Temperature Sensor Instrumentation Specification.docx

DOCX document 129 KB Posted

Attached to
CRYOMODULE TEMPERATURE SENSORS
Solicitation number
364237
Issued by

About this file

This document is a Temperature Sensor Instrumentation Specification detailing technical specifications for three specific temperature sensor models to be used in PIPII cryomodules: PT103 Platinum RTD and two Cernox models (CX-1030-CU-HT,SD-HT and CX-1030-CU-HT,CU-HT). The specifications comprehensively outline each sensor's performance characteristics, including temperature ranges, thermal response times, accuracy, reproducibility, radiation stability, and recommended excitation levels. Both sensor types are designed for cryogenic applications, with the Cernox models capable of measuring temperatures from 0.1K to 420K and the PT103 functioning between 14K and 873K.

The related federal contract opportunity, issued by the Department of Energy's Fermi National Accelerator Laboratory, is a pre-solicitation for procuring these specific temperature sensors. Fermi Research Alliance, LLC intends to issue a sole-source subcontract to Lake Shore Cryotronics Inc. for 421 CX-1030-SD-HT-1.4L sensors, 202 CX-1030-CU-HT-1.4L sensors, 45 PT-103 sensors, and 97 PT-103-AM sensors. Interested organizations may submit capabilities and qualifications by 4:00 p.m. Central Standard Time on 04/10/2025, with the government retaining discretion to determine competitive procurement approach.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Purpose This document contains the specifications for PIPII cryomodule temperature sensors.

Scope This specification document includes parameters for the internal cryomodule sensors:

Temperature sensors

· PT103 Platinum RTD

· Cernox CX-1030-CU-HT,SD-HT

· Cernox CX-1030-CU-HT,CU-HT Definitions

RTD
Resistance Temperature Detector

References

ED0002454
PIPII Cryomodule Instrumentation List

Responsibilities

Person(s) or Areas Responsible
Define Responsibility

Overview

PT103 Platinum RTDs.

Specification
Value
Implication
Type
PT103
Standard Curve
IEC 751
Not individually calibrated
Recommended excitation
1 mA
Temperature Range
14K to 873K
Suitable for cryogenic operation
Thermal Response Time
1.75 s at 77K
Reproducibility
±5 mK at 77 K
Short-term reproducibility data is obtained by subjecting sensor to repeated thermal shocks from 305 K to 77 K
Long-term stability
±10 mK at 77 K
Long-term stability data is obtained by subjecting sensor to 200 thermal shocks from 305 K to 77 K
Typical Accuracy
At 30K ±10 mK At 77K ±12 mK

At 305K ±23 mK

Dimensions
0.45” x 0.063”
Fits inside designated sensor mount
Lead Type
2, platinum
Material used
Platinum winding partially supported by a

high temperature alumina powder inside a ceramic tube, platinum lead wires Non-Magnetic

Radiation Stability
Δ75mK over full temperature range after 106 Rads exposure
Exceeds requirements (Δ250mK over full temperature range after 106 Rads exposure)

. Cernox CX-1030-CU-HT,SD-HT RTDs The following specifications are for the Cernox CX-1030-CU-HT,SD-HT RTD sensors.

Specification
Value
Implication
Type
Cernox CX-1030-CU-HT, Cernox CX-1030-SD-HT
CU Bobbin: Large thermal anchor

SD Package: Small, convenient pkg.

Recommended excitation
20 μV (0.1 K to 0.5 K)

63 μV (0.5 K to 1 K) 10 mV or less for T > 1.2 K

Temperature Range
0.1K to 420K
Suitable for liquid He temperatures
Thermal Response Time
0.25 s at 77K (SD)
Calibration
Non-Standard
Individually calibrated
Dissipation at recommended excitation
Typical 10-5 W at 300 K, 10-7 W at 4.2 K, 10-13 W at 0.3 K
Reproducibility
±3 mK at 4.2 K
Short-term reproducibility data is obtained by subjecting sensor to repeated thermal shocks from 305 K to 4.2 K
Calibrated Accuracy
1.4 K
±5 mK
4.2 K
±5 mK
10 K
±6 mK
20 K
±9 mK
30 K
±10mK
50 K
±13mK
77 K
±16mK
300K
±40mK

400K ±65mK

Long-term stability
1.4 K
±3 mK
4.2 K
±3 mK
10 K
±6 mK
20 K
±12mK
30 K
±18mK
50 K
±30mK
77 K
±46mK
300K
±180mK

Long-term stability data is obtained by subjecting sensor to 200 thermal shocks from 305 K to 77 K

Lead type
2, gold-plated copper
Radiation Stability
±10 mK at 1.4 K after 106 Rads exposure
Exceeds requirements (±250mK @ 1.4K after 106 Rads exposure)
EMI Susceptible
Yes
Leads should be shorted together when not in use to avoid damaging sensor

Magnetic Field Dependence Data for CX RTD Ionizing Radiation Effects image3.emf image4.emf image1.emf image2.png