Temperature humidity chamber spec sheet~1.docx

DOCX document 48 KB Posted

Attached to
EMA- Temp/Humidity Chamber - ARG
Solicitation number
22-588IEMA-MGMTA-B-27024
Issued by
St. Clair County, Illinois

About this file

This document contains the specifications for a Temperature/Humidity Chamber required by the Illinois Emergency Management Agency (EMA) for its Nuclear Calibration Laboratory located in Springfield, Illinois. The chamber is needed to conduct electrical leakage testing on direct reading electrostatic/electroscope type dosimeters loaded in test racks. The required chamber specifications include a size of 30" x 30" x 30", temperature range of -10°C to +50°C, and ambient conditions of 72°F and 20-60% relative humidity. The chamber must include various accessories such as casters, recirculating water, humidity water filtration system, stainless steel shelves, and a running time meter.

The EMA is seeking a vendor to provide this Temperature/Humidity Chamber under a state and local contract opportunity identified as "EMA- Temp/Humidity Chamber - ARG". The contract will require the vendor to deliver the system, provide onsite installation/startup, and offer training services. No further details on pricing, funding, or the contract term are provided in the document.

View the file

Other files

File Type Posted
Standard Certifications V.18.1~55.doc DOC document
Forms B Section~16.docx DOCX document
Standard Terms and Conditions v.20.1~70.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATIONS FOR Nuclear Calibration Laboratory 1301 Knotts Street | Springfield, IL 62703 EMA.CalibrationLab@illinois.gov 217.786.7221

TEMPERATURE/HUMIDITY CHAMBER

REQUIREMENTS

Purpose:
Electrical leakage testing as described in ANSI N322
Test Item Description:
Direct reading electrostatic/electroscope type dosimeters (loaded in test racks)

Rack diameter: 24” Rack height: 6” Weight (per rack): ~17 lbs

Chamber Size:
Z-Plus Model ZP(H)-16: 30” x 30” x 30”
Test Conditions:
-10 to +50 °C

Ramp time undefined (as quick as possible) Ambient conditions: 72 °F, 20 – 60% RH

Utilities:
208V 1PH 60Hz
Transportation:
Fork lift available, no dock
Accessories:
Casters

Recirculating water Humidity water filtration system Stainless steel shelves Running time meter Power cord

INSTALLATION AND ONSITE SERVICES

· Vendor will deliver systems to the agency, include freight quote

· Onsite installation/start-up and training services should be provided by vendor image1.jpeg