Temperature Humidity Chamber Specification Sheet.docx
DOCX document 48 KB Posted
- Attached to
- EMA- Temp/Humidity Chamber - ARG
- Solicitation number
- 22-588IEMA-MGMTA-B-28327
- Issued by
- St. Clair County, Illinois
About this file
This document is a specifications sheet for a Temperature/Humidity Chamber required by the Illinois Emergency Management Agency (EMA) for their Nuclear Calibration Laboratory located in Springfield, IL. The EMA is seeking a vendor to provide a Chamber that meets the detailed requirements outlined, including: chamber size of 30" x 30" x 30", temperature range of -10 to +50 C, ambient conditions of 72 F and 20-60% RH, and additional accessories such as casters, recirculating water, and a humidity water filtration system. The Chamber will be used for electrical leakage testing of direct reading electrostatic/electroscope type dosimeters loaded in 24" diameter, 6" height test racks weighing approximately 17 lbs each.
The EMA has a related state and local contract opportunity open for this Temperature/Humidity Chamber, with an opportunity ID of 66aaf2107cf5b14e190baa4c and the name "EMA- Temp/Humidity Chamber - ARG". No additional pricing, set-aside, incumbent, or funding details are provided in the specifications sheet.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Forms B Section~16.docx | DOCX document | |
| Standard Terms and Conditions v.20.1~70.pdf | ||
| Standard Certifications V.18.1~55.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS FOR Nuclear Calibration Laboratory 1301 Knotts Street | Springfield, IL 62703 EMA.CalibrationLab@illinois.gov 217.786.7221
TEMPERATURE/HUMIDITY CHAMBER
REQUIREMENTS
| Purpose: |
| Electrical leakage testing as described in ANSI N322 |
| Test Item Description: |
| Direct reading electrostatic/electroscope type dosimeters (loaded in test racks) |
Rack diameter: 24” Rack height: 6” Weight (per rack): ~17 lbs
| Chamber Size: |
| Z-Plus Model ZP(H)-16: 30” x 30” x 30” |
| Test Conditions: |
| -10 to +50 °C |
Ramp time undefined (as quick as possible) Ambient conditions: 72 °F, 20 – 60% RH
| Utilities: |
| 208V 1PH 60Hz |
| Transportation: |
| Forklift available, no dock |
| Accessories: |
| Casters |
Recirculating water Humidity water filtration system
(4) Stainless steel shelves Running time meter
INSTALLATION AND ONSITE SERVICES
· Vendor will deliver systems to the agency, include freight quote image1.jpeg
File details come from the government source that posted it. Updated .