This document outlines the technical specifications for a thermal furnace sulfur analyzer required for a state and local contract opportunity in Illinois. The requested instrument must be capable of analyzing sulfur content in samples ranging from 0.0015% to 35% within 3 minutes, and fit within a 3-foot by 3-foot lab bench space. The instrument must include a Windows computer, touch screen monitor, dedicated software, and remote access capability. Installation, onsite training, a dedicated balance, maintenance kit, and enough consumables for 5,000 analyses are also required. The use of a Leco sulfur analyzer or equivalent is specified, with the ability for "or equal" submissions that meet the stated requirements.

The contract opportunity provides for the purchase of a complete thermal sulfur analyzer system, including installation and training. No specific pricing terms, set asides, or funding details are provided. The contract will be awarded directly from the manufacturer, with no third-party contracting allowed, to ensure proper training, troubleshooting, and warranty support.

View the file

Other files

File Type Posted
standard-terms-and-conditions~97.pdf PDF
Standard Certifications V23.1~74.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Leco Sulfur Analyzer or Equivalent

• The instrument needs to be a thermal furnace sulfur analyzer capable of 1450°C

• The instrument must have dual IR cells capable of a detection range from 0.0015% to 35% sulfur in a sample

• The instrument must be able to analyze samples of 200-350mg within 3 minutes, nominal

• The unit must be able to fit within a 3 foot by three 3 foot space on a lab bench

• The instrument must come with a windows computer and touch screen monitor

• The instrument must have a dedicated software package for instrument control and data analysis

• The instrument software must be capable of remote access

• The instrument must include installation and onsite training

• The instrument must include a dedicated balance for the sulfur analyzer

• The balance must have a capacity of 100 grams and a readability of 0.lmilligrams

• The instrument must come with a maintenance kit including one spare heating element

• The instrument must come with enough consumables for 5000 analyses

• The instrument and all components must be in new condition

• Leco Sufur analyzer or similar instrument as listed in "Soil Science Society of America standard methods for soil sulfur analysis" is required

• The instrument must come direct from the manufacturer (no third party contracting) for training, trouble shooting, and warranty solutions

The use of brand name is to describe the standard of quality, performance and characteristics desired and is not intended to limit or restrict competition. “Or Equal” submissions will not be rejected because of minor differences in design, construction or features that do not affect the suitability of the product for its intended use. Burden of proof that the product is equal is on the bidder.

Vendor should provide product literature with the quote. Shipping costs to be included with bid.

Product must be new, unused and free from defects.