This is a technical specification document for a Pre-Solicitation (FA812525Q0027) issued by the Department of the Air Force Materiel Command Air Force Sustainment Center for four Water Cooled X-ray Systems.

The required system must automatically adapt to input line voltage of 100-130Vac or 200-250Vac at 50/60Hz, with operating capability of 10-160kV and 0.5-5.0mA. Key specifications include a Beryllium 0.8mm window with 40° directional radiation output of 14R/min at 50cm, radiation leakage less than 2.0 Roentgens per hour at 1 meter, and 100% duty cycle at 120°F. The system includes a liquid cooling unit (54 lbs), tube head (7.25" diameter, 30" length, 29 lbs), and control unit (26 lbs). Safety features include pressure relief valve, thermal cut-out, pressure gauge, low pressure cut-out, coolant flow sensor, and keyswitch. Standard accessories include operation manual, carrying case, 100' tubehead cable, 25' power cable, 50' coolant hose, and GUI control software for system operation and diagnostics.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

A24-007 Water Cooled Xray System 76 AMXG Request for the purchase of 4 ea.

Water Cooled Xray System to include the following:

Input Line Requirements:

Automatically adapts to input line voltage100-130Vac, 50/60Hz, 20 Amperes maximum.

200-250Vac, 50/60Hz, 10 Amperes maximum.

10 to 160kV, 0.5mA to 5.0mA (800 watts max.)

Constant potential, end-grounded anode, liquid cooled version.

X-Ray Tube Window: Beryllium 0.8mm (directional) 40° directional.

Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm).

aluminum at 160kV, 5mA.

Radiation Leakage: Less than 2.0 Roentgens per hour at 1 meter from the X-Ray tube target.

Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN125430.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336.

Ambient Temp: 100% duty cycle @ 120°F (49°C).

Duty Cycle: 100%.

Storage Temp: -30°F to 160°F (-35°C to 71°C).

Anode Cooling: Liquid coolant solution closed loop between X-Ray tube anode and cooling unit.

Liquid Cooling Unit Weight: 54lbs. (15.4kg) approx.

Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length.

Liquid Cooled: 29lbs. (13.15kg) X-Ray Control Unit: X-Ray Control Unit Weight: 26lbs. (11.8kg) approx.

Safety Devices:

Tubehead Pressure Relief Valve

Tubehead Thermal Cut-Out

Tubehead Pressure Gauge

Tubehead Low Pressure Cut-Out @ 25psi (1.72 Bar)

Coolant Flow Sensor (liquid cooled only)

Control Unit Safety Keyswitch

Microcomputer-based Self Diagnostics

Continuous Exposure Parameter Display

Standard Accessories:

Operation manual

Tubehead carrying case

Tubehead Cable- 100’ with strain relief

Extra key (1) for Control Unit Safety Lock

Power cable - 25’ with strain relief

Coolant hose - twin, 50’ with self-sealing terminations

(liquid cooled only)

Cooler power cable

GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user-friendly windows-based menu. Additionally, the GUI can be used as a diagnostic tool when the system is controlled via the front panel.