Temperature humidity chamber spec sheet~1.docx
DOCX document 48 KB Posted
- Attached to
- EMA- Temp/Humidity Chamber - JS
- Solicitation number
- 21-588IEMA-MGMTA-B-21656
- Issued by
- St. Clair County, Illinois
About this file
This document is a specification sheet for a temperature/humidity chamber required by the Illinois Emergency Management Agency (EMA) Calibration Laboratory located in Springfield, Illinois. The chamber is intended for electrical leakage testing of direct reading electrostatic/electroscope type dosimeters loaded in test racks. The required chamber size is 30" x 30" x 30" and must be able to maintain test conditions of -10°C to +50°C. Additional requirements include a recirculating water system, humidity water filtration, stainless steel shelves, and a running time meter. The vendor is responsible for delivering the chamber, providing onsite installation, start-up, and training services.
The related state and local contract opportunity is identified as EMA- Temp/Humidity Chamber - JS, with the State of Illinois as the jurisdiction. No other details regarding pricing terms, set asides, incumbents, funding, or contract duration are provided in the given information.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Standard Certifications V.18.1~31.doc | DOC document | |
| Standard Terms and Conditions v.20.1~20.pdf | ||
| Forms B Section~7.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS FOR Nuclear Calibration Laboratory 1301 Knotts Street | Springfield, IL 62703 EMA.CalibrationLab@illinois.gov 217.786.7221
TEMPERATURE/HUMIDITY CHAMBER
REQUIREMENTS
| Purpose: |
| Electrical leakage testing as described in ANSI N322 |
| Test Item Description: |
| Direct reading electrostatic/electroscope type dosimeters (loaded in test racks) |
Rack diameter: 24” Rack height: 6” Weight (per rack): ~17 lbs
| Chamber Size: |
| Z-Plus Model ZP(H)-16: 30” x 30” x 30” |
| Test Conditions: |
| -10 to +50 °C |
Ramp time undefined (as quick as possible) Ambient conditions: 72 °F, 20 – 60% RH
| Utilities: |
| 208V 1PH 60Hz |
| Transportation: |
| Fork lift available, no dock |
| Accessories: |
| Casters |
Recirculating water Humidity water filtration system Stainless steel shelves Running time meter Power cord
INSTALLATION AND ONSITE SERVICES
· Vendor will deliver systems to the agency, include freight quote
· Onsite installation/start-up and training services should be provided by vendor image1.jpeg