 0 | N/A | 608 | N/A | -478 | PASS |
| 334 | CELL CRYOPRESERVATION | - | - | 1764 | -0.045 | 15.0 | N/A | 39.12 | N/A | 134 | N/A | 0 | N/A | 1150 | N/A | -1,016 | PASS |
| 336 | CELL, TISSUE CULTURE | - | - | 1658 | 0.023 | 15.0 | N/A | 12.01 | N/A | 332 | N/A | 0 | N/A | 445 | N/A | -113 | FAIL |
| 337 | REARCH CLEAN ROOM | - | - | 3484 | -0.064 | 15.0 | N/A | 31.86 | N/A | 970 | N/A | 0 | N/A | 1850 | N/A | -880 | PASS |
| 345 | RESEARCH LAB | - | - | 2630 | -0.003 | 15.0 | N/A | 13.14 | N/A | 305 | N/A | 0 | N/A | 576 | N/A | -271 | FAIL |
| CORRIDOR | CORRIDOR | - | - | 20498 | 0.002 | 4.0 | N/A | 3.56 | N/A | 1,215 | N/A | 0 | N/A | 398 | N/A | 817 | FAIL |
| 6 |
Total Rooms = 6
BILL KLING CLINIC
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/20/23
| BILL KLING (BROWARD) CLINIC | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| SP139 | PREP AREA | 383 | 8.5 | 3,256 | 0.002 | 4.0 | N/A | 3.69 | N/A | 200 | N/A | 0 | N/A | 29 | N/A | 171 | FAIL | |
| DC112A | DENTAL CLEAN STORAGE | 43 | 8.3 | 355 | 0.001 | 4.0 | N/A | 19.28 | N/A | 114 | N/A | 65 | N/A | 0 | N/A | 49 | PASS | |
| SM123C | SCOPE STERILIZATION ROOM | 54 | 9.0 | 486 | -0.004 | 6.0 | N/A | 54.32 | N/A | 99 | N/A | 0 | N/A | 440 | N/A | -341 | PASS | |
| SM124 | SCOPE CLEAN ROOM | 57 | 9.0 | 517 | 0.001 | 4.0 | N/A | 5.81 | N/A | 50 | N/A | 0 | N/A | 0 | N/A | 50 | PASS | |
| PC127 | STORAGE ROOM | 77 | 9.0 | 693 | 0.001 | 4.0 | N/A | 4.33 | N/A | 50 | N/A | 0 | N/A | 0 | N/A | 50 | PASS | |
| DE107 | STORAGE ROOM | 66 | 9.0 | 594 | 0.001 | 4.0 | N/A | 4.44 | N/A | 44 | N/A | 0 | N/A | 0 | N/A | 44 | PASS | |
| AD108 | BOOTH 1 STAFF | 72 | 6.5 | 468 | -0.001 | 6.0 | N/A | 0.00 | N/A | 50 | N/A | 60 | N/A | 0 | N/A | -10 | FAIL | |
| AD108 | BOOTH 1 PATIENT | 72 | 6.5 | 468 | -0.001 | 6.0 | N/A | 0.00 | N/A | 50 | N/A | 60 | N/A | 0 | N/A | -10 | FAIL | |
| AD108 | BOOTH 2 STAFF | 107 | 6.5 | 696 | -0.006 | 6.0 | N/A | 0.00 | N/A | 52 | N/A | 101 | N/A | 0 | N/A | -49 | FAIL | |
| AD108 | BOOTH 2 PATIENT | 106 | 6.5 | 689 | 0 | 6.0 | N/A | 4.09 | N/A | 47 | N/A | 60 | N/A | 0 | N/A | -13 | FAIL | |
| AD108 | BOOTH 3 STAFF | 55 | 6.5 | 358 | -0.009 | 6.0 | N/A | 0.00 | N/A | 46 | N/A | 112 | N/A | 0 | N/A | -66 | FAIL | |
| AD108 | BOOTH 3 PATIENT | 55 | 6.5 | 358 | 0.003 | 6.0 | N/A | 20.31 | N/A | 121 | N/A | 61 | N/A | 0 | N/A | 60 | PASS | |
| AD101 | RECEPTION | 111 | 9.9 | 1,099 | 0.001 | 4.0 | N/A | 5.41 | N/A | 99 | N/A | 58 | N/A | 0 | N/A | 41 | PASS | |
| AD102 | OFFICE / TREATMENT | 135 | 9.0 | 1,215 | 0.001 | 4.0 | N/A | 2.17 | N/A | 44 | N/A | 23 | N/A | 0 | N/A | 21 | FAIL | |
| AD106 | OFFICE / TREATMENT | 111 | 9.0 | 999 | -0.009 | 4.0 | N/A | 0.00 | N/A | 58 | N/A | 130 | N/A | 0 | N/A | -72 | FAIL | |
| AD107 | OFFICE / TREATMENT | 146 | 9.0 | 1,314 | 0.003 | 4.0 | N/A | 2.74 | N/A | 60 | N/A | 16 | N/A | 0 | N/A | 44 | FAIL | |
| SM132 | EXAM ROOM 4 | 156 | 9.0 | 1,404 | 0.002 | 4.0 | N/A | 3.38 | N/A | 79 | N/A | 40 | N/A | 0 | N/A | 39 | FAIL | |
| SP138 | CLEAN STERILE STORAGE | 199 | 9.0 | 1,791 | 0.002 | 4.0 | N/A | 2.28 | N/A | 68 | N/A | 0 | N/A | 112 | N/A | -44 | FAIL | |
| SP137 | DECONTAMINATION | 182 | 9.0 | 1,641 | -0.022 | 6.0 | N/A | 9.33 | N/A | 73 | N/A | 0 | N/A | 255 | N/A | -182 | PASS | |
| RD104 | RADIOLOGY RIGHT ROOM | 108 | 9.5 | 1,026 | -0.002 | 6.0 | N/A | 23.80 | N/A | 263 | N/A | 0 | N/A | 407 | N/A | -144 | PASS | *No barrier between spaces |
| RD104 | RADIOLOGY LEFT ROOM | 240 | 9.0 | 2,160 | -0.002 | 6.0 | N/A | 14.64 | N/A | 164 | N/A | 0 | N/A | 527 | N/A | -363 | PASS | *No barrier between spaces |
| Site Address: Broward Clinic 9800 West Commercial Blvd Sunrise, 33351, FL | 21 |
Total Rooms = 21
PEMBROKE CLINIC
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/20/23
| PEMBROKE CLINIC | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| Corr | Audiology Booth Staff | 53 | 7.3 | 384 | 0 | 6.0 | N/A | 1.87 | N/A | 12 | N/A | 0 | N/A | 19 | N/A | -7 | FAIL |
| Corr | Audiology Booth Patient | 48 | 6.5 | 312 | 0 | 6.0 | N/A | 5.58 | N/A | 29 | N/A | 0 | N/A | 12 | N/A | 17 | FAIL |
| Site Address: Pembroke Pines 7369 Sheridan St, Hollywood, FL, 33024 | 2 |
Total Rooms = 2
HOLLYWOOD CLINIC
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/20/23
| HOLLYWOOD CLINIC | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 225 | Audiology | 204 | 8.5 | 1,734 | 0 | 6.0 | N/A | 6.16 | N/A | 178 | N/A | 155 | N/A | 0 | N/A | 23 | PASS |
| 225 | Audiology Booth | 45 | 6.5 | 293 | -0.001 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 71 | N/A | 0 | N/A | -71 | FAIL |
| 226 | Audiology | 108 | 8.5 | 918 | -0.003 | 6.0 | N/A | 0.00 | N/A | 80 | N/A | 208 | N/A | 0 | N/A | -128 | FAIL |
| 227 | Audiology | 152 | 8.5 | 1,292 | -0.001 | 6.0 | N/A | 0.00 | N/A | 212 | N/A | 235 | N/A | 0 | N/A | -23 | FAIL |
| 227 | Audiology Booth | 47 | 6.5 | 306 | 0 | 6.0 | N/A | 9.62 | N/A | 49 | N/A | 42 | N/A | 0 | N/A | 7 | PASS |
| 228 | Audiology | 97 | 8.5 | 825 | 0 | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 0 | N/A | 0 | N/A | 0 | FAIL |
| Site Address: Hollywood 3702 Washington St, Hollywood, FL, 33021 | 6 |
Total Rooms = 6
DEERFIELD CLINIC
AIR BALANCE SUMMARY SHEET
PROJECT NAME: 4/20/23
| DEERFIELD CLINIC | |||||||
| Pressure | Air Changes | Supply | Return | Exhaust | Inf. / Exf. | PASS |
or
FAIL
| Room Volume | ||
| Cu. Ft. | NEG |
POS.
EQUAL
| NR | CODE | |||||||||
| Req'd | Design | Actual | Design | Actual | Design | Actual | Design | Actual | Design | Actual |
Room # Room Description Area Sq. Ft. Ceiling Height
| 110 | Audiology | 93 | 10.0 | 930 | EQUAL | 6.0 | N/A | 4.84 | N/A | 75 | N/A | 0 | N/A | 0 | N/A | 23 | FAIL | *No return grille |
| 113 | Audiology | 113 | 10.0 | 1,130 | EQUAL | 6.0 | N/A | 0.27 | N/A | 5 | N/A | 0 | N/A | 0 | N/A | 24 | FAIL | *No return grille |
| 115 | Audiology | 176 | 10.0 | 1,760 | EQUAL | 6.0 | N/A | 9.72 | N/A | 285 | N/A | 70 | N/A | 0 | N/A | 25 | PASS | |
| 115 | Audiology Booth | 51 | 6.5 | 332 | EQUAL | 6.0 | N/A | 0.00 | N/A | 0 | N/A | 55 | N/A | 0 | N/A | 26 | FAIL | |
| Site Address: Deerfield Beach 2100 SW 10th St, Deerfield Beach, FL, 33442 | 4 |
Total Rooms = 4
ABBREVIATIONS TERMINOLOGIES
| ABBREVIATIONS & TERMINOLOGY | ABBREVIATIONS & TERMINOLOGY | ||||||
| A: | Nameplate Amps | EAT: | Entering Air Temperatures | HVAC: | Heating Ventilating & Air | RND: | Round |
| AHU: | Air Handling Unit | ECON: | Economizer - Controls & Components | Conditioning | RA: | Return Air | |
| AC: | Air Conditioner | that allow an Air Handler to logically | HWR: | Hot Water Return or Heating Water | RAT: | Return Air Temperature | |
| ACU: | Air Conditioning Unit | utilize outdoor air for cooling as | Return | REQ: | Required | ||
| ACCU: | Air Cooled Condensing Unit | opposed to the use of mechanical | HWS: | Hot Water Supply or Heating Water | RF: | Return Air Fan | |
| ACH: | Air Changers Per Hour | cooling. | Supply | RHC: | Re-Heat Coil | ||
| AMP: | Ampere | EDC: | Electric Duct Coil | HX: | Heat Exchanger | RPM: | Revolutions Per Minute |
| AVG: | Average | EDH: | Electric Duct Heater | Hydronic: | Indicates the Use or Conveyance | ||
| EF: | Exhaust Fan | of Liquid for Thermal Transfer | SA: | Supply Air | |||
| EH: | Exhaust Hood | SAT: | Supply Air Temperature | ||||
| BDD: | Back Draft Damper | EMS: | Energy Management System (s) | ID: | Inside Diameter | SCD: | Static Control Damper |
| BHP: | Brake Horsepower | EMCS: | Energy Management Control System(s) | IV: | Inlet Vanes | SEF: | Smoke Exhaust Fan |
| BTU: | British Thermal Unit | ENT: | Entering | SF: | Supply Fan (Air) | ||
| BTUH: | British Thermal Units Per Hour | ER: | Exhaust Register | K: | Correction Factor | SHC: | Steam Heating Coil |
| EWT: | Entering Water Temperature | SP: | Set Point | ||||
| CAV: | Constant Air Volume | LAT: | Leaving Air Temperature | SPF: | Stairwell Pressurization Fan | ||
| CC: | Cooling Coil | F: | Degrees Fahrenheit | LPS: | Low Pressure Steam | SP 'WC': | Static Pressure Resistance Measured |
| CD: | Ceiling Diffuser | FCU: | Fan Coil Unit | LV: | Leaving | in Inches of Water Column | |
| CFM: | Cubic Feet Per Minute | FD: | Fire Damper | LWG: | Low Wall Grille | SWD | Sidewall Diffuser |
| CG | Ceiling Grille | FH: | Fume Hood | LWR: | Low Wall Register | SWG: | Sidewall Grille |
| CH: | Chiller | FG: | Floor Grille | LWT: | Leaving Water Temperature | SWR: | Sidewall Register |
| CHW: | Chilled Water | FLA: | Full Load Amperage | ||||
| CHWR: | Chilled Water Return | FLTS: | Filters | MUAH: | Make-Up Air Handler | TAB: | Testing, Adjusting and Balancing |
| CHWS: | Chilled Water Supply | FPB: | Fan Powered Box | MAU: | Make-Up Air Unit | TD: | Differential Temperature, Net |
| CP: | Circulating Pump | FPM: | Feet Per Minute | MBH: | Thousand BTU's Per Hour | Temperature Decrease or Increase | |
| CR: | Ceiling Register | FR: | Floor Register | TP: | Traverse Point/Test Point | ||
| CT: | Cooling Tower | FSD: | Fire/Smoke Damper | N/A: | Not Applicable or Not Available | TSP: | Total Static Pressure |
| CU: | Condensing Unit | FTU: | Fan Terminal Unit | N/L: | Not Listed | ||
| CUH: | Cabinet Unit Heater | NVL: | No Valid Location | UH: | Unit Heater | ||
| CW: | Condensing Water | GPM: | Gallons Per Minute | ||||
| CWR: | Condenser Water Return | OD: | Outside Diameter | V: | Voltage/Volt/Volts | ||
| CWS: | Condenser Water Supply | HC: | Heating Coil | VAV: | Variable Air Volume | ||
| CX: | Commissioning | HD: | Head Pressure Measured | OSA or OA: | Outside Air | VD: | Volume Damper |
| in Inches or Feet of Water | OAT: | Outside Air Temperature | VEL: | Velocity | |||
| DAT: | Discharge Air Temperature | Heater OL: | Thermal Overload Protection for | VFD: | Variable Frequency Drive | ||
| DB: | Dry Bulb | Motors Located at the Motor Starter | P: | Circulating Pump | (Electric Motor Speed Controller) | ||
| DD: | Direct Drive | HEPA: | High Efficiency Particulate | PD: | Differential Pressure or Pressure Drop | VP: | Velocity Pressure |
| DP: | Delta Pressure, Difference, Net Decrease or Increase | Arrestance | PF: | Power Factor | |||
| DNA: | Data Not Available | HOA: | Hand/Off/Auto Switch | PHC: | Pre-Heat Coil | W: | Watts |
| DNG: | Data Not Given | HP: | Horse Power | PH: | Phase (s) | WB: | Wet Bulb |
| DNL: | Data Not Listed | HPS: | High Pressure Steam | PSI: | Pounds Per Square Inch | WG: | Water Gauge |
| DNP: | Data Not Provided | HRC: | Heat Recovery Coil or Heat | ||||
| Reclaim Coil |